SIST prEN 1518:2027
(Main)Non-destructive testing - Leak testing - Characterization of mass spectrometer leak detectors
- Abstract
This document defines terms and specifies procedures for the characterization of mass spectrometer leak detectors (MSLD). This document is not intended to provide a complete set of specifications for acceptance testing, but rather a description of procedures that can be applied without specialized calibration equipment.
The methods described in this document are applicable without restrictions to helium as the tracer gas. For other gases, additional precautions might be necessary.
These methods are applicable to commonly available MSLD, based on the present level of technology, which might be able to measure leakage rates down to 10−12 Pa⋅m3/s.
- Status
- Not Published
- Public Enquiry End Date
- 30-Jan-2025
- Technical Committee
- PKG - Testing of metallic materials
- Current Stage
- 5020 - Formal vote (FV) (Adopted Project)
- Start Date
- 17-Aug-2026
- Due Date
- 05-Oct-2026
- Completion Date
- 18-Aug-2026
SIST prEN 1518:2027 is a draft European standard for non-destructive testing and leak testing, published through SIST and developed by CEN. It specifies the terms and procedures used to characterize mass spectrometer leak detectors (MSLD), with helium as the tracer gas covered without restriction.
The document is aimed at manufacturers, test laboratories, and users who need to describe, compare, or verify MSLD performance. It is for characterization rather than full acceptance testing, and the procedures are designed to be usable without specialized calibration equipment.
What does SIST prEN 1518:2027 specify?
SIST prEN 1518:2027 specifies the vocabulary, test arrangements, operating conditions, and evaluation methods used to characterize mass spectrometer leak detectors. It explains what to measure, how to set up the instrument, and how to report the results.
The document is organized into:
- scope
- normative references
- terms and definitions
- description of an MSLD
- apparatus for characterization methods
- reference conditions
- test procedures
- results
- test report
The scope is limited to characterization. The methods are intended for commonly available MSLD, and they are described for helium tracer gas; other gases may require extra precautions.
What are the key requirements of SIST prEN 1518:2027?
SIST prEN 1518:2027 focuses on three practical questions: what an MSLD is, how to test it, and how to express the results. For users, this means the standard is not just descriptive - it tells you which operating conditions, reference conditions, and measurement records have to be controlled to get meaningful characterization data.
MSLD description and operating principles
Clause 4 describes the main parts of an MSLD: the mass spectrometer, the pumping system, controls, and the display. It also distinguishes direct flow and counterflow detectors, and gives formulas linking leakage rate to tracer-gas throughput, pumping speed, and compression ratio.
In practice, this matters because the calculation route depends on the detector design. A test lab needs to know whether the instrument is direct flow or counterflow before it can interpret the output correctly.
Apparatus and test arrangement
Clause 5 specifies the apparatus needed for characterization. It calls for a small known tracer-gas leak for minimum detectable leakage rate, a large air leak for minimum detectable concentration, a leak isolation valve, and a test vessel for inlet pumping speed.
The test vessel is described as a known-volume vacuum chamber, preferably cylindrical or spherical, with suitable flanges and metal construction. This matters because the vessel volume and conductance affect the time constant and pumping-speed evaluation.
Preparation and reference conditions
Clause 5 also sets preparation rules: the instrument must be powered according to the manufacturer’s specification, warmed up, adjusted for optimum tracer-gas detection, and held at constant pumping speed or constant compression ratio if those settings are adjustable.
Clause 6 gives reference conditions for ambient temperature, pressure, and humidity. If other conditions are used, they have to be stated in the report. This is important because MSLD behavior and leak-leak response can vary with the environment.
Test procedures near the detection limit
Clause 7 sets the procedures for measuring background drift and noise, auxiliary pumping system background, sensitivity, time constant, minimum detectable concentration, and inlet pumping speed for tracer gas.
The method starts by characterizing the background signal and then measuring sensitivity with a known leak. That sequence matters because a detector cannot be judged at its limit unless the background signal is known first.
Evaluation and expression of results
Clause 8 explains how to evaluate the records and calculate the final values. It uses regression-based evaluation for the minimum detectable leakage rate, correction for auxiliary-system background, and formulas for minimum detectable concentration and inlet pumping speed.
In practice, this means the report is not based on a single reading. The standard requires a processed result derived from the recorded signal, the sensitivity, and the time constant.
Reporting
Clause 9 lists the minimum content of the test report: reference to the document, the method used, test results with reference conditions, tracer gas, measurement uncertainty, mode of operation, deviations, unusual features, and the test date.
