ASTM D2593-93(2014)
(Test Method)Standard Test Method for Butadiene Purity and Hydrocarbon Impurities by Gas Chromatography
Standard Test Method for Butadiene Purity and Hydrocarbon Impurities by Gas Chromatography
SIGNIFICANCE AND USE
4.1 The trace hydrocarbon compounds listed can have an effect in the commercial use of butadiene. This test method is suitable for use in process quality control and in setting specifications.
SCOPE
1.1 This test method covers the determination of butadiene-1,3 purity and impurities such as propane, propylene, isobutane, n-butane, butene-1, isobutylene, propadiene, trans-butene-2, cis -butene-2, butadiene-1,2, pentadiene-1,4, and, methyl, dimethyl, ethyl, and vinyl acetylene in polymerization grade butadiene by gas chromatography. Impurities including butadiene dimer, carbonyls, inhibitor, and residue are measured by appropriate ASTM procedures and the results used to normalize the component distribution obtained by chromatography.
Note 1: Other impurities present in commercial butadiene must be calibrated and analyzed. Other impurities were not tested in the cooperative work on this test method.
Note 2: This test method can be used to check for pentadiene-1,4 and other C5s instead of Test Method D1088.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific warning statements, see 6.1 and 9.3.
General Information
Buy Standard
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: D2593 − 93 (Reapproved 2014)
Standard Test Method for
Butadiene Purity and Hydrocarbon Impurities by Gas
Chromatography
This standard is issued under the fixed designation D2593; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 2.2 Energy Institute Standards:
IP 194 Analysis of Butadiene-1,3 Polymerization Grade
1.1 This test method covers the determination of butadiene-
1,3 purity and impurities such as propane, propylene,
3. Summary of Test Method
isobutane, n-butane, butene-1, isobutylene, propadiene, trans-
3.1 A representative sample is introduced into a gas-liquid
butene-2, cis -butene-2, butadiene-1,2, pentadiene-1,4, and,
partition column. The butadiene and other components are
methyl, dimethyl, ethyl, and vinyl acetylene in polymerization
separated as they are transported through the column by an
grade butadiene by gas chromatography. Impurities including
inert carrier gas. Their presence in the effluent is measured by
butadienedimer,carbonyls,inhibitor,andresiduearemeasured
adetectorandrecordedasachromatogram.Thechromatogram
by appropriate ASTM procedures and the results used to
of the sample is interpreted by applying component attenuation
normalize the component distribution obtained by chromatog-
and detector response factors to the peak areas or peak heights
raphy.
and the relative concentration determined by relating indi-
NOTE 1—Other impurities present in commercial butadiene must be
vidual peak response to total peak response. Impurities includ-
calibrated and analyzed. Other impurities were not tested in the coopera-
ing butadiene dimer, carbonyls, inhibitor, and residue are
tive work on this test method.
NOTE 2—This test method can be used to check for pentadiene-1,4 and measured by appropriate ASTM procedures and the results
other C s instead of Test Method D1088.
5 used to normalize the distribution obtained by gas chromatog-
raphy.
1.2 The values stated in SI units are to be regarded as
standard. No other units of measurement are included in this
4. Significance and Use
standard.
4.1 The trace hydrocarbon compounds listed can have an
1.3 This standard does not purport to address all of the
effect in the commercial use of butadiene. This test method is
safety concerns, if any, associated with its use. It is the
suitable for use in process quality control and in setting
responsibility of the user of this standard to establish appro-
specifications.
priate safety and health practices and determine the applica-
bility of regulatory limitations prior to use. For specific
5. Apparatus
warning statements, see 6.1 and 9.3.
5.1 Chromatograph—Any chromatograph having either a
2. Referenced Documents
thermal-conductivity or flame ionization detector can be used
2.1 ASTM Standards: provided the system has sufficient sensitivity and stability to
D1088 Method of Test for Boiling Point Range of obtain a recorder deflection of at least 2 mm at signal-to-noise
Polymerization-Grade Butadiene (Withdrawn 1983) ratio of at least 5:1 for 0.01 weight % of impurity.
