ASTM G131-96(2023)e1
(Practice)Standard Practice for Cleaning of Materials and Components by Ultrasonic Techniques
Standard Practice for Cleaning of Materials and Components by Ultrasonic Techniques
SIGNIFICANCE AND USE
5.1 This practice is suitable for the removal of contaminants found on materials, parts, and components used in systems requiring a high level of cleanliness, such as oxygen. Parts shall have been precleaned to remove visible contaminants prior to using this procedure. Softgoods such as seals and valve seats may be cleaned without precleaning.
5.2 This procedure may also be used as the cleanliness verification technique for coupons used during cleaning effectiveness tests as in Test Method G122.
5.3 The cleaning efficiency has been shown to vary with the frequency and power density of the ultrasonic unit. Low frequencies in the 20 kHz to 25 kHz range have been found to damage soft metals such as aluminum and silver. Therefore, the specifications of the unit and the frequencies available must be considered in order to optimize the cleaning conditions without damaging the parts.
SCOPE
1.1 This practice covers a procedure for the cleaning of materials and components used in systems requiring a high level of cleanliness, such as oxygen, by ultrasonic techniques.
1.2 This practice may be used for cleaning small parts, components, softgoods, etc.
1.3 The values stated in SI units are standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in Note 1.
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
- Status
- Published
- Publication Date
- 30-Jun-2023
- Drafting Committee
- G04.02 - Recommended Practices
Relations
- Effective Date
- 01-Apr-2020
- Effective Date
- 01-Apr-2020
- Effective Date
- 01-Apr-2020
- Effective Date
- 01-Apr-2020
- Effective Date
- 01-Dec-2018
- Effective Date
- 01-Nov-2013
- Effective Date
- 01-Jun-2013
- Effective Date
- 01-Nov-2012
- Effective Date
- 01-Sep-2008
- Effective Date
- 01-May-2008
- Effective Date
- 01-May-2008
- Effective Date
- 01-Mar-2006
- Effective Date
- 01-Nov-2005
- Effective Date
- 01-Sep-2004
- Effective Date
- 01-Sep-2004
Overview
ASTM G131-96(2023)e1: Standard Practice for Cleaning of Materials and Components by Ultrasonic Techniques provides a detailed procedure for effectively removing contaminants from materials, parts, and components using ultrasonic cleaning. Developed by ASTM, this international standard is especially relevant for industries requiring exceptionally high cleanliness, such as those involving oxygen systems. It outlines practices for both cleaning and verification of cleanliness, emphasizing the need for optimized ultrasonic parameters to avoid damage to sensitive materials.
Key Topics
- Ultrasonic Cleaning Process: The standard details how ultrasonic energy, combined with suitable cleaning agents, can remove both molecular and particulate contaminants from items. It includes recommendations on operating frequencies, cleaning agent selection, and process parameters.
- Contaminant Removal: Guidance is provided on cleaning surfaces to a high level of cleanliness, which is especially critical for applications where contaminants could affect component performance or safety.
- Material Compatibility: Special attention is given to the impact of ultrasonic cleaning on softer metals like aluminum and silver, highlighting the importance of selecting the right frequency and power density.
- Cleanliness Verification: The practice can be used as a verification method for cleaning effectiveness, using control or witness coupons in accordance with ASTM cleaning evaluation methods.
- Safety and Environmental Considerations: Users are reminded to establish proper safety, health, and environmental practices and be aware of relevant regulatory limitations before use.
Applications
The ultrasonic cleaning method outlined in ASTM G131-96(2023)e1 is applicable in diverse sectors, including:
- Oxygen Service Equipment: Ensures removal of contaminants from fittings, valves, and piping used in oxygen-enriched atmospheres, minimizing the risk of ignition or explosion.
- Aerospace and Medical Devices: Critical for cleaning small, complex parts and softgoods where residual contamination can compromise functionality or sterility.
- Analytical Laboratories: Supports sample and component preparation where high cleanliness is necessary for analytical accuracy.
- Manufacturing and Maintenance: Used in production lines for the cleaning of precision components and in maintenance operations where cleanliness directly affects equipment reliability.
The standard is relevant for cleaning various items, such as small mechanical parts, softgoods like seals and valve seats, and test coupons used in effectiveness evaluations.
Related Standards
For enhanced cleaning processes and verification, the following ASTM standards are referenced or relevant:
- ASTM G121: Practice for Preparation of Contaminated Test Coupons for the Evaluation of Cleaning Agents
- ASTM G122: Test Method for Evaluating the Effectiveness of Cleaning Agents and Processes
- ASTM F311: Practice for Processing Aerospace Liquid Samples for Particulate Contamination Analysis Using Membrane Filtration
- ASTM F324 / F331 / E1235: Test Methods for Determining Nonvolatile Residues of Solvent Extracts and Environmentally Controlled Areas
- ASTM D1193: Specification for Reagent Water
These documents, together with ASTM G131-96(2023)e1, provide a comprehensive framework for ensuring material cleanliness in high-reliability and safety-critical environments.
