Abstract

This document specifies a method for the determination of boron (B), cobalt (Co), copper (Cu), iron (Fe), manganese (Mn), molybdenum (Mo), phosphorous (P) and zinc (Zn) in organic and organo-mineral fertilizers using inductively coupled plasma-atomic emission spectrometry (ICP-AES).
This method is applicable to extracts prepared according to EN 17766, EN 17767 and EN 17779.
This document is applicable to fertilizing product blends, where the blend is a mix of at least two of the following components: fertilizers, liming materials, soil improvers, growing media, inhibitors, plant biostimulants, and where the following category: organic fertilizers or organo-mineral fertilizers is the highest % in the blend by mass or volume, or in the case of liquid form by dry mass. If the organic fertilizer or the organo-mineral fertilizer is not the highest % in the blend, the European Standard for the highest % of the blend applies. In case a fertilizing product blend is composed of components in equal quantity, the user decides which standard to apply.
NOTE   Variations in analytical methods for fertilizing product blends can lead to differing results as some components or matrix interactions can affect the outcome. Validation procedures have shown that developed standard methods are robust and reliable across diverse product compositions, but possible interferences and unexpected results when analysing fertilizing product blends are possible.

Status
Not Published
Public Enquiry End Date
01-Apr-2026
Technical Committee
KAT - Soil quality
Current Stage
5020 - Formal vote (FV) (Adopted Project)
Start Date
17-Nov-2026
Due Date
05-Jan-2027

Buy Documents

Draft

oSIST prEN 17774:2026

English language (25 pages)
Preview
Preview
e-Library read for
×1 day

kSIST FprEN 17774:2026 is the Slovenian adoption of the CEN draft European Standard prEN 17774:2026 for organic and organo-mineral fertilizers. It specifies an inductively coupled plasma-atomic emission spectrometry (ICP-AES) method for determining boron, cobalt, copper, iron, manganese, molybdenum, phosphorous and zinc in fertilizer extracts. The document is used where these elements must be measured in organic fertilizers, organo-mineral fertilizers, and certain fertilizing product blends.

What does kSIST FprEN 17774:2026 specify?

kSIST FprEN 17774:2026 specifies the analytical method, the sample preparation link to related extraction standards, and the way results are calculated and reported. It is structured into Clauses 1 to 10, with Annex A giving supporting interlaboratory data.

The document applies to extracts prepared according to prEN 17766, prEN 17767 and prEN 17779. It also explains how to handle fertilizing product blends, where the relevant standard depends on which component category is present at the highest percentage by mass, volume, or dry mass for liquid forms.

AnnexWhat it covers
Annex AInformative results of the interlaboratory study, including the sample set and statistical evaluation for the specific elements

What are the key requirements of kSIST FprEN 17774:2026?

kSIST FprEN 17774:2026 is built around ICP-AES measurement of the target elements in prepared extracts, with direct measurement or measurement after dilution. In practice, the user must prepare the extract, set up the instrument, control interferences, calibrate correctly, and document the result in a consistent way.

Scope and applicability

Clause 1 limits the method to boron, cobalt, copper, iron, manganese, molybdenum, phosphorous and zinc in organic and organo-mineral fertilizers. The same clause extends the method to fertilizing product blends when the organic fertilizer or organo-mineral fertilizer category is the highest share in the blend; otherwise, the standard for the highest share applies. If the blend is present in equal quantity, the user decides which standard to apply.

Clause 1 also notes that ICP-MS may be used if the user proves it gives the same results. That matters because a laboratory may use different equipment, but it must show equivalence before relying on it for routine reporting.

Principle of measurement

Clause 4 says the method measures element concentrations in the extract by ICP-AES, using a nebulizer, a plasma source, a spectrometer and photosensitive detection. Sequential or simultaneous optical systems may be used, and both axial and radial viewing are allowed. In practice, the laboratory chooses the instrument setup that gives stable sensitivity while controlling interference.

Interferences and matrix control

Clause 5 groups the main problems as spectral, transport, excitation, chemical and memory interferences. The document describes how to reduce them by background correction, selection of another wavelength, dilution, matrix matching, standard addition, and rinse steps.

This matters especially for unknown or complex matrices, because the document warns that sample composition can change the response. The operator therefore has to check line interference, confirm the chosen wavelength, and react if the matrix changes the result.

