ASTM D7590-09(2014)
(Guide)Standard Guide for Measurement of Remaining Primary Antioxidant Content In In-Service Industrial Lubricating Oils by Linear Sweep Voltammetry
Standard Guide for Measurement of Remaining Primary Antioxidant Content In In-Service Industrial Lubricating Oils by Linear Sweep Voltammetry
SIGNIFICANCE AND USE
5.1 The quantitative determination of remaining antioxidants for in-service industrial oils by measuring the amount of these additives that have been added to the oil as protection against oxidation. Industrial lubricants, such as turbine oils, compressor oils, gear oils, hydraulic oils, bearing lubricants and greases can be formulated with a wide variety of antioxidants types such as phenols and amines (as primary antioxidants), which are working synergistically and therefore all important to be monitored individually. For in-service oils, the LSV determines and compares the amount of original primary antioxidants remaining after oxidation have reduced its initial concentration.
5.2 This guide covers procedures for primary antioxidants such as amines and phenols, as described by Test Method D6971 and D6810.
5.3 LSV is not designed or intended to detect all of the antioxidant intermediates formed during the thermal and oxidative stressing of the oils, which are recognized as having some contribution to the remaining useful life of the used or in-service oil. In order to measure the overall stability of an oil (including contribution of intermediates present), and before making final judgment on the remaining useful life of the used oil (which might result in the replacement of the oil reservoir), it is advised to perform additional analytical techniques (in accordance with Practice D4378 and Practice D6224).
5.4 This guide is applicable to a wide range of industrial oils, both mineral or synthetic based, which can contain rust and oxidation inhibitors, antiwear additives such as zinc dialkyl dithiophosphates on gear oils, circulating oils, transmission oils and other industrial lubricating oils.
5.5 The test is also suitable for manufacturing control and specification acceptance.
5.6 When a voltammetric analysis is obtained for a industrial lubricant inhibited with at least one type of antioxidant, there is an increase in the current of the produced v...
SCOPE
1.1 This guide covers the voltammetric analysis for qualitative measurements of primary antioxidants in new or in-service type industrial lubricants detectable in concentrations as low as 0.0075 mass percent up to concentrations found in new oils by measuring the amount of current flow at a specified voltage in the produced voltammogram.
1.2 This guide can be used as a resource for a condition monitoring program to track the oxidative health of a range of industrial lubricants which contain primary antioxidants. In order to avoid excessive degradation of the base-oil, these primary antioxidants play a major role to protect the lubricants against thermal-oxidative degradation. This guide can help users with interpretation and troubleshooting results obtained using linear sweep voltammetry (LSV).
1.3 When used as part of oil condition monitoring practices, it is important to apply trend analysis to monitor the antioxidant depletion rate relative to a baseline sample rather than use voltammetry for an absolute measurement of the antioxidant concentration. The trending pattern provides a proactive means to identify the level of oil degradation or abnormal changes in the condition of the in-service lubricant.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
Buy Standard
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: D7590 − 09 (Reapproved 2014)
Standard Guide for
Measurement of Remaining Primary Antioxidant Content In
In-Service Industrial Lubricating Oils by Linear Sweep
Voltammetry
This standard is issued under the fixed designation D7590; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
Under normal thermal and oxidative working conditions, which degrade the chemical composition
of the oil’s basestock and gradually deplete the oil’s additive package, good oil condition monitoring
procedures are necessary to determine and planning corrective actions before the oil properties
changes have passed their warning limits.Antioxidant monitoring practices are a vital part of modern
oil condition monitoring practices to achieve lubrication excellence. This guide addresses the correct
guidelines for voltammetric data interpretation.
1. Scope 1.5 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
1.1 This guide covers the voltammetric analysis for quali-
responsibility of the user of this standard to establish appro-
tative measurements of primary antioxidants in new or in-
priate safety and health practices and determine the applica-
service type industrial lubricants detectable in concentrations
bility of regulatory limitations prior to use.
as low as 0.0075 mass percent up to concentrations found in
newoilsbymeasuringtheamountofcurrentflowataspecified
2. Referenced Documents
voltage in the produced voltammogram.
