Standard Practice for Calibration of Routine Dosimetry Systems for Radiation Processing

SIGNIFICANCE AND USE
4.1 Ionizing radiation is used to produce various desired effects in products. Examples of applications include the sterilization of medical products, microbial reduction, modification of polymers and electronic devices, and curing of inks, coatings, and adhesives.
4.2 Absorbed-dose measurements, with statistical controls and documentation, are necessary to ensure that products receive the desired absorbed dose. These controls include a program that addresses requirements for calibration of routine dosimetry system.
4.3 A routine dosimetry system calibration procedure as described in this document provides the user with a dosimetry system whose dose measurements are traceable to national or international standards for the conditions of use (see Annex A4). The dosimetry system calibration is part of the user’s measurement management system.
SCOPE
1.1 This practice specifies the requirements for calibrating routine dosimetry systems for use in radiation processing, including establishing measurement traceability and estimating uncertainty in the measured dose using the calibrated dosimetry system.
NOTE 1—Regulations or other directives exist in many countries that govern certain radiation processing applications such as sterilization of healthcare products and radiation processing of food requiring that absorbed-dose measurements be traceable to national or international standards (ISO 11137-1, Refs (1-3)2).
1.2 The absorbed-dose range covered is up to 1 MGy.
1.3 The radiation types covered are photons and electrons with energies from 80 keV to 25 MeV.
1.4 This document is one of a set of standards that provides recommendations for properly implementing dosimetry in radiation processing, and describes a means of achieving compliance with the requirements of ASTM E2628 “Practice for Dosimetry in Radiation Processing” for the calibration of routine dosimetry systems. It is intended to be read in conjunction with ASTM E2628 and the relevant ASTM or ISO/ASTM standard practice for the dosimetry system being calibrated referenced in Section 2.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
28-Feb-2013
Technical Committee
Drafting Committee
Current Stage
Ref Project

Buy Standard

Standard
ASTM ISO/ASTM51261-13 - Standard Practice for Calibration of Routine Dosimetry Systems for Radiation Processing
English language
19 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
ISO/ASTM 51261:2013(E)
Standard Practice for
Calibration of Routine Dosimetry Systems for Radiation
1
Processing
This standard is issued under the fixed designation ISO/ASTM 51261; the number immediately following the designation indicates the
year of original adoption or, in the case of revision, the year of last revision.
1. Scope priate safety and health practices and determine the applica-
bility of regulatory limitations prior to use.
1.1 This practice specifies the requirements for calibrating
routine dosimetry systems for use in radiation processing,
2. Referenced documents
includingestablishingmeasurementtraceabilityandestimating
3
2.1 ASTM Standards:
uncertainty in the measured dose using the calibrated dosim-
E170Terminology Relating to Radiation Measurements and
etry system.
Dosimetry
NOTE 1—Regulations or other directives exist in many countries that
E178Practice for Dealing With Outlying Observations
govern certain radiation processing applications such as sterilization of
healthcare products and radiation processing of food requiring that E2628Practice for Dosimetry in Radiation Processing
absorbed-dose measurements be traceable to national or international
E2701Guide for Performance Characterization of Dosim-
2
standards (ISO 11137-1, Refs (1-3) ).
eters and Dosimetry Systems for Use in Radiation Pro-
1.2 The absorbed-dose range covered is up to 1 MGy.
cessing
3
2.2 ISO/ASTM Standards:
1.3 The radiation types covered are photons and electrons
51607Practice for Use of an Alanine-EPR Dosimetry Sys-
with energies from 80 keV to 25 MeV.
tem
1.4 This document is one of a set of standards that provides
51707Guide for Estimating Uncertainties in Dosimetry for
recommendations for properly implementing dosimetry in
Radiation Processing
radiation processing, and describes a means of achieving
2.3 International Commission on Radiation Units and Mea-
compliance with the requirements of ASTM E2628 “Practice 4
surements Reports:
for Dosimetry in Radiation Processing” for the calibration of
ICRU Report 85aFundamental Quantities and Units for
routinedosimetrysystems.Itisintendedtobereadinconjunc-
Ionizing Radiation
tion withASTM E2628 and the relevantASTM or ISO/ASTM 5
2.4 ISO Standards:
standard practice for the dosimetry system being calibrated
ISO 11137-1 Sterilization of health care products—
referenced in Section 2.
Radiation—Requirements for the development, validation
1.5 This standard does not purport to address all of the
and routine control of a sterilization process for medical
safety concerns, if any, associated with its use. It is the
devices
5
responsibility of the user of this standard to establish appro-
2.5 ISO/IEC Standards:
17025General Requirements for the Competence ofTesting
and Calibration Laboratories
1
This guide is under the jurisdiction of ASTM Committee E61 on Radiation
3
Processing and is the direct responsibility of Subcommittee E61.01 on Dosimetry, For referenced ASTM and ISO/ASTM standards, visit the ASTM website,
and is also under the jurisdiction of ISO/TC 85/WG 3. www.astm.org, or contact ASTM Customer Service at service@astm.org. For
Current edition approved Aug. 16, 2012. Published April 2013. Originally Annual Book of ASTM Standards volume information, refer to the standard’s
published asASTM E1261–88. Last previousASTM edition E1261–00.ASTM Document Summary page on the ASTM website.
ε1 4
E 1261–94 was adopted by ISO in 1998 with the intermediate designation ISO Available from International Commission on Radiation Units and
15556:1998(E). The present International Standard ISO/ASTM 51261:2013(E) is a Measurements, 7910 Woodmont Avenue, Suite 800, Bethesda, MD 20814, USA.
5
major revision of ISO/ASTM 51261:2002(E), which replaced ISO 15556. Available from International Organization for Standardization (ISO), 1, ch. de
2
The boldface numbers given in parentheses refer to the bibliography at the end la Voie-Creuse, Case postale 56, CH-1211, Geneva 20, Switzerland, http://
of this guide. www.iso.ch.
© ISO/ASTM International 2020 – All rights reserved
1

---------------------- Page: 1 ----------------------
ISO/ASTM 51261:2013(E)
2.6 Joint Committee for Guides in Metrology (JCGM) etry systems refers to irradiation of dosimeters along with
6
Reports: reference or transfer standard dosimeters, under operating
JCGM 100:2008, GUM 1995, with minor corrections, conditions that are representative of the routine processing
Evaluation of measurement data – Guide to the Expres- environment, for the purpose of developing a calibration curve
sion of Uncertainty in Measurement for the routine dosimetry systems.
3.1.11 measurand—specific quantity subject to measure-
3. Terminology
ment.
3.1 Definitions:
3.1.12 measurement management system—set of inter-
3.1.1 approved laboratory—laboratory that is a recognized
related or interacti
...