For buyers and quality teams, this makes the result traceable and comparable between instruments and laboratories.
What terms does SIST prEN 1518:2027 define?
- Mass spectrometer leak detector (MSLD) - An instrument that uses a mass spectrometer and a high-vacuum system to detect tracer-gas leakage.
- Mass spectrometer (MS) - The sensing element that measures the partial pressure of tracer gas in the detector.
- Compression ratio - The ratio used for counterflow detectors to relate tracer-gas partial pressure at the pump exhaust to that at the mass spectrometer inlet.
- Inlet pumping speed - The flow rate at the test port when the instrument operates below the MS maximum working pressure; it influences response time.
- Optimum working pressure - The MS pressure at which the minimum detectable concentration can be measured.
- Maximum inlet pressure - The highest test-port pressure at which the MSLD can still detect leaks in a given operating mode.
- Background signal drift - The slow change in background signal over time.
- Minimum detectable concentration ratio - The smallest tracer-gas concentration in an air mixture that can be detected unambiguously at optimum working pressure.
Who uses SIST prEN 1518:2027?
SIST prEN 1518:2027 is used by organizations that manufacture, calibrate, or operate mass spectrometer leak detectors. It is also relevant to non-destructive testing labs, vacuum-technology users, and quality teams that need to compare detector performance.
Typical tasks include:
- characterizing a new or repaired MSLD
- checking background drift and noise
- verifying sensitivity and time constant
- determining minimum detectable leakage rate
- determining minimum detectable concentration
- documenting inlet pumping speed for tracer gas
- preparing test reports for internal qualification or supplier assessment
What changed in SIST prEN 1518:2027 from the previous edition?
The document replaces EN 1518:1998. The foreword states these main technical changes:
- normative references were updated
- the terms and definitions clause was revised
- a new general subclause was added in Clause 4
- Clause 5.2 was updated
- the ambient temperature in Clause 6.1 was adapted
- Clauses 7 and 8 were revised
- the test report was revised
- the figures were integrated into the text
Which standards are used with SIST prEN 1518:2027?
The normative reference named in the document is:
- EN ISO 20484:2017 - provides the vocabulary for leak testing, so the terms used in this document align with the wider NDT terminology.
The document also refers to:
- ISO 3530 - used for the definition of 63 % rise time when evaluating the time constant.
What does the SIST prEN 1518:2027 document contain?
The document contains operating descriptions, test setups, calculation formulas, and reporting requirements for MSLD characterization. It includes figures for direct-flow and counterflow arrangements, a vessel arrangement for inlet pumping-speed determination, and a flowchart for result evaluation.
It also includes the practical procedures for background drift and noise measurements, sensitivity checks, tracer-gas concentration evaluation, and time-constant determination. The formulas are used to convert recorded signals into minimum detectable leakage rate, minimum detectable concentration, and inlet pumping speed.
For test work, the most useful parts are the apparatus description in Clause 5, the procedures in Clause 7, and the evaluation rules in Clause 8. These sections show how the detector should be arranged, how the signal should be recorded, and how the final performance values are derived.
Relations
- Effective Date
- 20-Dec-2023
Get Certified
Connect with accredited certification bodies for this standard

IMP NDT d.o.o.
Non-destructive testing services. Radiography, ultrasonic, magnetic particle, penetrant, visual inspection.

Inštitut za kovinske materiale in tehnologije
Institute of Metals and Technology. Materials testing, metallurgical analysis, NDT.

Q Techna d.o.o.
NDT and quality assurance specialist. 30+ years experience. NDT personnel certification per ISO 9712, nuclear and thermal power plant inspections, QA/
Sponsored listings
Frequently Asked Questions
SIST prEN 1518:2027 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Non-destructive testing - Leak testing - Characterization of mass spectrometer leak detectors". This standard covers: This document defines terms and specifies procedures for the characterization of mass spectrometer leak detectors (MSLD). This document is not intended to provide a complete set of specifications for acceptance testing, but rather a description of procedures that can be applied without specialized calibration equipment. The methods described in this document are applicable without restrictions to helium as the tracer gas. For other gases, additional precautions might be necessary. These methods are applicable to commonly available MSLD, based on the present level of technology, which might be able to measure leakage rates down to 10−12 Pa⋅m3/s.
