ISO 80601-2-74:2017
(Main)Medical electrical equipment - Part 2-74: Particular requirements for basic safety and essential performance of respiratory humidifying equipment
Medical electrical equipment - Part 2-74: Particular requirements for basic safety and essential performance of respiratory humidifying equipment
ISO 80601-2-74:2017 applies to the basic safety and essential performance of a humidifier, also hereafter referred to as me equipment, in combination with its accessories, the combination also hereafter referred to as me system.
 ISO 80601-2-74:2017 is also applicable to those accessories intended by their manufacturer to be connected to a humidifier where the characteristics of those accessories can affect the basic safety or essential performance of the humidifier.
 EXAMPLE 1 Heated breathing tubes (heated-wire breathing tubes) or me equipment intended to control these heated breathing tubes (heated breathing tube controllers).
 NOTE 1 Heated breathing tubes and their controllers are me equipment and are subject to the requirements of IEC 60601‑1.
 NOTE 2 ISO 5367 specifies other safety and performance requirements for breathing tubes.
 ISO 80601-2-74:2017 includes requirements for the different medical uses of humidification, such as invasive ventilation, non-invasive ventilation, nasal high-flow therapy, and obstructive sleep apnoea therapy, as well as humidification therapy for tracheostomy patients.
 NOTE 3 A humidifier can be integrated into other equipment. When this is the case, the requirements of the other equipment also apply to the humidifier.
 EXAMPLE 2 Heated humidifier incorporated into a critical care ventilator where ISO 80601‑2-12[12] also applies.
 EXAMPLE 3 Heated humidifier incorporated into a homecare ventilator for dependent patients where ISO 80601‑2-72[14] also applies.
 EXAMPLE 4 Heated humidifier incorporated into sleep apnoea therapy equipment where ISO 80601‑2-70[13] also applies.
 ISO 80601-2-74:2017 also includes requirements for an active hme (heat and moisture exchanger), me equipment which actively adds heat and moisture to increase the humidity level of the gas delivered from the hme to the patient. This document is not applicable to a passive hme, which returns a portion of the expired moisture and heat of the patient to the respiratory tract during inspiration without adding heat or moisture.
 NOTE 4 ISO 9360‑1[5] and ISO 9360‑2[6] specify the safety and performance requirements for a passive hme.
 If a clause or subclause is specifically intended to be applicable to me equipment only, or to me systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to me equipment and to me systems, as relevant.
 Hazards inherent in the intended physiological function of me equipment or me systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012, 7.2.13 and 8.4.1.
 NOTE 5 Additional information can be found in IEC 60601‑1:2005+AMD1:2012, 4.2.
 ISO 80601-2-74:2017 does not specify the requirements for cold pass-over or cold bubble-through humidification devices, the requirements for which are given in ISO 20789:?.[8]
 This document is not applicable to equipment commonly referred to as "room humidifiers" or humidifiers used in heating, ventilation and air conditioning systems, or humidifiers incorporated into infant incubators.
 ISO 80601-2-74:2017 is not applicable to nebulizers used for the delivery of drugs to patients.
 NOTE 6 ISO 27427[10] specifies the safety and performance requirements for nebulizers.
 ISO 80601-2-74:2017 is a particular standard in the IEC 60601‑1 and the ISO/IEC 80601 series.
Appareils électromédicaux - Partie 2-74: Exigences particulières pour la sécurité de base et les performances essentielles des équipements d'humidification respiratoire
ISO 80601-2-74:2017 s'applique à la sécurité de base et aux performances essentielles d'un humidificateur, ci-après également désigné par appareil EM, associé à ses accessoires, dont la combinaison est ci-après désignée par système EM.
 ISO 80601-2-74:2017 s'applique également aux accessoires conçus par leur fabricant pour être raccordés à un humidificateur, lorsque les caractéristiques de ces accessoires peuvent affecter la sécurité de base ou les performances essentielles de l'humidificateur.
 EXEMPLE 1 Tuyaux respiratoires chauffés (tuyaux respiratoires avec résistance chauffante) ou les appareils EM destinés à commander ces tuyaux respiratoires chauffés (commandes de tuyaux respiratoires chauffés).
 NOTE 1 Les tuyaux respiratoires chauffés et leurs commandes sont des appareils EM et sont soumis aux exigences de l'IEC 60601‑1.
