Abstract

This document provides requirements and guidelines for selecting the measuring device, measuring geometry and temperature control unit for tests involving rotational and oscillatory rheometry. The general principles of test performance are described and example evaluations of rotational and oscillatory rheometry are provided.

Status
Published
Public Enquiry End Date
01-Dec-2025
Publication Date
16-Nov-2026
Technical Committee
IBLP - Paints and varnishes
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
05-Nov-2026
Due Date
10-Jan-2027
Completion Date
17-Nov-2026

Buy Documents

Standard

SIST EN ISO 3219-3:2026

English language (45 pages)
Preview
Preview
e-Library read for
×1 day

SIST EN ISO 3219-3:2026 is the Slovenian standard for rheology published by SIST for rotational and oscillatory rheometry. It gives guidance for choosing the measuring device, measuring geometry and temperature control unit, and it explains how to perform basic tests and evaluate the results. The document is written for laboratories, engineers, quality staff and buyers who need comparable rheological data for materials such as coatings, polymers and other viscoelastic products.

What does SIST EN ISO 3219-3:2026 specify?

SIST EN ISO 3219-3:2026 specifies test procedure guidance and example evaluations for rotational and oscillatory rheometry. It focuses on how to select the measuring set-up, how to run the measurement, and how to interpret common rheological test results.

The document is organized into Clauses 1 to 9 and informative Annex A. Clause 1 gives the scope, Clause 2 lists normative references, Clause 3 gives terms and definitions, Clauses 4 to 6 cover selection, preconditions and measurement performance, Clauses 7 and 8 describe basic and combined tests, and Clause 9 sets out the test report content.

What are the key requirements of SIST EN ISO 3219-3:2026?

SIST EN ISO 3219-3:2026 is mainly about choosing the right rheometer setup, keeping the test conditions controlled, and using measurement profiles that match the material and the question being asked. The practical value is that users can compare results more reliably and document the same rheological property in a consistent way.

Selecting the instrument, geometry and temperature control

Clause 4 links the measuring device to the type of test. A rotational viscometer is used for viscosity measurements, while a rotational rheometer can also perform basic rotational and oscillatory tests and determine viscoelastic properties such as shear storage modulus and shear loss modulus. In practice, this means the user must choose the instrument based on whether the task is simple viscosity work or broader rheometry.

Clause 4 also gives guidance for measuring geometry. It ties the choice of coaxial cylinder, cone-plate, plate-plate, double gap or relative measuring geometries to sample properties such as viscosity, inhomogeneity, surface tension and sample volume. For absolute geometries, heterogeneous components should not exceed 10 % of the gap width, so users have a clear limit when dealing with particles, bubbles or other structure.

Temperature control is also part of the selection process. The document points out that the displayed temperature can differ from the actual sample temperature, so users need a suitable temperature control system and, where needed, a hood to reduce gradients. An inert gas atmosphere can be used when air would influence the sample through oxidation or moisture effects.

Preparing the sample and the test environment

Clause 5 requires ambient conditions to meet IEC 61010, and it recommends a climate-controlled room at 23 ± 2 °C and 50 ± 5 % relative humidity. This matters because the rheometer environment can influence repeatability and instrument safety.

Clause 5 also stresses sample preparation, filling and cleaning. The sample should be introduced with the smallest possible shear load, filled without air bubbles and trimmed properly in plate-plate and cone-plate geometries. After the test, the geometry must be cleaned completely and checked before the next measurement, because residue can affect the next result.

Defining the measuring profile

Clause 6 says the measuring system shall be adjusted, calibrated and verified before use. It also requires the measuring conditions and boundary conditions to be documented. In practice, this is the basis for reproducibility and for being able to explain later why two results differ.

The measuring profile may use constant values, continuous ramps or stepped ramps. The document notes that the adjustment time of each measuring point has to be considered, and it gives a practical rule for low-speed rotational tests: below 1 s⁻1, the measuring point duration should be at least the inverse of the shear rate. For oscillatory tests, at least one oscillation period should be used as the integration time.

Using the basic rotational tests

Clause 7 describes time-dependent tests in rotation, constant shear stress tests, constant shear strain tests, temperature-dependent tests and flow or viscosity curves. These are the core tools for describing viscosity, creep, relaxation and yield behavior.

The constant shear-rate test can be used to track viscosity over time, for example to follow gel time, pot life or changes caused by phase separation, shear heating or degradation. The constant shear-stress test gives creep curves, from which users can derive yield point information, zero shear viscosity and shear compliance. The constant shear-strain test gives relaxation curves, which can be used to study shear stress relaxation and, for non-crosslinked polymers, to infer average molar mass from the curve shape.

Using oscillatory tests and combined tests

Clause 7 also covers oscillatory tests. Time-dependent oscillatory tests are typically done within the linear-viscoelastic range (LVR), and they are used for gel point, gel time and stability-related observations. Temperature-dependent oscillatory tests are used for glass transition temperature, crystallization temperature, melting temperature and crosslinking-related behavior.