Reagents and apparatus

Clause 6 requires analytical-grade reagents with negligible amounts of the analytes being measured. It also gives the required water quality, the ammonium hydroxide solution, neutral ammonium citrate solution, formic acid solution, stock standards, argon purity and nitric acid preparation.

Clause 7 adds practical equipment requirements. For boron determination, the document requires avoiding borosilicate glass contact and using suitable plastic or silica ware instead. That is important because boron contamination can come from the labware itself.

Procedure, calibration and quality control

Clause 8 sets out the test solution, blank solution and calibration solution preparation. The extract is diluted so the final concentration sits roughly in the middle of the calibration range, which helps keep the measurement within the best part of the calibration curve. The document then gives separate calibration ranges for micronutrients and for phosphorous.

Clause 8 also requires the instrument to be stabilized, calibration to be verified during the run, and emission intensity to be read at least twice and averaged. A calibration blank and a calibration check solution are run every 20 samples or less and again at the end of the measurement. In practice, this is the routine control that shows whether the instrument drifted during the batch.

For unknown matrices, Clause 8 requires matrix checks by spiking, fivefold dilution, or measurement at a different wavelength. The stated acceptance checks are a spike recovery between 90 % and 110 % and a difference of less than 10 % between the original sample and the fivefold-diluted sample. If those checks fail, the standard addition method is used.

Calculation and reporting

Clause 9 gives the formula for calculating the content of each element and shows how to express results either as received or on a dry matter basis. It also defines the dilution factor and the dry matter correction factor. Clause 10 then sets the minimum contents of the test report, including the sample identification, extraction method, standard used, test method reference, results, sampling information and any unusual operating details.

What terms does kSIST FprEN 17774:2026 define?

kSIST FprEN 17774:2026 uses the vocabulary of EN 12944-1 and EN 12944-2, and several technical terms are central to the method.

  • Inductively coupled plasma-atomic emission spectrometry (ICP-AES) - The measurement technique used here, where excited atoms emit element-specific light that is measured by the instrument.
  • Inductively coupled plasma optical emission spectrometry (ICP-OES) - A synonym for ICP-AES in the introduction; the document treats both names as the same technique.
  • Fertilizing product blends - Mixtures of at least two listed fertilizing components, where the applicable standard depends on which component category dominates.
  • Blank test solution - A solution prepared by the same procedure as the sample, used to correct the measured result for background contribution.
  • Standard addition method - A check and correction approach where a known amount of analyte is added to the sample to assess matrix effects.
  • Dry matter content - The basis used to recalculate results for samples reported on a dry matter basis, using prEN 17773.
  • Calibration check solution - A calibration standard used during the run to verify that calibration remains stable.

Who uses kSIST FprEN 17774:2026?

kSIST FprEN 17774:2026 is used by laboratories that test organic fertilizers, organo-mineral fertilizers and fertilizing product blends containing them. It is also relevant to quality control teams in fertilizer production, because the method is designed for routine determination of micronutrients and phosphorous in prepared extracts.

The document is also useful to analytical staff who must validate the instrument, check for matrix effects, and report results in a standard format. That makes it relevant for product control, conformity assessment and laboratory comparison work.

What changed in kSIST FprEN 17774:2026 from the previous edition?

kSIST FprEN 17774:2026 supersedes CEN/TS 17774:2022. The European foreword states these significant changes:

  • the CEN Technical Specification was adopted as a European Standard;
  • the title was adapted;
  • the European foreword was updated;
  • Clause 1 was updated, including clarification for fertilizing product blends and applicability to prEN 17767 and prEN 17779;
  • the bibliography was updated;
  • Annex A was added.

Which standards are used with kSIST FprEN 17774:2026?

kSIST FprEN 17774:2026 refers to several companion standards that support the method:

  • prEN 17766 - provides extraction by water for subsequent element determination.
  • prEN 17767 - provides extraction of phosphorous by formic acid.
  • prEN 17779 - provides extraction of phosphorous soluble in a neutral ammonium citrate solution.
  • prEN 17773 - provides the dry matter content needed for result correction.
  • EN 12944-1 - supplies general fertilizer vocabulary.
  • EN 12944-2 - supplies fertilizer-specific vocabulary.

What does the kSIST FprEN 17774:2026 document contain?

kSIST FprEN 17774:2026 contains the full analytical workflow from principle to reporting. Clauses 5 to 8 give the practical laboratory content: interference handling, reagents, apparatus, sample preparation, calibration, measurement and matrix checks.