5.2 Column—Any column can be used that is capable of
resolving the components listed in 1.1 with the exception of
This test method is under the jurisdiction of ASTM Committee D02 on
butene-1 and isobutylene, which can be eluted together. The
Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of
Subcommittee D02.D0.04 on C4 Hydrocarbons.
componentsshouldberesolvedintodistinctpeakssuchthatthe
This test method was adopted as a joint ASTM-IP Standard, IP 194, in 1972.
ratio A/B will not be less than 0.5 where A is the depth of the
CurrenteditionapprovedMay1,2014.PublishedJuly2014.Originallyapproved
valley on either side of peak B and B is the height above the
in 1967. Last previous edition approved in 2009 as D2593–93(2009). DOI:
baseline of the smaller of any two adjacent peaks. In the case
10.1520/D2593-93R14.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
where the small component peak is adjacent to a large one, it
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website.
3 4
The last approved version of this historical standard is referenced on Obsolete. Contact Energy Institute, 61 New Cavendish St., London,WIG 7AR,
www.astm.org. U.K., http://www.energyinst.org.uk.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D2593 − 93 (2014)
can be necessary to construct a baseline of the small peak optimum resolution and analysis time. Optimum size ranges
tangent to the curve as shown in Fig. 1. cannot be predicted on purely theoretical grounds. For some
5.2.1 A description of columns that meet the requirements systems it has been found that a ratio of average particle
of this test method is tabulated in theAppendix. Persons using diameter to column inside diameter of 1:25 will result in
other column materials must establish that the column gives minimum retention time and minimum band widths.
results that meet the precision requirements of Section 11. 6.2.3 Tubing Material—Copper, stainless steel, Monel,
aluminum, and various plastic materials have been found to be
5.3 Sample Inlet System—Means shall be provided for
satisfactory for column tubing. The material must be nonreac-
introducing a measured quantity of representative sample into
tive with respect to substrate, sample, and carrier gas and of
the column. Pressure-sampling devices can be used to inject a
uniform internal diameter.
small amount of liquid directly into the carrier gas. Introduc-
tion can also be accomplished by use of a gas valve to charge 6.3 Hydrocarbons for Calibration and Identification—
the vaporized liquid. Hydrocarbon standards for all components present are needed
for identification by retention time and for calibration for
5.4 Recorder—A recording potentiometer with a full-scale
quantitative measurements.
deflection of 10 mV or less is suitable for obtaining the
chromatographic data. Full-scale response time should be 2 s
NOTE 4—Mixtures of hydrocarbons can be used provided there is no
uncertainty as to the identity or concentration of the compounds involved.
or less, and with sufficient sensitivity to meet the requirements
of 5.1.
7. Preparation of Apparatus
NOTE3—Othermethodsofrecordingdetectoroutputsuchascomputer-
7.1 Column Preparation—Thetechniqueusedtopreparethe
teletype systems can be used instead of a recorder, provided precision
column is not critical as long as the finished column produces
requirements of Section 11 are met.
the desired separation. Preparation of the packing is not
6. Reagents and Materials difficult once the support, partitioning liquid, and loading level
have been determined. The following general directions have
6.1 Carrier Gas—A carrier gas appropriate to the type of
been found to produce columns of acceptable characteristics.
detectorusedshouldbeemployed.Heliumorhydrogenmaybe
7.1.1 Weigh out the desired quantity of support, usually
used with thermal conductivity detectors. Nitrogen, helium, or
twice that required to fill the column.
argon may be used with ionization detectors. The minimum
7.1.2 Calculate and weigh out the required quantity of
purity of any carrier should be 99.95 mol %. (Warning—
partitioning agent. Dissolve the partitioning agent in a volume
Compressed gas. Hazardous pressure.) (Warning—Hydrogen
ofchemicallyinert,low-boilingsolventequaltoapproximately
flammable gas. Hazardous pressure.)
twice the volume of support.
6.1.1 Ifhydrogenisused,specialsafetyprecautionsmustbe
7.1.3 Graduallyaddthesupportmaterialtothesolutionwith
taken to ensure that the system is free from leaks and that the
gentle stirring.
effluent is properly vented.
7.1.4 Slowly evaporate the solvent while gently agitating
6.2 Column Materials:
the mixture until the packing is nearly dry and no free liquid is
6.2.1 Liquid Phase—The materials that have been used
apparent.
successfully in cooperative work as liquid phases are listed in
7.1.4.1 Some stationary phases such as benzyl cyanide
Table X1.1.
silvernitratearesusceptibletooxidationandmustbeprotected
6.2.2 Solid Support—The support for use in the packed
from excessive exposure to air during the evaporation and
column is usually crushed firebrick or diatomaceous earth.
drying steps.
