ASTM D6508-00(2005)e2
(Test Method)Standard Test Method for Determination of Dissolved Inorganic Anions in Aqueous Matrices Using Capillary Ion Electrophoresis and Chromate Electrolyte
Standard Test Method for Determination of Dissolved Inorganic Anions in Aqueous Matrices Using Capillary Ion Electrophoresis and Chromate Electrolyte
SIGNIFICANCE AND USE
Capillary ion electrophoresis provides a simultaneous separation and determination of several inorganic anions using nanolitres of sample in a single injection. All anions present in the sample matrix will be visualized yielding an anionic profile of the sample.
Analysis time is less than 5 minutes with sufficient sensitivity for drinking water and wastewater applications. Time between samplings is less than seven minutes allowing for high sample throughput.
Minimal sample preparation is necessary for drinking water and wastewater matrices. Typically, only a dilution with water is needed.
This test method is intended as an alternative to other multi-analyte methods and various wet chemistries for the determination of inorganic anions in water and wastewater. Compared to other multi-analyte methods the major benefits of CIE are speed of analysis, simplicity, and reduced reagent consumption and operating costs.
SCOPE
1.1 This test method cover the determination of the inorganic anions fluoride, bromide, chloride, nitrite, nitrate, ortho-phosphate, and sulfate in drinking water, wastewater, and other aqueous matrices using capillary ion electrophoresis (CIE) with indirect UV detection. See Figs. 1-6.
1.2 The test method uses a chromate-based electrolyte and indirect UV detection at 254 nm. It is applicable for the determination or inorganic anions in the range of 0.1 to 50 mg/L except for fluoride whose range is 0.1 to 25 mg/L.
1.3 It is the responsibility of the user to ensure the validity of this test method for other anion concentrations and untested aqueous matrices.
Note 1—The highest accepted anion concentration submitted for precision and bias extend the anion concentration range for the following anions: Chloride to 93 mg/L, Sulfate to 90 mg/L, Nitrate to 72 mg/L, and ortho-phosphate to 58 mg/L.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific hazard statements, see Section 9.
General Information
Relations
Buy Standard
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Please contact ASTM International (www.astm.org) for the latest information.
´2
Designation: D6508 – 00 (Reapproved 2005)
Standard Test Method for
Determination of Dissolved Inorganic Anions in Aqueous
Matrices Using Capillary Ion Electrophoresis and Chromate
Electrolyte
This standard is issued under the fixed designation D6508; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
´ NOTE—Warning notes were moved into the text in January 2005.
´ NOTE—Added research report reference to Section 17 editorially in March 2008.
1. Scope
1.1 This test method cover the determination of the inor-
ganic anions fluoride, bromide, chloride, nitrite, nitrate, ortho-
phosphate, and sulfate in drinking water, wastewater, and other
aqueous matrices using capillary ion electrophoresis (CIE)
with indirect UV detection. See Figs. 1-6.
1.2 The test method uses a chromate-based electrolyte and
indirect UV detection at 254 nm. It is applicable for the
determination or inorganic anions in the range of 0.1 to 50
mg/L except for fluoride whose range is 0.1 to 25 mg/L.
1.3 It is the responsibility of the user to ensure the validity
of this test method for other anion concentrations and untested
aqueous matrices.
NOTE 1—The highest accepted anion concentration submitted for
FIG. 1 Electropherogram of Mixed Anion Working Solution and
precision and bias extend the anion concentration range for the following
Added Common Organic Acids
anions: Chloride to 93 mg/L, Sulfate to 90 mg/L, Nitrate to 72 mg/L, and
ortho-phosphate to 58 mg/L.
1.4 This standard does not purport to address all of the
D2777 Practice for Determination of Precision and Bias of
safety concerns, if any, associated with its use. It is the
Applicable Test Methods of Committee D19 on Water
responsibility of the user of this standard to establish appro-
D3370 Practices for Sampling Water from Closed Conduits
priate safety and health practices and determine the applica-
bility of regulatory limitations prior to use. For specific hazard
statements, see Section 9.