 NOTE 2 L'ISO 5367 spécifie d'autres exigences de sécurité et de performance pour les tuyaux respiratoires.
 ISO 80601-2-74:2017 contient des exigences pour les différentes utilisations médicales de l'humidification, telles que la ventilation invasive, la ventilation non invasive, l'oxygénothérapie nasale à haut débit et le traitement de l'apnée obstructive du sommeil, ainsi que l'humidification pour les patients trachéotomisés.
 NOTE 3 Un humidificateur peut être intégré dans un autre appareil. Dans ce cas, les exigences de l'autre appareil s'appliquent aussi à l'humidificateur.
 EXEMPLE 2 Humidificateur chauffé intégré dans un ventilateur pulmonaire pour utilisation en soins intensifs où l'ISO 80601‑2‑12[12] s'applique aussi.
 EXEMPLE 3 Humidificateur chauffé intégré dans un ventilateur utilisé dans l'environnement des soins à domicile pour les patients dépendants où l'ISO 80601‑2‑72[14] s'applique aussi.
 EXEMPLE 4 Humidificateur chauffé intégré dans du matériel de traitement respiratoire de l'apnée du sommeil où l'ISO 80601‑2‑70[13] s'applique aussi.
 ISO 80601-2-74:2017 contient également des exigences pour les ECH actifs (échangeurs de chaleur et d'humidité), appareils EM qui ajoutent activement de la chaleur et de l'humidité afin d'augmenter le niveau d'humidité du gaz administré au patient par l'ECH. Le présent document ne s'applique pas aux ECH passifs qui renvoient une partie de l'humidité et de la chaleur de l'air expiré par le patient aux voies respiratoires lors de l'inspiration, sans ajout de chaleur ou d'humidité.
 NOTE 4 L'ISO 9360‑1[5] et l'ISO 9360‑2[6] spécifient les exigences de sécurité et de performances pour un ECH passif.
 Si un article ou un paragraphe est spécifiquement destiné à être applicable uniquement aux appareils EM ou uniquement aux systèmes EM, le titre et le contenu de cet article ou de ce paragraphe l'indiquent. Dans le cas contraire, l'article ou le paragraphe s'applique à la fois aux appareils EM et aux systèmes EM, selon le cas.
 Les dangers inhérents à la fonction physiologique prévue des appareils EM ou des systèmes EM dans le cadre du domaine d'application du présent document ne sont pas couverts par des exigences spécifiques contenues dans le présent document, à l'exception de l'IEC 60601‑1:2005+AMD1:2012, 7.2.13 et 8.4.1.
 NOTE 5 L'IEC 60601‑1:2005+AMD1:2012, 4.2 fournit des informations supplémentaires.
 ISO 80601-2-74:2017 ne spécifie pas les exigences pour les dispositifs d'humidification à froid par léchage («pass-over») ou à barbotage, dont les exigences sont données dans l'ISO 20789:?[8].
 ISO 80601-2-74:2017 ne s'applique pas aux appareils servant à humidifier des locaux et les humidificateurs utilisés pour le chauffage, la ventilation ou la climatisation, ni aux humidificateurs utilisés dans les couveuses.
 ISO 80601-2-74:2017 ne s'applique pas aux nébuliseurs utilisés pour administrer des médicaments aux patients.
 NOTE 6 L'ISO 27427[10] spécifie les exigences de sécurité et de performances pour les nébuliseurs.
 ISO 80601-2-74:2017 est une norme particulière dans les séries de normes IEC 60601‑1 et ISO/IEC 80601.
General Information
Relations
Standards Content (Sample)
INTERNATIONAL ISO
STANDARD 80601-2-74
First edition
2017-05
Medical electrical equipment —
Part 2-74:
Particular requirements for basic
safety and essential performance of
respiratory humidifying equipment
Appareils électromédicaux —
Partie 2-74: Exigences particulières pour la sécurité de base et
les performances essentielles des équipements d’humidification
respiratoire
Reference number
©
ISO 2017
© ISO 2017, Published in Switzerland
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.