Clause 8 combines basic tests to study structural change. Flow curves with hysteresis area show how the structure changes under increasing and decreasing shear, but the document warns not to use this method to assess thixotropy. Step tests and creep-recovery tests are used to evaluate structure recovery, thixotropy-related behavior and recovery after load. Frequency sweeps and time/temperature shift are used to build master curves and to extend the usable frequency range.

Recording the result

Clause 9 requires the test report to identify the sample, the measuring device, the geometry, the temperature control unit, the method, the evaluation type, the result, any deviations, unusual observations, the operator and the test date. For users, this means the report is not just a number - it is the full record needed to understand how the value was obtained.

What terms does SIST EN ISO 3219-3:2026 define?

SIST EN ISO 3219-3:2026 defines the key rheology terms used in the test procedure and result evaluation.

  • Crystallization temperature (T) - the temperature where crystallization begins and partial order forms as the melt solidifies.
  • Gel point - the oscillatory transition from sol to gel in the linear-viscoelastic range, where the loss angle δ is frequency-independent.
  • Gel time / pot life - the period during which a crosslinking material remains processable while a defined criterion is still met.
  • Glass transition temperature (Tg) - the approximate midpoint of the temperature range where the glass transition takes place.
  • Shear stress relaxation modulus (G) - the ratio of shear stress to shear strain during a relaxation test.
  • Sol/gel transition - the transition range from viscoelastic liquid to viscoelastic solid.
  • Trimming - removal of excess sample volume, especially relevant for plate-plate and cone-plate geometries.

Who uses SIST EN ISO 3219-3:2026?

SIST EN ISO 3219-3:2026 is used by people who measure and interpret rheological properties in laboratories and technical departments. Typical users include rheology labs, quality control teams, product development teams and suppliers or buyers who need comparable viscosity and viscoelastic data.

The document is relevant to paints and varnishes, plastics, coating materials, lubricating greases, cosmetics, powder coatings, hotmelt adhesives and asphalt binders, because its examples and test cases cover these kinds of materials. It is especially useful when a user needs to document flow behavior, gelation, recovery, processing window or temperature response.

Which standards are used with SIST EN ISO 3219-3:2026?

SIST EN ISO 3219-3:2026 uses ISO 3219-1:2021 for vocabulary and symbols, so readers use the same terms and notation across rotational and oscillatory rheometry. It also refers to IEC 61010 for safety requirements for electrical equipment used in measurement, control and laboratory work.

The document also points readers to related standards and reports for adjacent topics: ISO 3219-2 for general principles of rotational and oscillatory rheometry, ISO 6721-10 for complex shear viscosity by oscillatory rheometer, ISO/TR 20659-1 for yield point determination, and ISO/TR 20659-2 for time-dependent structural change.

What does the SIST EN ISO 3219-3:2026 document contain?

SIST EN ISO 3219-3:2026 contains selection tables, measurement profiles, example curves and evaluation routes for both rotational and oscillatory rheometry. The document uses these to show how to choose a geometry, set a test, and read the result for properties such as viscosity, yield point, gel point, glass transition temperature and recovery.

It includes tables that connect sample properties and applications to measuring geometries and measuring methods. It also includes figures for constant values, continuous and stepped ramps, flow curves, creep and recovery, amplitude sweeps, frequency sweeps, cycle tests and master-curve construction.

Annex A gives calculation examples for shear rate and shear stress in practical situations such as brush coating, pumping, spray application, sedimentation, pipe flow and sagging. That makes the document useful not only for test execution, but also for linking laboratory rheology to process conditions.

Buy Documents

Standard

SIST EN ISO 3219-3:2026

English language (45 pages)
Preview
Preview
e-Library read for
×1 day

Get Certified

Connect with accredited certification bodies for this standard

DIN CERTCO

DIN Group product certification.

DAKKS Germany Verified

Smithers Quality Assessments

US management systems and product certification.

ANAB United States Verified

Sponsored listings

Frequently Asked Questions

SIST EN ISO 3219-3:2026 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Rheology - Part 3: Test procedure and examples for the evaluation of results when using rotational and oscillatory rheometry (ISO 3219-3:2026)". This standard covers: This document provides requirements and guidelines for selecting the measuring device, measuring geometry and temperature control unit for tests involving rotational and oscillatory rheometry. The general principles of test performance are described and example evaluations of rotational and oscillatory rheometry are provided.

This document provides requirements and guidelines for selecting the measuring device, measuring geometry and temperature control unit for tests involving rotational and oscillatory rheometry. The general principles of test performance are described and example evaluations of rotational and oscillatory rheometry are provided.

SIST EN ISO 3219-3:2026 is classified under the following ICS (International Classification for Standards) categories: 83.080.01 - Plastics in general. The ICS classification helps identify the subject area and facilitates finding related standards.