2. Referenced Documents
2.1 ASTM Standards:
D1066 Practice for Sampling Steam
D1129 Terminology Relating to Water
D1193 Specification for Reagent Water
This test method is under the jurisdiction of ASTM Committee D19 on Water
and is the direct responsibility of Subcommittee D19.05 on Inorganic Constituents
in Water.
Current edition approved Jan. 13, 2005. Published April 2005.
Originally approved in 2000. Last previous edition approved in 2000 as
D6508 – 00. DOI: 10.1520/D6508-00R05E02.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on FIG. 2 Electropherogram of 0.2 mg/L Anions Used to Determine
the ASTM website. MDL
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Please contact ASTM International (www.astm.org) for the latest information.
´2
D6508 – 00 (2005)
FIG. 3 Electropherogram of Substitute Wastewater
FIG. 6 Electropherogram of Industrial Wastewater
F488 Test Method for On-Site Screening of Heterotrophic
Bacteria in Water
3. Terminology
3.1 Definitions—For definitions of terms used in this test
method, refer to Terminology D1129.
3.2 Definitions of Terms Specific to This Standard:
3.2.1 capillary ion electrophoresis, n—an electrophoretic
techniqueinwhichaUV-absorbingelectrolyteisplacedina50
µmto75µmfusedsilicacapillary.Voltageisappliedacrossthe
capillary causing electrolyte and anions to migrate towards the
anode and through the capillary’s UVdetector window.Anions
are separated based upon the their differential rates of migra-
tion in the electrical field.Anion detection and quantitation are
FIG. 4 Electropherogram of Drinking Water
based upon the principles of indirect UV detection.
3.2.2 electrolyte, n—a combination of a UV-absorbing salt
andanelectroosmoticflowmodifierplacedinsidethecapillary,
used as a carrier for the analytes, and for detection and
quantitation. The UV-absorbing portion of the salt must be
anionic and have an electrophoretic mobility similar to the
analyte anions of interest.
3.2.3 electroosmotic flow (EOF), n—the direction and ve-
locity of electrolyte solution flow within the capillary under an
applied electrical potential (voltage); the velocity and direction
of flow is determined by electrolyte chemistry, capillary wall
chemistry, and applied voltage.
3.2.4 electroosmotic flow modifier (OFM), n—a cationic
quaternary amine in the electrolyte that dynamically coats the
negatively charged silica wall giving it a net positive charge.
This reverses the direction of the electrolyte’s natural elec-
FIG. 5 Electropherogram of Municipal Wastewater Treatment
Plant Discharge troosmotic flow and directs it towards the anode and detector.
Thismodifieraugmentsanionmigrationandenhancesspeedof
analysis. Its concentration secondarily effects anion selectivity
and resolution, (see Fig. 7).
D3856 Guide for Good Laboratory Practices in Laborato-
3.2.5 electrophoretic mobility, n—the specific velocity of a
ries Engaged in Sampling and Analysis of Water
charged analyte in the electrolyte under specific electroosmotic
D5810 Guide for Spiking into Aqueous Samples
flow conditions. The mobility of an analyte is directly related
D5847 Practice for Writing Quality Control Specifications
for Standard Test Methods for Water Analysis
D5905 Practice for the Preparation of Substitute Wastewa-
Withdrawn. The last approved version of this historical standard is referenced
ter on www.astm.org.
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Please contact ASTM International (www.astm.org) for the latest information.
´2
D6508 – 00 (2005)
FIG. 7 Pictorial Diagram of Anion Mobility and ElectroOsomotic Flow Modifier
to the analyte’s equivalent ionic conductance and applied 4. Summary of Test Method
voltage, and is the primary mechanism of separation.
