Abstract

This document specifies methods for the determination of silicon, nitrogen, aluminium, iron, calcium, oxygen, carbon, fluorine, and chlorine in silicon nitride powders.
This document is applicable for the chemical analysis of silicon nitride powders used as raw material for fine ceramics.

Status
Published
Public Enquiry End Date
19-Mar-2026
Publication Date
18-Nov-2026
Technical Committee
I13 - Imaginarni 13
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
16-Nov-2026
Due Date
21-Jan-2027
Completion Date
19-Nov-2026

Buy Documents

Standard

SIST EN ISO 17947-1:2027

English language (40 pages)
Preview
Preview
e-Library read for
×1 day

SIST EN ISO 17947-1:2027 is the Slovenian adoption of ISO 17947-1 for chemical analysis of silicon nitride powders. It specifies methods for determining silicon, nitrogen, aluminium, iron, calcium, oxygen, carbon, fluorine, and chlorine in silicon nitride powders used as raw material for fine ceramics. Laboratories, producers, and buyers use it to obtain comparable results for material control, specification checking, and quality records.

What does SIST EN ISO 17947-1:2027 specify?

SIST EN ISO 17947-1:2027 specifies chemical analysis methods for silicon nitride powders used as raw material for fine ceramics. The document covers major constituents and trace elements, so it is aimed at routine control as well as lower-level impurity checking.

The scope is limited to silicon nitride powder analysis. The analytes and stated mass-fraction ranges are silicon, nitrogen, aluminium, iron, calcium, oxygen, carbon, fluorine, and chlorine, which helps the user see whether the standard fits the expected concentration level of the sample.

The document is organised into clauses on scope, normative references, analytes and ranges, sample preparation, apparatus and reagents, blank test, determination methods for each group of elements, reporting of analytical values, and the test report. It also includes three informative annexes that support method selection and interpretation.

AnnexWhat it covers
Annex AList of commercial certified reference materials
Annex BAnalytical results obtained from a round robin test
Annex CSpectral lines for ICP-OES

What are the key requirements of SIST EN ISO 17947-1:2027?

SIST EN ISO 17947-1:2027 is built around element-specific procedures, but the common practice is the same throughout: prepare the sample in a controlled way, run a blank, calibrate the instrument or titration system, and correct the result with the blank value. That matters because these powders are analysed by several different techniques, and the final value depends on consistent handling.

Sample preparation and general laboratory practice

Clause 4 requires sampling in accordance with ISO 8656-1 unless the analyser and customer agree otherwise. The sample is dried at 110 °C ± 5 °C for 2 h and then cooled in a desiccator, and the test portion is weighed to the nearest 0,1 mg. In practice, this controls moisture and mass error before analysis.

Clause 5 points the user to ordinary laboratory apparatus in ISO 26845 and reagents that meet ISO 6353-1, ISO 6353-2, and ISO 6353-3 as appropriate. This matters because the document relies on standard laboratory water, acids, salts, and reference reagents rather than ad hoc materials.

Clause 6 requires a blank test for each determination using the same reagents, conditions, and procedure as the sample. That blank is used to correct the reported analytical value.

Silicon and nitrogen

Clause 7 gives two options for silicon. Method A uses fusion, separation of insoluble and soluble silicon, gravimetry for the insoluble fraction, and ICP-OES for the soluble fraction. Method B uses XRF with the fused cast-bead method. The document says Method A is used when four figures are required, while Method B can be used when two or three figures are enough.

Clause 8 gives three nitrogen methods. Acid pressure decomposition, distillation, and acidimetric titration are used when four figures are required. Inert gas fusion with thermal conductivity is the faster instrumental route when two figures are required. Fusion-ammonia separation with acidimetric titration is also allowed. For the pressure-decomposition method, the recovery check with ammonium sulfate must be at least 99 %, so the user has to verify that the decomposition and titration system is working properly.

For the inert gas fusion method, the calibration material should have known nitrogen content and a similar oxygen level to the sample. That point matters because the document warns that a large oxygen difference can introduce error.

Aluminium, iron, calcium, oxygen, carbon, fluorine, and chlorine

Clause 9 determines aluminium, iron, and calcium after pressure decomposition with nitric acid and hydrofluoric acid, removal of silicon, and ICP-OES measurement. The standard tells the user to choose wavelengths with attention to spectral interference and sensitivity, which is why Annex C lists example spectral lines.