It also includes Table 1 with recommended wavelengths for each element and notes on likely interfering elements, which helps the analyst choose a usable line for the instrument. Clause 9 provides the calculation formula and correction factors, and Clause 10 defines what has to be recorded in the test report.

Annex A adds the interlaboratory study behind the method. It summarizes the sample set, the participating laboratories and the statistical evaluation for the elements measured in extracts prepared according to the linked extraction standards.

Relations

Effective Date
01-Mar-2026

Buy Documents

Draft

oSIST prEN 17774:2026

English language (25 pages)
Preview
Preview
e-Library read for
×1 day

Get Certified

Connect with accredited certification bodies for this standard

Control Union Certifications

Global certification for agriculture and sustainability.

RVA Netherlands Verified

ECOCERT

Organic and sustainability certification.

COFRAC France Verified

FSC International

Forest Stewardship Council certification.

ASI Germany Verified

Sponsored listings

Frequently Asked Questions

kSIST FprEN 17774:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Organic and organo-mineral fertilizers - Determination of the content of specific elements by ICP-AES after extraction by water and weak solvents". This standard covers: This document specifies a method for the determination of boron (B), cobalt (Co), copper (Cu), iron (Fe), manganese (Mn), molybdenum (Mo), phosphorous (P) and zinc (Zn) in organic and organo-mineral fertilizers using inductively coupled plasma-atomic emission spectrometry (ICP-AES). This method is applicable to extracts prepared according to EN 17766, EN 17767 and EN 17779. This document is applicable to fertilizing product blends, where the blend is a mix of at least two of the following components: fertilizers, liming materials, soil improvers, growing media, inhibitors, plant biostimulants, and where the following category: organic fertilizers or organo-mineral fertilizers is the highest % in the blend by mass or volume, or in the case of liquid form by dry mass. If the organic fertilizer or the organo-mineral fertilizer is not the highest % in the blend, the European Standard for the highest % of the blend applies. In case a fertilizing product blend is composed of components in equal quantity, the user decides which standard to apply. NOTE Variations in analytical methods for fertilizing product blends can lead to differing results as some components or matrix interactions can affect the outcome. Validation procedures have shown that developed standard methods are robust and reliable across diverse product compositions, but possible interferences and unexpected results when analysing fertilizing product blends are possible.

This document specifies a method for the determination of boron (B), cobalt (Co), copper (Cu), iron (Fe), manganese (Mn), molybdenum (Mo), phosphorous (P) and zinc (Zn) in organic and organo-mineral fertilizers using inductively coupled plasma-atomic emission spectrometry (ICP-AES). This method is applicable to extracts prepared according to EN 17766, EN 17767 and EN 17779. This document is applicable to fertilizing product blends, where the blend is a mix of at least two of the following components: fertilizers, liming materials, soil improvers, growing media, inhibitors, plant biostimulants, and where the following category: organic fertilizers or organo-mineral fertilizers is the highest % in the blend by mass or volume, or in the case of liquid form by dry mass. If the organic fertilizer or the organo-mineral fertilizer is not the highest % in the blend, the European Standard for the highest % of the blend applies. In case a fertilizing product blend is composed of components in equal quantity, the user decides which standard to apply. NOTE Variations in analytical methods for fertilizing product blends can lead to differing results as some components or matrix interactions can affect the outcome. Validation procedures have shown that developed standard methods are robust and reliable across diverse product compositions, but possible interferences and unexpected results when analysing fertilizing product blends are possible.

kSIST FprEN 17774:2026 is classified under the following ICS (International Classification for Standards) categories: 65.080 - Fertilizers. The ICS classification helps identify the subject area and facilitates finding related standards.

kSIST FprEN 17774:2026 has the following relationships with other standards: It is inter standard links to SIST-TS CEN/TS 17774:2023. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

kSIST FprEN 17774:2026 is associated with the following European legislation: EU Directives/Regulations: 2019/1009; Standardization Mandates: M/564, M/564 Amd 1, M/564 Amd 2. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.