2.1 ASTM Standards:
1.2 This guide can be used as a resource for a condition
D1193 Specification for Reagent Water
monitoring program to track the oxidative health of a range of
D4057 Practice for Manual Sampling of Petroleum and
industrial lubricants which contain primary antioxidants. In
Petroleum Products
order to avoid excessive degradation of the base-oil, these
D4378 Practice for In-Service Monitoring of Mineral Tur-
primary antioxidants play a major role to protect the lubricants
bine Oils for Steam, Gas, and Combined Cycle Turbines
against thermal-oxidative degradation. This guide can help
D6224 PracticeforIn-ServiceMonitoringofLubricatingOil
users with interpretation and troubleshooting results obtained
for Auxiliary Power Plant Equipment
using linear sweep voltammetry (LSV).
D6304 Test Method for Determination of Water in Petro-
1.3 When used as part of oil condition monitoring practices,
leum Products, Lubricating Oils, and Additives by Cou-
it is important to apply trend analysis to monitor the antioxi-
lometric Karl Fischer Titration
dant depletion rate relative to a baseline sample rather than use
D6810 Test Method for Measurement of Hindered Phenolic
voltammetry for an absolute measurement of the antioxidant
Antioxidant Content in Non-Zinc Turbine Oils by Linear
concentration.The trending pattern provides a proactive means
Sweep Voltammetry
to identify the level of oil degradation or abnormal changes in
D6971 Test Method for Measurement of Hindered Phenolic
the condition of the in-service lubricant.
and Aromatic Amine Antioxidant Content in Non-zinc
Turbine Oils by Linear Sweep Voltammetry
1.4 The values stated in SI units are to be regarded as
D7214 Test Method for Determination of the Oxidation of
standard. No other units of measurement are included in this
Used Lubricants by FT-IR Using Peak Area Increase
standard.
Calculation
This guide is under the jurisdiction of ASTM Committee D02 on Petroleum
Products, Liquid Fuels, and Lubricants and is the direct responsibility of Subcom-
mittee D02.09.0C on Oxidation of Turbine Oils. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
CurrenteditionapprovedMay1,2014.PublishedJuly2014.Originallyapproved contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
ε1
in 2009. Last previous edition approved in 2009 as D7590 – 09 . DOI: 10.1520/ Standards volume information, refer to the standard’s Document Summary page on
D7590-09R14. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D7590 − 09 (2014)
2.2 ISO Standards: standard test method has been developed for the detection of
ISO 4406.2 Hydraulic fluid power—Fluids—Method for other type of antioxidants by linear voltammetry, although
coding the level of contamination by solid particles
LSV also has detection capabilities for these types of second-
4 5
2.3 Other Standards: ary antioxidants (such as zinc dialkyl dithiophosphates).
VGB Guideline VGB-M 416 M In-Service Monitoring of
4.3 A measured quantity of sample is dispensed into a vial
Turbine Oils
containing a measured quantity of a selected test solution and
containing a layer of sand. When the vial is shaken, the
3. Oil Condition Monitoring Programs
antioxidants and other solution soluble oil components present
3.1 Mostindustriallubricantsconsistofmineralorsynthetic
in the sample are extracted into the electrolytic test solution
oilscompoundedwithoxidationandrustinhibitors.Depending
and the remaining droplets suspended in the test solution are
upon their application and the performance level desired,
agglomerated by the sand. The sand/droplet suspension is
specific required amounts of other additives such as metal
allowed to settle out and the antioxidants dissolved in the test
deactivators, pour depressants, extreme pressure additives, and
solution are quantified by voltammetric analysis. The results
foam suppressants can also be present.
