ASTM D7504-23
(Test Method)Standard Test Method for Trace Impurities in Monocyclic Aromatic Hydrocarbons by Gas Chromatography and Effective Carbon Number
Standard Test Method for Trace Impurities in Monocyclic Aromatic Hydrocarbons by Gas Chromatography and Effective Carbon Number
SIGNIFICANCE AND USE
5.1 Determining the type and amount of hydrocarbon impurities remaining from the manufacture of toluene, mixed xylenes, p-xylene, o-xylene, ethylbenzene, benzene, and styrene used as chemical intermediates and solvents is often required. This test method is suitable for setting specifications and for use as an internal quality control tool where these products are produced or are used. Typical impurities are: alkanes containing 1 to 10 carbons atoms, benzene, toluene, ethylbenzene (EB), xylenes, and aromatic hydrocarbons containing nine carbon atoms or more.
5.2 This method may not detect all components and there may be unknown components that would be assigned inappropriate correction factors and thus, the results may not be absolute.
SCOPE
1.1 This test method covers the determination of total nonaromatic hydrocarbons and monocyclic aromatic hydrocarbons in benzene, toluene, ethylbenzene, p-xylene, o-xylene, styrene and mixed xylenes by gas chromatography. The purity of benzene, toluene, ethylbenzene, p-xylene, o-xylene, styrene and mixed xylenes is also calculated. Similar test methods, using the internal standard calibration technique and the external standard calibration technique, are Test Methods D2360, D5060, D5135, and D5917 respectively.
1.2 The limit of detection (LOD) is 0.0002 mass % and limit of quantitation (LOQ) is 0.0006 mass % for impurities in toluene, mixed xylenes, p-xylene, o-xylene, ethylbenzene, benzene, and styrene.
1.3 In determining the conformance of the test results using this method to applicable specifications, results shall be rounded off in accordance with the rounding-off method of Practice E29.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
- Status
- Published
- Publication Date
- 30-Apr-2023
- Technical Committee
- D16 - Aromatic, Industrial, Specialty and Related Chemicals
- Drafting Committee
- D16.01 - Benzene, Toluene, Xylenes, Cyclohexane and Their Derivatives
Relations
- Effective Date
- 12-Jan-2018
- Replaces
ASTM D4492-10 - Standard Test Method for Analysis of Benzene by Gas Chromatography (Withdrawn 2018) - Effective Date
- 12-Jan-2018
- Refers
ASTM D5211-19 - Standard Specification for Xylenes for <emph type="ital">p</emph>-Xylene Feedstock - Effective Date
- 01-Nov-2019
- Effective Date
- 01-Sep-2019
- Effective Date
- 01-Jun-2019
- Effective Date
- 01-Jan-2019
- Effective Date
- 01-Jul-2017
- Refers
ASTM D5211-17 - Standard Specification for Xylenes for <emph type="ital">p</emph>-Xylene Feedstock - Effective Date
- 01-Jun-2017
- Effective Date
- 01-Jun-2017
- Effective Date
- 01-Nov-2016
- Refers
ASTM D5211-16 - Standard Specification for Xylenes for <emph type="ital">p</emph>-Xylene Feedstock - Effective Date
- 01-Jun-2016
- Effective Date
- 01-Jun-2016
- Effective Date
- 01-Feb-2015
- Effective Date
- 01-Jul-2014
- Effective Date
- 01-Feb-2014
Overview
ASTM D7504-23: Standard Test Method for Trace Impurities in Monocyclic Aromatic Hydrocarbons by Gas Chromatography and Effective Carbon Number defines a reliable procedure for assessing trace impurities in key monocyclic aromatic hydrocarbons. These include benzene, toluene, ethylbenzene, p-xylene, o-xylene, styrene, and mixed xylenes, which are commonly used as chemical intermediates and solvents. Using gas chromatography and effective carbon number (ECN) correction factors, this standard supports precise determination of the type and quantity of trace impurities-vital for ensuring product purity and quality control in the chemical and petrochemical industries.
Key Topics
Scope of Analysis
The method addresses both nonaromatic hydrocarbons and monocyclic aromatic hydrocarbon impurities present in industrial chemicals such as benzene, toluene, mixed xylenes, and styrene.Analytical Technique
Gas chromatography (GC) with flame ionization detection (FID) is used, leveraging ECN correction factors to normalize and accurately quantify impurity concentrations.Performance Criteria
- Limit of detection (LOD): 0.0002 mass %
- Limit of quantitation (LOQ): 0.0006 mass % Results are reported to specific significant figures in compliance with ASTM rounding practices.
Quality Control & Repeatability
Designed for high repeatability and reproducibility, the method serves as an internal specification tool and supports interlaboratory comparability. Quality control sample analysis and instrument calibration guidelines ensure consistent results.Limitations
It may not identify all possible components, particularly unknown substances without appropriate correction factors, meaning results are not absolute and should be interpreted within these constraints.
Applications
Purity Assessment
Widely used for verifying the purity of monocyclic aromatic hydrocarbons vital to chemical manufacturing, meeting both internal and external quality requirements.Specification Setting
Supports the creation and enforcement of product specifications for the sale and receipt of high-purity chemicals such as toluene, xylenes, benzene, ethylbenzene, and styrene.Process Control
Integral to in-house monitoring programs for manufacturers and end-users, detecting low-level impurities that may impact downstream processing or final product quality.Regulatory Compliance
Assists laboratories in fulfilling regulatory, safety, health, and environmental standards, particularly where trace-component documentation is required.Interlaboratory and Proficiency Testing
Provides a framework for proficiency testing and comparison across multiple laboratories, strengthening confidence in test result reliability for producers and consumers.