SIST EN ISO 3219-3:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


SLOVENSKI STANDARD
01-december-2026
Reologija - 3. del: Preskusni postopki in primeri za vrednotenje rezultatov,
dobljenih z rotacijsko in oscilacijsko reometrijo (ISO 3219-3:2026)
Rheology - Part 3: Test procedure and examples for the evaluation of results when using
rotational and oscillatory rheometry (ISO 3219-3:2026)
Rheologie - Teil 3: Versuchsdurchführung und beispielhafte Auswertungen der
Rotations- und Oszillationsrheometrie (ISO 3219-3:2026)
Rhéologie - Partie 3: Mode opératoire d'essai et exemples d'évaluation des résultats en
cas d'utilisation de la rhéométrie rotative et oscillatoire (ISO 3219-3:2026)
Ta slovenski standard je istoveten z: EN ISO 3219-3:2026
ICS:
83.080.01 Polimerni materiali na Plastics in general
splošno
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EN ISO 3219-3
EUROPEAN STANDARD
NORME EUROPÉENNE
October 2026
EUROPÄISCHE NORM
ICS 83.080.01
English Version
Rheology - Part 3: Test procedure and examples for the
evaluation of results when using rotational and oscillatory
rheometry (ISO 3219-3:2026)
Rhéologie - Partie 3: Mode opératoire d'essai et Rheologie - Teil 3: Versuchsdurchführung und
exemples d'évaluation des résultats en cas d'utilisation beispielhafte Auswertungen der Rotations- und
de la rhéométrie rotative et oscillatoire (ISO 3219- Oszillationsrheometrie (ISO 3219-3:2026)
3:2026)
This European Standard was approved by CEN on 28 September 2026.

CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this
European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references
concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN
member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by
translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management
Centre has the same status as the official versions.

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and
United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION

EUROPÄISCHES KOMITEE FÜR NORMUNG

CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2026 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN ISO 3219-3:2026 E
worldwide for CEN national Members.

Contents Page
European foreword . 3

European foreword
This document (EN ISO 3219-3:2026) has been prepared by Technical Committee ISO/TC 35 "Paints
and varnishes" in collaboration with Technical Committee CEN/TC 139 “Paints and varnishes” the
secretariat of which is held by DIN.
This European Standard shall be given the status of a national standard, either by publication of an
identical text or by endorsement, at the latest by April 2027, and conflicting national standards shall be
withdrawn at the latest by April 2027.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CEN shall not be held responsible for identifying any or all such patent rights.
Any feedback and questions on this document should be directed to the users’ national standards
body/national committee. A complete listing of these bodies can be found on the CEN website.
According to the CEN-CENELEC Internal Regulations, the national standards organizations of the
following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria,
Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland,
Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Republic of
North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and the
United Kingdom.
Endorsement notice
The text of ISO 3219-3:2026 has been approved by CEN as EN ISO 3219-3:2026 without any
modification.
International
Standard
ISO 3219-3
First edition
Rheology —
2026-10
Part 3:
Test procedure and examples for
the evaluation of results when
using rotational and oscillatory
rheometry
Rhéologie —
Partie 3: Mode opératoire d'essai et exemples d'évaluation
des résultats en cas d'utilisation de la rhéométrie rotative et
oscillatoire
Reference number
ISO 3219-3:2026(en) © ISO 2026

ISO 3219-3:2026(en)
© ISO 2026
All rights reserved.
This ISO publication is protected by copyright and is owned by ISO and/or its licensors.
The content of this ISO publication is provided under licence, not sold. Use is subject to the applicable licence terms issued by ISO,
an ISO member body, or an authorized third-party distributor.
Except as required for implementation or expressly permitted by a separate licence, no part of this ISO publication may be
reproduced, distributed, modified, used, or made available in any form or by any means – electronic or mechanical, including
photocopying, scanning, recording, or posting on internal or external digital platforms.
Any use beyond the scope of the granted rights is prohibited and may result in legal action.
ISO copyright office
CP 401 • CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii
ISO 3219-3:2026(en)
Contents Page
Foreword .iv
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 Guidelines and requirements for selection . 2
4.1 General .2
4.2 Selection of the measuring device.2
4.3 Selection of the measuring geometry .3
4.4 Selection of the temperature control system .3
4.5 Selection of the measuring method .4
5 Preconditions for a measurement . . 7
5.1 Ambient conditions.7
5.2 Sample preparation, filling and cleaning .7
6 Performance of the measurement . 8
6.1 General .8
6.2 Measuring profile .8
7 Basic tests and example evaluations . 10
7.1 General .10
7.2 Rotational tests .10
7.2.1 Time-dependent tests in rotation .10
7.2.2 Temperature-dependent tests in rotation . 13
7.2.3 Flow curves and viscosity curves .14
7.3 Oscillatory tests .18
7.3.1 Time-dependent tests in oscillation .18
7.3.2 Temperature-dependent tests in oscillation . 20
7.3.3 Amplitude sweeps . 23
7.3.4 Frequency sweeps .24
8 Combined basic tests .27
8.1 General .27
8.2 Flow curves and hysteresis area .27
8.3 Temperature tests as cycle tests . 29
8.4 Master curve by means of time/temperature shift of frequency sweeps. 29
8.5 Step tests for determination of the time-dependent structural change .31
8.6 Creep and recovery tests .32
9 Test report .34
Annex A (informative) Calculation examples .36
Bibliography .38