Clause 10 determines oxygen by inert gas fusion under helium with a graphite crucible, followed by infrared measurement of carbon monoxide or carbon dioxide. The calibration can use yttrium oxide or a nitride with known oxygen content. This matters because oxygen is measured by a gas-reaction method rather than a wet chemical one.

Clause 11 covers carbon. The main method is combustion with an induction furnace and IR absorption spectrometry, but resistance furnace methods with coulometry or thermal conductivity are also allowed through ISO 21068-2. The calibration can use calcium carbonate or a nitride with known carbon content.

Clause 12 determines fluorine and chlorine after pyrohydrolysis under steam and oxygen. The collected analytes can be measured by ion chromatography or, for separate determinations, by spectrophotometry. For chlorine, the spectrophotometric route uses the mercury(II)-thiocyanate and ammonium-iron(II)-sulfate reaction.

Reporting the result

Clause 13 requires two analyses on different days. If the two values agree within the tolerance in Table 1, report the average. If they do not, perform two more analyses and then report the average or, if needed, the median of four values. Results are reported as mass fractions in percent for the dried sample, with two decimal places for silicon and nitrogen and three decimal places for the other analytes.

Clause 14 defines the minimum test report content: sample identification, laboratory and test date, a reference to the standard, the method used, the results and expression format, unusual features, and any non-standard or optional operation that could affect the result.

What terms does SIST EN ISO 17947-1:2027 define?

Analyte - An analyte is the element being determined, such as silicon, nitrogen, or chlorine.

Mass fraction - Mass fraction is the way the results are expressed in this standard, as percent of the dried sample.

Blank test - A blank test is a run with the same reagents and procedure but without sample, used to correct the measured value.

Inductively coupled plasma optical emission spectrometry (ICP-OES) - ICP-OES measures element-specific light emitted from a plasma after the sample solution is introduced.

X-ray fluorescence (XRF) - XRF identifies and quantifies elements from their fluorescent X-rays, and this standard uses the fused cast-bead method for silicon.

Carrier-gas hot extraction (CGHE) - CGHE is one of the named method families in the title and refers to gas extraction under a carrier gas for elemental determination.

Pyrohydrolysis - Pyrohydrolysis is the steam-and-oxygen treatment used in this standard to release fluorine and chlorine for measurement.

Who uses SIST EN ISO 17947-1:2027?

SIST EN ISO 17947-1:2027 is used by analytical laboratories, fine ceramics producers, and quality-control teams working with silicon nitride raw powders. It is also relevant to suppliers and buyers who need agreed test methods for incoming inspection, product release, or contract specification checks.

The document is especially useful where a material needs several different analytical techniques in one control plan, for example ICP-OES for metals, combustion methods for carbon, and gas extraction methods for oxygen and nitrogen. It also supports laboratories that prepare certificates of analysis or compare results across multiple sites.

What changed in SIST EN ISO 17947-1:2027 from the previous edition?

The front matter identifies the document as ISO 17947-1 and shows the draft stage ISO/DIS 17947-1:2025, while the Slovene publication is marked as prEN ISO 17947-1. The introduction also says the International Standard was developed from JIS R 1603:2007 with reference to CEN ENV 14226:2002 and ASTM C1494-01:2007.

Which standards are used with SIST EN ISO 17947-1:2027?

SIST EN ISO 17947-1:2027 relies on several normative references that supply sampling, reagent quality, and companion analytical methods.

  • ISO 8656-1 - gives the sampling scheme for raw materials and unshaped products.
  • ISO 26845 - gives general requirements for wet chemical analysis, AAS, and ICP-AES/ICP-OES practice.
  • ISO 3696 - specifies laboratory water quality.
  • ISO 6353-1, ISO 6353-2, ISO 6353-3 - specify reagent test methods and reagent grades.
  • ISO 21068-2 - provides alternative procedures for silicon and carbon-related determinations.
  • ISO 21068-3 - provides alternative procedures for nitrogen and silicon-related determinations, including fused cast-bead XRF and fusion-ammonia separation.
  • ISO 21438-2 and ISO 21438-3 - support ion chromatographic determination of chlorine and fluorine.
  • ISO 2828 - supports the spectrophotometric method for fluorine.

What does the SIST EN ISO 17947-1:2027 document contain?

The document contains detailed analytical procedures, reagent lists, apparatus descriptions, calibration steps, blank corrections, and calculation formulae for each analyte. It also includes step-by-step set-ups for pressure decomposition vessels, steam distillation, pyrohydrolysis, and gas-analysis instruments.

