EN 60601-1-3:2008/A12:2026
(Amendment)Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
- Status
- Published
- Publication Date
- 05-Nov-2026
- Technical Committee
- CLC/TC 62 - Electrical equipment in medical practice
- Current Stage
- 6060 - Document made available - Publishing
- Start Date
- 06-Nov-2026
- Due Date
- 09-Nov-2028
- Completion Date
- 06-Nov-2026
EN 60601-1-3:2008/A12:2026 is a CLC amendment to EN 60601-1-3:2008, the collateral standard for radiation protection in diagnostic X-ray equipment. It updates the European reference list and the EU medical-device mapping used with the base standard. For engineers and compliance teams, it is the amendment to use when checking the European annexes and their connection to Regulation (EU) 2017/745.
What does EN 60601-1-3:2008/A12:2026 specify?
EN 60601-1-3:2008/A12:2026 specifies replacement text for Annex ZA and Annex ZZ of EN 60601-1-3:2008. The amendment is therefore focused on the European normative references and the regulatory relationship for medical devices, rather than on rewriting the full technical body of the collateral standard.
| Annex | What it covers |
|---|---|
| Annex ZA | Normative references to international publications and the corresponding European publications |
| Annex ZZ | Relationship between the European standard and the General Safety and Performance Requirements of Regulation (EU) 2017/745 |
Annex ZZ makes clear that the standard is intended as a voluntary means of supporting conformity with the EU medical-device regulation within the standard’s scope. It also states that, where a definition in the standard differs from the regulation, the regulation prevails for regulatory use.
What are the key requirements of EN 60601-1-3:2008/A12:2026?
EN 60601-1-3:2008/A12:2026 mainly changes how the standard is anchored to other documents and how it is used for EU conformity. The practical effect is that users must check the updated annexes, not just the base standard.
- Annex ZA replaces the reference list. It names the international publications that become part of the requirements when cited, and it gives the matching European editions. In practice, this tells manufacturers and test labs which editions to apply for design, verification, and conformity work.
- Dated and undated references are treated differently. Annex ZA states that dated references use only the cited edition, while undated references use the latest edition, including amendments. This matters when choosing the controlling document for a test or design rule.
- The reference set includes IEC 60336, IEC 60522, EN IEC 60522-1, IEC 60601-1, IEC/TR 60788 and ISO 497. These support focal spot dimensions, filtration, general safety, terminology, and preferred-number series. In practice, they provide the technical basis for the measurements and terms used by the standard.
- Annex ZZ maps clauses to Regulation (EU) 2017/745. It identifies which clauses support specific General Safety and Performance Requirements (GSPRs). This helps regulatory teams show where the standard supports the medical-device regulation.
- Several mappings are specific to radiation protection in X-ray equipment. The table covers risks, hazards, side-effects, patient and user safety, and aspects of combination with X-ray image receptors. That means compliance is not just about the X-ray source, but also system interaction and exposure-related risk.
- Some mappings depend on maintenance and supplied information. Annex ZZ links certain clauses to manufacturer instructions, user information, and other documentation requirements. In practice, the technical file and instructions for use are part of the compliance evidence.
- Risk management wording follows the regulation. Annex ZZ says that when the standard refers to risk management, the process must align with the wording of the relevant GSPRs. This matters because the acceptance of risk has to match the MDR language.
What terms does EN 60601-1-3:2008/A12:2026 define?
- Collateral standard - A standard that adds requirements for a specific safety topic alongside the general medical electrical equipment standard.
- Diagnostic X-ray equipment - The type of equipment addressed by the radiation-protection focus of this standard.
- Normative references - Publications cited in Annex ZA that provide requirements when the standard refers to them.
- General Safety and Performance Requirements (GSPR) - The requirements in Regulation (EU) 2017/745 that Annex ZZ maps to this standard.
- Regulation (EU) 2017/745 - The EU medical-device regulation used in Annex ZZ for the conformity relationship.