arecalculatedandreportedasmasspercentofantioxidantoras
3.2 With modern formulations of industrial lubricants, the
millimoles (mmol) of antioxidant per litre of sample for
antioxidants play a major role in protecting the base-oil against
prepared and fresh oils and as a percent remaining antioxidant
excessive degradation. To prevent this base-oil degradation,
for in-service oils.
resulting in the eventual build-up of deposits, varnish and
4.4 Voltammetric analysis is a technique that applies elec-
sludge, the monitoring of the antioxidants represents a proac-
troanalytic methods wherein a sample to be analyzed is mixed
tive information on the remaining oxidative health of the
with an electrolyte and a solvent (acetone or ethanol based),
in-service lubricant. Oxidation is a chemical reaction between
and placed within an electrolytic cell. Data is obtained by
oxygen atoms with the base oil hydrocarbon molecules, which
measuring the current passing through the cell as a function of
are converting the hydrocarbon molecules into oxidation prod-
the potential applied, and test results are based upon current,
ucts and subsequently weak organic acids. The rate of oxida-
tion depends on the presence of antioxidant additives, which voltage and time relationships at the cell electrodes. The cell
controls the speed of oxidation, but eventually the antioxidants consists of a fluid container into which is mounted a small,
are consumed. Consequently as part of modern proactive easily polarized working electrode, and a large non-polarizable
maintenance strategies, it is vital to know at any time during
reference electrode. The reference electrode should be massive
the operating cycle of the lubricants, its condition by assessing relative to the working electrode so that its behavior remains
the remaining activity of antioxidants, to prevent the oxidative
essentially constant with the passage of small current; that is, it
degradation of the base oil.
remains unpolarized during the analysis period. Additional
electrodes, auxiliary electrodes, can be added to the electrode
3.3 Antioxidant monitoring guidelines have been part of
system to eliminate the effects of resistive drop for high
International Standards such as Practice D4378, Practice
resistancesolutions.Inperformingavoltammetricanalysis,the
D6224, and VGB Guideline VGB-M 416 M, as well Interna-
potential across the electrodes is varied linearly with time, and
tional OEM Maintenance Specifications. This guide presents
the resulting current is recorded as a function of the potential.
guidelines for the lubricant professionals using voltammetric
As the increasing voltage is applied to the prepared sample
techniques as part of their regular maintenance strategies, such
within the cell, the various additive species under investigation
as data interpretation, oil analysis frequency, combination with
withintheoilarecausedtoelectrochemicallyoxidize.Thedata
other condition monitoring tests, etc.
recorded during this oxidation reaction can then be used to
4. Summary of Linear Sweep Voltammetric (LSV) Test
determine the remaining useful life of the oil type. A typical
Method
current-potential curve produced during the practice of the
voltammetric test can be seen by reference to Fig. 1. Initially
4.1 Linear Sweep Voltammetric (LSV) test can be per-
the applied potential produces an electrochemical reaction
formed on any type of industrial lubricant containing at least
havingaratesoslowthatvirtuallynocurrentflowsthroughthe
onetypeofantioxidant.Thevoltammetrictestisacomparative
cell. As the voltage is increased, as shown in Fig. 1, the
test method. By establishing a comparison between its refer-
ence oil (fresh oil or standard) and its used oil, this guide can electroactive species (for example, substituted phenols) begin
to oxidize at the working electrode surface, producing an
be used without the specific knowledge on the category to
which the antioxidants belong. anodic rise in the current.As the potential is further increased,
the decrease in the electroactive species concentration at the
4.2 ASTM International has two standards, Test Method
electrode surface and the exponential increase of the oxidation
D6810 and D6971, that shall enable the measurement of the
rate lead to a maximum in the current-potential curve shown in
remaining phenolic and aminic type of antioxidants. No
Fig. 1.
Available from International Organization for Standardization (ISO), 1, ch. de
la Voie-Creuse, Case postale 56, CH-1211, Geneva 20, Switzerland, http://
www.iso.org. “Remaining Useful Life Measurements of Diesel Engine Oils, Automotive
Available from VGB PowerTech e.V., P. O. Box 10 39 32, D-45039 Essen, Engine Oils, Hydraulic Fluids, and Greases Using Cyclic Voltammetric Methods,”
Klinkestraße 27 - 31, D-45136 Essen, http://www.vgb.org. STLE, Lubrication Engineering, Vol 51, 3, pp. 223 –229.
