ISO/FDIS 10555-5
(Main)Intravascular catheters — Sterile and single-use catheters — Part 5: Over-needle peripheral intravenous catheters
- Abstract
ISO 10555-5:2013 specifies requirements for over-needle peripheral intravascular catheters, intended for accessing the peripheral vascular system, supplied in the sterile condition and intended for single use.
- Status
- Not Published
- Technical Committee
- ISO/TC 84 - Devices for administration of medicinal products and catheters
- Drafting Committee
- ISO/TC 84/WG 9 - Catheters
- Current Stage
- 5000 - FDIS registered for formal approval
- Start Date
- 29-Sep-2026
- Completion Date
- 15-Oct-2026
Buy Documents
ISO/FDIS 10555-5 - Intravascular catheters — Sterile and single-use catheters — Part 5: Over-needle peripheral intravenous catheters
REDLINE ISO/FDIS 10555-5 - Intravascular catheters — Sterile and single-use catheters — Part 5: Over-needle peripheral intravenous catheters
Relations
- Effective Date
- 12-Feb-2026
- Consolidates
ISO 4530:2022 - Vitreous and porcelain enamelled manufactured articles — Determination of resistance to heat - Effective Date
- 20-Oct-2025
- Effective Date
- 04-May-2024
Buy Documents
ISO/FDIS 10555-5 - Intravascular catheters — Sterile and single-use catheters — Part 5: Over-needle peripheral intravenous catheters
REDLINE ISO/FDIS 10555-5 - Intravascular catheters — Sterile and single-use catheters — Part 5: Over-needle peripheral intravenous catheters
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
ISO/FDIS 10555-5 is a draft published by the International Organization for Standardization (ISO). Its full title is "Intravascular catheters — Sterile and single-use catheters — Part 5: Over-needle peripheral intravenous catheters". This standard covers: ISO 10555-5:2013 specifies requirements for over-needle peripheral intravascular catheters, intended for accessing the peripheral vascular system, supplied in the sterile condition and intended for single use.
ISO 10555-5:2013 specifies requirements for over-needle peripheral intravascular catheters, intended for accessing the peripheral vascular system, supplied in the sterile condition and intended for single use.
ISO/FDIS 10555-5 is classified under the following ICS (International Classification for Standards) categories: 11.040.25 - Syringes, needles an catheters. The ICS classification helps identify the subject area and facilitates finding related standards.
ISO/FDIS 10555-5 has the following relationships with other standards: It is inter standard links to prEN ISO 10555-5, ISO 4530:2022, ISO 10555-5:2013. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ISO/FDIS 10555-5 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
FINAL DRAFT
International
Standard
ISO/TC 84
Intravascular catheters — Sterile
Secretariat: DS
and single-use catheters —
Voting begins on:
2026-12-03
Part 5:
Over-needle peripheral intravenous
Voting terminates on:
2027-01-14
catheters
Cathéters intravasculaires — Cathéters stériles et non
réutilisables —
Partie 5: Cathéters intraveineux périphériques à aiguille interne
RECIPIENTS OF THIS DRAFT ARE INVITED TO SUBMIT,
WITH THEIR COMMENTS, NOTIFICATION OF ANY
RELEVANT PATENT RIGHTS OF WHICH THEY ARE AWARE
AND TO PROVIDE SUPPOR TING DOCUMENTATION.
IN ADDITION TO THEIR EVALUATION AS
BEING ACCEPTABLE FOR INDUSTRIAL, TECHNO-
ISO/CEN PARALLEL PROCESSING LOGICAL, COMMERCIAL AND USER PURPOSES, DRAFT
INTERNATIONAL STANDARDS MAY ON OCCASION HAVE
TO BE CONSIDERED IN THE LIGHT OF THEIR POTENTIAL
TO BECOME STAN DARDS TO WHICH REFERENCE MAY BE
MADE IN NATIONAL REGULATIONS.