- CENELEC - The European electrotechnical standardization body that approved this amendment.
Who uses EN 60601-1-3:2008/A12:2026?
EN 60601-1-3:2008/A12:2026 is used by manufacturers of diagnostic X-ray equipment, especially teams responsible for radiation protection, technical documentation, and EU conformity assessment. It is also used by quality managers, regulatory affairs staff, test laboratories, and certification bodies that need to check the updated European references and the Annex ZZ mapping to MDR. For organizations working on risk management, post-market surveillance, clinical evaluation, or clinical follow-up, the annex also clarifies how this collateral standard supports those regulatory processes.
What changed in EN 60601-1-3:2008/A12:2026 from the previous edition?
EN 60601-1-3:2008/A12:2026 modifies EN 60601-1-3:2008 and the earlier amendments EN 60601-1-3:2008/A1:2013, EN 60601-1-3:2008/A11:2016 and EN 60601-1-3:2008/A2:2021. The significant technical changes stated in the document are the replacement of the normative references in Annex ZA and the replacement of Annex ZZ.
Which standards are used with EN 60601-1-3:2008/A12:2026?
- IEC 60336 - Gives requirements for focal spot dimensions and related characteristics of X-ray tube assemblies for medical diagnosis.
- IEC 60522 - Covers determination of permanent filtration of X-ray tube assemblies.
- EN IEC 60522-1 - Covers determination of quality equivalent filtration and permanent filtration for diagnostic X-rays.
- IEC 60601-1 - Provides the general requirements for basic safety and essential performance for medical electrical equipment.
- IEC/TR 60788 - Provides the glossary of defined terms used in medical electrical equipment.
- ISO 497 - Gives guidance on preferred numbers and rounded preferred-number series.
What does the EN 60601-1-3:2008/A12:2026 document contain?
The document contains replacement Annex ZA and Annex ZZ text. Annex ZA gives the normative-reference table and the rule for using dated and undated references, while Annex ZZ gives the EU regulatory relationship and the clause-by-clause correspondence table to Regulation (EU) 2017/745. It also includes notes on how presumption of conformity is used, how MDR definitions prevail where they differ, and how other Union legislation may still apply.
Relations
- Effective Date
- 10-Nov-2026
- Effective Date
- 10-Nov-2026
- Effective Date
- 07-Jul-2026
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
EN 60601-1-3:2008/A12:2026 is a amendment published by CLC. Its full title is "Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment". This standard covers: Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
EN 60601-1-3:2008/A12:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.50 - Radiographic equipment; 13.280 - Radiation protection. The ICS classification helps identify the subject area and facilitates finding related standards.
EN 60601-1-3:2008/A12:2026 has the following relationships with other standards: It is inter standard links to EN 1789:2007+A1:2010, EN 13718-1:2008, EN 60601-1-3:2008. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
EN 60601-1-3:2008/A12:2026 is associated with the following European legislation: EU Directives/Regulations: 2017/745; Standardization Mandates: M/575, M/575 Amd 1, M/575 Amd 2. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.
