ASTM D3609-00(2014)
(Practice)Standard Practice for Calibration Techniques Using Permeation Tubes
Standard Practice for Calibration Techniques Using Permeation Tubes
SIGNIFICANCE AND USE
5.1 Most analytical methods used in air pollutant measurements are comparative in nature and require calibration or standardization, or both, often with known blends of the gas of interest. Since many of the important air pollutants are reactive and unstable, it is difficult to store them as standard mixtures of known concentration for extended calibration purposes. An alternative is to prepare dynamically standard blends as required. This procedure is simplified if a constant source of the gas of interest can be provided. Permeation tubes provide this constant source, if properly calibrated and if maintained at constant temperature. Permeation tubes have been specified as reference calibration sources, for certain analytical procedures, by the Environmental Protection Agency (3).
SCOPE
1.1 This practice describes a means for using permeation tubes for dynamically calibrating instruments, analyzers, and analytical procedures used in measuring concentrations of gases or vapors in atmospheres (1, 2).2
1.2 Typical materials that may be sealed in permeation tubes include: sulfur dioxide, nitrogen dioxide, hydrogen sulfide, chlorine, ammonia, propane, and butane (1).
1.3 The values stated in SI units are to be regarded as standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
Buy Standard
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: D3609 − 00 (Reapproved 2014)
Standard Practice for
Calibration Techniques Using Permeation Tubes
This standard is issued under the fixed designation D3609; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 4.3 Permeation tubes are held at constant temperature in a
carrier-gas stream of dry air or nitrogen to produce a gas
1.1 This practice describes a means for using permeation
concentration dependent on the permeation rate and the flow of
tubes for dynamically calibrating instruments, analyzers, and
the carrier gas.
analytical procedures used in measuring concentrations of
gases or vapors in atmospheres (1, 2).
5. Significance and Use
1.2 Typicalmaterialsthatmaybesealedinpermeationtubes
5.1 Most analytical methods used in air pollutant measure-
include: sulfur dioxide, nitrogen dioxide, hydrogen sulfide,
ments are comparative in nature and require calibration or
chlorine, ammonia, propane, and butane (1).
standardization, or both, often with known blends of the gas of
1.3 The values stated in SI units are to be regarded as
interest. Since many of the important air pollutants are reactive
standard.
andunstable,itisdifficulttostorethemasstandardmixturesof
known concentration for extended calibration purposes. An
1.4 This standard does not purport to address all of the
alternative is to prepare dynamically standard blends as re-
safety concerns, if any, associated with its use. It is the
quired. This procedure is simplified if a constant source of the
responsibility of the user of this standard to establish appro-
gas of interest can be provided. Permeation tubes provide this
priate safety and health practices and determine the applica-
constant source, if properly calibrated and if maintained at
bility of regulatory limitations prior to use.
constant temperature. Permeation tubes have been specified as
2. Referenced Documents
reference calibration sources, for certain analytical procedures,
by the Environmental Protection Agency (3).
2.1 ASTM Standards:
D1356 Terminology Relating to Sampling and Analysis of
6. Interferences and Precautions
Atmospheres
6.1 Permeation tubes are essentially devices to provide a
D3195 Practice for Rotameter Calibration
constant rate of emission of a specific gaseous substance over
3. Terminology
period of time.They consist of a two-phase (gas-liquid) system
to maintain a constant vapor pressure (at constant temperature)
3.1 Definitions—Refer to Terminology D1356.
which is the driving force for emission of the gas through a
4. Summary of Practice
semipermeable membrane (tube walls). They can be expected
to maintain a constant emission rate that is temperature
4.1 Aliquefiable gas, when enclosed in an inert plastic tube,
dependent as long as a significant amount of liquid is present
escapes by permeating the tubing wall at a constant,
inthedevice.Theliquidshallbepure,elseitscompositionmay
reproducible, temperature-dependent rate.
change during the life time of the tube, due to differential
4.2 Permeationtubesarecalibratedgravimetrically,withthe
evaporation, with consequent vapor pressure changes. Care
weight loss of the tube equated to the weight of the escaping
must also be exercised that the diffusion membrane (tube
material.
walls) is not damaged or altered during use. The contents of
permeation tubes are under relatively high pressure.