Reference number
FINAL DRAFT
International
Standard
ISO/TC 84
Intravascular catheters — Sterile
Secretariat: DS
and single-use catheters —
Voting begins on:
Part 5:
Over-needle peripheral intravenous
Voting terminates on:
catheters
Cathéters intravasculaires — Cathéters stériles et non
réutilisables —
Partie 5: Cathéters intraveineux périphériques à aiguille interne
© ISO 2026
All rights reserved.
RECIPIENTS OF THIS DRAFT ARE INVITED TO SUBMIT,
WITH THEIR COMMENTS, NOTIFICATION OF ANY
This ISO publication is protected by copyright and is owned by ISO and/or its licensors.
RELEVANT PATENT RIGHTS OF WHICH THEY ARE AWARE
The content of this ISO publication is provided under licence, not sold. Use is subject to the applicable licence terms issued by ISO,
AND TO PROVIDE SUPPOR TING DOCUMENTATION.
an ISO member body, or an authorized third-party distributor.
IN ADDITION TO THEIR EVALUATION AS
BEING ACCEPTABLE FOR INDUSTRIAL, TECHNO-
Except as required for implementation or expressly permitted by a separate licence, no part of this ISO publication may be
ISO/CEN PARALLEL PROCESSING
LOGICAL, COMMERCIAL AND USER PURPOSES, DRAFT
reproduced, distributed, modified, used, or made available in any form or by any means – electronic or mechanical, including
INTERNATIONAL STANDARDS MAY ON OCCASION HAVE
photocopying, scanning, recording, or posting on internal or external digital platforms. TO BE CONSIDERED IN THE LIGHT OF THEIR POTENTIAL
TO BECOME STAN DARDS TO WHICH REFERENCE MAY BE
Any use beyond the scope of the granted rights is prohibited and may result in legal action.
MADE IN NATIONAL REGULATIONS.
ISO copyright office
CP 401 • CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
Reference number
ii
Contents Page
Foreword .iv
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 Requirements . 3
4.1 General .3
4.2 Multilumen catheters .3
4.3 Physical requirements .3
4.3.1 Colour code .3
4.3.2 Catheter .4
4.3.3 Needle .4
4.3.4 Vent fitting .5
4.3.5 Vessel access indication .5
4.4 Information to be supplied by the manufacturer .5
Annex A (normative) Determination of strength of union of needle hub and needle tube . 6
Annex B (informative) Colours for opaque catheter hubs . 7
Annex C (informative) Needle point geometries . 8
Annex D (normative) Determination of liquid leakage from vent fitting . 9
Annex E (informative) Rationale and guidance .13
Bibliography .15
iii
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee
has been established has the right to be represented on that committee. International organizations,
governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely
with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types
of ISO documents should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent
rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a)
patent(s) which may be required to implement this document. However, implementers are cautioned that
this may not represent the latest information, which may be obtained from the patent database available at
www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO’s adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 84, Devices for administration of medicinal
products and catheters, in collaboration with the European Committee for Standardization (CEN) Technical
Committee CEN/TC 205, Non-active medical devices, in accordance with the Agreement on technical
cooperation between ISO and CEN (Vienna Agreement).
This third edition cancels and replaces the second edition (ISO 10555-5:2013), which has been technically
revised.
The main changes are as follows:
― scope modified to be only applicable to over-needle peripheral intravenous catheters;
― term “female” changed to “socket” to align with ISO 80369-7;
― new Annex E added to provide rationale and guidance for changes compared to the previous version;
― new test method option added in Annex D;
― modification made to test fluid preparation in Annex D;
― new requirement added for vessel access indication;
― changes made to upper limit of the distance between the heel of the needle point and the tip of the
catheter;
― added new terms and definitions.