ISO copyright office
Ch. de Blandonnet 8 • CP 401
CH-1214 Vernier, Geneva, Switzerland
Tel. +41 22 749 01 11
Fax +41 22 749 09 47
copyright@iso.org
www.iso.org
ii © ISO 2017 – All rights reserved
Contents Page
Foreword . v
Introduction . vii
201.1   Scope, object and related standards . 1
201.2   Normative references . 3
201.3   Terms and definitions . 5
201.4   General requirements . 7
201.5   General requirements for testing of ME EQUIPMENT . 10
201.6   Classification of ME EQUIPMENT and ME SYSTEMS . 11
201.7   ME EQUIPMENT identification, marking and documents . 11
201.8   Protection against electrical HAZARDS form ME EQUIPMENT . 19
201.9   Protection against mechanical hazards of ME EQUIPMENT and ME SYSTEMS . 19
201.10   Protection against unwanted and excessive radiation HAZARDS . 21
201.11   Protection against excessive temperatures and other HAZARDS . 21
201.12   Accuracy of controls and instruments and protection against
hazardous outputs . 24
201.13   HAZARDOUS SITUATIONS and fault conditions for ME EQUIPMENT . 27
201.14   Programmable electrical medical systems (pems) . 28
201.15   Construction of ME EQUIPMENT . 28
201.16   ME SYSTEMS . 29
201.17   Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 29
201.101   BREATHING SYSTEM connectors and ports . 29
201.102   Requirements for the BREATHING SYSTEM and ACCESSORIES . 32
201.103   LIQUID CONTAINER . 33
201.104   FUNCTIONAL CONNECTION . 34
202   Electromagnetic disturbances — Requirements and tests . 34
206   Usability . 35
208   General requirements, tests and guidance for alarm systems in medical
electrical equipment and medical electrical systems . 36
211   Requirements for medical electrical equipment and medical electrical
systems used in the home healthcare environment . 36
Annex C (informative)  Guide to marking and labelling requirements for
ME EQUIPMENT and ME SYSTEMS . 38
Annex D (informative)  Symbols on Marking . 44
Annex AA (informative)  Particular guidance and rationale . 45
Annex BB (normative)  * Determination of the accuracy of the displayed MEASURED
GAS TEMPERATURE . 61
Annex CC (normative)  * Determination of the HUMIDIFICATION OUTPUT . 63
Annex DD (normative)  * Specific enthalpy calculations . 67
©	ISO	2017	–	All	rights	reserved	 iii
Annex EE (normative)  Removable temperature sensors and mating ports . 69
Annex FF (normative)  * Standard temperature sensor . 73
Annex GG (informative)  Saturation vapour pressure . 76
Annex HH (informative)  Reference to the essential principles of safety and
[7]
performance of medical devices in accordance with ISO 16142-1:2016 . 77
Annex II (informative)  Terminology — Alphabetized index of defined terms . 81
Bibliography . 85
iv	 ©	ISO/IEC	2016	–	All	rights	reserved
Foreword
ISO	(the	International	Organization	for	Standardization)	is	a	worldwide	federation	of	national
standards	bodies	(ISO	member	bodies).	The	work	of	preparing	International	Standards	is
normally	carried	out	through	ISO	technical	committees.	Each	member	body	interested	in	a
subject	for	which	a	technical	committee	has	been	established	has	the	right	to	be	represented	on
that	committee.	International	organizations,	governmental	and	non‐governmental,	in	liaison
with	ISO,	also	take	part	in	the	work.	ISO	collaborates	closely	with	the	International
Electrotechnical	Commission	(IEC)	on	all	matters	of	electrotechnical	standardization.
The	procedures	used	to	develop	this	document	and	those	intended	for	its	further	maintenance
are	described	in	the	ISO/IEC	Directives,	Part	1.	In	particular	the	different	approval	criteria
needed	for	the	different	types	of	ISO	documents	should	be	noted.	This	document	was	drafted	in
accordance	 with	 the	 editorial	 rules	of	the	ISO/IEC	Directives,	 Part	2	 (see
www.iso.org/directives).
Attention	is	drawn	to	the	possibility	that	some	of	the	elements	of	this	document	may	be	the
subject	of	patent	rights.	ISO	shall	not	be	held	responsible	for	identifying	any	or	all	such	patent
rights.	Details	of	any	patent	rights	identified	during	the	development	of	the	document	will	be	in
the	 Introduction	 and/or	 on	 the	 ISO	 list	 of	 patent	 declarations	 received	 (see
www.iso.org/patents).
Any	trade	name	used	in	this	document	is	information	given	for	the	convenience	of	users	and
does	not	constitute	an	endorsement.
