Standard Test Method for Thermal Lag of Thermal Analysis Apparatus

SIGNIFICANCE AND USE
5.1 Differing temperature rates-of-change may be required for different measurements (for example, Test Method E698). Temperature calibration changes as a function of temperature rate-of-change. The use of the known thermal lag of an apparatus may be used to adjust the temperature calibration of the apparatus obtained at one temperature rate-of-change with that at another required for a given applications. This adjustment procedure for temperature calibration is described in 8.1.  
5.2 This test method may be used in research, quality assurance, and specification acceptance.
SCOPE
1.1 This test method addresses the dependence of temperature calibration on the temperature rate-of-change. This test method describes the determination of the thermal lag of thermal analysis apparatus and its application to the modification of the temperature calibration for that apparatus obtained at alternative linear temperature rates-of-change.  
1.2 This test method is applicable, but not limited to, the temperature calibration of differential thermal analyzers (DTAs), differential scanning calorimeters (DSCs), thermogravimetric analyzers (TGAs), thermomechanical analyzers (TMAs), and dynamic mechanical analyzers (DMAs).  
1.3 This test method is applicable only to apparatus demonstrating a linear relationship between indicated temperature and temperature rate-of-change.  
1.4 The values stated in SI units are to be regarded as the standard. No other units of measurement are included in this standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Historical
Publication Date
31-Jul-2018
Current Stage
Ref Project

Buy Standard

Standard
ASTM E3142-18a - Standard Test Method for Thermal Lag of Thermal Analysis Apparatus
English language
4 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM E3142-18a - Standard Test Method for Thermal Lag of Thermal Analysis Apparatus
English language
4 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E3142 − 18a
Standard Test Method for
1
Thermal Lag of Thermal Analysis Apparatus
This standard is issued under the fixed designation E3142; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
In thermal analysis, the temperature of a test specimen is changed while a physical property is
measured. The measured physical property is the dependent variable and temperature (also measured)
is the independent variable. In the majority of thermal analysis apparatus, temperature sensors cannot
be attached directly to the specimen but can only touch the surface or be placed adjacent or close to
the specimen such that the indicated temperature will be different to that of the specimen itself. In
consequence the specimen temperature will lag the indicated temperature upon heating and cooling
due to thermal resistance between sensor and specimen. The larger the test specimen, the greater the
thermal lag is likely to be. To obtain the correct specimen temperature, thermal analysis apparatus is
temperature calibrated so that the recorded temperature correctly indicates the specimen temperature.
Such temperature calibration compensates for the temperature offset (τ) between the specimen
temperature and that of the temperature sensor. This temperature offset changes linearly with
temperature rate-of-change (β) (heating or cooling). The slope of this linear relationship is known as
“thermal lag” (ΔT/Δβ). Knowing the thermal lag for an apparatus permits temperature calibration
determined at one temperature rate-of-change to be adjusted to that at other rates. It is the purpose of
this standard to aid the user to determine the thermal lag for an apparatus and to apply that thermal
lag to measurements made at temperature rates-of-change different from that at which the temperature
calibration is performed.
1. Scope 1.4 The values stated in SI units are to be regarded as the
standard. No other units of measurement are included in this
1.1 This test method addresses the dependence of tempera-
standard.
ture calibration on the temperature rate-of-change. This test
1.5 This standard does not purport to address all of the
method describes the determination of the thermal lag of
safety concerns, if any, associated with its use. It is the
thermal analysis apparatus and its application to the modifica-
responsibility of the user of this standard to establish appro-
tion of the temperature calibration for that apparatus obtained
priate safety, health, and environmental practices and deter-
at alternative linear temperature rates-of-change.
mine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accor-
1.2 This test method is applicable, but not limited to, the
dance with internationally recognized principles on standard-
temperature calibration of differential thermal analyzers
ization established in the Decision on Principles for the
(DTAs), differential scanning calorimeters (DSCs), thermogra-
Development of International Standards, Guides and Recom-
vimetric analyzers (TGAs), thermomechanical analyzers
mendations issued by the World Trade Organization Technical
(TMAs), and dynamic mechanical analyzers (DMAs).
Barriers to Trade (TBT) Committee.
1.3 This test method is applicable only to apparatus dem-
onstrating a linear relationship between indicated temperature 2. Referenced Documents
2
and temperature rate-of-change.
2.1 ASTM Standards:
E473 Terminology Relating to Thermal Analysis and Rhe-
ology
1
This test method is under the jurisdiction of ASTM Committee E37 on Thermal
Measurements and is the direct responsibility of Subcommittee E37.10 on
2
Fundamental, Statistical and Mechanical Properties. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved Aug. 1, 2018. Published August 2018. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 2018. Last previous edition approved in 2018 as E3142 – 18. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/E3142-18A. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E3142 − 18a
E698 Test Method for Kinetic Parameters for Thermally 5. Significance and Use
Unstable Materials Using Differential Scanning Calorim-
5.1 Differing temperature rates-of-change may be required
etry and the Flynn/Wall/Ozawa Method
for different measurements (for exa
...