kSIST FprEN 17774:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


SLOVENSKI STANDARD
oSIST prEN 17774:2026
01-marec-2026
Organska in organsko-mineralna gnojila - Določanje specifičnih elementov z
atomsko emisijsko spektrometrijo z induktivno sklopljeno plazmo (ICP-AES) po
ekstrakciji z vodo in šibkimi topili
Organic and organo-mineral fertilizers - Determination of the content of specific elements
by ICP-AES after extraction by water and weak solvents
Organische und organisch-mineralische Düngemittel - Bestimmung des Gehalts
spezifischer Elemente mittels ICP-AES nach Extraktion mit Wasser und schwachen
Lösungsmitteln
Engrais organiques et organo-minéraux - Détermination de la teneur en éléments
spécifiques par ICP-AES après extraction à l’eau et aux solvants faibles
Ta slovenski standard je istoveten z: prEN 17774
ICS:
65.080 Gnojila Fertilizers
oSIST prEN 17774:2026 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

oSIST prEN 17774:2026
oSIST prEN 17774:2026
DRAFT
EUROPEAN STANDARD
prEN 17774
NORME EUROPÉENNE
EUROPÄISCHE NORM
January 2026
ICS 65.080 Will supersede CEN/TS 17774:2022
English Version
Organic and organo-mineral fertilizers - Determination of
the content of specific elements by ICP-AES after
extraction by water and weak solvents
Engrais organiques et organo-minéraux - Organische und organisch-mineralische Düngemittel -
Détermination de la teneur en éléments spécifiques par Bestimmung des Gehalts spezifischer Elemente mittels
ICP-AES après extraction à l'eau et aux solvants faibles ICP-AES nach Extraktion mit Wasser und schwachen
Lösungsmitteln
This draft European Standard is submitted to CEN members for enquiry. It has been drawn up by the Technical Committee
CEN/TC 260.
If this draft becomes a European Standard, CEN members are bound to comply with the CEN/CENELEC Internal Regulations
which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.

This draft European Standard was established by CEN in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC
Management Centre has the same status as the official versions.

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and
United Kingdom.
Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are
aware and to provide supporting documentation.

Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without
notice and shall not be referred to as a European Standard.

EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION

EUROPÄISCHES KOMITEE FÜR NORMUNG

CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2026 CEN All rights of exploitation in any form and by any means reserved Ref. No. prEN 17774:2026 E
worldwide for CEN national Members.

oSIST prEN 17774:2026
prEN 17774:2026 (E)
Contents Page
European foreword . 3
Introduction . 4
1 Scope . 5
2 Normative references . 5
3 Terms and definitions . 6
4 Principle . 6
5 Interferences . 6
5.1 General . 6
5.2 Spectral interferences . 6
5.3 Transport interferences . 6
5.4 Excitation interferences . 7
5.5 Chemical interferences . 7
5.6 Memory interferences . 7
6 Reagents . 7
7 Apparatus . 8
7.1 Common laboratory glassware . 8
7.2 Inductively coupled plasma atomic emission spectrometer . 8
8 Procedure. 9
8.1 Preparation of test and blank solution . 9
8.2 Preparation of the calibration solutions . 9
8.3 Measurement . 9
9 Calculation and expression of the results . 12
10 Test report . 13
Annex A (informative) Results of the interlaboratory study . 14
A.1 Interlaboratory tests . 14
A.2 Statistical results for the determination of specific elements . 14
Bibliography . 25