For	an	explanation	on	the	meaning	of	ISO	specific	terms	and	expressions	related	to	conformity
assessment,	as	well	as	information	about	ISO's	adherence	to	the	World	Trade	Organization
(WTO)	 principles	 in	 the	 Technical	Barriers	to	Trade	(TBT)	see	the	following	URL:
www.iso.org/iso/foreword.html.
The	committee	responsible	for	this	document	is	ISO/TC	121,	 Anaesthetic and respiratory
equipment,	Subcommittee	SC	3,	Lung ventilators and related equipment	and	IEC/TC	62,	Electrical
equipment in medical practice,	Subcommittee	SC	62D,	Electromedical equipment.
[1]
This	first	edition	of	ISO	80601‐2‐74	cancels	and	replaces	the	third	edition	of	ISO	8185:2007 ,
which	has	been	technically	revised.	It	also	incorporates	the	third	edition	of	IEC	60601‐1,
including	amendment	1,	the	fourth	edition	of	IEC	60601‐1‐2,	the	third	edition	of	IEC	60601‐1‐6,
including	amendment	1,	the	second	edition	of	IEC	60601‐1‐8,	including	amendment	1,	and	the
second	edition	of	IEC	60601‐1‐11.
The	most	significant	changes	are	the	following	modifications:
—	 extending	the	scope	to	include	the	HUMIDIFIER	and	its	ACCESSORIES,	where	the	characteristics
of	those	ACCESSORIES	can	affect	the	BASIC	SAFETY	and	ESSENTIAL	PERFORMANCE	of	the	HUMIDIFIER,
and	thus	not	only	the	HUMIDIFIER	itself;
—	identification	of	ESSENTIAL	PERFORMANCE	for	a	HUMIDIFIER	and	its	ACCESSORIES;
—	modification	of	the	humidification	test	PROCEDURE	and	the	disclosure	of	humidification
performance;
©	ISO	2017	–	All	rights	reserved	 v
—	more	fully	dimensioning	the	removable	temperature	sensor	port	and	sensor;
—	removal	of	requirements	for	so‐called	“bubble”	HUMIDIFIERS	as	a	separate	document	is	being
[8]
prepared	for	them ;
and	the	following	additions:
—	requirements	for	mechanical	strength	(via	IEC	60601‐1‐11);
—	new	symbols;
—	requirements	for	a	HUMIDIFIER	as	a	component	of	an	ME	SYSTEM;
—	requirements	for	ENCLOSURE	integrity	(water	ingress	via	IEC	60601‐1‐11);
—	requirements	for	cleaning	and	disinfection	PROCEDURES	(via	IEC	60601‐1‐11);
—	requirements	for	BIOCOMPATIBILITY;
—	requirements	for	USABILITY.
vi	 ©	ISO	2017	–	All	rights	reserved
Introduction
This	document	specifies	requirements	for	respiratory	humidifying	equipment	intended	for	use
on	PATIENTS	in	HOME	HEALTHCARE	ENVIRONMENT	and	in	healthcare	facilities.	HUMIDIFIERS	are	used	to
raise	the	water	content	of	gases	delivered	to	PATIENTS.	Gases	available	for	medical	use	do	not
contain	sufficient	moisture	and	can	damage	or	irritate	the	respiratory	tract	or	desiccate
secretions	of	PATIENTS	whose	upper	airways	have	been	bypassed.	Inadequate	humidity	at	the
PATIENT‐CONNECTION	PORT	can	cause	drying	of	the	upper	airway,	or	desiccation	of	tracheo‐
bronchial	secretions	in	the	tracheal	or	tracheostomy	tube,	which	can	cause	narrowing	or	even
[19][20]
obstruction	of	the	airway .	Heat	is	employed	to	increase	the	water	output	of	the	HUMIDIFIER.
In	addition,	many	HUMIDIFIERS	utilize	heated	BREATHING	TUBES	in	order	to	increase	operating
efficiency	and	reduce	water	loss	(condensate)	as	well	as	heat	loss	in	the	BREATHING	TUBE.
Ventilator	and	anaesthesia	BREATHING	TUBES	in	common	use	might	not	withstand	the	heat
generated	by	HUMIDIFIERS	and	BREATHING	TUBE	heating	mechanisms.