Keywords: ultrasonic cleaning, removal of contaminants, material cleanliness, ASTM G131, oxygen systems, cleaning verification, nonvolatile residue, cleaning agents, industrial cleaning standards, high cleanliness systems
Buy Documents
ASTM G131-96(2023)e1 - Standard Practice for Cleaning of Materials and Components by Ultrasonic Techniques
Get Certified
Connect with accredited certification bodies for this standard
Great Wall Tianjin Quality Assurance Center
Established 1993, first batch to receive national accreditation with IAF recognition.
Hong Kong Quality Assurance Agency (HKQAA)
Hong Kong's leading certification body.

Innovative Quality Certifications Pvt. Ltd. (IQCPL)
Known for integrity, providing ethical & impartial Assessment & Certification. CMMI Institute Partner.
Sponsored listings
Frequently Asked Questions
ASTM G131-96(2023)e1 is a standard published by ASTM International. Its full title is "Standard Practice for Cleaning of Materials and Components by Ultrasonic Techniques". This standard covers: SIGNIFICANCE AND USE 5.1 This practice is suitable for the removal of contaminants found on materials, parts, and components used in systems requiring a high level of cleanliness, such as oxygen. Parts shall have been precleaned to remove visible contaminants prior to using this procedure. Softgoods such as seals and valve seats may be cleaned without precleaning. 5.2 This procedure may also be used as the cleanliness verification technique for coupons used during cleaning effectiveness tests as in Test Method G122. 5.3 The cleaning efficiency has been shown to vary with the frequency and power density of the ultrasonic unit. Low frequencies in the 20 kHz to 25 kHz range have been found to damage soft metals such as aluminum and silver. Therefore, the specifications of the unit and the frequencies available must be considered in order to optimize the cleaning conditions without damaging the parts. SCOPE 1.1 This practice covers a procedure for the cleaning of materials and components used in systems requiring a high level of cleanliness, such as oxygen, by ultrasonic techniques. 1.2 This practice may be used for cleaning small parts, components, softgoods, etc. 1.3 The values stated in SI units are standard. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in Note 1. 1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
SIGNIFICANCE AND USE 5.1 This practice is suitable for the removal of contaminants found on materials, parts, and components used in systems requiring a high level of cleanliness, such as oxygen. Parts shall have been precleaned to remove visible contaminants prior to using this procedure. Softgoods such as seals and valve seats may be cleaned without precleaning. 5.2 This procedure may also be used as the cleanliness verification technique for coupons used during cleaning effectiveness tests as in Test Method G122. 5.3 The cleaning efficiency has been shown to vary with the frequency and power density of the ultrasonic unit. Low frequencies in the 20 kHz to 25 kHz range have been found to damage soft metals such as aluminum and silver. Therefore, the specifications of the unit and the frequencies available must be considered in order to optimize the cleaning conditions without damaging the parts. SCOPE 1.1 This practice covers a procedure for the cleaning of materials and components used in systems requiring a high level of cleanliness, such as oxygen, by ultrasonic techniques. 1.2 This practice may be used for cleaning small parts, components, softgoods, etc. 1.3 The values stated in SI units are standard. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in Note 1. 1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
ASTM G131-96(2023)e1 is classified under the following ICS (International Classification for Standards) categories: 03.080.10 - Maintenance services. Facilities management. The ICS classification helps identify the subject area and facilitates finding related standards.
ASTM G131-96(2023)e1 has the following relationships with other standards: It is inter standard links to ASTM F331-13(2020), ASTM E1235-12(2020), ASTM F311-08(2020), ASTM E1235-12(2020)e1, ASTM G121-18, ASTM F311-08(2013), ASTM F331-13, ASTM E1235-12, ASTM G122-96(2008), ASTM E1235-08, ASTM F311-08, ASTM D1193-06, ASTM F331-05, ASTM G121-98(2015)e1, ASTM G121-98(2010)e1. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ASTM G131-96(2023)e1 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
´1
Designation: G131 − 96 (Reapproved 2023)
Standard Practice for
Cleaning of Materials and Components by Ultrasonic
Techniques
This standard is issued under the fixed designation G131; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
ε NOTE—Editorially updated formatting in 7.1 in July 2023.