4.1 Capillary ion electrophoresis, see Figs. 7-10, is a free
3.2.6 electropherogram, n—a graphical presentation of UV- zone electrophoretic technique optimized for the determination
detector response versus time of analysis; the x axis is
of anions with molecular weight less than 200. The anions
migration time, which is used to qualitatively identify the migrate and are separated according to their mobility in the
anion, and the y axis is UV response, which can be converted
electrolyte when an electrical field is applied through the open
to time corrected peak area for quantitation. tubular fused silica capillary. The electrolyte’s electroosmotic
low modifier dynamically coats the inner wall of the capillary
3.2.7 hydrostatic sampling, n—a sample introduction tech-
changing the surface to a net positive charge. This reversal of
nique in which the capillary with electrolyte is immersed in the
wall charge reverses the natural EOF. The modified EOF in
sample, and both are elevated to a specific height, typically 10
combination with a negative power supply augments the
cm, above the receiving electrolyte reservoir for a preset
mobility of the analyte anions towards the anode and detector
amount of time, typically less than 60 s. Nanolitres of sample
achieving rapid analysis times. Cations migrate in the opposite
aresiphonedintothecapillarybydifferentialheadpressureand
directiontowardsthecathodeandareremovedfromthesample
gravity.
during analysis. Water and other neutral species move toward
3.2.8 indirect UV detection, n—a form of UV detection in
the detector at the same rate as the EOF. The neutral species
which the analyte displaces an equivalent net charge amount of
migrate slower than the analyte anions and do not interfere
the highly UV-absorbing component of the electrolyte causing
with anion analysis (see Figs. 7 and 8).
anetdecreaseinbackgroundabsorbance.Themagnitudeofthe
4.2 The sample is introduced into the capillary using hydro-
decreased absorbance is directly proportional to analyte con-
staticsampling.Theinletofthecapillarycontainingelectrolyte
centration. Detector output polarity is reversed in order to
is immersed in the sample and the height of the sample raised
obtain a positive mV response.
3.2.9 midpoint of peak width, n—CIE peaks typically are
asymmetrical with the peak apex shifting with increasing
concentration, and the peak apex may not be indicative of true
analyte migration time. Midpoint of peak width is the midpoint
between the analyte peak’s start and stop integration, or the
peak center of gravity.
3.2.10 migration time, n—the time required for a specific
analyte to migrate through the capillary to the detector. The
migration time in capillary ion electrophoresis is analogous to
retention time in chromatography.
3.2.11 time corrected peak area, n—normalized peak area;
peak area divided by migration time. CE principles state that
peak area is dependent upon migration time, that is, for the
same concentration of analyte, as migration time increases
(decreases) peak area increases (decreases). Time corrected
FIG. 8 Selectivity Diagram of Anion Mobility Using Capillary Ion
peak area accounts for these changes. Electrophoresis
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Please contact ASTM International (www.astm.org) for the latest information.
´2
D6508 – 00 (2005)
quantitation can be achieved when the concentration differen-
tial is less than 100:1. As the resolution between two anion
peaks increase so does the tolerated concentration differential.
In samples containing 1000 mg/L Cl, 1 mg/L SO can be
resolved and quantitated, however, the high Cl will interfere
with Br and NO quantitation.
-1
6.2 Dissolved carbonate, detected as HCO , is an anion
present in all aqueous samples, especially alkaline samples.
Carbonate concentrations greater than 500 mg/L will interfere
with PO quantitation.
6.3 Monovalent organic acids, except for formate, and
neutral organics commonly found in wastewater migrate later
in the electropherogram, after carbonate, and do not interfere.
Formate, a common organic acid found in environmental
samples, migrates shortly after fluoride but before phosphate.