Many	HUMIDIFIER	MANUFACTURERS	use	off‐the‐shelf	electrical	connectors	for	their	electrically
heated	BREATHING	TUBES.	However,	since	different	MANUFACTURERS	have	used	the	same	electrical
connector	for	different	power	outputs,	electrically	heated	BREATHING	TUBES	can	be	physically,	but
not	electrically,	interchangeable.	Use	of	improper	electrically	heated	BREATHING	TUBES	has	caused
overheating,	circuit	melting,	PATIENT	and	OPERATOR	burns	and	fires.	It	was	not	found	practical	to
specify	the	interface	requirements	for	electrical	connectors	to	ensure	compatibility	between
HUMIDIFIERS	and	BREATHING	TUBES	produced	by	different	MANUFACTURERS.
Since	the	safe	use	of	a	HUMIDIFIER	depends	on	the	interaction	of	the	HUMIDIFIER	with	its	many
ACCESSORIES,	this	document	sets	total	system	performance	requirements	up	to	the	PATIENT‐
CONNECTION	PORT.	These	requirements	are	applicable	to	ACCESSORIES	such	as	BREATHING	TUBES
(both	heated	and	non‐heated),	temperature	sensors	and	equipment	intended	to	control	the
environment	within	these	BREATHING	TUBES.
Humidification	can	also	be	used	by	respiratory	support	ME	EQUIPMENT	to	increase	PATIENT
comfort	and	compliance	with	the	therapy.	Examples	are	obstructive	sleep	apnoea	and	nasal	high
flow	therapy	equipment.	The	HUMIDIFICATION	OUTPUT	requirements	of	such	ME	EQUIPMENT	is	less
demanding	as	the	PATIENT’S	upper	airway	is	not	bypassed.
HUMIDIFIERS	are	commonly	used	with	air	and	air‐oxygen	mixtures	and	any	HUMIDIFIER	should	be
able	to	operate	with	these	gases.	Care	should	be	taken	if	using	other	gas	mixes	such	as	helium‐
oxygen	mixtures,	as	the	different	physical	and	thermal	properties	of	these	gases	may	disturb	the
operation	of	the	HUMIDIFIER.
In	this	document,	the	following	print	types	are	used:
—			Requirements	and	definitions:	roman	type;
—			Test specifications: italic type;
—			Informative	material	appearing	outside	of	tables,	such	as	notes,	examples	and	references:	in
smaller	type.	Normative	text	of	tables	is	also	in	a	smaller	type;
—			Terms	defined	in	CLAUSE	3	of	the	general	standard,	in	this	document	or	as	noted:	small
capitals;
©	ISO	2017	–	All	rights	reserved	 vii
In	referring	to	the	structure	of	this	document,	the	term
—			“clause”	means	one	of	the	five	numbered	divisions	within	the	table	of	contents,	inclusive	of
all	subdivisions	(e.g.	Clause	201	includes	subclauses	201.7,	201.8,	etc.);
—			“subclause”	means	a	numbered	subdivision	of	a	clause	(e.g.	201.7,	201.8	and	201.9	are	all
subclauses	of	Clause	201).
References	to	clauses	within	this	document	are	preceded	by	the	term	“Clause”	followed	by	the
clause	number.	References	to	subclauses	within	this	document	are	by	number	only.
In	this	document,	the	conjunctive	“or”	is	used	as	an	“inclusive	or”	so	a	statement	is	true	if	any
combination	of	the	conditions	is	true.
The	verbal	forms	used	in	this	document	conform	to	usage	described	in	Annex	H	of	the	ISO/IEC
Directives,	Part	2.	For	the	purposes	of	this	document,	the	auxiliary	verb:
—	“shall”	means	that	compliance	with	a	requirement	or	a	test	is	mandatory	for	c
 ...