Figures show example apparatus for nitrogen decomposition and distillation and block diagrams for the instrumental systems used in nitrogen, oxygen, and carbon analysis. Table 1 gives tolerance limits for analytical values, while Annex B presents round robin results that support the method’s performance history.

Annex A lists commercial certified reference materials, which is useful when laboratories need materials for calibration or control. Annex C lists example ICP-OES spectral lines for aluminium, calcium, iron, and silicon, which helps the user select a wavelength while managing spectral interference.

Relations

Effective Date
15-Nov-2023

Buy Documents

Standard

SIST EN ISO 17947-1:2027

English language (40 pages)
Preview
Preview
e-Library read for
×1 day

Frequently Asked Questions

SIST EN ISO 17947-1:2027 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Fine ceramics (advanced ceramics, advanced technical ceramics) - Methods for chemical analysis of silicon nitride powders - Part 1: Wet chemical methods, X-ray fluorescence (XRF) using the fused cast-bead method, carrier-gas hot extraction (CGHE) and combustion methods (ISO 17947-1:2026)". This standard covers: This document specifies methods for the determination of silicon, nitrogen, aluminium, iron, calcium, oxygen, carbon, fluorine, and chlorine in silicon nitride powders. This document is applicable for the chemical analysis of silicon nitride powders used as raw material for fine ceramics.

This document specifies methods for the determination of silicon, nitrogen, aluminium, iron, calcium, oxygen, carbon, fluorine, and chlorine in silicon nitride powders. This document is applicable for the chemical analysis of silicon nitride powders used as raw material for fine ceramics.

SIST EN ISO 17947-1:2027 is classified under the following ICS (International Classification for Standards) categories: 81.060.30 - Advanced ceramics. The ICS classification helps identify the subject area and facilitates finding related standards.

SIST EN ISO 17947-1:2027 has the following relationships with other standards: It is inter standard links to SIST EN ISO 17947:2023. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

SIST EN ISO 17947-1:2027 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


SLOVENSKI STANDARD
01-januar-2027
Fina keramika (sodobna keramika, sodobna tehnična keramika) - Metode za
kemijsko analizo finih praškov silicijevega nitrida - 1. del: Mokre kemijske metode,
rentgenska fluorescenca (XRF) z metodo taljenih krogel, vroča ekstrakcija z
nosilnim plinom (CGHE) in metode zgorevanja (ISO 17947-1:2026)
Fine ceramics (advanced ceramics, advanced technical ceramics) - Methods for
chemical analysis of silicon nitride powders - Part 1: Wet chemical methods, X-ray
fluorescence (XRF) using the fused cast-bead method, carrier-gas hot extraction
(CGHE) and combustion methods (ISO 17947-1:2026)
Hochleistungskeramik - Verfahren zur chemischen Analyse von Pulvern aus
Siliciumnitrid - Teil 1: Nasschemische Verfahren, Röntgenfluoreszenzanalyse (RFA)
unter Anwendung des Schmelzaufschluss-Verfahrens, Trägergasheißextraktion (CGHE)
und Verbrennungsverfahren (ISO 17947-1:2026)
Céramiques techniques - Méthodes pour l’analyse chimique de poudres de nitrure de
silicium - Partie 1: Méthodes chimiques par voie humide, fluorescence de rayons X
(XRF) utilisant la méthode de la perle fondue, extraction à chaud sous gaz vecteur
(CGHE) et méthodes de combustion (ISO 17947-1:2026)
Ta slovenski standard je istoveten z: EN ISO 17947-1:2026
ICS:
81.060.30 Sodobna keramika Advanced ceramics
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EN ISO 17947-1
EUROPEAN STANDARD
NORME EUROPÉENNE
November 2026
EUROPÄISCHE NORM
ICS 81.060.30 Supersedes EN ISO 17947:2023
English Version
Fine ceramics (advanced ceramics, advanced technical
ceramics) - Methods for chemical analysis of silicon nitride
powders - Part 1: Wet chemical methods, X-ray
fluorescence (XRF) using the fused cast-bead method,
carrier-gas hot extraction (CGHE) and combustion
methods (ISO 17947-1:2026)
Céramiques techniques - Méthodes pour l'analyse Hochleistungskeramik - Verfahren zur chemischen
chimique de poudres de nitrure de silicium - Partie 1: Analyse von Pulvern aus Siliciumnitrid - Teil 1:
Méthodes chimiques par voie humide, fluorescence de Nasschemische Verfahren, Röntgenfluoreszenzanalyse
rayons X (XRF) utilisant la méthode de la perle fondue, (RFA) unter Anwendung des Schmelzaufschluss-
extraction à chaud sous gaz vecteur (CGHE) et Verfahrens, Trägergasheißextraktion (CGHE) und
méthodes de combustion (ISO 17947-1:2026) Verbrennungsverfahren (ISO 17947-1:2026)
This European Standard was approved by CEN on 25 April 2026.

CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this
European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references
concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN
member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by
translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management
Centre has the same status as the official versions.

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and
United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION

EUROPÄISCHES KOMITEE FÜR NORMUNG

CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2026 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN ISO 17947-1:2026 E
worldwide for CEN national Members.

Contents Page
European foreword . 3

European foreword
This document (EN ISO 17947-1:2026) has been prepared by Technical Committee ISO/TC 206 "Fine
ceramics" in collaboration with Technical Committee CEN/TC 184 “Advanced technical ceramics” the
secretariat of which is held by DIN.
This European Standard shall be given the status of a national standard, either by publication of an
identical text or by endorsement, at the latest by May 2027, and conflicting national standards shall be
withdrawn at the latest by May 2027.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CEN shall not be held responsible for identifying any or all such patent rights.
This document supersedes EN ISO 17947:2023.
Any feedback and questions on this document should be directed to the users’ national standards
body/national committee. A complete listing of these bodies can be found on the CEN website.
According to the CEN-CENELEC Internal Regulations, the national standards organizations of the
following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria,
Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland,
Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Republic of
North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and the
United Kingdom.
Endorsement notice
The text of ISO 17947-1:2026 has been approved by CEN as EN ISO 17947-1:2026 without any
modification.
International
Standard
ISO 17947-1
First edition
Fine ceramics (advanced ceramics,
2026-10
advanced technical ceramics) —
Methods for chemical analysis of
silicon nitride powders —
Part 1:
Wet chemical methods, X-ray
fluorescence (XRF) using the fused
cast-bead method, carrier-gas hot
extraction (CGHE) and combustion
methods
Céramiques techniques — Méthodes pour l’analyse chimique de
poudres de nitrure de silicium —
Partie 1: Méthodes chimiques par voie humide, fluorescence de
rayons X (XRF) utilisant la méthode de la perle fondue, extraction
à chaud sous gaz vecteur (CGHE) et méthodes de combustion
Reference number
ISO 17947-1:2026(en) © ISO 2026

ISO 17947-1:2026(en)
© ISO 2026
All rights reserved.
This ISO publication is protected by copyright and is owned by ISO and/or its licensors.
The content of this ISO publication is provided under licence, not sold. Use is subject to the applicable licence terms issued by ISO,
an ISO member body, or an authorized third-party distributor.
Except as required for implementation or expressly permitted by a separate licence, no part of this ISO publication may be
reproduced, distributed, modified, used, or made available in any form or by any means – electronic or mechanical, including
photocopying, scanning, recording, or posting on internal or external digital platforms.
Any use beyond the scope of the granted rights is prohibited and may result in legal action.
ISO copyright office
CP 401 • CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii
ISO 17947-1:2026(en)
Contents Page
Foreword .v
Introduction .vi
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 2
4 Analytes and ranges . 2
5 Preparation of test sample . 2
5.1 General .2
5.2 Sampling .2
5.3 Drying .2
5.4 Weighing .3
6 Apparatus and reagents . 3
7 Blank test . 3
8 Determination of silicon . 3
8.1 Classification of determination methods .3
8.2 Fusion-dehydration/insolubilization separation-gravimetry and ICP-OES .3
8.2.1 Principle .3
8.2.2 Reagents .3
8.2.3 Apparatus and instruments .4
8.2.4 Procedure .4
8.2.5 Blank test .5
8.2.6 Calibration . . .5
8.2.7 Calculation .5
8.3 XRF using fused cast-bead method .5
9 Determination of nitrogen . 6
9.1 Classification of determination methods .6
9.2 Acid pressure decomposition-distillation separation-acidimetric titration method .6
9.2.1 Principle .6
9.2.2 Reagents .6
9.2.3 Apparatus .7
9.2.4 Procedure .10
9.2.5 Recovery measurement.11
9.2.6 Calculation .11
9.3 Inert gas fusion-thermal conductivity method .11
9.4 Fusion-ammonia separation-acidimetric titration method .14
10 Determination of aluminium, iron, and calcium . 14
10.1 Principle .14
10.2 Reagents .14
10.3 Apparatus .14
10.4 Procedure . 15
10.5 Blank test . 15
10.6 Calibration . 15
10.7 Calculation . 15
11 Determination of oxygen. 16
11.1 Principle .16
11.2 Reagents .16
11.3 Apparatus .16
11.4 Instrument .16
11.5 Procedure .16
11.6 Blank test .16