Sieve size will depend on the diameter of the column used and
7.1.5 Spread the packing in thin layers on a nonabsorbent
liquid-phase loading, and should be such as would give
surface and air- or oven-dry as required to remove all traces of
solvent.
7.1.6 Resieve the packing to remove fines and agglomerates
produced in the impregnation step.
7.1.7 Fill the column tubing with packing by plugging one
end with glass wool and pouring the packing into the other end
throughasmallfunnel.Vibratethetubingcontinuouslyoverits
entirelengthwhilefilling.Whenthepackingceasestoflow,tap
the column gently on the floor or bench-top while vibrating is
continued. Add packing as necessary until no further settling
occurs during a 2-min period. Remove a small amount of
packingfromtheopenend,plugwithglasswool,andshapethe
column to fit the chromatograph.
7.2 Chromatograph—Mount the column in the chromato-
graph and establish the operating conditions required to give
thedesiredseparation(AppendixX1).Allowsufficienttimefor
the instrument to reach equilibrium as indicated by a stable
FIG. 1 Illustration of A/B Ratio for Small-Component Peak base line. Control the oven temperature so that it is constant to
D2593 − 93 (2014)
within 0.5°C without thermostat cycling which causes an forpentadiene-1,4butadiene-1,2,dimethylacetylene,ethyland
uneven base line. Set the carrier-gas flow rate, measured with vinyl acetylene in this study to obtain the precision listed in
a soap film meter, so that it is constant to within 1 mL/min of Section 11. It is permissible to use the above established
the selected value. response factors instead of calibration when using thermal-
conductivity detectors with helium-carrier gas. With other
8. Calibration detectors or carrier gas, or both, it is necessary to calibrate
(Note 5).
8.1 Identification—Select the conditions of column tem-
8.2.2 Measurements can be made using peak heights as
perature and carrier gas flow that will give the necessary
criteria for calculations instead of peak areas. If peak heights
separation.Determinetheretentiontimeforeachcompoundby
are used, care must be taken so that chromatograph-operating
injecting small amounts of the compound either separately or
parameters such as column temperature and carrier-gas flow
in mixtures. Recommended sample sizes for retention data are
rate are kept at the same conditions as when the unit was
1 µL for liquids and 1 cm or less for gases.
calibrated. The chromatograph can be calibrated using known
8.2 Standardization—The area under the peak of the chro-
blends or by establishing relative-response data using peak
matogram is considered a quantitative measure of the amount
heights in the same manner as listed above.
of the corresponding compound. The relative area is propor-
NOTE 6—Use of a hydrogen-flame detector gives essentially equal
tional to the concentration if the detector responses of the
relative response to hydrocarbons. On a weight basis, the sensitivity of the
sample components are equal. The recommended procedure
flame detector for hydrocarbons is essentially independent of the hydro-
for quantitative calibration is as follows: with all equipment at
carbonsstructure.Onamolarbasis,thesensitivityappearstobeafunction
equilibrium at operating conditions, inject constant volume of the carbon content, giving essentially equal relative response to
hydrocarbons containing the same number of carbon atoms.
samples of high-purity components. Each compound should be
injected at least three times. The areas of the corresponding
8.2.3 Becausedetectororamplifieroutputneednotbelinear
peaks should agree within 1 %.When a recorder is used, adjust
with component concentration, this must be checked by inject-
the attenuation in all cases to keep the peak on-scale and with
ing constant volumes of pure butadiene at a series of decreas-
a height of at least 50 % of full scale. Measure the area of the
ing pressures from ambient down to 20 mm Hg (torr) or by
peaks by any reliable method (Note 7). To obtain component
using synthetic standards with vapor sample valves at ambient
weight % response data from the area response of the volume
or at decreasing pressures or by using synthetic standards with
injections, it is necessary to consider the density and purity of
liquid sample valves. If on plotting the results the response is
the compounds used for calibration. The average volume area
linear, then the calibration procedure given above is satisfac-
response of each component is divided by the density multi-
tory. If not, the relative responses of the minor components
pliedbytheweightpercentpurityofthecomponentasfollows:
must be determined in the linear response region.