Related Standards
- ASTM D2360: Trace Impurities in Monocyclic Aromatic Hydrocarbons by Gas Chromatography (withdrawn)
- ASTM D5060: Determining Impurities in High-Purity Ethylbenzene by Gas Chromatography
- ASTM D5135: Analysis of Styrene by Capillary Gas Chromatography
- ASTM D5917: Trace Impurities in Monocyclic Aromatic Hydrocarbons by Gas Chromatography and External Calibration
- ASTM E29: Practice for Using Significant Digits in Test Data
- ASTM D3437: Practice for Sampling and Handling Liquid Cyclic Products
- ASTM D6809: Guide for Quality Control and Quality Assurance Procedures for Aromatic Hydrocarbons and Related Materials
Practical Value
Adopting ASTM D7504-23 enhances the ability of laboratories and manufacturers to accurately monitor and control the purity of essential aromatic hydrocarbons. The method’s low detection limits and robust procedural controls ensure the detection and quantification of even trace amounts of contamination, playing a critical role in upholding product integrity across global supply chains. Leveraging this standard benefits organizations by improving quality assurance, facilitating regulatory compliance, and enabling consistent communication regarding product quality among trading partners.
Keywords: gas chromatography, aromatic hydrocarbons, trace impurities, benzene, toluene, xylenes, styrene, purity testing, ECN, industrial chemicals, ASTM D7504-23.
Buy Documents
ASTM D7504-23 - Standard Test Method for Trace Impurities in Monocyclic Aromatic Hydrocarbons by Gas Chromatography and Effective Carbon Number
REDLINE ASTM D7504-23 - Standard Test Method for Trace Impurities in Monocyclic Aromatic Hydrocarbons by Gas Chromatography and Effective Carbon Number
Get Certified
Connect with accredited certification bodies for this standard

ECOCERT
Organic and sustainability certification.

Eurofins Food Testing Global
Global leader in food, environment, and pharmaceutical product testing.

Intertek Bangladesh
Intertek certification and testing services in Bangladesh.
Sponsored listings
Frequently Asked Questions
ASTM D7504-23 is a standard published by ASTM International. Its full title is "Standard Test Method for Trace Impurities in Monocyclic Aromatic Hydrocarbons by Gas Chromatography and Effective Carbon Number". This standard covers: SIGNIFICANCE AND USE 5.1 Determining the type and amount of hydrocarbon impurities remaining from the manufacture of toluene, mixed xylenes, p-xylene, o-xylene, ethylbenzene, benzene, and styrene used as chemical intermediates and solvents is often required. This test method is suitable for setting specifications and for use as an internal quality control tool where these products are produced or are used. Typical impurities are: alkanes containing 1 to 10 carbons atoms, benzene, toluene, ethylbenzene (EB), xylenes, and aromatic hydrocarbons containing nine carbon atoms or more. 5.2 This method may not detect all components and there may be unknown components that would be assigned inappropriate correction factors and thus, the results may not be absolute. SCOPE 1.1 This test method covers the determination of total nonaromatic hydrocarbons and monocyclic aromatic hydrocarbons in benzene, toluene, ethylbenzene, p-xylene, o-xylene, styrene and mixed xylenes by gas chromatography. The purity of benzene, toluene, ethylbenzene, p-xylene, o-xylene, styrene and mixed xylenes is also calculated. Similar test methods, using the internal standard calibration technique and the external standard calibration technique, are Test Methods D2360, D5060, D5135, and D5917 respectively. 1.2 The limit of detection (LOD) is 0.0002 mass % and limit of quantitation (LOQ) is 0.0006 mass % for impurities in toluene, mixed xylenes, p-xylene, o-xylene, ethylbenzene, benzene, and styrene. 1.3 In determining the conformance of the test results using this method to applicable specifications, results shall be rounded off in accordance with the rounding-off method of Practice E29. 1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
SIGNIFICANCE AND USE 5.1 Determining the type and amount of hydrocarbon impurities remaining from the manufacture of toluene, mixed xylenes, p-xylene, o-xylene, ethylbenzene, benzene, and styrene used as chemical intermediates and solvents is often required. This test method is suitable for setting specifications and for use as an internal quality control tool where these products are produced or are used. Typical impurities are: alkanes containing 1 to 10 carbons atoms, benzene, toluene, ethylbenzene (EB), xylenes, and aromatic hydrocarbons containing nine carbon atoms or more. 5.2 This method may not detect all components and there may be unknown components that would be assigned inappropriate correction factors and thus, the results may not be absolute. SCOPE 1.1 This test method covers the determination of total nonaromatic hydrocarbons and monocyclic aromatic hydrocarbons in benzene, toluene, ethylbenzene, p-xylene, o-xylene, styrene and mixed xylenes by gas chromatography. The purity of benzene, toluene, ethylbenzene, p-xylene, o-xylene, styrene and mixed xylenes is also calculated. Similar test methods, using the internal standard calibration technique and the external standard calibration technique, are Test Methods D2360, D5060, D5135, and D5917 respectively. 1.2 The limit of detection (LOD) is 0.0002 mass % and limit of quantitation (LOQ) is 0.0006 mass % for impurities in toluene, mixed xylenes, p-xylene, o-xylene, ethylbenzene, benzene, and styrene. 1.3 In determining the conformance of the test results using this method to applicable specifications, results shall be rounded off in accordance with the rounding-off method of Practice E29. 1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
ASTM D7504-23 is classified under the following ICS (International Classification for Standards) categories: 71.080.15 - Aromatic hydrocarbons. The ICS classification helps identify the subject area and facilitates finding related standards.