1 Accordingly, there is the possibility of violent rupture of tube
This practice is under the jurisdiction ofASTM Committee D22 on Air Quality
and is the direct responsibility of Subcommittee D22.01 on Quality Control.
walls under high temperature exposure. Permeation rates have
Current edition approved Sept. 1, 2014. Published September 2014. Originally
temperature coefficients up to 10 % per degree Celsius. When
approved in 1977. Last previous edition approved in 2010 as D3609 – 00 (2010).
temperature coefficients are large, above 3 % per degree
DOI: 10.1520/D3609-00R14.
Celsius, stringent temperature control is required. Furthermore
The boldface numbers in parentheses refer to a list of references at the end of
this standard.
permeation tubes exhibit temperature hysteresis so that they
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
must be temperature equilibrated from 2 to 24 h before use,
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
depending upon the temperature differential between storage
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website. and use (4). It is important that permeation tubes are filled with
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D3609 − 00 (2014)
NOTE 1—This system has the advantage of smaller uncertainty of the temperature of the permeation tube.
FIG. 1 Optional System for Laboratory Use of a Permeation Tube
anhydrous constituents of high purity. They shall be handled being held at constant temperature. If lower concentrations are
with care to minimize contact with moisture, oil, and foreign desired, a second gas supply (diluent gas) with its control and
substances. measurement devices may be needed to mix with the gas from
the permeation tube chamber. Equipment of this kind is
6.2 Sulfur dioxide (SO ) permeation tubes are relatively
available commercially. A typical system contains a thermo-
insensitive to interferences.
electrically temperature-controlled permeation tube chamber
6.3 Nitrogen dioxide (NO ) permeation tubes are sensitive
with temperature control within 60.1°C over the range from
to moisture, hence they should be stored in dry atmospheres
15 to 35°C. Such equipment is well suited to field usage.
and used with relatively dry carrier gases (<10 % relative
7.3 A typical system for laboratory use that can be as-
humidity). Permeation of moisture into the contents of a tube
sembled from readily available parts is shown schematically in
may damage the walls and also cause progressive decreases in
Fig. 1. The parts required are described in the following
the permeation rate. Moisture incorporated in the contents
subsections.
during manufacture can cause the same effect (4).
7.3.1 Flowmeters—Several, sufficient to cover the range
6.4 Hydrogen sulfide (H S) permeation tubes may turn
from 0 to 15 L/min, calibrated by Practice D3195.
white during use in the presence of oxygen because of inverse
7.3.2 Copper Tubing—Approximately 1 m long [3 ft] by
permeation and formation of collodial sulfur. This phenom-
6.25 mm [0.25 in.] in outside diameter for use as a heat
enon may affect the permeation rate, if severe, hence is a
exchanger in the water bath.
reason for recalibration. However, in an inert gas stream, the
7.3.3 Ball Joints (Ungreased) and Tubing, for the necessary
tubes are relatively stable.
connections. Butt seals may also be used made with inert
6.5 Materials of construction shall be compatible with the
materials such as polyethylene.
contents of the tube. For instance, some fluorocarbons may
7.3.4 Mixing Bulb, to ensure adequate mixing of the perme-
cause FEP tubes to swell and possibly to rupture.
ated gas and the diluent gas stream. A Kjeldahl trap is
recommended.
7. Apparatus
7.3.5 Long Condenser, with large bore in which a thermom-
7.1 Permeation Tube sized in accordance with and cali-
eter and a permeation tube can be inserted.
brated to concentrations needed or expected for the analysis
7.3.6 Temperature Controlled Water Bath—About 8-L [2-
method. The user should check calibration as described in
gal] capacity, capable of 60.1°C or better water temperature
Section 9.1.
control, with a variable temperature control range from about
7.2 Flow and Temperature Control System—Prepare or 15 to 35°C, preferably equipped with a positive displacement
purchase a system that will dry the carrier gas, and control and type recirculating pump with at least 1-L/min liquid flow rate
measure its flow as it passes over the permeation tube that is to supply water to the condenser.
D3609 − 00 (2014)
NOTE 1—This system is constructed from readily available laboratory equipment.
NOTE 2—Warning—If the room temperature is significantly different from that of the water bath, a small difference in temperature between the bath
and the condensor containing the permeation tube can exist. In this event, the temperature indicated by the thermometer in the condensor should be used
as that of the permeation tube, rather than that of the water bath.