EN 60601-1-3:2008/A12:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
SLOVENSKI STANDARD
01-januar-2027
Medicinska električna oprema - 1-3. del: Splošne zahteve za osnovno varnost in
bistvene lastnosti - Spremljevalni standard: Varstvo pred sevanjem pri rentgenski
diagnostični opremi - Dopolnilo A12
Medical electrical equipment - Part 1-3: General requirements for basic safety and
essential performance - Collateral Standard: Radiation protection in diagnostic X-ray
equipment
Medizinische elektrische Geräte - Teil 1-3: Allgemeine Festlegungen für die Sicherheit
einschließlich der wesentlichen Leistungsmerkmale - Ergänzungsnorm: Strahlenschutz
von diagnostischen Röntgengeräten
Appareils électromédicaux - Partie 1-3: Exigences générales pour la sécurité de base et
les performances essentielles - Norme collatérale: Radioprotection dans les appareils à
rayonnement X de diagnostic
Ta slovenski standard je istoveten z: EN 60601-1-3:2008/A12:2026
ICS:
11.040.50 Radiografska oprema Radiographic equipment
13.280 Varstvo pred sevanjem Radiation protection
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EUROPEAN STANDARD EN 60601-1-3:2008/A12
NORME EUROPÉENNE
EUROPÄISCHE NORM November 2026
ICS 11.040.50; 13.280
English Version
Medical electrical equipment - Part 1-3: General requirements for
basic safety and essential performance - Collateral Standard:
Radiation protection in diagnostic X-ray equipment
Appareils électromédicaux - Partie 1-3: Exigences Medizinische elektrische Geräte - Teil 1-3: Allgemeine
générales pour la sécurité de base et les performances Festlegungen für die Sicherheit einschließlich der
essentielles - Norme collatérale: Radioprotection dans les wesentlichen Leistungsmerkmale - Ergänzungsnorm:
appareils à rayonnement X de diagnostic Strahlenschutz von diagnostischen Röntgengeräten
This amendment A12 modifies the European Standard EN 60601-1-3:2008; it was approved by CENELEC on 2026-06-24. CENELEC
members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this amendment the
status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This amendment exists in three official versions (English, French, German). A version in any other language made by translation under the
responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the same status as
the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Türkiye and the United Kingdom.
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2026 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN 60601-1-3:2008/A12:2026 E
European foreword
This document (EN 60601-1-3:2008/A12:2026) has been prepared by CLC/TC 62 “Electrical equipment
in medical practice”.
The following dates are fixed:
• latest date by which this document has to be (dop) 2027-11-30
implemented at national level by publication of an
identical national standard or by endorsement
• latest date by which the national standards (dow) 2029-11-30
conflicting with this document have to be withdrawn
This document modifies EN 60601-1-3:2008, EN 60601-1-3:2008/A1:2013, EN 60601-1-3:2008/
A11:2016 and EN 60601-1-3:2008/A2:2021.
— the replacement of normative references in Annex ZA:
— the replacement of Annex ZZ.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
This document has been prepared under a standardization request addressed to CENELEC by the
European Commission. The Standing Committee of the EFTA States subsequently approves these
requests for its Member States.
For the relationship with EU Legislation, see informative Annex ZZ, which is an integral part of this
document.
Any feedback and questions on this document should be directed to the users’ national committee. A
complete listing of these bodies can be found on the CENELEC website.
1 Replacement of Annex ZA, “Normative references to international
publications with their corresponding European publications” and
Annex ZZ, “Coverage of Essential Requirements of EU Directives”
Replace Annex ZA and Annex ZZ with the following:
“
Annex ZA
(normative)
Normative references to international publications with their
corresponding European publications
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
NOTE 1 Where an International Publication has been modified by common modifications, indicated by (mod),
the relevant EN/HD applies.
NOTE 2 Up-to-date information on the latest versions of the European Standards listed in this annex is available
here: www.cencenelec.eu.
Publication Year Title EN/HD Year
IEC 60336 - Medical electrical equipment – X-ray tube EN IEC 60336 2021
assemblies for medical diagnosis – Focal spot
dimensions and related characteristics
IEC 60522 1999 Determination of the permanent filtration of X- EN 60522 1999
ray tube assemblies
Medical electrical equipment – Diagnostic X- EN IEC 60522-1 2021
rays – Part 1: Determination of quality
equivalent filtration and permanent filtration
IEC 60601-1 2005 Medical electrical equipment – Part 1: General EN 60601-1 2006
requirements for basic safety and essential
performance
+ corrigendum March
+ A1 2012 + A1 2013
+ COR1 2012 + AC 2014
+ A12 2014
+ A2 2020 + A2 2021
+ AC 2022
+ A13 2024
IEC/TR 60788 2004 Medical electrical equipment – Glossary of - -
defined terms
ISO 497 1973 Guide to the choice of series of preferred - -
numbers and of series containing more rounded
values of preferred n
...