ISO
NORME
80601-2-74
INTERNATIONALE
Première édition
2017-05
Appareils électromédicaux —
Partie 2-74:
Exigences particulières pour la
sécurité de base et les performances
essentielles des équipements
d’humidification respiratoire
Medical electrical equipment —
Part 2-74: Particular requirements for basic safety and essential
performance of respiratory humidifying equipment
Numéro de référence
©
ISO 2017
DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2017, Publié en Suisse
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni utilisée
sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie, l’affichage sur
l’internet ou sur un Intranet, sans autorisation écrite préalable. Les demandes d’autorisation peuvent être adressées à l’ISO à
l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
Ch. de Blandonnet 8 • CP 401
CH-1214 Vernier, Geneva, Switzerland
Tel. +41 22 749 01 11
Fax +41 22 749 09 47
copyright@iso.org
www.iso.org
ii © ISO 2017 – Tous droits réservés
Sommaire Page
Avant-propos . 4
Introduction . 6
201.1 Domaine d’application, objet et normes connexes . 1
201.2 Références normatives . 4
201.3 Termes et définitions. 5
201.4 Exigences générales . 9
201.5 Exigences générales relatives aux essais des APPAREILS EM . 11
201.6 Classification des APPAREILS EM et des SYSTÈMES EM . 12
APPAREILS EM . 13
201.7 Identification, marquage et documentation des
201.8 Protection contre les DANGERS d’origine électrique provenant des APPAREILS EM . 20
201.9 Protection contre les dangers mécaniques des APPAREILS EM et SYSTÈMES EM . 21
201.10   Protection contre les DANGERS dus aux rayonnements involontaires ou excessifs . 23
201.11   Protection contre les températures excessives et les autres DANGERS . 23
201.12   Précision des commandes, des instruments et de la protection contre les
caractéristiques de sortie présentant des risques . 26
201.13   SITUATIONS DANGEREUSES et conditions de défaut pour les APPAREILS EM . 29
201.14   Systèmes électromédicaux programmables (SEMP) . 30
201.15   Construction de l’APPAREIL EM . 31
201.16   SYSTÈMES EM . 31
201.17   Compatibilité électromagnétique des APPAREILS EM et des SYSTÈMES EM . 31
201.101 Raccords et orifices des SYSTÈMES RESPIRATOIRES . 31
201.102 Exigences pour le SYSTÈME RESPIRATOIRE et ses ACCESSOIRES . 34
201.103 CONTENEUR DE LIQUIDE . 35
201.104 CONNEXION FONCTIONNELLE . 35
202  Perturbation électromagnétique — Exigences et essais . 36
206  Aptitude à l’utilisation . 37
208  Exigences générales, essais et guide pour les systèmes d’alarme des appareils et des
systèmes électromédicaux . 38
211  Exigences pour les appareils électromédicaux et les systèmes électromédicaux
utilisés dans l’environnement des soins à domicile . 39
Annexe C (informative) Guide pour les exigences en matière de marquage et d’étiquetage
des APPAREILS EM et SYSTÈMES EM . 40
Annexe D (informative) Symboles des marquages . 45
Annexe AA (informative) Préconisations particulières et justifications . 46
Annexe BB (normative) Détermination de la précision de la TEMPÉRATURE MESURÉE DU GAZ
affichée . 63
Annexe CC (normative) * Détermination de la QUANTITÉ D’HUMIDIFICATION DÉLIVRÉE . 65
Annexe DD (normative) * Calculs de l’enthalpie spécifique . 69
Annexe EE (normative) Capteurs de température amovibles et orifices de raccordement . 71
Annexe FF (normative) * Capteur de température normalisé . 75
Annexe GG (informative) Pression de vapeur saturante . 77
©	ISO	2017	–	Tous	droits	réservés
ii-1
Annexe HH (informative) Référence aux principes essentiels de sécurité et de performance
[7]
des dispositifs médicaux conformément avec l’ISO 16142-1 :2016 . 78
Annexe II (informative) Terminologie — Index alphabétique des termes définis . 82
Bibliographie . 86
©	ISO	2017	–	Tous	droits	réservés
iii
Avant-propos
L'ISO	(Organisation	internationale	de	normalisation)	est	une	fédération	mondiale	d'organismes
nationaux	de	normalisation	(comités	membres	de	l'ISO).	L'élaboration	des	Normes	internationales	est
en	général	confiée	aux	comités	techniques	de	l'ISO.	Chaque	comité	membre	intéressé	par	une	étude	a	le
droit	de	faire	partie	du	comité	technique	créé	à	cet	effet.	Les	organisations	internationales,
gouvernementales	et	non	gouvernementales,	en	liaison	avec	l'ISO	participent	également	aux	travaux.
L'ISO	collabore	étroitement	avec	la	Commission	électrotechnique	internationale	(IEC)	en	ce	qui
concerne	la	normalisation	électrotechnique.
Les	procédures	utilisées	pour	élaborer	le	présent	document	et	celles	destinées	à	sa	mise	à	jour	sont
décrites	dans	les	Directives	ISO/IEC,	Partie	1.	Il	convient,	en	particulier	de	prendre	note	des	différents
critères	d'approbation	requis	pour	les	différents	types	de	documents	ISO.	Le	présent	document	a	été
rédigé	 conformément	 aux	 règles	 de	 rédaction	 données	 dans	 les	 Directives	 ISO/IEC,	 Partie	2
(voir	www.iso.org/directives).