FIG. 9 Pictorial Diagram of Indirect UV Detection
Formate concentrations greater than 5 mg/Lwill interfere with
fluoride identification and quantitation. Inclusion of 2 mg/L
10 cm for 30 s where low nanolitre volumes are siphoned into
formate into the mixed anion working solution aids in fluoride
thecapillary.Aftersampleloading,thecapillaryisimmediately
and formate identification and quantitation.
immersed back into the electrolyte. The voltage is applied
6.4 Divalent organic acids usually found in wastewater
initiating the separation process.
migrate after phosphate. At high concentrations, greater than
4.3 Anion detection is based upon the principles of indirect
10 mg/L, they may interfere with phosphate identification and
UVdetection.The UV-absorbing electrolyte anion is displaced
quantitation.
charge-for-charge by the separated analyte anion. The analyte
6.5 Chlorate also migrates after phosphate and at concen-
anion zone has a net decrease in background absorbance. This
trations greater than 10 mg/L will interfere with phosphate
decrease in UV absorbance in quantitatively proportional to
identification and quantitation. Inclusion of 5 mg/L chlorate
analyte anion concentration (see Fig. 9). Detector output
into the mixed anion working solution aids in phosphate and
polarityisreversedtoprovidepositivemVresponsetothedata
chlorate identification and quantitation.
system, and to make the negative absorbance peaks appear
6.6 As analyte concentration increases, analyte peak shape
positive.
becomes asymmetrical. If adjacent analyte peaks are not
4.4 The analysis is complete once the last anion of interest
baseline resolved, the data system will drop a perpendicular
is detected. The capillary is vacuum purged automatically by
between them to the baseline. This causes a decrease in peak
the system of any remaining sample and replenished with fresh
area for both analyte peaks and a low bias for analyte amounts.
electrolyte. The system now is ready for the next analysis.
For optimal quantitation, insure that adjacent peaks are fully
5. Significance and Use resolved, if they are not, dilute the sample 1:1 with water.
5.1 Capillary ion electrophoresis provides a simultaneous
7. Apparatus
separation and determination of several inorganic anions using
nanolitres of sample in a single injection.All anions present in
7.1 Capillary Ion Electrophoresis System—the system con-
the sample matrix will be visualized yielding an anionic profile
sists of the following components, as shown in Fig. 10 or
of the sample.
equivalent:
5.2 Analysis time is less than 5 minutes with sufficient
7.1.1 High Voltage Power Supply, capable of generating
sensitivity for drinking water and wastewater applications.
voltage (potential) between 0 and minus 30 kV relative to
Time between samplings is less than seven minutes allowing
ground with the capability working in a constant current mode.
for high sample throughput.
7.1.2 Covered Sample Carousel, to prevent environmental
5.3 Minimal sample preparation is necessary for drinking
contamination of the samples and electrolytes during a multi-
water and wastewater matrices. Typically, only a dilution with
sample batch analysis.
water is needed.
7.1.3 Sample Introduction Mechanism, capable of hydro-
5.4 This test method is intended as an alternative to other
static sampling technique, using gravity, positive pressure, or
multi-analyte methods and various wet chemistries for the
equivalent.
determination of inorganic anions in water and wastewater.
7.1.4 Capillary Purge Mechanism, to purge the capillary
Compared to other multi-analyte methods the major benefits of
after every analysis with fresh electrolyte to eliminate any
CIE are speed of analysis, simplicity, and reduced reagent
interference from the previous sample matrix, and to clean the
consumption and operating costs.
capillary with other reagent, such as sodium hydroxide.
6. Interferences
7.1.5 UV Detector, having the capability of monitoring 254
nm, or equivalent, with a time constant of 0.3 s.
6.1 Analyte identification, quantitation, and possible comi-
gration occur when one anion is in significant
...
This document is not anASTM standard and is intended only to provide the user of anASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
e1 e1
Designation: D 6508 – 00 (Reapproved 2005) D 6508 – 00 (2005) e2
Standard Test Method for
Determination of Dissolved Inorganic Anions in Aqueous
Matrices Using Capillary Ion Electrophoresis and Chromate
Electrolyte
This standard is issued under the fixed designation D 6508; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
e NOTE—Warning notes were moved into the text in January 2005.
—Warning notes were moved into the text in January 2005.
e NOTE—Added research report reference to Section 17 editorially in March 2008.
