SIST EN 285:2006/oprA1:2006
Sterilization - Steam sterilizers - Large sterilizers
Sterilization - Steam sterilizers - Large sterilizers
This amendment specifies a test method to demonstrate that at the levels at which the controls are set, air dilution from within the test piece is sufficient to permit even steam penetration into it.
Sterilisation - Dampf-Sterilisatoren - Groß-Sterilisatoren
Stérilisation - Stérilisateurs a la vapeur d'eau - Grands stérilisateurs
Sterilizacija – Parni sterilizatorji – Veliki sterilizatorji
General Information
- Status
- Not Published
- Technical Committee
- VAZ - Healthcare
- Current Stage
- 98 - Abandoned project (Adopted Project)
- Start Date
- 02-Oct-2014
- Due Date
- 07-Oct-2014
- Completion Date
- 02-Oct-2014
- Directive
- Not Harmonized93/42/EEC - Medical devices
Relations
- Effective Date
- 28-Jan-2026
- Effective Date
- 01-Jul-2008
- Effective Date
- 22-Dec-2008
Overview
SIST EN 285:2006/oprA1:2006 - Sterilization - Steam Sterilizers - Large Sterilizers is a draft amendment developed by the European Committee for Standardization (CEN), focusing on the performance validation of large steam sterilizers used in healthcare and laboratory environments. This amendment introduces a specific test method, the hollow load test, to assess whether air inside test items is sufficiently diluted to ensure even steam penetration at the preset control levels of the sterilization process. Effective steam penetration is critical for achieving reliable sterilization results, especially with increasingly complex instrument designs.
Key Topics
Hollow Load Test:
The core addition of this amendment is the hollow load test. This test is designed to confirm that air removal in steam sterilizers is adequate, allowing steam to reach all internal surfaces of hollow medical instruments and devices. The test specifically assesses conditions relevant for complex medical instruments with long lumens or internal channels, which are challenging to sterilize using traditional textile load tests.Test Method Specification:
- Requires a process challenge device and indicator system compliant with EN 867-5.
- The process challenge device is preconditioned to specified temperature and humidity ranges to simulate real-use conditions.
- After the sterilization process, the indicator must show evidence of complete steam penetration, confirming efficient air removal and overall sterilization efficacy.
Updated Table References and Clauses:
The amendment modifies and deletes certain clauses and table lines relating to load types, aligning them with new requirements for hollow load testing.Increased Relevance for Modern Healthcare:
The document recognizes the ongoing shift from textile-based sterilization to the use of intricate medical instruments, making the hollow load test an essential tool for meeting contemporary sterilization challenges.
Applications
Healthcare Facilities:
Large steam sterilizers are widely used in hospitals, clinics, and dental practices. The hollow load test ensures that sterilizers are validated for their ability to effectively sterilize items with narrow lumens, such as endoscopes, surgical instruments, and tubing.Medical Device Manufacturing:
Sterilization process qualification and quality assurance programs in medical device manufacturing benefit from standardized hollow load testing to validate steam penetration for product safety and regulatory compliance.Service Providers:
Sterilization service providers use these standards to maintain consistency and compliance in reprocessing reusable medical instruments, minimizing the risk of infection and complying with healthcare regulations.
Related Standards
EN 867-5:
Specifies non-biological indicator systems and process challenge devices used in sterilizer performance testing, directly referenced for determining compliance in the hollow load test.EN 285:2006:
The parent standard to which this amendment applies, providing comprehensive requirements for large steam sterilizers.ISO 17665:
International standard for sterilization of healthcare products by moist heat, providing harmonized requirements that complement the European approach.
Practical Value
Risk Mitigation:
This amendment helps healthcare facilities and sterilization providers reduce the risk of suboptimal sterilization, especially for complex medical instruments.Regulatory Compliance:
Incorporating the hollow load test into routine validation supports compliance with EU and national regulatory requirements for medical device sterilization.Quality Assurance:
Operationalizing this standard ensures that large steam sterilizers are consistently achieving high levels of microbiological safety, ultimately protecting patient health and enhancing service reliability.
By aligning with SIST EN 285:2006/oprA1:2006, organizations bolster their sterilization processes, address evolving instrument complexities, and ensure thorough steam penetration for optimal sterilization outcomes.
Frequently Asked Questions
SIST EN 285:2006/oprA1:2006 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Sterilization - Steam sterilizers - Large sterilizers". This standard covers: This amendment specifies a test method to demonstrate that at the levels at which the controls are set, air dilution from within the test piece is sufficient to permit even steam penetration into it.
This amendment specifies a test method to demonstrate that at the levels at which the controls are set, air dilution from within the test piece is sufficient to permit even steam penetration into it.
SIST EN 285:2006/oprA1:2006 has the following relationships with other standards: It is inter standard links to SIST EN ISO 17665-1:2006, SIST EN 285:2006+A1:2008, SIST EN 285:2006. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
SIST EN 285:2006/oprA1:2006 is associated with the following European legislation: EU Directives/Regulations: 93/42/EEC. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.
SIST EN 285:2006/oprA1:2006 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
SLOVENSKI SIST EN 285:2006/oprA1:2006
PREDSTANDARD
september 2006
Sterilizacija – Parni sterilizatorji – Veliki sterilizatorji
Sterilization - Steam sterilizers - Large sterilizers
ICS 11.080.10 Referenčna številka
© Standard je založil in izdal Slovenski inštitut za standardizacijo. Razmnoževanje ali kopiranje celote ali delov tega dokumenta ni dovoljeno
EUROPEAN STANDARD
DRAFT
EN 285:2006
NORME EUROPÉENNE
EUROPÄISCHE NORM
prA1
July 2006
ICS
English Version
Sterilization - Steam sterilizers - Large sterilizers
Stérilisation - Stérilisateurs à la vapeur d'eau - Grands Sterilisation - Dampf-Sterilisatoren - Groß-Sterilisatoren
stérilisateurs
This draft amendment is submitted to CEN members for enquiry. It has been drawn up by the Technical Committee CEN/TC 102.
This draft amendment A1, if approved, will modify the European Standard EN 285:2006. If this draft becomes an amendment, CEN
members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for inclusion of this amendment
into the relevant national standard without any alteration.
This draft amendment was established by CEN in three official versions (English, French, German). A version in any other language made
by translation under the responsibility of a CEN member into its own language and notified to the Management Centre has the same status
as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Cyprus, Czech Republic, Denmark, Estonia, Finland, France,
Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania,
Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are aware and to
provide supporting documentation.
: This document is not a European Standard. It is distributed for review and comments. It is subject to change without notice and
Warning
shall not be referred to as a European Standard.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
Management Centre: rue de Stassart, 36 B-1050 Brussels
© 2006 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN 285:2006/prA1:2006: E
worldwide for CEN national Members.
Contents Page
Foreword.3
Revised text.3
2 Clause.3
8.3.5 Hollow load test .3
15 Hollow load test .4
15.1 General.4
15.2 Apparatus.4
15.3 Procedure.4
Foreword
This document (EN 285:2006/prA1:2006) has been prepared by Technical Committee CEN/TC 102
“Sterilizers for medical purposes”, the secretariat of which is held by DIN.
This document is currently submitted to the CEN Enquiry.
Revised text
2 Clause
Add
EN 867-5, Non-biological systems for use in sterilizers – Part 5: Specification for indicator systems and
process challenge devices for use in performance testing for small sterilizers type B and type S
7.1.13
Delete at the end of the first sentence "(e.g. the sterilization of instruments, textiles, rubber)"
8.1.3
replace the text of "NOTE 2" by the following
...