iii
ISO 17947-1:2026(en)
11.7 Calculation of calibration coefficient .17
11.8 Calculation .17
12 Determination of carbon . 17
12.1 Classification of determination methods .17
12.2 Combustion (induction furnace)-IR absorption spectrometry .18
12.2.1 Principle .18
12.2.2 Reagents .18
12.2.3 Apparatus .18
12.2.4 Instrument .18
12.2.5 Procedure .19
12.2.6 Blank test .19
12.2.7 Calculation of calibration coefficient .19
12.2.8 Calculation . 20
12.3 Combustion (resistance furnace)-coulometry . 20
12.4 Combustion (resistance furnace)-thermal conductivity. 20
13 Determination of fluorine and chlorine.20
13.1 Principle . 20
13.2 Reagents . 20
13.3 Apparatus .21
13.4 Procedure .21
13.4.1 Extraction of fluorine and chlorine from the sample .21
13.4.2 Determination of fluorine and chlorine .21
13.5 Blank test . 22
13.6 Calibration . 22
13.7 Calculation . 22
14 Reporting analytical values .23
14.1 Number of analyses . 23
14.2 Evaluation of analytical values . 23
14.3 Expression of analytical values . 23
15 Test report .23
Annex A (informative) List of commercial certified reference materials .25
Annex B (informative) Analytical results obtained from a round robin test .26
Annex C (informative) Spectral lines for ICP-OES .30
Bibliography .31

iv
ISO 17947-1:2026(en)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee
has been established has the right to be represented on that committee. International organizations,
governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely
with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types
of ISO document should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent
rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a)
patent(s) which may be required to implement this document. However, implementers are cautioned that
this may not represent the latest information, which may be obtained from the patent database available at
www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO's adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 206, Fine ceramics, in collaboration with
the European Committee for Standardization (CEN) Technical Committee CEN/TC 184, Advanced technical
ceramics, in accordance with the Agreement on technical cooperation between ISO and CEN (Vienna
Agreement).
This first edition of ISO 17947-1 cancels and replaces ISO 17947:2014, which has been technically revised.
A list of all parts in the ISO 17947 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.

v
ISO 17947-1:2026(en)
Introduction
[1]
This document has been developed from Japanese Industrial Standard JIS R 1603:2007 with reference to
[2] [3]
CEN ENV 14226:2002 and ASTM C1494-01:2007 , and is applicable to the chemical analysis of silicon
nitride raw powders for fine ceramics use. This document covers both major and minor constituents, such as
silicon, nitrogen, and some trace metallic and non-metallic elements.

vi
International Standard ISO 17947-1:2026(en)
Fine ceramics (advanced ceramics, advanced technical
ceramics) — Methods for chemical analysis of silicon nitride
powders —
Part 1:
Wet chemical methods, X-ray fluorescence (XRF) using the
fused cast-bead method, carrier-gas hot extraction (CGHE)
and combustion methods
1 Scope
This document specifies methods for the determination of silicon, nitrogen, aluminium, iron, calcium,
oxygen, carbon, fluorine, and chlorine in silicon nitride powders.
This document is applicable for the chemical analysis of silicon nitride powders used as raw material for fine
ceramics.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes
requirements of this document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
ISO 2828, Aluminium oxide primarily used for the production of aluminium — Determination of fluorine content
— Alizarin complexone and lanthanum chloride spectrophotometric method
ISO 3696, Water for analytical laboratory use — Specification and test methods
ISO 6353-1, Reagents for chemical analysis — Part 1: General test methods
ISO 6353-2, Reagents for chemical analysis — Part 2: Specifications — First series
ISO 6353-3, Reagents for chemical analysis — Part 3: Specifications — Second series
ISO 8656-1, Refractory products — Sampling of raw materials and unshaped products — Part 1: Sampling
scheme
ISO 21068-2:2024, Chemical analysis of raw materials and refractory products containing silicon-carbide,
silicon-nitride, silicon-oxynitride and sialon — Part 2: Determination of volatile components, total carbon, free
carbon, silicon carbide, total and free silicon, free and surface silica
ISO 21068-3:2024, Chemical analysis of raw materials and refractory products containing silicon-carbide,
silicon-nitride, silicon-oxynitride and sialon — Part 3: Determination of nitrogen, oxygen and metallic and oxidic
constituents
ISO 21438-2, Workplace atmospheres — Determination of inorganic acids by ion chromatography — Part 2:
Volatile acids, except hydrofluoric acid (hydrochloric acid, hydrobromic acid and nitric acid)
ISO 21438-3, Workplace atmospheres — Determination of inorganic acids by ion chromatography — Part 3:
Hydrofluoric acid and particulate fluorides