iii
ISO 3219-3:2026(en)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee
has been established has the right to be represented on that committee. International organizations,
governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely
with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types
of ISO document should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent
rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a)
patent(s) which may be required to implement this document. However, implementers are cautioned that
this may not represent the latest information, which may be obtained from the patent database available at
www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO's adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 35, Paints and varnishes, Subcommittee
SC 9, General test methods for paints and varnishes, in cooperation with ISO/TC 61, Plastics, SC 5, Physical-
chemical properties, andin collaboration with the European Committee for Standardization (CEN) Technical
Committee CEN/TC 139, Paints and varnishes, in accordance with the Agreement on technical cooperation
between ISO and CEN (Vienna Agreement),
A list of all parts in the ISO 3219 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.

iv
International Standard ISO 3219-3:2026(en)
Rheology —
Part 3:
Test procedure and examples for the evaluation of results
when using rotational and oscillatory rheometry
1 Scope
This document provides requirements and guidelines for selecting the measuring device, measuring
geometry and temperature control unit for tests involving rotational and oscillatory rheometry. The
general principles of test performance are described and example evaluations of rotational and oscillatory
rheometry are provided.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes
requirements of this document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
ISO 3219-1:2021, Rheology — Part 1: Vocabulary and symbols for rotational and oscillatory rheometry
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 3219-1 and the following apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at https:// www .electropedia .org/
3.1
crystallization temperature
T
c
temperature at which crystallization starts and partial order is established when the melt solidifies
Note 1 to entry: The unit of the crystallization temperature, T , is degrees Celsius (°C).
c
3.2
gel point
transition point from sol to gel in oscillation in the linear-viscoelastic range, where the loss angle, δ, is
independent of frequency
Note 1 to entry: The gel point can be a time or a temperature.
Note 2 to entry: Contrary to this definition, the gel point is usually measured in practice at one frequency and
determined on the basis of the loss angle δ = 45° or on the basis of the loss factor tan δ = 1.

ISO 3219-3:2026(en)
3.3
gel time
pot life
period of time for which a crosslinking material remains processable while a user-specified criterion is
satisfied
Note 1 to entry: The gel time can be determined in oscillation in the nonlinear range or in rotation.
Note 2 to entry: In practice, relative measuring geometries or alternative methods are also used, such as flow cups.
Note 3 to entry: Limiting values for the viscosity or moduli can be used as user-defined criteria.
3.4
glass transition temperature
T
g
characteristic temperature of the temperature range in which the glass transformation takes place
Note 1 to entry: The glass transition temperature depends on the test procedure and the measuring conditions.
3.5
sample preparation
handling, conditioning and load applied to the sample prior to measurement
3.6
shear relaxation modulus
G
ratio of shear stress τ to shear strain γ during the relaxation test
Note 1 to entry: The unit for the shear relaxation modulus G is Pascal (Pa).
3.7
sol/gel transition
transition range from viscoelastic liquid (sol) to viscoelastic solid (gel)
3.8
trimming
removal of excess sample volume
Note 1 to entry: With plate-plate and cone-plate measuring geometries, a trimming position is often selected being
larger than the gap width and at which the excess volume is removed so that the sample is flushed to the edge of the
measuring geometry. When the measuring position is subsequently approached, this ensures optimum filling of the
gap.
4 Guidelines and requirements for selection
4.1 General
The choice of measuring device, measuring geometry, temperature control unit, measuring method and
further accessories depends on the properties of the sample and the resulting rheological requirements.
4.2 Selection of the measuring device
In ISO 3219-2, the differences between the different measuring devices for the determination of rheological
properties and their setups are described. A rotational viscometer can be used to determine the shear
viscosity. With a rotational rheometer, it is in addition possible to carry out all basic rotational and oscillatory
tests (see Clause 7). For this reason, in addition to the viscosity function, this can be used to determine the
viscoelastic properties, e.g. shear storage modulus and shear loss modulus.
For viscometric measurements, all viscometers are principally suitable, regardless of the type of bearing.
For oscillatory measurements, creep tests and relaxation tests, devices with low internal friction should
be used, e.g. devices with air bearings or magnetic bearings. Some devices with ball bearings also offer the