ASTM D7504-23 has the following relationships with other standards: It is inter standard links to ASTM D6526-12, ASTM D4492-10, ASTM D5211-19, ASTM E260-96(2019), ASTM D5917-15(2019), ASTM D5136-19, ASTM D5136-17a, ASTM D5211-17, ASTM D5136-17, ASTM D6809-02(2016), ASTM D5211-16, ASTM D5135-16, ASTM D5917-15, ASTM D4790-14, ASTM D5135-14. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ASTM D7504-23 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: D7504 − 23
Standard Test Method for
Trace Impurities in Monocyclic Aromatic Hydrocarbons by
Gas Chromatography and Effective Carbon Number
This standard is issued under the fixed designation D7504; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope* mendations issued by the World Trade Organization Technical
Barriers to Trade (TBT) Committee.
1.1 This test method covers the determination of total
nonaromatic hydrocarbons and monocyclic aromatic hydrocar-
2. Referenced Documents
bons in benzene, toluene, ethylbenzene, p-xylene, o-xylene,
2.1 ASTM Standards:
styrene and mixed xylenes by gas chromatography. The purity
D1555M Test Method for Calculation of Volume and Weight
of benzene, toluene, ethylbenzene, p-xylene, o-xylene, styrene
of Industrial Aromatic Hydrocarbons and Cyclohexane
and mixed xylenes is also calculated. Similar test methods,
[Metric]
using the internal standard calibration technique and the
D2360 Test Method for Trace Impurities in Monocyclic
external standard calibration technique, are Test Methods
Aromatic Hydrocarbons by Gas Chromatography (With-
D2360, D5060, D5135, and D5917 respectively.
drawn 2016)
1.2 The limit of detection (LOD) is 0.0002 mass % and limit
D3437 Practice for Sampling and Handling Liquid Cyclic
of quantitation (LOQ) is 0.0006 mass % for impurities in
Products
toluene, mixed xylenes, p-xylene, o-xylene, ethylbenzene,
D4790 Terminology of Aromatic Hydrocarbons and Related
benzene, and styrene.
Chemicals
1.3 In determining the conformance of the test results using
D5060 Test Method for Determining Impurities in High-
this method to applicable specifications, results shall be
Purity Ethylbenzene by Gas Chromatography
rounded off in accordance with the rounding-off method of
D5135 Test Method for Analysis of Styrene by Capillary
Practice E29.
Gas Chromatography
D5136 Specification for High Purity p-Xylene
1.4 The values stated in SI units are to be regarded as
D5211 Specification for Xylenes for p-Xylene Feedstock
standard. No other units of measurement are included in this
D5917 Test Method for Trace Impurities in Monocyclic
standard.
Aromatic Hydrocarbons by Gas Chromatography and
1.5 This standard does not purport to address all of the
External Calibration
safety concerns, if any, associated with its use. It is the
D6809 Guide for Quality Control and Quality Assurance
responsibility of the user of this standard to establish appro-
Procedures for Aromatic Hydrocarbons and Related Ma-
priate safety, health, and environmental practices and deter-
terials
mine the applicability of regulatory limitations prior to use.
E29 Practice for Using Significant Digits in Test Data to
1.6 This international standard was developed in accor-
Determine Conformance with Specifications
dance with internationally recognized principles on standard-
E260 Practice for Packed Column Gas Chromatography
ization established in the Decision on Principles for the
E355 Practice for Gas Chromatography Terms and Relation-
Development of International Standards, Guides and Recom-
ships
This test method is under the jurisdiction of ASTM Committee D16 on
Aromatic, Industrial, Specialty and Related Chemicals and is the direct responsi- For referenced ASTM standards, visit the ASTM website, www.astm.org, or
bility of Subcommittee D16.01 on Benzene, Toluene, Xylenes, Cyclohexane and contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Their Derivatives. Standards volume information, refer to the standard’s Document Summary page on
Current edition approved May 1, 2023. Published May 2023. Originally the ASTM website.
approved in 2009. Last previous edition approved in 2021 as D7504 – 21. DOI: The last approved version of this historical standard is referenced on
10.1520/D7504-23. www.astm.org.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D7504 − 23
TABLE 1 Recommended Method Parameters
E691 Practice for Conducting an Interlaboratory Study to
Determine the Precision of a Test Method Inlet Split
Temperature, °C 270
E1510 Practice for Installing Fused Silica Open Tubular
Column:
Capillary Columns in Gas Chromatographs
Tubing fused silica
Length, m 60
2.2 Other Document:
Internal diameter, mm 0.32
OSHA Regulations, 29 CFR paragraphs 1910.1000 and
Stationary phase crosslinked polyethylene
1910.1200
glycol
Film thickness, μm 0.25
Column temperature program
3. Terminology
Initial temperature, °C 60
3.1 See Terminology D4790 for definitions of terms used in Initial time, min 10
Programming rate, °C/min 5
this test method.