D7590 − 09 (2014)
FIG. 1 Zinc Dialkyl Dithiophosphate (ZDDP) Voltammetric Response in the Neutral Test Solution with Blank Response Zeroed
5. Significance and Use 5.6 When a voltammetric analysis is obtained for a indus-
trial lubricant inhibited with at least one type of antioxidant,
5.1 The quantitative determination of remaining antioxi-
there is an increase in the current of the produced voltammo-
dants for in-service industrial oils by measuring the amount of
gram between 5 to 8 s (or 0.5 to 0.8 V applied voltage) (see
these additives that have been added to the oil as protection
Note1)forthezincdialkyldithiophosphatetypeofantioxidant
against oxidation. Industrial lubricants, such as turbine oils,
(Fig. 1), an increase in the current of the produced voltammo-
compressor oils, gear oils, hydraulic oils, bearing lubricants
gram between 8 to 12 s (or 0.8 to 1.2 V applied voltage) (Fig.
and greases can be formulated with a wide variety of antioxi-
2) (see Note 1) for the aromatic amines, and increase in the
dants types such as phenols and amines (as primary
current of the produced voltammogram between 13 and 16 s
antioxidants), which are working synergistically and therefore
(or 1.3 to 1.6 V applied voltage) (see Note 1) for the hindered
all important to be monitored individually. For in-service oils,
phenols or carbamates in the neutral acetone solution (Fig. 2:
the LSV determines and compares the amount of original
x-axis1s=0.1V),or both. Hindered phenol antioxidants
primary antioxidants remaining after oxidation have reduced
detected by voltammetric analysis include, but are not limited
its initial concentration.
to, 2,6-di-tert -butyl-4-methylphenol; 2,6-di-tert-butylphenol
5.2 This guide covers procedures for primary antioxidants
and 4,4’-Methylenebis(2,6-di- tert-butylphenol). Aromatic
such as amines and phenols, as described by Test Method
amine antioxidants detected by voltammetric analysis include,
D6971 and D6810.
but are not limited to, phenyl alpha naphthylamines, and
5.3 LSV is not designed or intended to detect all of the
alkylated diphenylamines.
antioxidant intermediates formed during the thermal and oxi-
NOTE 1—Voltages listed with respect to reference electrode. The
dative stressing of the oils, which are recognized as having
voltammograms shown in Figs. 1-6 were obtained with a platinum
some contribution to the remaining useful life of the used or
reference electrode and a voltage scan rate of 0.1 V/s.
in-service oil. In order to measure the overall stability of an oil
(including contribution of intermediates present), and before 5.7 For industrial lubricants containing zinc dialkyl dithio-
making final judgment on the remaining useful life of the used
phosphate type of antioxidants, there is an increase in the
oil (which might result in the replacement of the oil reservoir), currentoftheproducedvoltammogrambetween5to8s(or0.5
it is advised to perform additional analytical techniques (in
to 0.8 V applied voltage) (see Note 1) by using the neutral
accordance with Practice D4378 and Practice D6224).
acetone test solution ( see Fig. 1). There is no corresponding
ASTM International standard describing the test method pro-
5.4 This guide is applicable to a wide range of industrial
cedures for measuring zinc dialkyl dithiophosphates type of
oils, both mineral or synthetic based, which can contain rust
antioxidants in industrial lubricants.
and oxidation inhibitors, antiwear additives such as zinc
dialkyl dithiophosphates on gear oils, circulating oils, trans-
5.8 For industrial lubricants containing only aromatic
mission oils and other industrial lubricating oils.