ISO 17947-1:2026(en)
ISO 26845:2008, Chemical analysis of refractories — General requirements for wet chemical analysis, atomic
absorption spectrometry (AAS) and inductively coupled plasma atomic emission spectrometry (ICP-AES)
methods
3 Terms and definitions
No terms and definitions are listed in this document.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at https:// www .electropedia .org/
4 Analytes and ranges
Analytes and ranges specified in this document shall be as follows.
a) Silicon (Si), range of 30 % to 70 % (mass fraction)
b) Nitrogen (N), range of 30 % to 45 % (mass fraction)
c) Aluminium (Al), range of 0,001 % to 0,6 % (mass fraction)
d) Iron (Fe), range of 0,001 % to 0,6 % (mass fraction)
e) Calcium (Ca), range of 0,001 % to 0,03 % (mass fraction)
f) Oxygen (O), range of 0,05 % to 5 % (mass fraction)
g) Carbon (C), range of 0,01 % to 6 % (mass fraction)
h) Fluorine (F), range of 0,001 % to 0,2 % (mass fraction)
i) Chlorine (Cl), range of 0,001 % to 0,2 % (mass fraction)
NOTE The results of an interlaboratory test conducted using three silicon nitride powder samples are listed in
Annex B.
5 Preparation of test sample
5.1 General
The method of preparing samples shall be in accordance with ISO 8656-1 unless otherwise mutually agreed
upon between the analyser and the customer.
5.2 Sampling
Sampling shall be in accordance with ISO 8656-1.
5.3 Drying
The following procedure shall be used: Take about 10 g of the sample into a flat-type weighing bottle
(60 mm × 30 mm) and spread it uniformly over the bottom of the bottle. Place the bottle in a drying oven
at 110 °C ± 5 °C for 2 h without a lid, and then cool in a desiccator (desiccant: magnesium perchlorate for
drying) with a lid for 1 h.
ISO 17947-1:2026(en)
5.4 Weighing
The required quantity of the sample shall be weighed to the nearest 0,1 mg using a balance.
6 Apparatus and reagents
Unless otherwise specified for each determination, ordinary laboratory apparatus for chemical analysis as
listed in ISO 26845:2008, Clause 4 may be used. Reagents shall conform to the requirements of ISO 6353-1,
ISO 6353-2, and ISO 6353-3. Unless otherwise specified for each determination, reagents of analytical grade
listed in ISO 26845:2008, Clause 5, shall be used.
7 Blank test
Blank test shall be carried out without the sample by using identical quantities of reagents, conditions, and
procedures throughout each determination to correct the analytical values obtained.
8 Determination of silicon
8.1 Classification of determination methods
Silicon shall be determined by either of the following methods. If analytical results with four figures are
required, use the method A. If analytical results with two or three figures are required, the method B can be
used.
— Method A: Fusion–dehydration/insolubilization separation–gravimetry and inductively coupled plasma
optical emission spectrometry (ICP-OES)
— Method B: XRF using fused cast-bead method
8.2 Fusion-dehydration/insolubilization separation-gravimetry and ICP-OES
8.2.1 Principle
A sample is fused with alkaline carbonate and the melt is treated with an acid to separate into two parts of
silicon, insoluble silicon and soluble silicon, by filtration. Insoluble silicon is determined gravimetrically as
silicon dioxide after ignition of the insoluble residue, whereas soluble silicon is determined in the filtrate
using ICP-OES. The sum of them represents the content of silicon in the sample.
8.2.2 Reagents
Reagents of analytical grade shall be used. Reagent solutions shall be preserved in plastic bottles.
8.2.2.1 Water, of grade 1 or superior specified in ISO 3696.
8.2.2.2 Sodium carbonate, Na CO , anhydrous, specified in ISO 6353-3 or that of higher grade.
2 3
8.2.2.3 Hydrochloric acid (1+1), (1+4), (1+50), HCl, concentrated hydrochloric acid (c(HCl) = 32 %, mass
fraction) diluted (1+1), (1+4) and (1+50) by volume with water (8.2.2.1).
8.2.2.4 Sulfuric acid (1+1), (1+4), H SO , concentrated sulfuric acid (c(H SO ) = 96 %, mass fraction)
2 4 2 4
diluted (1+1) and (1+4) by volume with water (8.2.2.1).
8.2.2.5 Cellulose powder.
ISO 17947-1:2026(en)
8.2.2.6 Polyethylene glycol (PEG) solution, (C H O) H O, 0,05 % mass fraction of PEG, prepared by
2 4 n 2
dissolving solid PEG in water (8.2.2.1).
8.2.2.7 Hydrofluoric acid, HF, c(HF) = 48 %, mass fraction.
8.2.3 Apparatus and instruments
Ordinary laboratory apparatus and instruments for chemical analysis in accordance with ISO 26845:2008,
4 may be used.
8.2.3.1 Platinum dish.
8.2.3.2 Platinum crucible.
8.2.3.3 Burner, capable of heating at 1 100 °C.
8.2.3.4 Muffle furnace, capable of being operated at 1 100 °C.
8.2.3.5 Balance, readable to 0,1 mg.
8.2.3.6 Inductively coupled plasma optical emission spectrometer (ICP-OES).
8.2.4 Procedure
8.2.4.1 General
The procedure shall be as follows. The procedure described in ISO 21068-2:2024, Clause 8 may be
alternatively used.
8.2.4.2 Fusion of sample
The following procedure shall be used: Weigh 0,30 g of the sample and 2,0 g of sodium carbonate, anhydrous,
into a platinum dish and mix well. Start to heat carefully and increase the temperature gradually to 1 000 °C
to completely fuse the sample using a burner or a muffle furnace.
8.2.4.3 Separation of silicon
The following procedure shall be used: Add 20 ml of hydrochloric acid (1+1) to dissolve the melt on a hot
plate. Silicon dioxide will appear to be jellified and precipitated at this stage. There are two methods to
separate the precipitated silicon dioxide.
a) Dehydrate carefully the precipitate to dryness in order to prevent it from spattering and add 5 ml of
hydrochloric acid (1+1) and 20 ml of water to dissolve hydrochloric acid soluble components of the
precipitate. Filtrate the precipitate with a filter paper and wash with hot hydrochloric acid (1+50) several
times and then with hot water sufficiently until it contains no salt. Collect the filtrate and washings
together in a volumetric flask and dilute to volume with water. Preserve this precipitate for gravimetry
of insoluble silicon and the solution for the ICP-OES determination of soluble silicon, respectively.
b) After formation of jellified silicon dioxide, add 0,05 g of cellulose powder and 10 ml of PEG solution
(8.2.2.6) to agglomerate silicon dioxide for easy filtration. Filtrate and wash using the same procedure
as described in a), and then preserve this precipitate for gravimetry of insoluble silicon and the solution
for the ICP-OES determination of soluble silicon, respectively.