Final, °C 150
Time 2, min 10
4. Summary of Test Method
Carrier gas helium or hydrogen
Linear velocity, cm/s at 145 °C 20 helium or 45 hydrogen
4.1 The specimen to be analyzed is injected into a gas
Split ratio 100:1
chromatograph equipped with a flame ionization detector (FID)
Sample size, μL 0.6
Detector: flame ionization
and a capillary column. The peak area of each component is
5 Temperature, °C 300
measured and adjusted using effective carbon number (ECN)
Analysis time, min 38
correction factors. The concentration of each component is
calculated based on its relative percentages of total adjusted
peak area and normalized to 100.0000 %.
used successfully and shall be used as a referee in cases of
5. Significance and Use
dispute. A 60 m column with an internal diameter of 0.32 mm
and a film thickness of 0.50 μm crosslinked polyethylene
5.1 Determining the type and amount of hydrocarbon im-
glycol may improve resolution.
purities remaining from the manufacture of toluene, mixed
xylenes, p-xylene, o-xylene, ethylbenzene, benzene, and sty-
7.3 Gas Chromatograph—Any instrument having a flame
rene used as chemical intermediates and solvents is often
ionization detector and a splitter injector suitable for use with
required. This test method is suitable for setting specifications
a fused silica capillary column may be used, provided the
and for use as an internal quality control tool where these
system has sufficient sensitivity, linearity, and range to deter-
products are produced or are used. Typical impurities are:
mine 0.0001 mass %, while not exceeding the full scale of
alkanes containing 1 to 10 carbons atoms, benzene, toluene,
either the detector or the electronic integration for the major
ethylbenzene (EB), xylenes, and aromatic hydrocarbons con-
component. It shall have a split injection system that will not
taining nine carbon atoms or more. discriminate over the boiling range of the samples analyzed.
The system should be capable of operating at conditions given
5.2 This method may not detect all components and there
in Table 1.
may be unknown components that would be assigned inappro-
priate correction factors and thus, the results may not be
7.4 Injector—The specimen must be precisely and repeat-
absolute. ably injected into the gas chromatograph. An automatic sample
injection devise is highly recommended.
6. Interferences
7.5 Syringe—Chromatographic, capable of delivering ap-
6.1 The complete separation of p-xylene from ethylbenzene,
propriate μL volumes.
or ethylbenzene and m-xylene from p-xylene can be difficult
when either ethylbenzene or p-xylene is analyzed, respectively.
8. Reagents and Materials
The separation can be considered adequate if the distance from
8.1 Purity of Reagent—Reagent grade chemicals shall be
the baseline to the valley between the two peaks is not greater
used in all tests. Unless otherwise indicated, it is intended that
than 50 % of the peak height of lower of the two peaks.
all reagents shall conform to the specifications of the Commit-
tee on Analytical Reagents of the American Chemical Society,
7. Apparatus
where such specifications are available. Reagents with an
7.1 Chromatographic data system is required.
establish purity greater than ACS reagent grade may be used.
7.2 Columns—The choice of column is based on resolution 8.1.1 Alpha-methylstyrene (AMS) CAS 98-83-9,
8.1.2 Benzene CAS 71-43-2,
requirements. Any column may be used that is capable of
resolving all significant impurities from the major component. 8.1.3 Cumene (isopropylbenzene) CAS 98-82-8,
8.1.4 1,4-dioxane CAS 123-91-1,
The column and conditions described in Table 1 have been
4 6
Available from U.S. Government Printing Office Superintendent of Documents, ACS Reagent Chemicals, Specifications and Procedures for Reagents and
732 N. Capitol St., NW, Mail Stop: SDE, Washington, DC 20401, http:// Standard-Grade Reference Materials, American Chemical Society, Washington,
www.access.gpo.gov. DC. For suggestions on the testing of reagents not listed by the American Chemical
Scanlon, J. T. and Willis, D. E., “Calculation of Flame Ionization Detector Society, see Analar Standards for Laboratory Chemicals, BDH Ltd., Poole, Dorset,
Relative Response Factors Using the Effective Carbon Number Concept,” Journal U.K., and the United States Pharmacopeia and National Formulary, U.S. Pharma-
of Chromatographic Science, Vol. 23, August 1985, pp. 333–339. copeial Convention, Inc. (USPC), Rockville, MD.
D7504 − 23
8.1.5 Ethylbenzene CAS 100-41-4, 8.4.4.2 Pipette 1 mL of the peak identification material
prepared in 8.4.2 into the 500 mL volumetric flask.
8.1.6 Ethyltoluene CAS 622-96-8,
8.1.7 m-xylene CAS 108-38-3, 8.4.4.3 Use a 10 mL burette to add 1.1 mL of the sensitivity
check material prepared in 8.4.3 to the 500 mL volumetric
8.1.8 n-propylbenzene CAS 103-65-1,
flask.
8.1.9 o-xylene CAS 95-47-6,
8.4.4.4 Dilute to volume with high purity p-xylene and mix
8.1.10 p-diethylbenzene (PDEB) CAS 105-05-5,
thoroughly.
8.1.11 p-xylene CAS 106-42-3,
8.4.5 Impurities that are not present in the samples being
8.1.12 Phenylacetylene (ethynylbenzene) CAS 536-74-3,
analyzed may be omitted from the check sample. Impurities
8.1.13 2-propenylbenzene CAS 300-57-2,
not included in the check sample may be added. However,
8.1.14 sec-butylbenzene CAS 135-98-8,
m-xylene must be included.
8.1.15 Styrene CAS 100-42-5,
8.4.6 The equipment set-up check sample may be purchased
8.1.16 tert-butylbenzene CAS 98-06-6,
if available.
8.1.17 Toluene CAS 108-88-3, and
8.4.7 The purpose of the set-up check sample is to help
8.1.18 Trimethylbenzene CAS 108-67-8.
determine the retention time of the various components and
8.2 Carrier Gas, Makeup Gas and Detector Gases
that the p-xylene and m-xylene are adequately separated. This
99.999 % Pure. Oxygen in carrier gas less than 1 ppm, less
sample should not be used for calibration.
than 0.5 ppm is preferred. Purify carrier, makeup and detector
NOTE 1—The toluene, cumene, 1,4-dioxane, alpha methylstyrene,
gases to remove oxygen, water, and hydrocarbons.
phenylacetylene, o-xylene, and styrene are included to aid in peak
8.3 Air for the FID should contain less than 0.1 ppm total identification.
NOTE 2—The m-xylene is included to determine that adequate resolu-
hydrocarbon.
tion of m-xylene from p-xylene occurs.