Weight percent response of component (1)
9. Procedure
average component peak area
9.1 Attach the sample cylinder to the instrument-sampling
density 3weight percent purity of component
valve so that the sample is obtained from the liquid phase. If
introduction is through a liquid valve the sample cylinders
Component weight percent detector correction factors are
should be pressured with a suitable gas, such as helium, to a
then obtained by selecting a reference component such as
pressuresufficienttoensurethatsampleflashingdoesnotoccur
butadiene, and dividing the individual component weight
inthelinetothesamplingvalveorinthevalveitself.Ifavapor
responses into the reference weight response.
valve is used, care must be taken to completely vaporize a
8.2.1 Factors derived on a thermal-conductivity detector
small liquid sample, allowing the vapor to flow through the
using helium-carrier gas are as follows:
sample loop at a flow rate of 5 to 10 mL/min until at least ten
Component Mol wt Thermal Weight Factor Weight
times the volume of the sample loop has been flushed through.
Response Factor,
Butadiene-
If a vacuum-sampling system is used with a vapor valve, the
1,3=1.00
sampleloopshouldbefilledandevacuatedatleasttwicebefore
Butadiene-1,3 54 80 0.68 1.00
introduction of sample.
Propane 44 65 0.68 1.00
Propylene 42 63 0.67 0.98
9.2 Charge sufficient sample to ensure a
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: D2593 − 93 (Reapproved 2009) D2593 − 93 (Reapproved 2014)
Standard Test Method for
Butadiene Purity and Hydrocarbon Impurities by Gas
Chromatography
This standard is issued under the fixed designation D2593; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This test method covers the determination of butadiene-1,3 purity and impurities such as propane, propylene, isobutane,
n-butane, butene-1, isobutylene, propadiene, trans-butene-2, cis -butene-2, butadiene-1,2, pentadiene-1,4, and, methyl, dimethyl,
ethyl, and vinyl acetylene in polymerization grade butadiene by gas chromatography. Impurities including butadiene dimer,
carbonyls, inhibitor, and residue are measured by appropriate ASTM procedures and the results used to normalize the component
distribution obtained by chromatography.
NOTE 1—Other impurities present in commercial butadiene must be calibrated and analyzed. Other impurities were not tested in the cooperative work
on this test method.
NOTE 2—This test method can be used to check for pentadiene-1,4 and other C s instead of Test Method D1088.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use. For specific warning statements, see 6.1 and 9.3.
2. Referenced Documents
2.1 ASTM Standards:
D1088 Method of Test for Boiling Point Range of Polymerization-Grade Butadiene (Withdrawn 1983)
2.2 Energy Institute Standards:
IP 194 Analysis of Butadiene-1,3 Polymerization Grade
3. Summary of Test Method
3.1 A representative sample is introduced into a gas-liquid partition column. The butadiene and other components are separated
as they are transported through the column by an inert carrier gas. Their presence in the effluent is measured by a detector and
recorded as a chromatogram. The chromatogram of the sample is interpreted by applying component attenuation and detector
response factors to the peak areas or peak heights and the relative concentration determined by relating individual peak response
to total peak response. Impurities including butadiene dimer, carbonyls, inhibitor, and residue are measured by appropriate ASTM
procedures and the results used to normalize the distribution obtained by gas chromatography.
4. Significance and Use
4.1 The trace hydrocarbon compounds listed can have an effect in the commercial use of butadiene. This test method is suitable
for use in process quality control and in setting specifications.
This test method is under the jurisdiction of ASTM Committee D02 on Petroleum Products Products, Liquid Fuels, and Lubricants and is the direct responsibility of
Subcommittee D02.D0.04 on C4 Hydrocarbons.
This test method was adopted as a joint ASTM-IP Standard, IP 194, in 1972.
Current edition approved July 1, 2009May 1, 2014. Published November 2009July 2014. Originally approved in 1967. Last previous edition approved in 20042009 as
ε1
D2593–93(2004)D2593–93(2009). . DOI: 10.1520/D2593-93R09.10.1520/D2593-93R14.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
The last approved version of this historical standard is referenced on www.astm.org.
Obsolete. Contact Energy Institute, 61 New Cavendish St., London, WIG 7AR, U.K., http://www.energyinst.org.uk.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D2593 − 93 (2014)
5. Apparatus
5.1 Chromatograph—Any chromatograph having either a thermal-conductivity or flame ionization detector can be used
provided the system has sufficient sensitivity and stability to obtain a recorder deflection of at least 2 mm at signal-to-noise ratio
of at least 5:1 for 0.01 weight % of impurity.