L'attention	est	appelée	sur	le	fait	que	certains	des	éléments	du	présent	document	peuvent	faire	l'objet
de	droits	de	propriété	intellectuelle	ou	de	droits	analogues.	L'ISO	ne	saurait	être	tenue	pour
responsable	de	ne	pas	avoir	identifié	de	tels	droits	de	propriété	et	averti	de	leur	existence.	Les	détails
concernant	les	références	aux	droits	de	propriété	intellectuelle	ou	autres	droits	analogues	identifiés
lors	de	l'élaboration	du	document	sont	indiqués	dans	l'Introduction	et/ou	dans	la	liste	des	déclarations
de	brevets	reçues	par	l'ISO	(voir	www.iso.org/brevets).
Les	appellations	commerciales	éventuellement	mentionnées	dans	le	présent	document	sont	données
pour	information,	par	souci	de	commodité,	à	l’intention	des	utilisateurs	et	ne	sauraient	constituer	un
engagement.
Pour	une	explication	de	la	signification	des	termes	et	expressions	spécifiques	de	l'ISO	liés	à	l'évaluation
de	la	conformité,	ou	pour	toute	information	au	sujet	de	l'adhésion	de	l'ISO	aux	principes	de	l'OMC
concernant	 les	 obstacles	 techniques	au	commerce	(OTC),	voir	le	 lien
suivant:	www.iso.org/iso/fr/foreword.html.
Le	comité	chargé	de	l'élaboration	du	présent	document	est	l'ISO/TC	121,	Matériel d’anesthésie et de
réanimation respiratoire,	Sous‐comité	SC	3,	 Ventilateurs pulmonaires et équipements connexes,	et
IEC/TC	62,	 Équipements électriques dans la pratique médicale,	Sous‐comité	SC	62D,	 Appareils
électromédicaux.
Cette	 première	 édition	 de	 l’ISO	80601‐2‐74	 annule	 et	 remplace	 la	 troisième	 édition	 de
[1]
l’ISO	8185:2007, 	qui	a	fait	l’objet	d’une	révision	technique.	Elle	intègre	également	la	troisième	édition
de	la	norme	IEC	60601‐1,	y	compris	l’amendement	1,	la	quatrième	édition	de	la	norme	IEC	60601‐1‐2,
la	troisième	édition	de	la	norme	IEC	60601‐1‐6,	y	compris	l’amendement	1,	la	deuxième	édition	de	la
norme	IEC	60601‐1‐8,	y	compris	l’amendement	1,	et	la	deuxième	édition	de	la	norme	IEC	60601‐1‐11.
Les	principales	modifications	apportées	sont	les	suivantes:
―	 extension	du	domaine	d’application	pour	inclure	non	seulement	l’HUMIDIFICATEUR,	mais	également
ses	ACCESSOIRES,	lorsque	les	caractéristiques	de	ces	ACCESSOIRES	peuvent	affecter	la	SÉCURITÉ	DE
BASE	 et	 les	 PERFORMANCES	 ESSENTIELLES	de	l’HUMIDIFICATEUR	et	donc	pas	uniquement
l’HUMIDIFICATEUR	lui‐même;
―	 identification	des	PERFORMANCES	ESSENTIELLES	d’un	HUMIDIFICATEUR	et	de	ses	ACCESSOIRES;
©	ISO	2017	–	Tous	droits	réservés
iv
―	 modification	 de	 la	 PROCEDURE	 d’essai	 d’humidification	 et	 indication	 des	 performances
d’humidification;
―	 dimensionnement	plus	complet	de	l’orifice	pour	capteur	de	température	amovible	et	du	capteur;
―	 suppression	des	exigences	pour	les	HUMIDIFICATEURS	à	«barbotage»,	étant	donné	qu’un	document
[8]
distinct	leur	étant	consacré	est	en	cours	d’élaboration ;
et	les	ajouts	suivants:
―	 exigences	de	résistance	mécanique	(via	l’IEC	60601‐1‐11);
―	 nouveaux	symboles;
―	 exigences	pour	un	HUMIDIFICATEUR	en	tant	que	composant	d’un	SYSTEME	EM;
―	 exigences	d’intégrité	de	l’ENVELOPPE	(pénétration	d’eau	via	l’IEC	60601‐1‐11);
―	 exigences	pour	les	PROCEDURES	de	nettoyage	et	de	désinfection	(via	l’IEC	60601‐1‐11);
―	 exigences	de	BIOCOMPATIBILITE;
―	 exigences	d’APTITUDE	A	L’UTILISATION.