FIG. 2 Typical System for Laboratory Use
7.3.7 Thermometer, ASTM No. 91C or equivalent, cali- pick up moisture on exposure to air, they may need to be
brated to 60.1°C. weighed on a rigid time schedule to reproduce any mass
7.3.8 Mercury Barometer.
changes as a result of this cause (5). Repeat the weighing
operation at scheduled intervals and plot the gross weight
7.4 An alternate system is shown in Fig. 2. It has the
against elapsed time in minutes. The slope expressed as
advantage of lower uncertainty of the temperature of the
micrograms per minute represents the output of the tube. The
permeation tube.The required parts are described in the figure.
total time usually needed to calibrate at a given temperature
8. Reagents and Materials
should not be less than five days. Alternatively, linear regres-
sion analysis may be used to determine the permeation r
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: D3609 − 00 (Reapproved 2010) D3609 − 00 (Reapproved 2014)
Standard Practice for
Calibration Techniques Using Permeation Tubes
This standard is issued under the fixed designation D3609; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This practice describes a means for using permeation tubes for dynamically calibrating instruments, analyzers, and
analytical procedures used in measuring concentrations of gases or vapors in atmospheres (1, 2).
1.2 Typical materials that may be sealed in permeation tubes include: sulfur dioxide, nitrogen dioxide, hydrogen sulfide,
chlorine, ammonia, propane, and butane (1).
1.3 The values stated in SI units are to be regarded as standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Referenced Documents
2.1 ASTM Standards:
D1356 Terminology Relating to Sampling and Analysis of Atmospheres
D3195 Practice for Rotameter Calibration
3. Terminology
3.1 Definitions—referRefer to Terminology D1356.
4. Summary of Practice
4.1 A liquefiable gas, when enclosed in an inert plastic tube, escapes by permeating the tubing wall at a constant, reproducible,
temperature-dependent rate.
4.2 Permeation tubes are calibrated gravimetrically, with the weight loss of the tube equated to the weight of the escaping
material.
4.3 Permeation tubes are held at constant temperature in a carrier-gas stream of dry air or nitrogen to produce a gas
concentration dependent on the permeation rate and the flow of the carrier gas.
5. Significance and Use
5.1 Most analytical methods used in air pollutant measurements are comparative in nature and require calibration or
standardization, or both, often with known blends of the gas of interest. Since many of the important air pollutants are reactive
and unstable, it is difficult to store them as standard mixtures of known concentration for extended calibration purposes. An
alternative is to prepare dynamically standard blends as required. This procedure is simplified if a constant source of the gas of
interest can be provided. Permeation tubes provide this constant source, if properly calibrated and if maintained at constant
temperature. Permeation tubes have been specified as reference calibration sources, for certain analytical procedures, by the
Environmental Protection Agency (3).
This practice is under the jurisdiction of ASTM Committee D22 on Air Quality and is the direct responsibility of Subcommittee D22.01 on Quality Control.
Current edition approved April 1, 2010Sept. 1, 2014. Published June 2010September 2014. Originally approved in 1977. Last previous edition approved in 20052010 as
D3609 - 00 (2005).D3609 – 00 (2010). DOI: 10.1520/D3609-00R10.10.1520/D3609-00R14.
The boldface numbers in parentheses refer to a list of references at the end of this standard.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D3609 − 00 (2014)
NOTE 1—This system has the advantage of smaller uncertainty of the temperature of the permeation tube.
FIG. 1 Optional System for Laboratory Use of a Permeation Tube
6. Interferences and Precautions
6.1 Permeation tubes are essentially devices to provide a constant rate of emission of a specific gaseous substance over period
of time. They consist of a two-phase (gas-liquid) system to maintain a constant vapor pressure (at constant temperature) which is
the driving force for emission of the gas through a semipermeable membrane (tube walls). They can be expected to maintain a
constant emission rate that is temperature dependent as long as a significant amount of liquid is present in the device. The liquid
shall be pure, else its composition may change during the life time of the tube, due to differential evaporation, with consequent
vapor pressure changes. Care must also be exercised that the diffusion membrane (tube walls) is not damaged or altered during
use. The contents of permeation tubes are under relatively high pressure. Accordingly, there is the possibility of violent rupture
of tube walls under high temperature exposure. Permeation rates have temperature coefficients up to 10 % per degree Celsius.