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E3142 − 18 E3142 − 18a
Standard Test Method for
1
Thermal Lag of Thermal Analysis Apparatus
This standard is issued under the fixed designation E3142; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
In thermal analysis, the temperature of a test specimen is changed while a physical property is
measured. The measured physical property is the dependent variable and temperature (also measured)
is the independent variable. In the majority of thermal analysis apparatus, temperature sensors cannot
be attached directly to the specimen but can only touch the surface or be placed adjacent or close to
the specimen such that the indicated temperature will be different to that of the specimen itself. In
consequence the specimen temperature will lag the indicated temperature upon heating and cooling
due to thermal resistance between sensor and specimen. The larger the test specimen, the greater the
thermal lag is likely to be. To obtain the correct specimen temperature, thermal analysis apparatus is
temperature calibrated so that the recorded temperature correctly indicates the specimen temperature.
Such temperature calibration compensates for the temperature offset (τ) between the specimen
temperature and that of the temperature sensor. This temperature offset changes linearly with
temperature rate-of-change (β) (heating or cooling). The slope of this linear relationship is known as
“thermal lag” (ΔT/Δβ). The Knowing the thermal lag for an apparatus permits temperature calibration
determined at one temperature rate-of-change to be adjusted to that at other rates. It is the purpose of
this standard to aid the user to determine the thermal lag for an apparatus and to apply that thermal
lag to measurements made at temperature rates-of-change different from that at which the temperature
calibration is performed.
1. Scope
1.1 This test method addresses the dependence of temperature calibration on the temperature rate-of-change. This test method
describes the determination of the thermal lag of thermal analysis apparatus and its application to the modification of the
temperature calibration for that apparatus obtained at alternative linear temperature rates-of-change.
1.2 This test method is applicable, but not limited to, the temperature calibration of differential thermal analyzers (DTAs),
differential scanning calorimeters (DSCs), thermogravimetric analyzers (TGAs), thermomechanical analyzers (TMAs), and
dynamic mechanical analyzers (DMAs).
1.3 This test method is applicable only to apparatus demonstrating a linear relationship between indicated temperature and
temperature rate-of-change.
1.4 The values stated in SI units are to be regarded as the standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of
regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
1
This test method is under the jurisdiction of ASTM Committee E37 on Thermal Measurements and is the direct responsibility of Subcommittee E37.10 on Fundamental,
Statistical and Mechanical Properties.
Current edition approved Jan. 15, 2018Aug. 1, 2018. Published February 2018August 2018. Originally approved in 2018. Last previous edition approved in 2018 as E3142
– 18. DOI: 10.1520/E3142-18.10.1520/E3142-18A.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E3142 − 18a
2. Referenced Documents
2
2.1 ASTM Standards:
E473 Terminology Relating to Thermal Analysis and Rheology
E698 Test Method for Kinetic Parameters for Thermally Unstable Materials Using Differential Scanning Calorimetry and the
Flynn/Wall/Ozawa Method
E967 Test Method for Temperature Calibration of Di
...