ISO 17947-1:2026(en)
8.2.4.4 Gravimetry for insoluble silicon
The following procedure shall be used: Transfer the filter paper with the precipitate to a platinum crucible.
Heat the platinum crucible slowly using a burner (8.2.3.3) or a muffle furnace (8.2.3.4) until complete ashing
of the filter paper and then calcine the residue at 1 100 °C. Place the platinum crucible in a desiccator to
cool to room temperature. Weigh the platinum crucible. Moisten the residue in the platinum crucible with
a few drops of water (8.2.2.1) and sulfuric acid (1+1) (8.2.2.4) and add 10 ml of hydrofluoric acid (8.2.2.7).
Then evaporate to dryness on a hot plate to remove all silicon dioxide, calcine the residue at 1 100 °C using
a burner or in a muffle furnace and place the platinum crucible in a desiccator to cool to room temperature.
Weigh the platinum crucible again. The loss of mass after hydrofluoric acid treatment corresponds to the
amount of insoluble silicon dioxide.
8.2.4.5 ICP-OES for soluble silicon
The following procedure shall be used: Nebulize the test solution obtained in 8.2.4.3 into the plasma of an
ICP-OES to determine soluble silicon in the sample.
NOTE An example for a suited silicon emission line is listed in Annex C.
8.2.5 Blank test
Blank determinations shall be performed according to 8.2.4.2 to 8.2.4.5 without the sample.
8.2.6 Calibration
For ICP-OES, calibration solutions shall be prepared to span the range of concentration of silicon in the test
solution. Each calibration solution shall have a similar matrix to the test solution.
With those calibration solutions, the calibration function for soluble silicon shall be created to establish the
relation between the emission intensity measured with the test solution and the concentration of silicon in
the test solution.
8.2.7 Calculation
With the amount of insoluble silicon (8.2.4.4), soluble silicon (8.2.4.5) and the blank test (8.2.5), the content
of silicon shall be calculated according to Formula (1).
 VAA 

w  mm 0,4674 (1)
 