oSIST prEN 17774:2026
prEN 17774:2026 (E)
European foreword
This document (prEN 17774:2026) has been prepared by Technical Committee CEN/TC 260 “Fertilizers
and liming materials”, the secretariat of which is held by DIN.
This document is currently submitted to the CEN Enquiry.
This document will supersede CEN/TS 17774:2022.
Compared to the previous edition, prEN 17774:2026 includes the following significant technical changes:
— the CEN Technical Specification has been adopted as a European Standard;
— title adapted;
— European foreword updated;
— Scope (Clause 1) updated including the clarification on fertilizing product blends and the applicability
to prEN 17767, Organo-mineral fertilizers — Extraction of phosphorus by formic acid and prEN 17779,
Organo-mineral fertilizers — Extraction of phosphorus, soluble in a neutral ammonium citrate solution;
— Bibliography updated;
— Annex A (Results of the inter-laboratory study) added.
This document has been prepared under a Standardization Request addressed to CEN by the European
Commission. The Standing Committee of the EFTA States subsequently approves these requests for its
Member States.
oSIST prEN 17774:2026
prEN 17774:2026 (E)
Introduction
Water is applied to extract available fraction of micronutrients from different matrices for their
subsequent determination by ICP-AES. A similar procedure was applied for extraction of micronutrients
from inorganic fertilizers [2].
Inductively coupled plasma atomic emission spectrometry (ICP-AES) is nowadays widely used and a well-
established method in many laboratories.
Inductively coupled plasma atomic emission spectrometry (ICP-AES) is often called also inductively
coupled plasma optical emission spectrometry (ICP-OES). Both these terms are synonyms for the same
analytical technique.
WARNING — Persons using this document should be familiar with usual laboratory practice. This
document does not purport to address all of the safety issues, if any, associated with its use. It is the
responsibility of the user to establish appropriate health and safety practices and to ensure compliance
with any national regulatory conditions.
IMPORTANT — It is absolutely essential that tests conducted according to this document are carried out
by suitably trained staff.
oSIST prEN 17774:2026
prEN 17774:2026 (E)
1 Scope
This document specifies a method for the determination of boron (B), cobalt (Co), copper (Cu), iron (Fe),
manganese (Mn), molybdenum (Mo), phosphorous (P) and zinc (Zn) in organic and organo-mineral
fertilizers using inductively coupled plasma-atomic emission spectrometry (ICP-AES).
NOTE 1 Alternatively, inductively coupled plasma mass spectrometry (ICP-MS) can be used for the measurement
if the user proves that the method gives the same results.
This method is applicable to extracts prepared according to prEN 17766, prEN 17767 and prEN 17779.
NOTE 2 The method can be used for the determination of other elements, provided the user has verified the
applicability.
This document is applicable to the fertilizing products blends where a blend is a mix of at least two of the
following components: fertilizers, liming materials, soil improvers, growing media, inhibitors, plant
biostimulants, and where the following category: organic fertilizers, organo-mineral fertilizers is the
highest % in the blend by mass or volume, or in the case of liquid form by dry mass. If the organic fertilizer
or the organo-mineral fertilizer is not the highest % in the blend, the European Standard for the highest
% of the blend applies. In case a fertilizing product blend is composed of components in equal quantity,
the user decides which standard to apply. Variations in analytical methods for fertilizing product blends
can lead to differing results as some components or matrix interactions can affect the outcome. Validation
procedures have shown that developed standard methods are robust and reliable across diverse product
compositions, but possible interferences and unexpected results when analysing fertilizing product
blends are possible.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
prEN 17766 , Organic and organo-mineral fertilizers — Extraction by water for subsequent determination
of elements
prEN 17767, Organo-mineral fertilizers — Extraction of phosphorus by formic acid
prEN 17779, Organo-mineral fertilizers — Extraction of phosphorus soluble in a neutral ammonium
citrate solution
prEN 17773, Organic and organo-mineral fertilizers — Determination of the dry matter content
EN 12944-1, Fertilizers and liming materials — Vocabulary — Part 1: General terms
EN 12944-2, Fertilizers and liming materials — Vocabulary — Part 2: Terms relating to fertilizers