A list of all parts in the ISO 10555 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.
iv
FINAL DRAFT International Standard ISO/FDIS 10555-5:2026(en)
Intravascular catheters — Sterile and single-use catheters —
Part 5:
Over-needle peripheral intravenous catheters
1 Scope
This document specifies requirements for over-needle peripheral intravenous catheters, intended for
accessing the peripheral venous system, supplied in the sterile condition and intended for single use.
NOTE 1 Attention is drawn to ISO 11070, which specifies requirements for accessory devices for use with
intravenous catheters, and to ISO 14972, which specifies requirements for sterile obturators for use with over-needle
peripheral intravenous catheters.
NOTE 2 See Annex E for additional information.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes
requirements of this document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
ISO 80369-7, Small-bore connectors for liquids and gases in healthcare applications — Part 7: Connectors for
intravascular or hypodermic applications
ISO 9626, Stainless steel needle tubing for the manufacture of medical devices
ISO 10555-1, Intravascular catheters — Sterile and single-use catheters — Part 1: General requirements
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 10555-1 and the following apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at https:// www .electropedia .org/
3.1
over-needle peripheral intravenous catheter
catheter designed for the introduction or withdrawal of liquids or devices into or from the peripheral venous
system
Note 1 to entry: See Figure 1.
3.2
needle
assembly comprising at least a needle tube (3.3) attached to a needle hub (3.4)
Note 1 to entry: See Figure 1.
3.3
needle tube
rigid tube with one end sharpened with a bevel to facilitate entry into body tissue
Note 1 to entry: See Figure 1.
3.4
needle hub
fitting attached to the needle tube (3.3)
Note 1 to entry: See Figure 1.
3.5
vent fitting
fixed or removable fitting permitting venting of air while restricting or preferably preventing the escape of
blood
Note 1 to entry: See Figure 1.
3.6
catheter
assembly comprising the catheter tube, catheter hub and any integral fittings
Note 1 to entry: See Figure 1.
3.7
flashback
blood flow into the needle (3.2) or catheter tube
Note 1 to entry: See Annex E for additional information.
3.8
needle point
entire sharpened section of the distal end of the needle tube (3.3)
Note 1 to entry: See Figure 1.
Key
a distance between heel of needle point and catheter tip (see 4.3.2)
l effective length
1 catheter tube
2 catheter hub
3 needle tube (3.3)
4 needle hub (3.4)
5 vent fitting (3.5)
6 heel of needle point
7 catheter (3.6)
8 needle point (3.8)
NOTE Other design features can include wings, injection ports integral with the catheter hub, other means of
connecting to the fluid path, protection against accidental needle stick injury, etc. The catheter tube can have a single
lumen or multiple lumens.
Figure 1 — Typical over-needle peripheral intravenous catheter
4 Requirements
4.1 General
Unless otherwise specified in this document, over-needle peripheral intravenous catheters shall conform
with ISO 10555-1.
4.2 Multilumen catheters
For multilumen catheters, identification of each lumen shall be apparent to the user.
4.3 Physical requirements
4.3.1 Colour code
The catheter shall be colour coded in accordance with Table 1 to indicate the nominal outside diameter of
the catheter tube.
Table 1 — Colour coding and corresponding sizes of catheter
a,b c
Nominal outside diameter of Range of actual outside diameter Colour Gauge
catheter tube
mm
mm
0,6 0,550 to 0,649 Violet 26
0,7 0,650 to 0,749 Yellow 24
0,8; 0,9 0,750 to 0,949 Deep blue 22
1,0; 1,1 0,950 to 1,149 Pink 20
1,2; 1,3 1,150 to 1,349 Deep green 18
1,4; 1,5 1,350 to 1,549 White 17
1,6; 1,7; 1,8 1,550 to 1,849 Medium grey 16
1,9; 2,0; 2,1; 2,2 1,850 to 2,249 Orange 14
2,3; 2,4; 2,5 2,250 to 2,549 Red 13
2,6; 2,7; 2,8 2,550 to 2,849 Pale blue 12
3,3; 3,4 3,250 to 3,549 Light brown 10
a
The colour may be opaque or translucent. Suggested colour references for opaque materials are given in Annex B.
b
The colour coding is usually applied to the catheter hub or to an integral fitting.
c
The use of gauge number is optional.