8.4 Equipment Set-up Check Sample
NOTE 3—Benzene is included to determine that the equipment set-up
8.4.1 High Purity p-xylene (99.9999 mass % or greater has the sensitivity specified in this standard.
purity)—Most p-xylene is available commercially at a purity
less than 99.9 mass %, but should be purified by recrystalliza- 9. Hazards
tion. To prepare 2 L of high-purity p-xylene, begin with
9.1 Consult current OSHA regulations, supplier’s Safety
approximately 6 L of reagent-grade p-xylene and cool in an
Data Sheets, and local regulations for all materials used in this
explosion-proof freezer at between –10 °C to +10 °C until
test method.
1 3
approximately ⁄2 to ⁄4 of the p-xylene has frozen. Remove the
sample and decant the liquid portion. Allow the p-xylene to
10. Sampling
thaw and repeat the crystallization step on the remaining
10.1 Sample the material in accordance with Practice
sample until the p-xylene is free of contamination (no peaks
D3437.
detected other than p-xylene) as indicated by gas chromatog-
raphy.
11. Preparation of Apparatus
8.4.2 Prepare peak identification material by pipetting the
following amounts into a 250 mL flask:
11.1 Follow manufacturer’s instructions for mounting and
100 mL m-xylene,
conditioning the column into the chromatograph and adjusting
10 mL alpha methylstyrene,
the instrument to the conditions described in Table 1, allowing
10 mL toluene,
sufficient time for the equipment to reach equilibrium. See
10 mL cumene,
Practices E260, E355, and E1510 for additional information on
10 mL 1,4-dioxane,
gas chromatography practices and terminology.
10 mL ethylbenzene,
10 mL phenylacetylene,
12. Calibration
10 mL styrene,
12.1 Prior to implementation of the ECN method, a labora-
10 mL o-xylene,
tory should demonstrate that the equipment is set up properly
Mix thoroughly.
using an equipment set-up check sample. This sample should
8.4.3 Prepare sensitivity check material by:
be used to determine retention times of each component, and
8.4.3.1 Partially fill a 500 mL volumetric flask with high
that the separation of m-xylene from p-xylene is satisfactory.
purity p-xylene.
See 6.1 for the definition of an adequate separation.
8.4.3.2 Add 1 mL of benzene to the 500 mL volumetric
flask. 12.2 The LOD for this standard is 0.0002 mass %. The
equipment set-up check sample contains 0.0004 mass %
8.4.3.3 Fill the 500 mL volumetric flask to the mark with
high purity p-xylene and mix thoroughly. benzene. Acceptable results are 0.0001 mass % to 0.0008 mass
%.
8.4.4 Preparation of the final equipment set-up check
sample:
NOTE 4—Benzene was chosen because this is the trace impurity that
8.4.4.1 Partially fill a 500 mL volumetric flask with high
typically is the lowest level impurity and is of significant environmental
purity p-xylene. concern.
D7504 − 23
TABLE 2 Effective Carbon Number Correction Factors and
13. Procedure
Density
13.1 Warning—It is important to check the column reso-
ECN
lution. When different types of samples, particularly low purity
Component Correction Density at 20 °C
A
Factor
samples are injected, the front of the column will become
B
Non-Aromatics 1.0000 0.7255 (average)
contaminated. Eventually the column resolution will be re-
C
Benzene 0.9095 0.8780
duced to the point that the integration system will not separate
C
Toluene 0.9195 0.8658
D
some peaks, particularly m-xylene from p-xylene. 1,4-dioxane 3.0774 1.0329
C
Ethylbenzene 0.9271 0.8658
NOTE 5—Contamination on the front end of the column can be
C
p-xylene 0.9271 0.8597
minimized by baking out the column on a periodic basis or eliminated by
C
m-xylene 0.9271 0.8630
cutting off a few inches from the front end.
C
Cumene 0.9329 0.8605
C
NOTE 6—It is helpful to dedicate GCs for each high purity product and
o-xylene 0.9271 0.8786
H
others GCs for samples with significant impurities.
n-propylbenzene 0.9329 0.8620
E
C Aromatics 0.9329 0.8715 average
I
13.2 Bring the sample to room temperature.
tert-butylbenzene 0.9376 0.8669
J
sec-butylbenzene 0.9376 0.8580
C
13.3 Analyze the equipment set-up check sample as needed
Styrene 0.9210 0.9048
K
to ensure adequate resolution of m-xylene from p-xylene, that Ethyltoluene 0.9329 0.861
E
C Aromatics 0.9376 0.8694 average
the gas chromatograph has the sensitivity specified by this
E
p-diethylbenzene 0.9376 0.8620
standard, and that all the peaks are properly identified.
(PDEB)
F
Phenylacetylene 0.8296 0.9300
13.4 Inject an appropriate amount of sample into the instru- F
Alpha-methylstyrene 0.9276 0.9077
(AMS)
ment.
L
Trimethylbenzene 0.9329 0.8758
2-propenylbenzene 0.9276 0.893
13.5 Review the chromatographic data system result. Mea-
G
Unknown 0.931
sure the area of all peaks. The non-aromatics fraction includes
A
Correction factors are relative to n-heptane.
all peaks up to ethylbenzene except for the peaks assigned to
B
DS # 4A Physical Constants of Hydrocarbons C through C , ASTM, 1971.