Under preparation.
Under preparation.
Under preparation.
Under preparation.
oSIST prEN 17774:2026
prEN 17774:2026 (E)
3 Terms and definitions
For the purposes of this document, the terms and definitions given in EN 12944-1 and EN 12944-2 apply.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https://www.iso.org/obp
— IEC Electropedia: available at https://www.electropedia.org/
4 Principle
The method is based on the ICP-AES measurement of the concentration of boron (B), cobalt (Co), copper
(Cu), iron (Fe), manganese (Mn), molybdenum (Mo) and zinc (Zn) in fertilizer extracts prepared
according to prEN 17766 and can also be used for determination of phosphorous in the extracts prepared
according to prEN 17767 and prEN 17779. The elements are determined directly or after appropriate
dilution of the extract. The solution is dispersed by a nebulizer of the ICP-AES instrument and the
resulting aerosol is transported into the plasma. Element specific emission spectra are produced by a
radio-frequency inductively coupled argon plasma where atoms or ions are excited at high temperature.
The emission spectra are dispersed by a spectrometer, and the intensities of the emission lines are
monitored by photosensitive devices. Multi-element determinations using sequential or simultaneous
optical systems and axial or radial viewing of the plasma may be used.
The method may be used for the determination of other elements, provided the user has verified the
applicability.
5 Interferences
5.1 General
Interferences and matrix effects shall be recognized and appropriate measures to minimize them shall be
made. There are several types of interferences (see 5.2 to 5.6).
5.2 Spectral interferences
Spectral interferences are due to incomplete isolation of the radiation emitted by the analyte from other
detected radiation sources. Spectral interferences are caused by background emission from continuous
or recombination phenomena, by stray light which causes background increase or overlap of a spectral
line from another element, or unresolved overlap of molecular band spectra. Background emission and
stray light can usually be compensated for by subtracting the background emission measured adjacent to
the analyte wavelength peak. To correct a sloping background shift two background correction points on
each side of the peak are used. Increase of background is more intensive with axial-view instruments.
Background correction is not required in cases of line broadening where a background correction
measurement would actually degrade the analytical result. A spectral line overlap usually leads to the
choice of an alternative line. If this is not possible, mathematical correction procedures (e.g. inter-element
correction technique, multi-component spectral fitting) can be used to compensate for the interference.
These correction procedures are usually a part of the instrument software.
5.3 Transport interferences
Transport interferences are caused by differences in the properties between the sample solutions and
the calibration solutions (viscosity, surface tension, density, dissolved solid content, type and
concentration of acids). As a consequence, the supply of solution to the nebuliser, the efficiency of
nebulisation and the droplet size distribution of the aerosol is altered, resulting in a change of sensitivity.
Errors due to these interferences can be overcome by dilution of the solutions, by matrix matching, by
standard addition or by internal standard.
oSIST prEN 17774:2026
prEN 17774:2026 (E)
5.4 Excitation interferences
Excitation interferences are attributed to a change in the excitation conditions in the plasma, especially
by the presence of easily ionisable elements. The interference depends on the operating conditions of the
plasma (e.g. power, sample introduction, gas flowrate or observation height) and differ from element to
element. Improvement of the plasma conditions can therefore reduce excitation interferences. Other
possibilities are dilution of the solutions, matrix matching or the standard addition technique.
5.5 Chemical interferences
Chemical interferences are not significant with the ICP-AES technique, but if observed, they can be
minimized by a careful selection of operating conditions (e.g. radio frequency power, observation
position, gas flow rate and so forth).
5.6 Memory interferences
Memory interferences result when analytes in a previous sample contribute to the signals measured in a
new sample. This type of interference can be caused by sample deposits or the accumulation in pump
tubing, nebulizer, spray chamber or plasma torch. The possibility of memory interferences should be
recognized within an analytical run and suitable rinse steps and rinse times should be used.
6 Reagents
All reagents shall be of recognized analytical grade and they shall have negligible concentration of the
element to be determined if compared to the lowest concentration of that element in the sample solution.
All solutions should be stored in clean boron-free vessels to avoid contamination during storage.
6.1 Water, with a specific conductivity not higher than 0,2 mS/m at 25 °C, free from the elements to be
determined.
6.2 Ammonium hydroxide solution, containing 28 % to 29 % of NH .
6.3 Neutral ammonium citrate solution(C H O ·H O), pH = 7, mass concentration,
6 8 7 2
ρ(C H O ·H O) = 185 g/l, relative density = 1,09 at 20 °C.
6 8 7 2
Prepare the reagent as follows:
H O⋅H O) in about 1,5 l of water and make an approximately
Dissolve 370 g of crystalline citric acid (C6 8 7 2
neutral solution by adding 345 ml of ammonium hydroxide solution (6.2).
Cool and make exactly neutral by keeping the electrodes of a pH-meter immersed in the solution. Add the