5.2 Column—Any column can be used that is capable of resolving the components listed in 1.1 with the exception of butene-1
and isobutylene, which can be eluted together. The components should be resolved into distinct peaks such that the ratio A/B will
not be less than 0.5 where A is the depth of the valley on either side of peak B and B is the height above the baseline of the smaller
of any two adjacent peaks. In the case where the small component peak is adjacent to a large one, it can be necessary to construct
a baseline of the small peak tangent to the curve as shown in Fig. 1.
5.2.1 A description of columns that meet the requirements of this test method is tabulated in the Appendix. Persons using other
column materials must establish that the column gives results that meet the precision requirements of Section 11.
5.3 Sample Inlet System—Means shall be provided for introducing a measured quantity of representative sample into the
column. Pressure-sampling devices can be used to inject a small amount of liquid directly into the carrier gas. Introduction can
also be accomplished by use of a gas valve to charge the vaporized liquid.
5.4 Recorder—A recording potentiometer with a full-scale deflection of 10 mV or less is suitable for obtaining the
chromatographic data. Full-scale response time should be 2 s or less, and with sufficient sensitivity to meet the requirements of
5.1.
NOTE 3—Other methods of recording detector output such as computer-teletype systems can be used instead of a recorder, provided precision
requirements of Section 11 are met.
6. Reagents and Materials
6.1 Carrier Gas—A carrier gas appropriate to the type of detector used should be employed. Helium or hydrogen may be used
with thermal conductivity detectors. Nitrogen, helium, or argon may be used with ionization detectors. The minimum purity of any
carrier should be 99.95 mol %. (Warning—Compressed gas. Hazardous pressure.) (Warning—Hydrogen flammable gas.
Hazardous pressure.)
6.1.1 If hydrogen is used, special safety precautions must be taken to ensure that the system is free from leaks and that the
effluent is properly vented.
6.2 Column Materials:
6.2.1 Liquid Phase—The materials that have been used successfully in cooperative work as liquid phases are listed in Table
X1.1.
6.2.2 Solid Support—The support for use in the packed column is usually crushed firebrick or diatomaceous earth. Sieve size
will depend on the diameter of the column used and liquid-phase loading, and should be such as would give optimum resolution
and analysis time. Optimum size ranges cannot be predicted on purely theoretical grounds. For some systems it has been found
that a ratio of average particle diameter to column inside diameter of 1:25 will result in minimum retention time and minimum
band widths.
6.2.3 Tubing Material—Copper, stainless steel, Monel, aluminum, and various plastic materials have been found to be
satisfactory for column tubing. The material must be nonreactive with respect to substrate, sample, and carrier gas and of uniform
internal diameter.
FIG. 1 Illustration of A/B Ratio for Small-Component Peak
D2593 − 93 (2014)
6.3 Hydrocarbons for Calibration and Identification —Identification—Hydrocarbon standards for all components present are
needed for identification by retention time and for calibration for quantitative measurements.
NOTE 4—Mixtures of hydrocarbons can be used provided there is no uncertainty as to the identity or concentration of the compounds involved.
7. Preparation of Apparatus
7.1 Column Preparation—The technique used to prepare the column is not critical as long as the finished column produces the
desired separation. Preparation of the packing is not difficult once the support, partitioning liquid, and loading level have been
determined. The following general directions have been found to produce columns of acceptable characteristics.
7.1.1 Weigh out the desired quantity of support, usually twice that required to fill the column.
7.1.2 Calculate and weigh out the required quantity of partitioning agent. Dissolve the partitioning agent in a volume of
chemically inert, low-boiling solvent equal to approximately twice the volume of support.
7.1.3 Gradually add the support material to the solution with gentle stirring.
7.1.4 Slowly evaporate the solvent while gently agitating the mixture until the packing is nearly dry and no free liquid is
apparent.
7.1.4.1 Some stationary phases such as benzyl cyanide silver nitrate are susceptible to oxidation and must be protected from
excessive exposure to air during the evaporation and drying steps.
7.1.5 Spread the packing in thin layers on a nonabsorbent surface and air- or oven-dry as required to remove all traces of solvent.
7.1.6 Resieve the packing to remove fines and agglomerates produced in the impregnation step.
7.1.7 Fill the column tubing with packing by plugging one end with glass wool and pouring the packing into the other end
through a small funnel. Vibrate the tubing continuously over its entire length while filling. When the packing ceases to flow, tap
the column gently on the floor or bench-top while vibrating is continued. Add packing as necessary until no further settling occurs
during a 2-min period. Remove a small amount of packing from the open end, plug with glass wool, and shape the column to fit
the chromatograph.