ISO 3219-3:2026(en)
option of performing oscillatory measurements, but only within a restricted measuring range. During the
measurement execution, it shall be ensured that the work is carried out within the specifications of the
measuring device, e.g. taking into account the torque limits and data acquisition rate.
To cover the broadest possible range of applications, the viscometer or rheometer shall be capable of working
with different measuring geometries. The ranges of torque and angular displacement, and therefore the
accessible measuring range, depend on the measuring device.
4.3 Selection of the measuring geometry
The measuring geometry shall be selected taking into account the available measuring device, the sample
properties and the measuring profile. Table 1 contains recommendations on selecting the different measuring
geometries. Here, CC stands for coaxial cylinder, CP for cone-plate, PP for plate-plate, DG for double gap
and RMG for relative measuring geometries (e.g. vane sensor, for details see ISO 3219-2). In general, when
selecting an absolute measuring geometry, the maximum size of heterogeneities, e.g. the particle size in the
sample, shall not exceed 10 % of the gap width. If, as a result, it is not possible to use absolute measuring
geometries, then it is necessary to select relative measuring geometries. Optional accessories, such as a
hood with a solvent trap to avoid solvent loss are described in ISO 3219-2.
Table 1 — Recommendations for selection of the measuring geometry
Criteria Recommended measuring geometry
Coaxial cylin- Cone-plate Plate-plate Double gap Relative
ders
CC CP PP DG RMG
Sample properties that determine the selection of the measuring geometry
a
Low viscosity x x x
High viscosity x x
Pasty, gel-like x
Low surface tension x x
Inhomogeneity, e.g. air bubbles,   x
foaming
Further criteria for the selection of a measuring geometry
Availability of a small sample vol- x x
ume
b
Cleaning of the measuring geometry x x x
difficult
b
Cleaning of the measuring geometry  x
not possible
c c c
Prevention of sedimentation/sepa- x x x
ration/phase separation
d d d d
Avoidance of wall slip x x x x
a
For homogeneous samples using small cone angles.
b
Use of disposable measuring geometries.
c
Surface-treated (e.g. helical spiral groove).
d
Surface-treated (e.g. sandblasting, profiling). The surface roughness, R , of the sandblasted geometry should be at least 10
a
times smaller than the measuring gap.
4.4 Selection of the temperature control system
Figure 1 illustrates the best-suited temperature control system for each measurement. It is important to
keep in mind that the temperature display can deviate from the true sample temperature. The temperature
gradient within the sample can be minimized by using an active or passive hood. Any chemical reactions
that take place with air (e.g. oxygen, humidity) can be avoided under an inert gas atmosphere.