1 10
benzene and toluene. Sum together all the non-aromatic peaks
Average of hexane, methylcyclopentane, methylcyclohexane, heptane, and
as a total area. When either benzene or toluene is analyzed and ethylcyclopentane.
C
Test Method D1555M.
1,4-dioxane is required to be reported, the non-aromatic
D
Keith, L. H., Walters, D. B., Compendium of Safety Data Sheets for Research
fraction does not in
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: D7504 − 21 D7504 − 23
Standard Test Method for
Trace Impurities in Monocyclic Aromatic Hydrocarbons by
Gas Chromatography and Effective Carbon Number
This standard is issued under the fixed designation D7504; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope*
1.1 This test method covers the determination of total nonaromatic hydrocarbons and monocyclic aromatic hydrocarbons in
benzene, toluene, ethylbenzene, p-xylene, o-xylene, styrene and mixed xylenes by gas chromatography. The purity of benzene,
toluene, ethylbenzene, p-xylene, o-xylene, styrene and mixed xylenes is also calculated. Similar test methods, using the internal
standard calibration technique and the external standard calibration technique, are Test Methods D2360, D3797, D4492, D5060,
D5135, D5917, andand D5917 D6563respectively.
1.2 The limit of detection (LOD) is 0.0002 mass % and limit of quantitation (LOQ) is 0.0006 mass % for impurities in toluene,
mixed xylenes, p-xylene, o-xylene, ethylbenzene, benzene, and styrene.
1.3 In determining the conformance of the test results using this method to applicable specifications, results shall be rounded off
in accordance with the rounding-off method of Practice E29.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of
regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2.1 ASTM Standards:
D1555M Test Method for Calculation of Volume and Weight of Industrial Aromatic Hydrocarbons and Cyclohexane [Metric]
D2360 Test Method for Trace Impurities in Monocyclic Aromatic Hydrocarbons by Gas Chromatography (Withdrawn 2016)
D3437 Practice for Sampling and Handling Liquid Cyclic Products
D3797 Test Method for Analysis of o-Xylene by Gas Chromatography (Withdrawn 2014)
D4492 Test Method for Analysis of Benzene by Gas Chromatography (Withdrawn 2018)
This test method is under the jurisdiction of ASTM Committee D16 on Aromatic, Industrial, Specialty and Related Chemicals and is the direct responsibility of
Subcommittee D16.01 on Benzene, Toluene, Xylenes, Cyclohexane and Their Derivatives.
Current edition approved July 1, 2021May 1, 2023. Published August 2021May 2023. Originally approved in 2009. Last previous edition approved in 20202021 as
D7504 – 20.D7504 – 21. DOI: 10.1520/D7504-21.10.1520/D7504-23.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
The last approved version of this historical standard is referenced on www.astm.org.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D7504 − 23
D4790 Terminology of Aromatic Hydrocarbons and Related Chemicals
D5060 Test Method for Determining Impurities in High-Purity Ethylbenzene by Gas Chromatography
D5135 Test Method for Analysis of Styrene by Capillary Gas Chromatography
D5136 Specification for High Purity p-Xylene
D5211 Specification for Xylenes for p-Xylene Feedstock
D5917 Test Method for Trace Impurities in Monocyclic Aromatic Hydrocarbons by Gas Chromatography and External
Calibration
D6563 Test Method for Benzene, Toluene, Xylene (BTX) Concentrates Analysis by Gas Chromatography (Withdrawn 2018)
D6809 Guide for Quality Control and Quality Assurance Procedures for Aromatic Hydrocarbons and Related Materials
E29 Practice for Using Significant Digits in Test Data to Determine Conformance with Specifications
E260 Practice for Packed Column Gas Chromatography
E355 Practice for Gas Chromatography Terms and Relationships
E691 Practice for Conducting an Interlaboratory Study to Determine the Precision of a Test Method
E1510 Practice for Installing Fused Silica Open Tubular Capillary Columns in Gas Chromatographs
2.2 Other Document:
OSHA Regulations, 29 CFR paragraphs 1910.1000 and 1910.1200
3. Terminology
3.1 See Terminology D4790 for definitions of terms used in this test method.
4. Summary of Test Method
4.1 The specimen to be analyzed is injected into a gas chromatograph equipped with a flame ionization detector (FID) and a
capillary column. The peak area of each component is measured and adjusted using effective carbon number (ECN) correction
factors. The concentration of each component is calculated based on its relative percentages of total adjusted peak area and
normalized to 100.0000 %.
5. Significance and Use
5.1 Determining the type and amount of hydrocarbon impurities remaining from the manufacture of toluene, mixed xylenes,
p-xylene, o-xylene, ethylbenzene, benzene, and styrene used as chemical intermediates and solvents is often required. This test
method is suitable for setting specifications and for use as an internal quality control tool where these products are produced or
are used. Typical impurities are: alkanes containing 1 to 10 carbons atoms, benzene, toluene, ethylbenzene (EB), xylenes, and
aromatic hydrocarbons containing nine carbon atoms or more.
5.2 This method may not detect all components and there may be unknown components that would be assigned inappropriate
correction factors and thus, the results may not be absolute.
6. Interferences
6.1 The complete separation of p-xylene from ethylbenzene, or ethylbenzene and m-xylene from p-xylene can be difficult when
either ethylbenzene or p-xylene is analyzed, respectively. The separation can be considered adequate if the distance from the
baseline to the valley between the two peaks is not greater than 50 % of the peak height of lower of the two peaks.