Si 10
m
 
 
where
w is the mass fraction of silicon in the sample, in percent;
Si
m is the mass of insoluble silicon dioxide in the sample (8.2.4.4), in grams;
m is the mass of insoluble silicon dioxide in the blank test (8.2.5), in grams;
A is the concentration of soluble silicon in the test solution (8.2.4.3), in milligrams per litre;
A is the concentration of soluble silicon in the blank solution (8.2.5), in milligrams per litre;
V is the volume of the test and blank solution, in millilitres;
m is the mass of the sample (8.2.4.2), in grams.
8.3 XRF using fused cast-bead method
The procedure shall be in accordance with ISO 21068-3:2024, 8.2.

ISO 17947-1:2026(en)
9 Determination of nitrogen
9.1 Classification of determination methods
Nitrogen shall be determined by either of the following methods. Methods A and C provide a higher precision
than method B.
NOTE Based on practical experience, methods A and C provide a precision of two valid decimal places. Method B
usually provides a precision of one valid decimal place.
— Method A: Acid pressure decomposition–distillation separation–acidimetric titration method
— Method B: Inert gas fusion–thermal conductivity method
— Method C: Fusion–ammonia separation–acidimetric titration method
9.2 Acid pressure decomposition-distillation separation-acidimetric titration method
9.2.1 Principle
The sample is decomposed in a pressure decomposition vessel with a mixture of hydrofluoric acid and
sulfuric acid to convert nitrogen into ammonia. Boric acid is added and the solution is transferred into a
distillation flask. After addition of sodium hydroxide, the ammonia is separated by steam distillation
and the distillate is collected in a receiving vessel containing a known amount of amidosulfuric acid. The
distilled ammonia reacts with amidosulfuric acid and the excess of amidosulfuric acid is back-titrated with
a standardized sodium hydroxide solution.
9.2.2 Reagents
Reagents of analytical grade shall be used. Reagent solutions shall be preserved in plastic bottles.
9.2.2.1 Water, of grade 1 or superior specified in ISO 3696.
9.2.2.2 Hydrofluoric acid, HF, c(HF) = 48 %, mass fraction.
9.2.2.3 Sulfuric acid (1+1), H SO , concentrated sulfuric acid (c(H SO ) = 96 %, mass fraction) diluted
2 4 2 4
(1+1) by volume with water (9.2.2.1).
9.2.2.4 Sodium hydroxide, NaOH, more than 97,0 % (mass fraction) of purity.
9.2.2.5 Sodium hydroxide solution (500 g/l), prepared by dissolving sodium hydroxide (9.2.2.4) in
water (9.2.2.1).
9.2.2.6 Amidosulfuric acid, H NO S, more than 99,0 % (mass fraction) of purity.
3 3
9.2.2.7 0,1mol/l amidosulfuric acid solution, prepared by dissolving amidosulfuric acid (9.2.2.6) in
water (9.2.2.1). Weigh 10,0 g of amidosulfuric acid into a 1 000 ml volumetric flask, dissolve in approximately

ISO 17947-1:2026(en)
200 ml water, dilute with water to volume and mix well. The factor of this solution shall be calculated
according to Formula (2).
F = m × P/(9,709 5 × 100) (2)
where
F is the factor of the 0,1 mol/l amidosulfuric acid solution;
m is the mass of amidosulfuric acid, g;
P is the purity of amidosulfuric acid, % (mass fraction).
9.2.2.8 0,1 mol/l sodium hydroxide solution, prepared by dissolving sodium hydroxide (9.2.2.4) in
water (9.2.2.1). Weigh 4,0 g of sodium hydroxide into a 1 000 ml volumetric flask, dissolve in approximately
200 ml water, dilute with water to volume and mix well.
The factor of this solution shall be determined using the following procedure: Pour exactly 50 ml of the
0,1 mol/l amidosulfuric acid solution into a beaker (200 ml) and dilute with water to about 100 ml. Titrate
this solution with the 0,1 mol/l sodium hydroxide solution using a pH meter. Take the end point as pH 5,5
and determine the volume of the titrant consumed. The factor of this solution shall be calculated according
to Formula (3).
F = F × 50,00/V (3)
where
F is the factor of the 0,1 mo
...