This document defines terms and specifies procedures for the characterization of mass spectrometer leak detectors (MSLD). This document is not intended to provide a complete set of specifications for acceptance testing, but rather a description of procedures that can be applied without specialized calibration equipment. The methods described in this document are applicable without restrictions to helium as the tracer gas. For other gases, additional precautions might be necessary. These methods are applicable to commonly available MSLD, based on the present level of technology, which might be able to measure leakage rates down to 10−12 Pa⋅m3/s.
SIST prEN 1518:2027 is classified under the following ICS (International Classification for Standards) categories: 19.100 - Non-destructive testing. The ICS classification helps identify the subject area and facilitates finding related standards.
SIST prEN 1518:2027 has the following relationships with other standards: It is inter standard links to SIST EN 1518:2000. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
SIST prEN 1518:2027 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
SLOVENSKI STANDARD
oSIST prEN 1518:2025
01-januar-2025
Neporušitveno preskušanje - Preskus tesnosti – Karakterizacija masno-
spektrometričnih detektorjev puščanja
Non-destructive testing - Leak testing - Characterization of mass spectrometer leak
detectors
Zerstörungsfreie Prüfung - Dichtheitsprüfung - Charakterisierung von
massenspektrometrischen Leckdetektoren
Essais non destructifs - Contrôle d’étanchéité - Caractérisation des détecteurs de fuite à
spectrométrie de masse
Ta slovenski standard je istoveten z: prEN 1518
ICS:
19.100 Neporušitveno preskušanje Non-destructive testing
oSIST prEN 1518:2025 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
oSIST prEN 1518:2025
oSIST prEN 1518:2025
DRAFT
EUROPEAN STANDARD
prEN 1518
NORME EUROPÉENNE
EUROPÄISCHE NORM
November 2024
ICS 19.100 Will supersede EN 1518:1998
English Version
Non-destructive testing - Leak testing - Characterization of
mass spectrometer leak detectors
Essais non destructifs - Contrôle d'étanchéité - Zerstörungsfreie Prüfung - Dichtheitsprüfung -
Caractérisation des détecteurs de fuite à spectrométrie Charakterisierung von massenspektrometrischen
de masse Leckdetektoren
This draft European Standard is submitted to CEN members for enquiry. It has been drawn up by the Technical Committee
CEN/TC 138.
If this draft becomes a European Standard, CEN members are bound to comply with the CEN/CENELEC Internal Regulations
which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
This draft European Standard was established by CEN in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC
Management Centre has the same status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and
United Kingdom.
Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are
aware and to provide supporting documentation.
Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without
notice and shall not be referred to as a European Standard.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2024 CEN All rights of exploitation in any form and by any means reserved Ref. No. prEN 1518:2024 E
worldwide for CEN national Members.
oSIST prEN 1518:2025
prEN 1518:2024 (E)
Contents Page
European foreword . 4
1 Scope . 5
2 Normative references . 5
3 Terms and definitions . 5
Terms . 5
3.1.1 Terms related to the MSLD . 5
3.1.2 Terms related to the operation of the MSLD . 6
3.1.3 Terms related to the specification of the MSLD . 7
3.1.4 Terms relating to the tracer gas background signal . 7
3.1.5 Terms relating to the detection limit. 7
3.1.6 Terms relating to the display resolution . 7
4 Description of an MSLD . 8
General. 8
Main parts of an MSLD . 8
Operating principles of MSLD . 9
4.3.1 Direct flow leak detectors . 9
4.3.2 Counterflow leak detectors . 9
MLSD specification . 9
5 Apparatus for methods of characterization . 10
Leaks . 10
5.1.1 General. 10
5.1.2 Small tracer gas leaks . 10
5.1.3 Large air leak . 10
Leak isolation valve. 10
Test vessel . 11
Arrangement for test . 12
5.4.1 Auxiliary pumping system . 12
5.4.2 Materials . 13
Preparation for test . 13
5.5.1 Power supply . 13
5.5.2 Warming up . 13
5.5.3 Adjustment . 13
5.5.4 Constant pumping speed . 13
5.5.5 Constant compression ratio of the high vacuum pump . 13
6 Reference conditions for characterization . 13
Reference temperature . 13
Reference pressure . 13
Reference humidity . 13
7 Test procedures . 14
General. 14
Minimum detectable leakage rate . 14
7.2.1 Background drift and noise . 14
7.2.2 Auxiliary pumping system background . 14
7.2.3 Sensitivity . 14
oSIST prEN 1518:2025
prEN 1518:2024 (E)
7.2.4 Time constant . 15
Minimum detectable concentration . 15
7.3.1 General . 15
7.3.2 Background drift and noise . 17
7.3.3 Background signal . 17
7.3.4 Sensitivity . 17
Inlet pumping speed for tracer gas . 18
8 Results . 18
Minimum detectable leakage rate . 18
8.1.1 Evaluation. 18
8.1.2 Expression of minimum detectable leakage rate . 20
Minimum detectable concentration . 20
8.2.1 Evaluation. 20
8.2.2 Expression of minimum detectable concentration . 21
Inlet pumping speed . 21
8.3.1 Evaluation. 21
8.3.2 Expression of inlet pumping speed . 22
9 Test report . 22
Bibliography . 24
oSIST prEN 1518:2025
prEN 1518:2024 (E)
European foreword
This document (prEN 1518:2024) has been prepared by Technical Committee CEN/TC 138 “Non-
destructive testing”, the secretariat of which is held by AFNOR.