7.2 Chromatograph—Mount the column in the chromatograph and establish the operating conditions required to give the
desired separation (Appendix X1). Allow sufficient time for the instrument to reach equilibrium as indicated by a stable base line.
Control the oven temperature so that it is constant to within 0.5°C without thermostat cycling which causes an uneven base line.
Set the carrier-gas flow rate, measured with a soap film meter, so that it is constant to within 1 mL/min of the selected value.
8. Calibration
8.1 Identification—Select the conditions of column temperature and carrier gas flow that will give the necessary separation.
Determine the retention time for each compound by injecting small amounts of the compound either separately or in mixtures.
Recommended sample sizes for retention data are 1 μL for liquids and 1 cm or less for gases.
8.2 Standardization—The area under the peak of the chromatogram is considered a quantitative measure of the amount of the
corresponding compound. The relative area is proportional to the concentration if the detector responses of the sample components
are equal. The recommended procedure for quantitative calibration is as follows: with all equipment at equilibrium at operating
conditions, inject constant volume samples of high-purity components. Each compound should be injected at least three times. The
areas of the corresponding peaks should agree within 1 %. When a recorder is used, adjust the attenuation in all cases to keep the
peak on-scale and with a height of at least 50 % of full scale. Measure the area of the peaks by any reliable method (Note 7). To
obtain component weight % response data from the area response of the volume injections, it is necessary to consider the density
and purity of the compounds used for calibration. The average volume area response of each component is divided by the density
multiplied by the weight percent purity of the component as follows:
Weight percent response of component (1)
average component peak area
density 3weight percent purity of component
Component weight percent detector correction factors are then obtained by selecting a reference component such as butadiene,
and dividing the individual component weight responses into the reference weight response.
8.2.1 Factors derived on a thermal-conductivity detector using helium-carrier gas are as follows:
Component Mol wt Thermal Weight Factor Weight
Response Factor,
Butadiene-
1,3 = 1.00
Butadiene-1,3 54 80 0.68 1.00
Propane 44 65 0.68 1.00
Propylene 42 63 0.67 0.98
Isobutane 58 82 0.71 1.04
n-Butane 58 85 0.68 1.00
Butene-1 56 81 0.69 1.01
Isobutylene 56 82 0.68 1.00
D2593 − 93 (2014)
Component Mol wt Thermal Weight Factor Weight
Response Factor,
Butadiene-
1,3 = 1.00
trans-Butene-2 56 85 0.66 0.97
cis-Butene-2 56 87 0.64 0.94
Propadiene 40 53 0.75 1.10
Methyl acetylene 40 58 0.69 1.01
NOTE 5—Response based on data represented by Messner, A. E., Rosie, D. M., and Argabright, P. A., Analytical Chemistry, Vol 31, 1959, pp. 230–233,
and Dietz, W. A., Journal of Gas Chromatography, Vol 5, No. 2, 1967, pp. 68–71.
8.2.1.1 Although not determined with standards, weight factors of 1.00 (compared to butadiene 1,3 as 1.00) were used for
pentadiene-1,4 butadiene-1,2, dimethyl acetylene, ethyl and vinyl acetylene in this study to obtain the precision listed in Section
11. It is permissible to use the above established response factors instead of calibration when using thermal-conductivity detectors
with helium-carrier gas. With other detectors or carrier gas, or both, it is necessary to calibrate (Note 5).
8.2.2 Measurements can be made using peak heights as criteria for calculations instead of peak areas. If peak heights are used,
care must be taken so that chromatograph-operating parameters such as column temperature and carrier-gas flow rate are kept at
the same conditions as when the unit was calibrated. The chromatograph can be calibrated using known blends or by establishing
relative-response data using peak heights in the same manner as listed above.
NOTE 6—Use of a hydrogen-flame detector gives essentially equal relative response to hydrocarbons. On a weight basis, the sensitivity of the flame
detector for hydrocarbons is essentially independent of the hydrocarbons structure. On a molar basis, the sensitivity appears to be a function of the carbon
content, giving essentially equal relative response to hydrocarbons containing the same number of carbon atoms.
8.2.3 Because detector or amplifier output need not be linear with component concentration, this must be checked by injecting
c
...










Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.