1. Scope Cleaning Solvents Using the Solvent Purity Meter (With-
drawn 1987)
1.1 This practice covers a procedure for the cleaning of
F331 Test Method for Nonvolatile Residue of Solvent Ex-
materials and components used in systems requiring a high
tract from Aerospace Components (Using Flash Evapora-
level of cleanliness, such as oxygen, by ultrasonic techniques.
tor)
1.2 This practice may be used for cleaning small parts,
G121 Practice for Preparation of Contaminated Test Cou-
components, softgoods, etc.
pons for the Evaluation of Cleaning Agents
G122 Test Method for Evaluating the Effectiveness of
1.3 The values stated in SI units are standard.
Cleaning Agents and Processes
1.4 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
3. Terminology
responsibility of the user of this standard to establish appro-
3.1 Definitions of Terms Specific to This Standard:
priate safety, health, and environmental practices and deter-
3.1.1 contaminant (contamination), n—unwanted molecular
mine the applicability of regulatory limitations prior to use.
and particulate matter that could affect or degrade the perfor-
Specific precautionary statements are given in Note 1.
mance of the components upon which they reside.
1.5 This international standard was developed in accor-
dance with internationally recognized principles on standard-
3.1.2 contaminate, v—a process of applying a contaminant.
ization established in the Decision on Principles for the
3.1.3 control coupon (witness coupon), n—a coupon made
Development of International Standards, Guides and Recom-
from the same material and prepared in exactly the same way
mendations issued by the World Trade Organization Technical
as the test coupons, which is used to verify the validity of the
Barriers to Trade (TBT) Committee.
method or part thereof.
3.1.3.1 Discussion—In this practice, the control coupon will
2. Referenced Documents
be contaminated in the same manner as the test coupons and
2.1 ASTM Standards:
will be subjected to the identical cleaning procedure.
D1193 Specification for Reagent Water
3.1.4 degas, v—the process of removing gases from a liquid.
E1235 Test Method for Gravimetric Determination of Non-
3.1.5 nonvolatile residue (NVR), n—residual molecular and
volatile Residue (NVR) in Environmentally Controlled
particulate matter remaining following the filtration and con-
Areas for Spacecraft
trolled evaporation of a solvent containing contaminants.
F311 Practice for Processing Aerospace Liquid Samples for
Particulate Contamination Analysis Using Membrane Fil- 3.1.6 particle (particulate contaminant), n—a piece of mat-
ters ter in a solid state with observable length, width, and thickness.
F324 Test Method for Nonvolatile Residue of Volatile
3.1.6.1 Discussion—The size of a particle is usually defined
by its greatest dimension and is specified in micrometres.
This practice is under the jurisdiction of ASTM Committee G04 on Compat-
4. Summary of Practice
ibility and Sensitivity of Materials in Oxygen Enriched Atmospheres and is the
direct responsibility of Subcommittee G04.02 on Recommended Practices. 4.1 A part, material or component is placed in a container
Current edition approved July 1, 2023. Published July 2023. Originally approved
containing the cleaning agent. This container is then placed in
ɛ1
in 1995. Last previous edition approved in 2016 as G131 – 96 (2016) . DOI:
an ultrasonic cleaner and treated for a given period of time at
10.1520/G0131-96R23E01.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on The last approved version of this historical standard is referenced on
the ASTM website. www.astm.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
´1
G131 − 96 (2023)
the recommended temperature for the cleaning agent. This safe handling procedures, particularly in open tanks. Many
results in a solution if the contaminant is soluble in the test solvents are not considered to be compatible with oxygen and
fluid or a emulsion if the contaminant is not soluble in the test must be completely removed from cleaned components prior to
fluid. The cleaning solution combined with the rinse solutions the use of these components in oxygen systems. The preferred
may then be analyzed for particulate, NVR, or total carbon method of removal shall be determined by the user.
(TC).
7.2 Purity of Reagents—Reagent grade chemicals shall be
4.1.1 In the case of aqueous-based agents, the parts are
used in all tests. Unless otherwise indicated, it is intended that
rinsed after the removal from the cleaning bath and ultrasoni-
all reagents conform to the specifications of the Committee on
cally cleaned in reagent water to provide a solution for TC
Analytical Reagents of the American Chemical Society where
analysis using G TC.
such specifications are available. Other grades may be used,
4.1.2 In the case of solvent-based agents, the parts are rinsed
provided it is first ascertained that the reagent is of sufficiently
with fresh solvent, which is collected and combined with the
high purity to permit its use without lessening the accuracy of
solvent used in the cleaning process, and the NVR determined
the determination. Detergents used shall be identified by
using Test Method E1235, Test Method F324, or Test Method
manufacturer and name (registered trademark, if any).
F331, as appropriate.
7.3 Purity of Water—The water used shall meet the require-
4.1.3 Particulate analyses may be performed by filtering the
ments of Specification D1193, Type II.
final cleaning solution. The particles captured by the filter are
then counted using Practice F311.