ISO 3219-3:2026(en)
Figure 1 — Selection of the temperature control system
4.5 Selection of the measuring method
In Table 2, typical rheological requirements are listed with recommendations for meaningful measuring
methods. Here, CR stands for controlled rate, CS for controlled stress and CD for controlled deformation
(also called controlled strain, see ISO 3219-1:2021, 3.47 and Table 1). During execution of the measurement,
it shall be ensured that the work is carried out within the specifications of the measuring device used.
Table 2 — Selection of measuring methods for viscometers and rheometers based on the example of
coating materials
a a b
Shelf life/sedimentation/phase separation x x x x x x
Starting pumping (yield point) x x x
Sagging/levelling  x x
Pumping/mixing/stirring x x
Painting/brushing/blade application (manual) x x  x
a
Evaluation via zero shear viscosity or yield point (if applicable, adaptation via rheological model function).
b
For determination of the linear-viscoelastic range (LVR).
c
Only in oscillation.
Application
Flow curve and viscosity
curve CR
Flow curve and viscosity curve CS
Amplitude sweep CD or CS
Frequency sweep CD or CS
Creep test
Step test
Rotation - rotation - rotation or
Oscillation – oscillation – oscilla-
tion or
Oscillation - rotation - oscillation
Temperature sweep/time-de-
pendent test in rotation or oscil-
lation
ISO 3219-3:2026(en)
TTabablele 2 2 ((ccoonnttiinnueuedd))
Spraying x x  x
High-speed coating x
Gel time/open time/processing time   x
c
Gel point   x
c
Temperature behaviour (glass transition tem-   x
perature, crystallization temperature, melting
temperature, crosslinking temperature)
Hardening/crosslinking   x
a
Evaluation via zero shear viscosity or yield point (if applicable, adaptation via rheological model function).
b
For determination of the linear-viscoelastic range (LVR).
c
Only in oscillation.
The shear rate range plays an important role in rotational measurements. Representative shear rate ranges
are shown in Figure 2 for selected applications from Table 2. The shear rates may, for example, be calculated
as described in Annex A.
Figure 2 — Typical shear rate ranges for different applications based on the example of coating
materials
Typical rheological parameters are shown in Table 3 with reference to the corresponding subclause in this
document.
Application
Flow curve and viscosity
curve CR
Flow curve and viscosity curve CS
Amplitude sweep CD or CS
Frequency sweep CD or CS
Creep test
Step test
Rotation - rotation - rotation or
Oscillation – oscillation – oscilla-
tion or
Oscillation - rotation - oscillation
Temperature sweep/time-de-
pendent test in rotation or oscil-
lation
ISO 3219-3:2026(en)
Table 3 — Rheological parameters and references
Rheological parameter Description in subclauses
Flow behaviour 7.2.3 Flow curves and viscosity curves
7.2.1.2 Specification of a constant shear stress (evaluation via the creep curve)
7.2.3.1 Specification of a variable shear rate (evaluation via curve fitting; or via the
Yield point viscosity maximum method)
7.2.3.2 Specification of a variable shear stress (analogous to 7.2.3.1; as a plateau value
or by means of regression lines)
7.3.1 Time-dependent tests in oscillation
Gel point
7.3.2 Temperature-dependent tests in oscillation
7.2.1.1 Specification of a constant shear rate
Gel time
7.2.1.2 Specification of a constant shear stress
Processing time
7.2.2 Temperature-dependent tests in rotation
Processing temperature
7.3.1 Time-dependent tests in oscillation
Crosslinking temperature
7.3.2 Temperature-dependent tests in oscillation
Glass transition temperature 7.3.2 Temperature-dependent tests in oscillation
Hysteresis area 8.2 Flow curves and hysteresis area
Crystallization temperature 7.3.2 Temperature-dependent tests in oscillation
Linearity limit 7.3.3 Amplitude sweeps (end of the LVR)
LVR (linear-viscoelastic
7.3.3 Amplitude sweeps
range)
Master curve 8.4. Master curve by means of time/temperature shift of frequency sweeps
7.2.1.3 Specification of a constant shear strain (value of the initial shear relaxation
modulus)
7.3.4 Frequency sweeps (position of the crossover point of the shear storage modulus
Average molar mass
and shear loss modulus)
8.4 Master curve by means of time/temperature shift of frequency sweeps (position
of the crossover point of the shear storage modulus and shear loss modulus)
7.2.1.3 Specification of a constant shear strain (via the time-based curve of the shear
relaxation modulus)
7.3.4 Frequency sweeps (position of the crossover point of the shear storage modulus
Molar mass distribution
and shear loss modulus)
8.4. Master curve by means of time/temperature shift of frequency sweeps (position
of the crossover point of the shear storage modulus and shear loss modulus)
7.2.1.2 Specification of a constant shear stress (via the creep curve)
7.2.3.1 Specification of a variable shear rate (as a plateau value at low shear rates)
Zero shear viscosity
7.2.3.2 Specification of a variable shear stress (analogous to 7.2.3.1)
7.3.4 Frequency sweeps (plateau of the viscosity at low frequencies)
Recovery 8.6 Creep and recovery test
7.2.1.2 Specification of a constant shear stress (via creep test)
Shear compliance
8.6 Creep and recovery test (analogous to 7.2.1.2)
Melting temperature 7.3.2 Temperature-dependent tests in oscillation
7.3.1 Time-dependent tests in oscillation
Sol/gel transition
7.3.2 Temperature-dependent tests in oscillation
Structure recovery 8.5 Step tests for determination of the time-dependent structural change
Thixotropy 8.5 Step tests for determination of the time-dependent structural change

ISO 3219-3:2026(en)
5 Preconditions for a measurement
5.1 Ambient conditions
The ambient conditions under which the rheometer is operated shall meet the requirements of the
manufacturer of the apparatus.
It is recommended that the measuring device is placed in a climate-controlled room [temperature:
(23 ± 2) °C, relative humidity: (50 ± 5) %].
5.2 Sample preparation, filling and cleaning
The sample preparation can play a decisive role for the reproducibility of the measurement. In order to
obtain comparable results, it is necessary to coordinate all aspects of the sample preparation between the
interested parties, to adhere to these and to log any deviations (Clause 9).
Before the measurement, the measuring geometry shall be completely cleaned, with no residue left behind.
In the ideal scenario, the sample should be filled in such a way that the smallest possible shear load is applied
so that any damage to the structure is kept as small as possible. After the approach to the measuring gap, the
sample shall fill the measuring gap completely and shall be free of air bubbles.
With plate-plate and cone-plate measuring geometries, the sample is trimmed, which means that any excess
leaked from the measuring gap is removed using a suitable tool. For specific sample types, such as asphalt
binders or hotmelt adhesives, it can be necessary to adjust the temperature of the trimming tool, but to use
a certain temperature range for the trimming tool. For the removal of the excess sample, first a trimming
position above the measuring position is approached, the sample is trimmed, and then the measuring
position is approached, which produces a slightly curved sample edge. Experience shows that a gap is
chosen for the trimming position that is around 1 % to 5 % larger than the gap width. During trimming, it
is recommended to block the upper measuring geometry so that any structure present in the sample cannot
be damaged due to twisting of the geometry. The excess sample can be scraped off e.g. using a spatula (see
Figure 3). Use of a lower recess plate makes the trimming process easier. Figure 3 a) through to Figure 3 d)
show the complete process of filling a plate-plate measuring geometry, including trimming.