This document is currently submitted to the CEN Enquiry.
This document will supersede EN 1518:1998.
prEN 1518:2024 includes the following significant technical changes with respect to EN 1518:1998:
a) normative references have been updated;
b) clause “Terms and definitions” has been revised;
c) new Subclause 4.1, General, has been added;
d) Clause 5.2 has been updated;
e) in Clause 6.1, the ambient temperature has been adapted;
f) Clause 7 and 8 have been revised;
g) the test report has been revised;
h) the figures have been integrated in the text.
oSIST prEN 1518:2025
prEN 1518:2024 (E)
1 Scope
This document specifies terms and procedures for the characterization of mass spectrometer leak
detectors (MSLD). It is not intended to give a complete set of specifications for an acceptance test but a
description of procedures that can be used without particular calibration equipment.
The methods described in this document are applicable without restrictions to helium as the tracer gas.
For other gases, additional precautions may be necessary.
These methods are applicable to commonly available MSLD, based on the present level of technology,
−12 3
which may be able to measure leakage rates down to 10 Pa⋅m /s.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
EN ISO 20484:2017, Non-destructive testing — Leak testing — Vocabulary (ISO 20484:2017)
3 Terms and definitions
For the purposes of this document, the terms and definitions given in EN ISO 20484:2017 and the
following apply.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
• IEC Electropedia: available at https://www.electropedia.org/
• ISO Online browsing platform: available at https://www.iso.org/obp
Terms
3.1.1 Terms related to the MSLD
3.1.1.1
compression ratio
ratio of partial pressure of tracer gas with zero flow at the exhaust of the counterflow stages of the high
vacuum pump to the partial pressure at the inlet of the mass spectrometer (MS)
3.1.1.2
display
device which indicates visually the leakage rate measured
Note 1 to entry: The units in which the leakage rate is expressed can be selectable.
Note 2 to entry: The display can be analogue or quasi-analogue (continuous scale with definite divisions) or digital
(numbers with a definite number of digits) or a combination of both.
3.1.1.3
inlet pumping speed
volume rate of flow at the test port of an MSLD when the instrument is operating with the MS below its
maximum working pressure
Note 1 to entry: The inlet pumping speed can be different for different gases and different modes of operation. The
inlet pumping speed for the tracer gas determines the response time for the volume under test.
oSIST prEN 1518:2025
prEN 1518:2024 (E)
3.1.1.4
internal leak port
flange directly behind the inlet valve, used to connect a small leak for zero drift determination
3.1.1.5
intrinsic pumping speed
volume rate of flow of tracer gas at the inlet of the MS in a direct flow leak detector
3.1.1.6
ion collector
part of the MS where ions are collected and neutralized producing a current in the collector which is a
measure of the number of neutralized ions
3.1.1.7
sensitivity control
electrical hardware or software control which may be used to adjust the sensitivity of the instrument so
that a calibrated leak is indicated with its true leakage rate
3.1.1.8
sensitivity
ratio of the ion current at the output of the MS to the corresponding partial pressure of tracer gas inside
the MS
3.1.1.9
zero control
electrical hardware or software control which may be used to shift the output indication of the leak
detector, to a determined point of the scale range in use, usually zero
3.1.2 Terms related to the operation of the MSLD
3.1.2.1
peak
trace showing a maximum when the leak detector is scanned with respect to mass with gas
present, usually the tracer gas, to which the detector is tuned
3.1.2.2
peak
to set the scanning control (see 3.1.2.3) of a leak detector so that the output due to a given tracer
gas input is maximized
Note 1 to entry: It is a form of tuning (see 3.1.2.4).