1. Scope
1.1 This test method cover the determination of the inorganic anions fluoride, bromide, chloride, nitrite, nitrate, ortho-
phosphate, and sulfate in drinking water, wastewater, and other aqueous matrices using capillary ion electrophoresis (CIE) with
indirect UV detection. See Figs. 1-6.
1.2 The test method uses a chromate-based electrolyte and indirect UV detection at 254 nm. It is applicable for the
determination or inorganic anions in the range of 0.1 to 50 mg/L except for fluoride whose range is 0.1 to 25 mg/L.
1.3 It is the responsibility of the user to ensure the validity of this test method for other anion concentrations and untested
aqueous matrices.
NOTE 1—The highest accepted anion concentration submitted for precision and bias extend the anion concentration range for the following anions:
Chloride to 93 mg/L, Sulfate to 90 mg/L, Nitrate to 72 mg/L, and ortho-phosphate to 58 mg/L.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use. For specific hazard statements, see Section 9.
2. Referenced Documents
2.1 ASTM Standards:
D 1066 Practice for Sampling Steam
D 1129 Terminology Relating to Water
D 1193 Specification for Reagent Water
D 2777 Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water
D 3370 Practices for Sampling Water from Closed Conduits
D 3856 Guide for Good Laboratory Practices in Laboratories Engaged in Sampling and Analysis of Water
D 5810 Guide for Spiking into Aqueous Samples
D 5847 Practice for Writing Quality Control Specifications for Standard Test Methods for Water Analysis
D 5905 Practice for the Preparation of Substitute Wastewater
F 488 Test Method for On-Site Screening of Heterotrophic Bacteria in Water
3. Terminology
3.1 Definitions— For definitions of terms used in this test method, refer to Terminology D 1129.
3.2 Definitions of Terms Specific to This Standard:
3.2.1 capillary ion electrophoresis, n— an electrophoretic technique in which a UV-absorbing electrolyte is placed in a 50 µm
to 75 µm fused silica capillary. Voltage is applied across the capillary causing electrolyte and anions to migrate towards the anode
and through the capillary’s UV detector window. Anions are separated based upon the their differential rates of migration in the
electrical field. Anion detection and quantitation are based upon the principles of indirect UV detection.
ThistestmethodisunderthejurisdictionofASTMCommitteeD19onWaterandisthedirectresponsibilityofSubcommitteeD19.05onInorganicConstituentsinWater.
Current edition approved Jan. 13, 2005. Published April 2005.
Originally approved in 2000. Last previous edition approved in 2000 as D 6508 – 00.
For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
e2
FIG. 1 Electropherogram of Mixed Anion Working Solution and
Added Common Organic Acids
FIG. 2 Electropherogram of 0.2 mg/L Anions Used to Determine
MDL
FIG. 3 Electropherogram of Substitute Wastewater
3.2.2 electrolyte, n—a combination of a UV-absorbing salt and an electroosmotic flow modifier placed inside the capillary, used
as a carrier for the analytes, and for detection and quantitation. The UV-absorbing portion of the salt must be anionic and have
an electrophoretic mobility similar to the analyte anions of interest.
3.2.3 electroosmotic flow (EOF), n—the direction and velocity of electrolyte solution flow within the capillary under an applied
electrical potential (voltage); the velocity and direction of flow is determined by electrolyte chemistry, capillary wall chemistry,
and applied voltage.
3.2.4 electroosmotic flow modifier (OFM), n— a cationic quaternary amine in the electrolyte that dynamically coats the
e2
FIG. 4 Electropherogram of Drinking Water
FIG. 5 Electropherogram of Municipal Wastewater Treatment
Plant Discharge
FIG. 6 Electropherogram of Industrial Wastewater
negatively charged silica wall giving it a net positive charge. This reverses the direction of the electrolyte’s natural electroosmotic
flow and directs it towards the anode and detector. This modifier augments anion migration and enhances speed of analysis. Its
concentration secondarily effects anion selectivity and resolution, (see Fig. 7).