ISO 3219-3:2026(en)
a) Sample in the trimming position b) Scraping off the excess sample in the trimming
position, e.g. with a spatula
c) Trimmed sample in the trimming position d) Optimum sample filling in the measuring posi-
tion
Figure 3 — Filling a plate-plate measuring geometry
6 Performance of the measurement
6.1 General
Depending on the sample and the rheological issue being investigated, a suitable measuring system and
a meaningful measuring method shall be selected from the guidelines and requirements (Clause 4) and,
where applicable, with the aid of preliminary tests. The measuring system used shall be adjusted, calibrated
and verified. Prior to the start of the measurement, it can be necessary to allow a longer waiting time for
recovery of the sample structure or to carry out a specified pre-shear, independent of the temperature
control of the sample.
All measuring conditions and boundary conditions shall be reported (see Clause 9).
6.2 Measuring profile
Once a measuring method has been chosen, the measuring profile is set. It may contain various measuring
segments (see Clause 7), each of which is specified by the preset parameter(s) and by the duration and
number of measuring points.
The preset parameters of a measuring segment may be constant (see Figure 4) or may be varied over
time. The variation may be continuous [continuous ramp, see Figure 5 a)] or discrete [stepped ramp, see
Figure 5 b)].
ISO 3219-3:2026(en)
Key
X time, t

Y preset parameter: e.g. shear rate, γ , shear stress, τ, shear strain, γ, temperature, T
t end time to switch off the preset parameter
n
Figure 4 — Illustration of a constant preset parameter
a) Continuous ramp b) Stepped ramp
Key
X time, t

Y preset parameter: e.g. shear rate, γ , shear stress, τ, shear strain, γ, temperature, T
Figure 5 — Illustration of the continuous or discrete variation of the preset parameter
Detailed specifications of the measuring profile shall be provided to ensure reproducible measuring results
and comparable evaluations.
In the case of continuous ramps, deviations will increase with increasing rates of change of the specified
conditions (e.g. the shear load or temperature).
The adjustment time of each measuring point shall be considered.
−1
For rotational tests with shear rates below 1 s , a measuring point duration of at least the inverse shear rate
(i.e. 1/shear rate) should be specified. For oscillatory tests, at least one oscillation period should be used for
the integration time, i.e. the measuring time to calculate a data point.

ISO 3219-3:2026(en)
7 Basic tests and example evaluations
7.1 General
A basic test consists of a single measuring segment.
The basic tests in 7.2 and 7.3 are described in terms of the following aspects:
a) measuring profiles;
b) examples of measuring curves;
c) typical evaluations.
7.2 Rotational tests
7.2.1 Time-dependent tests in rotation
7.2.1.1 Specification of a constant shear rate
a) Measuring profiles

A constant, application-relevant value is specified for the shear rate, γ . All other values are also kept
constant (see Figure 4).
b) Examples of measuring curves
Figure 6 shows examples of measuring curves for time-dependent tests.

Key
X time, t
Y shear viscosity, η
1 result that is independent of time
2 initial decrease in the shear viscosity as a function of time (e.g. in the case of shear-induced disentangling of
polymer molecules or during the orientation of particles or if the sample is in a non-relaxed/non-equilibrium
state before the start of the measurement)
3 initial increase in the shear viscosity as a function of time (e.g. in the case of a shear-induced increase in
interactions or if the sample is in a non-relaxed/non-equilibrium state before the start of the measurement)
4 increase in the shear viscosity as a function of time (e.g. due to gelation, chemical crosslinking)
5 decrease in the shear viscosity as a function of time (e.g. due to phase separation, shear heating, degradation
of polymer molecules)
Figure 6 — Example of measuring curves for time-dependent tests
c) Typical evaluations
1) preliminary test to determine parameters for following tests (e.g. maximum test duration, limiting
values of the shear load);
ISO 3219-3:2026(en)
2) mean value of the viscosity (single-point measurement) for process control and quality assurance;
3) determination of the time during the viscosity increase when reaching a specified viscosity (as gel
time, maximum processing time);
4) determination of a ratio of two viscosity values, η and η , at the time points, t and t .
1 2 1 2
7.2.1.2 Specification of a constant shear stress
a) Measuring profiles
A constant, application-relevant value is specified for the shear stress, τ . All other values are also kept
constant (see Figure 4).
b) Examples of measuring curves
Figure 6 shows examples of viscosity measuring curves for time-dependent tests.
In addition, Figure 7 shows examples of measured time-dependent shear strain results.