3.1.2.3
scan
to adjust the accelerating voltage (or other equivalent operating parameter) of a leak detector,
particularly across that range of voltage which includes the voltage necessary to produce a tracer gas
peak
3.1.2.4
tune
in leak detection, to adjust one or more of the controls of a leak detector so that its response to a
tracer gas is maximized
Note 1 to entry: Tuning by means of the scanning control only is called “peaking” (see 3.1.2.2).
oSIST prEN 1518:2025
prEN 1518:2024 (E)
3.1.3 Terms related to the specification of the MSLD
3.1.3.1
optimum working pressure
pressure in the MS, at which the minimum detectable concentration can be measured
3.1.3.2
maximum working pressure
pressure in the MS above which operation is no longer possible
3.1.3.3
maximum inlet pressure
maximum pressure at the test port at which the MSLD is able to detect leaks in a given mode of operation
Note 1 to entry: For an MSLD to be connected directly to a system under test, the total pressure in the system shall
be less than the maximum inlet pressure of the MSLD.
3.1.3.4
maximum gas load
maximum pV-throughput of all gases emerging from the test specimen that the MSLD can pump during
leak detection in a given mode of operation
Note 1 to entry: For component testing, the MSLD is normally ready for measurement when the desorption of
water vapour from the inner surfaces is less than the maximum gas load.
3.1.4 Terms relating to the tracer gas background signal
3.1.4.1
background signal drift
relatively slow change in background signal, given by the maximum change in a given period of time
3.1.4.2
background signal noise
relatively rapid change in background signal given by an average measure of scatter in a specified period
of time
3.1.5 Terms relating to the detection limit
3.1.5.1
minimum detectable concentration ratio
smallest concentration of a given tracer gas in an air mixture that can be detected unambiguously when
the mixture is fed into the MSLD at such a rate as that is at its optimum working pressure
3.1.5.2
resolving power
ratio of a given mass number to the peak width measured at a specified (for example 10 %) height of the
peak (in units of mass numbers)
3.1.6 Terms relating to the display resolution
3.1.6.1
display resolution
quantitative expression of the ability of the leakage rate-display device to distinguish meaningfully
between closely adjacent values of the leakage rate indicated
oSIST prEN 1518:2025
prEN 1518:2024 (E)
3.1.6.2
linear display resolution
constant difference between adjacent scale intervals expressed in % full scale indication
3.1.6.3
logarithmic display resolution
constant ratio between two adjacent scale intervals expressed as a percentage of the indicated value
4 Description of an MSLD
General
An MSLD has an integral high vacuum system for maintaining the sensing element (mass spectrometer)
at low operating pressure and for establishing a partial pressure related to the incoming gas flow. This
pressure is measured quantitatively by the mass spectrometer. Such instruments are able to selectively
measure the flow of a tracer gas. In most cases the tracer gas will be helium, flowing in/out of an object
through a leak.
MSLD are commonly operated in either direct or counterflow configurations. These are both illustrated
in Figure 2.
Main parts of an MSLD
An MSLD (see definition given in EN ISO 20484) consists basically of a mass spectrometer (MS) and a
high vacuum pumping system for:
— maintaining the MS under appropriate vacuum conditions;
— producing a definite partial pressure of tracer gas when a specific throughput of tracer gas enters the
leak detector.
An MSLD includes also a number of valves and pressure gauges to ensure the appropriate vacuum
conditions within the system.
The leakage rate output can be displayed in a number of ways, for example an electrical meter, or digital
displays of different types, such as with a logarithmic or linear display resolution. In addition, a digital
and/or analogue output is usually available, which shall be used for the test procedures in this document.
The general structure of an MSLD is described by the following list:
a) mass spectrometer:
— ion source;
— separation system;
— ion collector.
b) pumping system:
— inlet system;
— inlet line;
— inlet valve;
— pump valve;
oSIST prEN 1518:2025
prEN 1518:2024 (E)
— test port;
— vent valve;
— internal leak port.
c) sensitivity zero control.
d) display.
Operating principles of MSLD
4.3.1 Direct flow leak detectors
The leakage rate is given by the formula:
q
x
(1)
q pS⋅
x xx
where
is the pV-throughput of tracer gas;
q
x
is the intrinsic pumping speed;
S
x
is the partial pressure of tracer
...