©	ISO	2017	–	Tous	droits	réservés
v
Introduction
Le	présent	document	spécifie	des	exigences	pour	les	équipements	d’humidification	respiratoire
destinés	à	être	utilisés	sur	des	PATIENTS	dans	l’ENVIRONNEMENT	DES	SOINS	A	DOMICILE	et	en	établissements
de	soins.	Les	HUMIDIFICATEURS	sont	employés	pour	augmenter	la	teneur	en	eau	des	gaz	administrés	aux
PATIENTS.	Les	gaz	disponibles	pour	usage	médical	ne	sont	pas	suffisamment	humides	et	peuvent
endommager,	irriter	les	voies	respiratoires	ou	dessécher	les	sécrétions	des	PATIENTS	dont	les	voies
respiratoires	supérieures	ont	été	dérivées.	Une	humidité	inadéquate	au	niveau	de	l’ORIFICE	DE
RACCORDEMENT	COTE	PATIENT	peut	provoquer	une	sécheresse	des	voies	respiratoires	supérieures,	ou	un
dessèchement	des	sécrétions	trachéo‐bronchiques	dans	la	sonde	trachéale	ou	le	tube	de	trachéotomie,
[19][20]
et	ultérieurement	un	rétrécissement	ou	même	une	obstruction	de	la	voie	respiratoire. 	La	chaleur
est	utilisée	pour	augmenter	le	débit	d’eau	de	l’HUMIDIFICATEUR.
En	outre,	de	nombreux	HUMIDIFICATEURS	utilisent	des	TUYAUX	RESPIRATOIRES	chauffés	afin	d’augmenter
l’efficacité	de	fonctionnement	et	de	réduire	les	pertes	d’eau	(condensat)	et	de	chaleur	dans	le	TUYAU
RESPIRATOIRE.	Le	ventilateur	et	les	TUYAUX	RESPIRATOIRES	d’anesthésie	couramment	employés	peuvent	ne
pas	résister	à	la	chaleur	produite	par	les	HUMIDIFICATEURS	et	les	mécanismes	de	chauffage	des	TUYAUX
RESPIRATOIRES.
De	nombreux	FABRICANTS	D’HUMIDIFICATEURS	utilisent	des	raccords	électriques	du	commerce	pour	leurs
TUYAUX	RESPIRATOIRES	chauffés	électriquement.	Cependant,	comme	les	différents	FABRICANTS	ont	utilisé	le
même	raccord	électrique	pour	différentes	sorties	d’alimentation,	les	TUYAUX	RESPIRATOIRES	chauffés
électriquement	peuvent	être	physiquement	interchangeables	sans	l’être	électriquement.	Un	usage
impropre	de	TUYAUX	RESPIRATOIRES	chauffés	électriquement	a	été	à	l’origine	de	problèmes	de	surchauffe,
de	fusion	du	circuit,	de	brûlures	du	PATIENT	et	de	l’OPERATEUR,	et	d’incendies.	Il	n’a	pas	été	jugé	pratique
de	spécifier	les	exigences	d’interface	applicables	aux	raccords	électriques	pour	assurer	la	compatibilité
entre	les	HUMIDIFICATEURS	et	les	TUYAUX	RESPIRATOIRES	produits	par	différents	FABRICANTS.
Comme	l’utilisation	sans	risque	d’un	HUMIDIFICATEUR	dépend	de	l’interaction	entre	l’HUMIDIFICATEUR	et
ses	nombreux	ACCESSOIRES,	le	présent	document	établit	des	exigences	de	performance	s’appliquant	à
l’ensemble	du	système	jusqu’à	l’ORIFICE	DE	RACCORDEMENT	COTE	PATIENT.	Ces	exigences	sont	applicables
aux	 ACCESSOIRES	tels	que	les	 TUYAUX	 RESPIRATOIRES	(chauffés	et	non	chauffés),	les	capteurs	de
température	et	les	dispositifs	destinés	à	maîtriser	l’environnement	à	l’intérieur	de	ces	TUYAUX
RESPIRATOIRES.
L’humidification	peut	éga
 ...