8. Procedure
5. Significance and Use
8.1 Sample Preparation:
8.1.1 If cleanliness verification is to be performed on control
5.1 This practice is suitable for the removal of contaminants
or test coupons, prepare the coupons in accordance with
found on materials, parts, and components used in systems
Practice G121.
requiring a high level of cleanliness, such as oxygen. Parts
8.1.2 If cleanliness verification is to be performed on small
shall have been precleaned to remove visible contaminants
parts, measure the total surface area (S) in square centimetres
prior to using this procedure. Softgoods such as seals and valve
or the mass in grams, or both, as applicable, to the nearest tenth
seats may be cleaned without precleaning.
of a milligram (M1). Record the surface area (S) and mass
5.2 This procedure may also be used as the cleanliness
(M1).
verification technique for coupons used during cleaning effec-
8.2 Preliminary Procedure:
tiveness tests as in Test Method G122.
8.2.1 If a cleaning agent is being used that requires dilution
5.3 The cleaning efficiency has been shown to vary with the
or special preparation, carefully follow the manufacturer’s
frequency and power density of the ultrasonic unit. Low
instructions. Use Type II water to prepare the aqueous cleaning
frequencies in the 20 kHz to 25 kHz range have been found to
agent solutions or as the actual cleaning agent.
damage soft metals such as aluminum and silver. Therefore, the
NOTE 2—It has been found that many common hydrocarbon based
specifications of the unit and the frequencies available must be
lubricants are effectively removed to acceptable levels using Type II water
considered in order to optimize the cleaning conditions without
at 50 °C to 55 °C. Contaminants more difficult to remove, such as
damaging the parts.
fluorinated or silicone based lubricants, have typically been found to
require the use of surface active agents. Use Test Method G122 to evaluate
6. Apparatus
the cleaning effectiveness of the proposed cleaning agent.
6.1 Ultrasonic Cleaner, with an operating frequency range
8.2.2 Fill the ultrasonic bath to the level specified by the
between 25 kHz and 90 kHz, a typical power range between
manufacturer with water. Place the support bracket in the
10 W ⁄L and 25 W ⁄L, and a temperature-controlled bath ca-
ultrasonic bath, heat the ultrasonic bath to the desired
pable of maintaining a temperature between ambient and 70 °C
temperature, and degas the water for 10 min.
with an accuracy of 2 °C.
8.2.3 Clean the stainless steel sample parts pan to be used.
Conduct the sampling procedure using the selected cleaning
6.2 Parts Pans, stainless steel container with volumes
agent without parts to verify the cleanliness of the parts pan.
between 1 L and 4 L.
Use the same sampling and analysis procedures that will be
6.3 Bracket, stainless steel device capable of supporting the
used on the actual parts. Determine the contamination level of
parts pans in the ultrasonic bath.
the parts pan, the blank value (B), which shall be less than the
NOTE 1—The bracket should be designed to hang in the ultrasonic bath allowable contamination level for the items being cleaned or
without contact with the bottom.
extracted. If the contamination level of the parts pan is greater
than that specified for the parts, reclean the parts pan until the
7. Reagents
7.1 Solvents such as the following may be used: tetrachlo-
roethylene (perchloroethylene), trichloroethylene, methylene
ACS Reagent Chemicals, Specifications and Procedures for Reagents and
Standard-Grade Reference Materials, American Chemical Society, Washington,
chloride, and perfluorinated carbon fluids. Warning—Solvents
DC. For suggestions on the testing of reagents not listed by the American Chemical
such as tetrachloroethylene (perchloroethylene),
Society, see Analar Standards for Laboratory Chemicals, BDH Ltd., Poole, Dorset,
trichloroethylene, and methylene chloride have relative low
U.K., and the United States Pharmacopeia and National Formulary, U.S. Pharma-
threshold limit values and the user should refer to appropriate copeial Convention, Inc. (USPC), Rockville, MD.
´1
G131 − 96 (2023)
contamination level is less than the allowable contamination 8.4.4 Determine and record the mass (M ) of the nonvolatile
specified for the parts. residue in milligrams to the nearest tenth of a milligram using
Test Method E1235, Test Methods F324, or F331. Ensure that
8.3 Cleaning Procedure:
the reported NVR is adjusted to subtract the NVR of an
8.3.1 Place the material or part(s) being cleaned in the
equivalent volume of “blank” solvent.
stainless steel parts pan.
8.5 Sampling Procedure for Aqueous Cleaned Materials
8.3.2 Pour a measured amount of the cleaning agent into the
and Parts
stainless steel cleaning pan sufficient to cover the parts. Cover
8.5.1 Remove the parts pan from the ultrasonic bath and
the parts pan with aluminum foil or a stainless steel lid, place
remove the cover. Swirl the parts pan to mix the Type II water.
the parts pan in the bracket in the ultrasonic bath, adjust the
8.5.2 After swirling, quickly decant the Type II water from
water level in the bath such that it is above the cleaning agent
the parts pan.
level in the parts pan, an
...