When temperature coefficients are large, above 3 % per degree Celsius, stringent temperature control is required. Furthermore
permeation tubes exhibit temperature hysteresis so that they must be temperature equilibrated from 2 to 24 h before use, depending
upon the temperature differential between storage and use (4). It is important that permeation tubes are filled with anhydrous
constituents of high purity. They shall be handled with care to minimize contact with moisture, oil, and foreign substances.
6.2 Sulfur dioxide (SO ) permeation tubes are relatively insensitive to interferences.
6.3 Nitrogen dioxide (NO ) permeation tubes are sensitive to moisture, hence they should be stored in dry atmospheres and used
with relatively dry carrier gases (<10 % relative humidity). Permeation of moisture into the contents of a tube may damage the
walls and also cause progressive decreases in the permeation rate. Moisture incorporated in the contents during manufacture can
cause the same effect (4).
6.4 Hydrogen sulfide (H S) permeation tubes may turn white during use in the presence of oxygen because of inverse
permeation and formation of collodial sulfur. This phenomenon may affect the permeation rate, if severe, hence is a reason for
recalibration. However, in an inert gas stream, the tubes are relatively stable.
6.5 Materials of construction shall be compatible with the contents of the tube. For instance, some fluorocarbons may cause FEP
tubes to swell and possibly to rupture.
7. Apparatus
7.1 Permeation Tube sized in accordance with and calibrated to concentrations needed or expected for the analysis method. The
user should check calibration as described in Section 9.1.
7.2 Flow and Temperature Control System—Prepare or purchase a system that will dry the carrier gas, and control and measure
its flow as it passes over the permeation tube that is being held at constant temperature. If lower concentrations are desired, a
second gas supply (diluent gas) with its control and measurement devices may be needed to mix with the gas from the permeation
D3609 − 00 (2014)
tube chamber. Equipment of this kind is available commercially. A typical system contains a thermoelectrically temperature-
controlled permeation tube chamber with temperature control within 60.1 °C 60.1°C over the range from 15 to 35 °C. 35°C. Such
equipment is well suited to field usage.
7.3 A typical system for laboratory use that can be assembled from readily available parts is shown schematically in Fig. 1. The
parts required are described in the following subsections.
7.3.1 Flowmeters—Several, sufficient to cover the range from 0 to 15 L/min, calibrated by Practice D3195.
7.3.2 Copper Tubing—Approximately 1 m long [3 ft] by 6.25 mm [0.25 in.] in outside diameter for use as a heat exchanger in
the water bath.
7.3.3 Ball Joints (Ungreased) and Tubing, for the necessary connections. Butt seals may also be used made with inert materials
such as polyethylene.
7.3.4 Mixing Bulb—Bulb, to ensure adequate mixing of the permeated gas and the diluent gas stream. A Kjeldahl trap is
recommended.
7.3.5 Long Condenser, with large bore in which a thermometer and a permeation tube can be inserted.
NOTE 1—This system is constructed from readily available laboratory equipment.
NOTE 2—Warning—If the room temperature is significantly different from that of the water bath, a small difference in temperature between the bath
and the condensor containing the permeation tube can exist. In this event, the temperature indicated by the thermometer in the condensor should be used
as that of the permeation tube, rather than that of the water bath.Warning: If the room temperature is significantly different from that of the water bath,
a small difference in temperature between the bath and the condensor containing the permeation tube can exist. In this event, the temperature indicated
by the thermometer in the condensor should be used as that of the permeation tube, rather than that of the water bath.
FIG. 2 Typical System for Laboratory Use
7.3.6 Temperature Controlled Water Bath—About 8-L [2-gal] capacity, capable of 60.1 °C 60.1°C or better water temperature
control, with a variable temperature control range from about 15 to 35 °C, 35°C, preferably equipped with a positive displacement
type recirculating pump with at least 1-L/min liquid flow rate to supply water to the condenser.
7.3.7 Thermometer, ASTM No. 91C or equivalent, calibrated to 60.1 °C.60.1°C.
7.3.8 Mercury Barometer.
7.4 An alternate system is shown in Fig. 2. It has the advantage of lower uncertainty of the temperature of the permeation tube.
The required parts are described in the figure.
8. Reagents and Materials
8.1 Carrier Gas or Diluent Gas for Flow Over Permeation Tube—Cylinder of dry nitrogen or pure, dry air, or purified room
air (charcoal and drying agent—inert air mover).
8.1.1 Drier, indicating type and should be discarded wh
...










Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.