4.3.2 Catheter
The distal end shall be tapered for ease of insertion and shall fit closely to the needle. When the needle is
fully inserted into the catheter:
1) the catheter tube shall not extend beyond the heel of the needle point;
2) the distance between the heel of the needle point and the tip of the catheter shall not be more than 1
mm (see dimension a in Figure 1) unless supported by clinical justification and risk assessment.
NOTE See Annex E for additional information.
4.3.3 Needle
4.3.3.1 Needle tube material
If a steel tube is used, it shall conform with ISO 9626, except for the roundness requirement in areas where
roundness is intentionally modified (e.g. crimp, notch, groove).
4.3.3.2 Needle point
When examined by
...
ISO/DISFDIS 10555-5:2026(en)
ISO/TC 84/WG 9 N 1068
Replaces N 1067
Secretariat: DS
Date: 2026-09-2211-19
Intravascular catheters — Sterile and single-use catheters — —
Part 5:
Over-needle peripheral intravenous catheters
Cathéters intravasculaires — Cathéters stériles et non réutilisables — —
Partie 5: Cathéters intraveineux périphériques à aiguille interne
TTTTTThhhihhihiiiis drs drs drsss d d drrraaaaftaafffffttttt i i i i iis s s sss sssusususubububbbbmimmmmmiiiiitttttttttttteeeeeeddd d d d ttttttoooooo aaaaaa ppppppaaaaaarrrrrraaaaaallellellellellellel l l l l l vvvvvvooooootttttteeeeee i i i i i innnnnn IIIIIISSSSSSOOOOOO,,,,,, CCCCCCEEEEEEN.N.N.N.N.N.
© FDIS stage
ISO/DISFDIS 10555-5:2026(en) (N 1068)
All rights reserved. Unless otherwise specified,
This ISO publication is protected by copyright and is owned by ISO and/or its licensors.
The content of this ISO publication is provided under licence, not sold. Use is subject to the applicable licence terms issued
by ISO, an ISO member body, or an authorized third-party distributor.
Except as required in the context of its for implementation or expressly permitted by a separate licence, no part of this
ISO publication may be reproduced or utilized otherwise, distributed, modified, used, or made available in any form or
by any means, – electronic or mechanical, including photocopying, scanning, recording, or posting on internal or external
digital platforms.
Any use beyond the internet or an intranet, without prior written permission. Permission can be requested from either
ISO at the address below or ISO’s member body in the countryscope of the requestergranted rights is prohibited and may
result in legal action.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: + 41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
iii
Contents
Foreword . v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 Requirements . 3
4.1 General . 3
4.2 Multilumen catheters . 3
4.3 Physical requirements . 4
4.4 Information to be supplied by the manufacturer . 5
Annex A (normative) Determination of strength of union of needle hub and needle tube . 6
Annex B (informative) Colours for opaque catheter hubs . 7
Annex C (informative) Needle point geometries . 8
Annex D (normative) Determination of liquid leakage from vent fitting . 11
Annex E (informative) Rationale and guidance . 16
Bibliography . 18
iv
ISO/DISFDIS 10555-5:2026(en) (N 1068)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types of
ISO documents should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent rights
in respect thereof. As of the date of publication of this document, ISO had not received notice of (a) patent(s)
which may be required to implement this document. However, implementers are cautioned that this may not
represent the latest information, which may be obtained from the patent database available at
www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO'sISO’s adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 84, Devices for administration of medicinal
products and intravascular catheters, in collaboration with the European Committee for Standardization (CEN)
Technical Committee CEN/TC 205, Non-active medical devices, in accordance with the Agreement on technical
cooperation between ISO and CEN (Vienna Agreement).