Key
X time, t
Y shear strain, γ
1 ideal elastic behaviour
2 ideal viscous behaviour
3 viscoelastic behaviour
t end time to switch off the preset parameter
n
Figure 7 — Examples of measuring curves for creep tests
c) Typical evaluations
See 7.2.1.1
d) Additional evaluations:
1) yield point determination via creep curves (for details, see ISO/TR 20659-1);
2) maximum deformation value, γ , at the end of the load phase at a constant shear stress, τ , see
max
Figure 8 a);
ISO 3219-3:2026(en)
3) determination of the zero shear viscosity, η , with n creep curves (with n≥ 2 ), via the slope of the

shear strain, γ , after reaching a steady-state. This results in a shear rate, γ , according to
Formula (1) and Formula (2); see Figure 8 a):

  (1)
t

  (2)


4) shear compliance, J, (unit in 1/Pa) according to Formula (3):
 t

Jt  (3)


5) determination of the shear compliance, J, with n creep curves [with n≥ 2), see Figure 8 b)], after
reaching a constant slope. The creep curves shall be measured at different specified shear stress
values.
a) Time profile of shear strain b) Time profile of shear compliance
Key
X time, t
Y1 shear strain, γ
Y2 shear compliance, J
∆t time period after reaching a constant slope value
 section of the change in the shear strain after reaching a constant slope value
γ maximum value of the shear strain at the end of the creep curve
max
Figure 8 — Example of evaluations of creep tests
7.2.1.3 Specification of a constant shear strain
a) Measuring profiles
A constant, application-relevant value is specified over a specified measuring time for the shear strain,
γ . All other values are also kept constant.

ISO 3219-3:2026(en)
b) Examples of measuring curves
Figure 9 a) and b) shows examples of measuring results of relaxation tests.

a) time profile of the shear stress relaxation b) time profile of the shear relaxation modulus on
on a linear time scale a logarithmic time scale, exemplary for curve 3 in
Figure 9a)
Key
X time, t
Y1 shear stress, τ
Y2 lg shear relaxation modulus, G
G initial shear relaxation modulus
1 ideal elastic behaviour
2 ideal viscous behaviour
3 with delayed, complete shear relaxation of a viscoelastic fluid
4 with delayed, incomplete shear stress relaxation of a viscoelastic solid (the constant end value is referred to
as equilibrium shear stress)
Figure 9 — Relaxation tests
c) Typical evaluations
Calculation of the time-dependent function of the shear relaxation modulus, G, (unit in Pa) is according to
Formula (4):
 t

Gt  (4)


The molar mass distribution for non-crosslinked polymers can be derived via the shape of the G(t) curve.
The initial shear relaxation modulus, G , is the constant value of the G(t) function that can be displayed in
the short term range in a logarithmic scale; see Figure 9 b). The absolute value of the G value describes the
average molar mass.
7.2.2 Temperature-dependent tests in rotation
a) Measuring profiles
The temperature is varied between the start and the end value, either in a continuous ramp with
a specified heating or cooling rate, or alternatively in a stepped ramp with specified temperature
increments and thermal equilibrium time for each step (see Figure 5). All other values are kept constant.
It shall be stated in the test report, if shear stress or shear rate has been kept constant.

ISO 3219-3:2026(en)
In the case of temperature ramps, a maximum rate of temperature change of 1 K/min is recommended
to ensure that measurements are performed approximately at the temperature equilibrium.
b) Examples of measuring curves
Figure 10 shows examples of measuring curves for temperature-dependent tests.

Key
X temperature, T
Y shear viscosity, η
1 decrease in viscosity with increasing temperature (e.g. softening) or increase in viscosity with decreasing
temperature (e.g. solidification, crystallization)
2 increase in viscosity, possibly after an initial decrease, with increasing temperature (e.g. chemical crosslinking,
gelation, drying/evaporation)
Figure 10 — Examples of measuring curves for temperature-dependent tests
c) Typical evaluations
1) as a preliminary test for determining parameters for subsequent performance of the testing
(limiting value for the thermal load);
2) viscosity minimum (e.g. as criterion of levelling behaviour or as crosslinking temperature);
3) determination of the temperature until a multiple of the initial viscosity is reached (as the maximum
processing temperature);
4) curve fitting for rheological model functions for description of the temperature dependency of the
viscosity, e.g. according to Arrhenius.
7.2.3 Flow curves and viscosity curves
7.2.3.1 Specification of a variable shear rate
a) Measuring profiles
The value of the shear rate, γ , is changed either continuously or in steps (see Figure 5). All other values
are kept constant.
ISO 3219-3:2026(en)
b) Examples of measuring curves
Figure 11 shows examples of flow and viscosity curves without yield points (with yield point; see 7.2.3.2).
a) Flow curves b) Viscosity curves
Key

X shear rate, γ
Y1 shear stress, τ
Y2 shear viscosity, η
1 Newtonian flow behaviour (ideal viscous behaviour)
2 shear-thinning flow behaviour
3 shear-thickening flow behaviour
Figure 11 — Examples of flow and viscosity curves
c) Typical evaluations
Details for the evaluation of flow curves are described in ISO/TR 20659-1 and are presented in brief
below. The result of the evaluation is different depending on whether the shear rates were varied as
stepped or continuous valu
...