ammonium hydroxide solution (6.2), drop by drop, stirring continuously (e.g. with a mechanical stirrer)
until obtaining exactly a pH of 7 at a temperature of 20 °C. At this point make up the volume with water
(6.1) to 2 l and check the pH again. Keep the reagent in a closed container and check the pH at regular
intervals.
Alternatively, a commercially certified solution can be used.
6.4 Formic acid 2 %, (ρ ≈ 20 g/l).
Make 82 ml of formic acid (concentration 98 % to 100 %; mass concentration at 20 °C ρ20 = 1,22 g/ml)
up to 5 l with distilled water.
6.5 Standard stock solutions for boron, cobalt, copper, iron, manganese, molybdenum and zinc,
mass concentration ρ = 1 000 mg/l for each element.
oSIST prEN 17774:2026
prEN 17774:2026 (E)
Both single-element stock solutions and multi-element stock solutions with adequate specification,
stating the acid used and the preparation technique, are commercially available. Multi-element stock
solutions are usually available at the individual mass concentration ρ = 100 mg/l for each element. These
solutions are considered to be stable for more than one year, but in reference to guaranteed stability, the
recommendations of the manufacturer should be considered. Alternatively, the stock solutions may be
prepared by dissolution of high purity metals.
NOTE If additional elements are to be determined, they can be included in the stock solution as well.
6.6 Standard solution for boron, cobalt, copper, iron, manganese, molybdenum and zinc, mass
concentration ρ = 100 mg/l of element.
Use commercially available multi-element solution of this concentration for each element or pipette
10 ml of the appropriate element stock solution (6.5) into a 100 ml volumetric flask, add 5 ml of diluted
nitric acid (6.9.1), fill to the mark with water and mix well. This solution is used to prepare spiked test
solutions and calibration solutions.
6.7 Standard stock solution for phosphorous, ρ = 1000 mg/l of P.
This single-element stock solution is commercially available.
6.8 Argon, purity 99,995 % or higher.
6.9 Nitric acid, substance concentration c(HNO ) ≈ 14,3 mol/l; ρ ≈ 1,4 g/ml.
6.9.1 Diluted nitric acid solution, c(HNO ) = 5 mol/l.
Add 350 ml of nitric acid (6.9) to 650 ml of water.
7 Apparatus
7.1 Common laboratory glassware
If boron content is to be determined, it is necessary to avoid the contact of all solutions with borosilicate
glassware. Suitable plastic or silica ware shall be used. Glass volumetric flasks may be used for making
up to volume but not for storage of digests, reagents, and solutions.
7.2 Inductively coupled plasma atomic emission spectrometer
WARNING — It is essential that the manufacturer’s safety instructions are strictly observed when using
this apparatus.
The ICP atomic-emission spectrometer consists of a sample introduction system, the plasma as an
excitation source, an optical system, a detector and a computer with suitable software. The sample is
transported by the introduction system (rotation tube pump, nebulizer and a spray chamber) to the
plasma torch. Around the torch a water-cooled radio frequency (RF) coil is placed. A frequency from
27 MHz to 56 MHz with a power of 600 W to 2 000 W is usually used. The emission from the plasma can
be observed either from the side (radial) or from the torch central symmetrical axis (axial). Axial viewing
gives more signal intensity due to the increased observation path length of the normal analytical zone of
the plasma but an increase of interference is also commonly observed. Spectral lines are measured and
registered either in a sequential or a simultaneous manner.
oSIST prEN 17774:2026
prEN 17774:2026 (E)
8 Procedure
8.1 Preparation of test and blank solution
Dilute an aliquot portion of the extract, obtained according to prEN 17766, prEN 17767 or prEN 17779
in one or more steps so that the final concentration of the element to be determined is approximately in
the middle of the given calibration range (8.2). In the final diluting step add a suitable volume of the
extract or diluted extract to a 100 ml volumetric flask, add 5 ml diluted nitric acid (6.9.1), fill to the mark
with water and mix well.
A blank test solution is prepared for the measurement following the same procedure as for test sample
solutions. If dilution is necessary, dilute an aliquot portion of the extract in one or more steps so that the
final concentration of the element to be determined is approximately in the middle of the given
calibration range (8.2). In the final diluting step fill to the mark with water (6.1) or extracting solutions
for phosphorus determination (6.3; 6.4) and mix well. Prepare a diluted blank test solution by pipetting
a blank test solution and dilute in the same way as the test sample solution
8.2 Preparation of the calibration solutions
8.2.1 Calibration solutions for micro-nutrients
Pipette volumes of 0 ml; 0,5 ml; 2 ml; 5 ml and 10 ml of the standard solution (6.6) and 5 ml of diluted
nitric acid (6.9.1) into five 100 ml volumetric flasks respectively. Fill to the mark with water. The mass
concentrations of each element in the calibration solutions are: 0 mg/l; 0,5 mg/l; 2 mg/l; 5 mg/l and
10 mg/l. Mixed calibration solutions for all elements or individual sets of calibration solutions for each
element may be used. If necessary, calibration solutions of higher or lower concentrations may be
prepared in the same way.
NOTE It is possible to calibrate the instrument for higher concentrations of the micronutrients if the calibration
curve is linear.
8.2.2 Calibration solutions for phosphorus
Pipette volumes of 0 ml; 10 ml; 20 ml, 30 ml, 40 ml and 50
...