This third edition cancels and replaces the second edition (ISO 10555-5:2013), which has been technically
revised.
The main changes are as follows:
— Scope isscope modified to be only applicable to over the -needle peripheral intravenous catheters.;
— Termterm “female” changed to “socket” to align with ISO 80369-7.;
— New Annex Enew Annex E added to provide rationale and guidance for changes compared to the previous
version.;
— Newnew test method option added in Annex DAnnex D.;
— Modificationmodification made to test fluid preparation in Annex DAnnex D.;
— Newnew requirement added for vessel access indication.;
— Changeschanges made to upper limit of the distance between the heel of the needle point and the tip of
the catheter.;
— Addedadded new terms and definitions.
v
A list of all parts in the ISO 10555 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.
vi
DRAFT International Standard ISO/DIS 10555-5:2026(en)
Intravascular catheters — Sterile and single-use catheters — —
Part 5:
Over-needle peripheral intravenous catheters
1 Scope
This document specifies requirements for over-needle peripheral intravenous catheters, intended for
accessing the peripheral venous system, supplied in the sterile condition and intended for single use.
NOTE 1 Attention is drawn to ISO 11070, which specifies requirements for accessory devices for use with intravenous
catheters, and to ISO 14972, which specifies requirements for sterile obturators for use with over-needle peripheral
intravenous catheters.
NOTE 2 See Annex EAnnex E for additional information.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes
requirements of this document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
ISO 80369--7, Small-bore connectors for liquids and gases in healthcare applications — Part 7: Connectors for
intravascular or hypodermic applications
ISO 9626, Stainless steel needle tubing for the manufacture of medical devices
ISO 10555--1, Intravascular catheters — Sterile and single-use catheters — Part 1: General requirements
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 10555-1 and the following apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— — ISO Online browsing platform: available at https://www.iso.org/obp
— — IEC Electropedia: available at https://www.electropedia.org/
3.1 3.1
over-needle peripheral intravenous catheter
catheter designed for the introduction or withdrawal of liquids or devices into or from the peripheral venous
system
Note 1 to entry: See Figure 1Figure 1.
3.2 3.2
needle
assembly comprising at least a needle tube (3.3(3.3)) attached to a needle hub (3.4(3.4))
Note 1 to entry: See Figure 1Figure 1.
3.3 3.3
needle tube
rigid tube with one end sharpened with a bevel to facilitate entry into body tissue
Note 1 to entry: See Figure 1Figure 1.
3.4 3.4
needle hub
fitting attached to the needle tube (3.3(3.3))
Note 1 to entry: See Figure 1Figure 1.
3.5 3.5
vent fitting
fixed or removable fitting permitting venting of air while restricting or preferably preventing the escape of
blood
Note 1 to entry: See Figure 1Figure 1.
3.6 3.6
catheter
assembly comprising the catheter tube, catheter hub and any integral fittings
Note 1 to entry: See Figure 1Figure 1.
3.7 3.7
flashback
blood flow into the needle (3.2(3.2)) or catheter tube
Note 1 to entry: See Annex EAnnex E for additional information.
3.8 3.8
needle point
entire sharpened section of the distal end of the needle tube (3.3(3.3))
Note 1 to entry: See Figure 1Figure 1.
ISO/DISFDIS 10555-5:2026(en) (N 1068)
Key
a distance between heel of needle point and catheter tip (see 4.3.24.3.2) )
l1 effective length
1 catheter tube
2 catheter hub
3 needle tube (3.3(3.3))
4 needle hub (3.4(3.4))
5 vent fitting (3.5(3.5))
6 heel of needle point
7 catheter (3.6(3.6))
8 needle point (3.8(3.8))
NOTE Other design features can include wings, injection ports integral with the catheter hub, other means of
connecting to the fluid path, protection against accidental needle stick injury, etc. The catheter tube can have a single
lumen or multiple lumens.