7. Apparatus
7.1 Chromatographic data system is required.
7.2 Columns—The choice of column is based on resolution requirements. Any column may be used that is capable of resolving
all significant impurities from the major component. The column and conditions described in Table 1 have been used successfully
and shall be used as a referee in cases of dispute. A 60 m column with an internal diameter of 0.32 mm and a film thickness of
0.50 μm crosslinked polyethylene glycol may improve resolution.
Available from U.S. Government Printing Office Superintendent of Documents, 732 N. Capitol St., NW, Mail Stop: SDE, Washington, DC 20401, http://
www.access.gpo.gov.
Scanlon, J. T. and Willis, D. E., “Calculation of Flame Ionization Detector Relative Response Factors Using the Effective Carbon Number Concept,” Journal of
Chromatographic Science, Vol. 23, August 1985, pp. 333–339.
D7504 − 23
TABLE 1 Recommended Method Parameters
Inlet Split
Temperature, °C 270
Column:
Tubing fused silica
Length, m 60
Internal diameter, mm 0.32
Stationary phase crosslinked polyethylene
glycol
Film thickness, μm 0.25
Column temperature program
Initial temperature, °C 60
Initial time, min 10
Programming rate, °C/min 5
Final, °C 150
Time 2, min 10
Carrier gas helium or hydrogen
Linear velocity, cm/s at 145°C 20 helium or 45 hydrogen
Linear velocity, cm/s at 145 °C 20 helium or 45 hydrogen
Split ratio 100:1
Sample size, μL 0.6
Detector: flame ionization
Temperature, °C 300
Analysis time, min 38
7.3 Gas Chromatograph—Any instrument having a flame ionization detector and a splitter injector suitable for use with a fused
silica capillary column may be used, provided the system has sufficient sensitivity, linearity, and range to determine 0.0001 mass %,
while not exceeding the full scale of either the detector or the electronic integration for the major component. It shall have a split
injection system that will not discriminate over the boiling range of the samples analyzed. The system should be capable of
operating at conditions given in Table 1.
7.4 Injector—The specimen must be precisely and repeatably injected into the gas chromatograph. An automatic sample injection
devise is highly recommended.
7.5 Syringe—Chromatographic, capable of delivering appropriate μL volumes.
8. Reagents and Materials
8.1 Purity of Reagent—Reagent grade chemicals shall be used in all tests. Unless otherwise indicated, it is intended that all
reagents shall conform to the specifications of the Committee on Analytical Reagents of the American Chemical Society, where
such specifications are available. Reagents with an establish purity greater than ACS reagent grade may be used.
8.1.1 Alpha-methylstyrene (AMS) CAS 98-83-9,
8.1.2 Benzene CAS 71-43-2,
8.1.3 Cumene (isopropylbenzene) CAS 98-82-8,
8.1.4 1,4-dioxane CAS 123-91-1,
8.1.5 Ethylbenzene CAS 100-41-4,
8.1.6 Ethyltoluene CAS 622-96-8,
8.1.7 m-xylene CAS 108-38-3,
8.1.8 n-propylbenzene CAS 103-65-1,
ACS Reagent Chemicals, Specifications and Procedures for Reagents and Standard-Grade Reference Materials, American Chemical Society, Washington, DC. For
suggestions on the testing of reagents not listed by the American Chemical Society, see Analar Standards for Laboratory Chemicals, BDH Ltd., Poole, Dorset, U.K., and
the United States Pharmacopeia and National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville, MD.
D7504 − 23
8.1.9 o-xylene CAS 95-47-6,
8.1.10 p-diethylbenzene (PDEB) CAS 105-05-5,
8.1.11 p-xylene CAS 106-42-3,
8.1.12 Phenylacetylene (ethynylbenzene) CAS 536-74-3,
8.1.13 2-propenylbenzene CAS 300-57-2,
8.1.14 sec-butylbenzene CAS 135-98-8,
8.1.15 Styrene CAS 100-42-5,
8.1.16 tert-butylbenzene CAS 98-06-6,
8.1.17 Toluene CAS 108-88-3, and
8.1.18 Trimethylbenzene CAS 108-67-8.
8.2 Carrier Gas, Makeup Gas and Detector Gases 99.999 % Pure. Oxygen in carrier gas less than 1 ppm, less than 0.5 ppm is
preferred. Purify carrier, makeup and detector gases to remove oxygen, water, and hydrocarbons.
8.3 Air for the FID should contain less than 0.1 ppm total hydrocarbon.
8.4 Equipment Set-up Check Sample
8.4.1 High Purity p-xylene (99.9999 mass % or greater purity)—Most p-xylene is available commercially at a purity less than 99.9
mass %, but should be purified by recrystallization. To prepare 2 L of high-purity p-xylene, begin with approximately 6 L of
1 3
reagent-grade p-xylene and cool in an explosion-proof freezer at between –10 °C to +10 °C until approximately ⁄2 to ⁄4 of the
p-xylene has frozen. Remove the sample and decant the liquid portion. Allow the p-xylene to thaw and repeat the crystallization
step on the remaining sample until the p-xylene is free of contamination (no peaks detected other than p-xylene) as indicated by
gas chromatography.
8.4.2 Prepare peak identification material by pipetting the following amounts into a 250 mL flask:
100 mL m-xylene,
10 mL alpha methylstyrene,
10 mL benzene,toluene,
10 mL cumene,
10 mL 1,4-dioxane,
10 mL ethylbenzene,
10 mL phenylacetylene,
10 mL styrene,
10 mL o-xylene,
Mix thoroughly.