3.2.5 electrophoretic mobility, n—the specific velocity of a charged analyte in the electrolyte under specific electroosmotic flow
conditions. The mobility of an analyte is directly related to the analyte’s equivalent ionic conductance and applied voltage, and
is the primary mechanism of separation.
3.2.6 electropherogram, n—a graphical presentation of UV-detector response versus time of analysis; the x axis is migration
e2
FIG. 7 Pictorial Diagram of Anion Mobility and ElectroOsomotic Flow Modifier
time, which is used to qualitatively identify the anion, and the y axis is UV response, which can be converted to time corrected
peak area for quantitation.
3.2.7 hydrostatic sampling, n—a sample introduction technique in which the capillary with electrolyte is immersed in the
sample, and both are elevated to a specific height, typically 10 cm, above the receiving electrolyte reservoir for a preset amount
of time, typically less than 60 s. Nanolitres of sample are siphoned into the capillary by differential head pressure and gravity.
3.2.8 indirect UV detection, n—a form of UV detection in which the analyte displaces an equivalent net charge amount of the
highly UV-absorbing component of the electrolyte causing a net decrease in background absorbance. The magnitude of the
decreased absorbance is directly proportional to analyte concentration. Detector output polarity is reversed in order to obtain a
positive mV response.
3.2.9 midpoint of peak width, n—CIE peaks typically are asymmetrical with the peak apex shifting with increasing
concentration, and the peak apex may not be indicative of true analyte migration time. Midpoint of peak width is the midpoint
between the analyte peak’s start and stop integration, or the peak center of gravity.
3.2.10 migration time, n—the time required for a specific analyte to migrate through the capillary to the detector.The migration
time in capillary ion electrophoresis is analogous to retention time in chromatography.
3.2.11 time corrected peak area, n—normalized peak area; peak area divided by migration time. CE principles state that peak
area is dependent upon migration time, that is, for the same concentration of analyte, as migration time increases (decreases) peak
area increases (decreases). Time corrected peak area accounts for these changes.
4. Summary of Test Method
4.1 Capillary ion electrophoresis, see Figs. 7-10, is a free zone electrophoretic technique optimized for the determination of
anions with molecular weight less than 200. The anions migrate and are separated according to their mobility in the electrolyte
when an electrical field is applied through the open tubular fused silica capillary. The electrolyte’s electroosmotic low modifier
dynamicallycoatstheinnerwallofthecapillarychangingthesurfacetoanetpositivecharge.Thisreversalofwallchargereverses
FIG. 8 Selectivity Diagram of Anion Mobility Using Capillary Ion
Electrophoresis
e2
FIG. 9 Pictorial Diagram of Indirect UV Detection
FIG. 10 General Hardware Schematic of a Capillary Ion Electrophoresis System
the natural EOF. The modified EOF in combination with a negative power supply augments the mobility of the analyte anions
towards the anode and detector achieving rapid analysis times. Cations migrate in the opposite direction towards the cathode and
are removed from the sample during analysis. Water and other neutral species move toward the detector at the same rate as the
EOF. The neutral species migrate slower than the analyte anions and do not interfere with anion analysis (see Figs. 7 and 8).
4.2 The sample is introduced into the capillary using hydrostatic sampling. The inlet of the capillary containing electrolyte is
immersed in the sample and the height of the sample raised 10 cm for 30 s where low nanolitre volumes are siphoned into the
capillary. After sample loading, the capillary is immediately immersed back into the electrolyte. The voltage is applied initiating
the separation process.
4.3 Anion detection is based upon the principles of indirect UV detection. The UV-absorbing electrolyte anion is displaced
charge-for-charge by the separated analyte anion. The analyte anion zone has a net decrease in background absorbance. This
decrease in UV absorbance in quantitatively proportional to analyte anion concentration (see Fig. 9). Detector output polarity is
reversed to provide positive mV response to the data system, and to make the negative absorbance peaks appear positive.