Figure 1 — Typical over-needle peripheral intravenous catheter
4 Requirements
4.1 General
Unless otherwise specified in this document, over-needle peripheral intravenous catheters shall conform with
ISO 10555--1.
4.2 Multilumen catheters
For multilumen catheters, identification of each lumen shall be apparent to the user.
4.3 Physical requirements
4.3.1 Colour code
The catheter shall be colour coded in accordance with Table 1Table 1 to indicate the nominal outside diameter
of the catheter tube.
Table 1 — Colour coding and corresponding sizes of catheter
a,b c
Nominal outside diameter Range of actual outside diameter Colour Gauge
of catheter tube
mm
mm
0,6 0,550 to 0,649 Violet 26
0,7 0,650 to 0,749 Yellow 24
0,8; 0,9 0,750 to 0,949 Deep blue 22
1,0; 1,1 0,950 to 1,149 Pink 20
1,2; 1,3 1,150 to 1,349 Deep green 18
1,4; 1,5 1,350 to 1,549 White 17
1,6; 1,7; 1,8 1,550 to 1,849 Medium grey 16
1,9; 2,0; 2,1; 2,2 1,850 to 2,249 Orange 14
2,3; 2,4; 2,5 2,250 to 2,549 Red 13
2,6; 2,7; 2,8 2,550 to 2,849 Pale blue 12
3,3; 3,4 3,250 to 3,549 Light brown 10
a
The colour may be opaque or translucent. Suggested colour references for opaque materials are given in Annex BAnnex B.
b
The colour coding is usually applied to the catheter hub or to an integral fitting.
c
The use of gauge number is optional.
4.3.2 Catheter
The distal end shall be tapered for ease of insertion and shall fit closely to the needle. When the needle is fully
inserted into the catheter:
1) the catheter tube shall not extend beyond the heel of the needle point; and
2) the distance between the heel of the needle point and the tip of the catheter shall not be more than 1 mm
(see dimension a in Figure 1Figure 1)) unless supported by clinical justification and risk assessment.
NOTE See Annex EAnnex E for additional information.
4.3.3 Needle
4.3.3.1 Needle tube material
If a steel tube is used, it shall conform with ISO 9626, except for the roundness requirement in areas where
roundness is intentionally modified (e.g. crimp, notch, groove).
4.3.3.2 Needle point
When examined by normal or corrected-to-normal vision with ×2,5 magnification, the needle point shall
appear sharp and free from feather edges, burrs and hooks.
ISO/DISFDIS 10555-5:2026(en) (N 1068)
NOTE Annex CAnnex C shows examples of typical needle point geometries.
4.3.3.3 Needle hub
If the needle is provided with a removable vent fitting, the needle hub shall terminate in a socket fitting
conforming with ISO 80369-7.
NOTE See Annex EAnnex E for additional information.
4.3.3.4 Strength of union between needle hub and needle tube
When tested in accordance with Annex AAnnex A,, the needle tube shall not be loosened in the needle hub.
4.3.4 Vent fitting
A vent fitting shall be provided. When tested in accordance with Annex DAnnex D,, fluid shall not leak out of
the vent fitting within 15 s. In Annex DAnnex D,, either test A or B shall be used.
4.3.5 Vessel access indication
Means shall be provided to indicate when the needle tip enters the vein (e.g. flashback).
NOTE See Annex EAnnex E for additional information.
4.4 Information to be supplied by the manufacturer
Information supplied by the manufacturer shall conform with ISO 10555--1 and shall also include the
following:
a) a) the flowrate for each lumen;
b) b) a warning against attempting to re-insert a partially or completely withdrawn needle;
c) c) on each primary packaging, the corresponding product colour (see Table 1Table 1),), unless
the colour on the product is visible through the primary packaging, and the outside diameter, as defined
in Table 1Table 1.
Units of measurement systems other than those specified may additionally be given.
...