8.4.3 Prepare sensitivity check material by:
8.4.3.1 Partially fill a 500 mL volumetric flask with high purity p-xylene.
8.4.3.2 Add 1 mL of toluenebenzene to the 500 mL volumetric flask.
8.4.3.3 Fill the 500 mL volumetric flask to the mark with high purity p-xylene and mix thoroughly.
8.4.4 Preparation of the final equipment set-up check sample:
D7504 − 23
8.4.4.1 Partially fill a 500 mL volumetric flask with high purity p-xylene.
8.4.4.2 Pipette 1 mL of the peak identification material prepared in 8.4.2 into the 500 mL volumetric flask.
8.4.4.3 Use a 10 mL burette to add 1.1 mL of the sensitivity check material prepared in 8.4.3 to the 500 mL volumetric flask.
8.4.4.4 Dilute to volume with high purity p-xylene and mix thoroughly.
8.4.5 Impurities that are not present in the samples being analyzed may be omitted from the check sample. Impurities not included
in the check sample may be added. However, m-xylene and toluene must be included.
8.4.6 The equipment set-up check sample may be purchased if available.
8.4.7 The purpose of the set-up check sample is to help determine the retention time of the various components and that the
p-xylene and m-xylene are adequately separated. This sample should not be used for calibration.
NOTE 1—The benzene,toluene, cumene, 1,4-dioxane, alpha methylstyrene, phenylacetylene, o-xylene, and styrene are included to aid in peak
identification.
NOTE 2—The amount of benzene was increased to help ensure the retention time of benzene is properly determined. There are a number of non-aromatics
with retention times near benzene.
NOTE 2—The m-xylene is included to determine that adequate resolution of m-xylene from p-xylene occurs.
NOTE 3—TolueneBenzene is included to determine that the equipment set-up has the sensitivity specified in this standard.
9. Hazards
9.1 Consult current OSHA regulations, supplier’s Safety Data Sheets, and local regulations for all materials used in this test
method.
10. Sampling
10.1 Sample the material in accordance with Practice D3437.
11. Preparation of Apparatus
11.1 Follow manufacturer’s instructions for mounting and conditioning the column into the chromatograph and adjusting the
instrument to the conditions described in Table 1, allowing sufficient time for the equipment to reach equilibrium. See Practices
E260, E355, and E1510 for additional information on gas chromatography practices and terminology.
12. Calibration
12.1 Prior to implementation of the ECN method, a laboratory should demonstrate that the equipment is set up properly using an
equipment set-up check sample. This sample should be used to determine:determine retention times of each component, and that
the separation of m-xylene from p-xylene is satisfactory. See 6.1 for the definition of an adequate separation.
12.2 The LOD for this standard is 0.0002 mass %. The equipment set-up check sample contains 0.0004 mass % toluene.benzene.
Acceptable results are 0.0001 mass % to 0.0008 mass %.
NOTE 4—TolueneBenzene was chosen because there are no other peaks near the toluene peak.this is the trace impurity that typically is the lowest level
impurity and is of significant environmental concern.
13. Procedure
13.1 Warning—It is important to check the column resolution. When different types of samples, particularly low purity samples
are injected, the front of the column will become contaminated. Eventually the column resolution will be reduced to the point that
the integration system will not separate some peaks, particularly m-xylene from p-xylene.
D7504 − 23
NOTE 5—Contamination on the front end of the column can be minimized by baking out the column on a periodic basis or eliminated by cutting off a
few inches from the front end.
NOTE 6—It is helpful to dedicate GCs for each high purity product and others GCs for samples with significant impurities.
13.2 Bring the sample to room temperature.
13.3 Analyze the equipment set-up check sample as needed to ensure adequate resolution of m-xylene from p-xylene, that the gas
chromatograph has the sensitivity specified by this standard, and that all the peaks are properly identified.
13.4 Inject an appropriate amount of sample into the instrument.
13.5 Review the chromatographic data system result. Measure the area of all peaks. The non-aromatics fraction includes all peaks
TABLE 2 Effective Carbon Number Correction Factors and
Density
ECN
Component Correction Density at 20°C20 °C
A
Factor
B
Non-Aromatics 1.0000 0.7255 (average)
C
Benzene 0.9095 0.8780
C
Toluene 0.9195 0.8658
D
1,4-dioxane 3.0774 1.0329
C
Ethylbenzene 0.9271 0.8658
C
p-xylene 0.9271 0.8597
C
m-xylene 0.9271 0.8630
C
Cumene 0.9329 0.8605
C
o-xylene 0.9271 0.8786
H
n-propylbenzene 0.9329 0.8620
E
C Aromatics 0.9329 0.8715 average
tert-butylbenzene 0.9376 0.867
I
tert-butylbenzene 0.9376 0.8669
sec-butylbenzene 0.9376 0.863
J
sec-butylbenzene 0.9376 0.8580
C
Styrene 0.9210 0.9048
K
Ethyltoluene 0.9329 0.861
E
C Aromatics 0.9376 0.8694 average
E
p-diethylbenzene 0.9376 0.8620
(PDEB)
F
Phenylacetylene 0.8296 0.9300
F
Alpha-methylstyrene 0.9276 0.9077
(AMS)
Trimethylbenzene 0.9329 0.8637
L
Trimethylbenzene 0.9329 0.8758
2-propenylbenzene 0.9276 0.893
G
Unknown 0.931
A
Correction factors are relative to n-heptane.
B
DS # 4A Physical Constants of Hydrocarbons C through C , ASTM, 1971.
1 10
Average of hexane, methylcyclopentane, methylcyclohexane, heptane, and
ethylcyclopentane.
C
Test Method D1555M.
D
Keith, L. H., Walters, DD. B., Compendium of Safety Data Sheets for Res
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...