4.4 The analysis is complete once the last anion of interest is detected. The capillary is vacuum purged automatically by the
system of any remaining sample and replenished with fresh electrolyte. The system now is ready for the next analysis.
5. Significance and Use
5.1 Capillary ion electrophoresis provides a simultaneous separation and determination of several inorganic anions using
nanolitres of sample in a single injection.All anions present in the sample matrix will be visualized yielding an anionic profile of
the sample.
5.2 Analysistimeislessthan5minuteswithsufficientsensitivityfordrinkingwaterandwastewaterapplications.Timebetween
samplings is less than seven minutes allowing for high sample throughput.
e2
5.3 Minimal sample preparation is necessary for drinking water and wastewater matrices. Typically, only a dilution with water
is needed.
5.4 This test method is intended as an alternative to other multi-analyte methods and various wet chemistries for the
determination of inorganic anions in water and wastewater. Compared to other multi-analyte methods the major benefits of CIE
are speed of analysis, simplicity, and reduced reagent consumption and operating costs.
6. Interferences
6.1 Analyte identification, quantitation, and possible comigration occur when one anion is in significant excess to other anions
in the sample matrix. For two adjacent peaks, reliable quantitation can be achieved when the concentration differential is less than
100:1.As the resolution between two anion peaks increase so does the tolerated concentration differential. In samples containing
1000 mg/L Cl, 1 mg/L SO can be resolved and quantitated, however, the high Cl will interfere with Br and NO quantitation.
4 2
-1
6.2 Dissolvedcarbonate,detectedasHCO ,isananionpresentinallaqueoussamples,especiallyalkalinesamples.Carbonate
concentrations greater than 500 mg/L will interfere with PO quantitation.
6.3 Monovalent organic acids, except for formate, and neutral organics commonly found in wastewater migrate later in the
electropherogram,aftercarbonate,anddonotinterfere.Formate,acommonorganicacidfoundinenvironmentalsamples,migrates
shortly after fluoride but before phosphate. Formate concentrations greater than 5 mg/L will interfere with fluoride identification
and quantitation. Inclusion of 2 mg/Lformate into the mixed anion working solution aids in fluoride and formate identification and
quantitation.
6.4 Divalent organic acids usually found in wastewater migrate after phosphate.At high concentrations, greater than 10 mg/L,
they may interfere with phosphate identification and quantitation.
6.5 Chlorate also migrates after phosphate and at concentrations greater than 10 mg/L will interfere with phosphate
identification and quantitation. Inclusion of 5 mg/Lchlorate into the mixed anion working solution aids in phosphate and chlorate
identification and quantitation.
6.6 As analyte concentration increases, analyte peak shape becomes asymmetrical. If adjacent analyte peaks are not baseline
resolved, the data system will drop a perpendicular between them to the baseline. This causes a decrease in peak area for both
analyte peaks and a low bias for analyte amounts. For optimal quantitation, insure that adjacent peaks are fully resolved, if they
are not, dilute the sample 1:1 with water.
7. Apparatus
7.1 Capillary Ion Electrophoresis System—the system consists of the following components, as shown in Fig. 10 or equivalent:
7.1.1 High Voltage Power Supply, capable of generating voltage (potential) between 0 and minus 30 kVrelative to ground with
the capability working in a constant current mode.
7.1.2 Covered Sample Carousel, to prevent environmental contamination of the samples and electrolytes during a multisample
batch analysis.
7.1.3 Sample Introduction Mechanism , capable of hydrostatic sampling technique, using gravity, positive pressure, or
equivalent.
7.1.4 Capillary Purge Mechanism, to purge the capillary after every analysis with fresh electrolyte to eliminate any interference
from the previous sample matrix, and to clean the capillary with other reagent, such as sodium hydroxide.
7.1.5 UV Detector, having the capability of monitoring 254 nm, or equivalent, with a time constant of 0.3 s.
7.1.6 Fused Silica Capillary—A75 µm (inner diameter) x 375 µm (outer diameter) x 60 cm
...










Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.