aminesasantioxidants,thereisanincreaseinthecurrentofthe
produced voltammogram between 8 to 12 s (or 0.8 to 1.2 V
5.5 The test is also suitable for manufacturing control and
specification acceptance. applied voltage) (see Note 1) for the aromatic amines, by using
D7590 − 09 (2014)
FIG. 2 Aromatic Amine and Hindered Phenol Voltammetric Response in the Neutral Test
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
´1
Designation: D7590 − 09 D7590 − 09 (Reapproved 2014)
Standard Guide for
Measurement of Remaining Primary Antioxidant Content In
In-Service Industrial Lubricating Oils by Linear Sweep
Voltammetry
This standard is issued under the fixed designation D7590; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
ε NOTE—Corrected figure placement and numbering editorially in March 2011.
INTRODUCTION
Under normal thermal and oxidative working conditions, which degrade the chemical composition
of the oil’s basestock and gradually deplete the oil’s additive package, good oil condition monitoring
procedures are necessary to determine and planning corrective actions before the oil properties
changes have passed their warning limits. Antioxidant monitoring practices are a vital part of modern
oil condition monitoring practices to achieve lubrication excellence. This guide addresses the correct
guidelines for voltammetric data interpretation.
1. Scope
1.1 This guide covers the voltammetric analysis for qualitative measurements of primary antioxidants in new or in-service type
industrial lubricants detectable in concentrations as low as 0.0075 mass percent up to concentrations found in new oils by
measuring the amount of current flow at a specified voltage in the produced voltammogram.
1.2 This guide can be used as a resource for a condition monitoring program to track the oxidative health of a range of industrial
lubricants which contain primary antioxidants. In order to avoid excessive degradation of the base-oil, these primary antioxidants
play a major role to protect the lubricants against thermal-oxidative degradation. This guide can help users with interpretation and
troubleshooting results obtained using linear sweep voltammetry (LSV).
1.3 When used as part of oil condition monitoring practices, it is important to apply trend analysis to monitor the antioxidant
depletion rate relative to a baseline sample rather than use voltammetry for an absolute measurement of the antioxidant
concentration. The trending pattern provides a proactive means to identify the level of oil degradation or abnormal changes in the
condition of the in-service lubricant.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Referenced Documents
2.1 ASTM Standards:
D1193 Specification for Reagent Water
D4057 Practice for Manual Sampling of Petroleum and Petroleum Products
D4378 Practice for In-Service Monitoring of Mineral Turbine Oils for Steam, Gas, and Combined Cycle Turbines
D6224 Practice for In-Service Monitoring of Lubricating Oil for Auxiliary Power Plant Equipment
This guide is under the jurisdiction of ASTM Committee D02 on Petroleum Products Products, Liquid Fuels, and Lubricants and is the direct responsibility of
Subcommittee D02.09.0C on Oxidation of Turbine Oils.
Current edition approved Dec. 1, 2009May 1, 2014. Published February 2010July 2014. Originally approved in 2009. Last previous edition approved in 2009 as
ε1
D7590 – 09 . DOI: 10.1520/D7590-09R14.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D7590 − 09 (2014)
D6304 Test Method for Determination of Water in Petroleum Products, Lubricating Oils, and Additives by Coulometric Karl
Fischer Titration
D6810 Test Method for Measurement of Hindered Phenolic Antioxidant Content in Non-Zinc Turbine Oils by Linear Sweep
Voltammetry
D6971 Test Method for Measurement of Hindered Phenolic and Aromatic Amine Antioxidant Content in Non-zinc Turbine Oils
by Linear Sweep Voltammetry
D7214 Test Method for Determination of the Oxidation of Used Lubricants by FT-IR Using Peak Area Increase Calculation
2.2 ISO Standards:
ISO 4406.2 Hydraulic fluid power -- Fluids -- Method power—Fluids—Method for coding the level of contamination by solid
particles
2.3 Other Standards:
VGB Guideline VGB-M 416 M In-Service Monitoring of Turbine Oils
3. Oil Condition Monitoring Programs
3.1 Most industrial lubricants consist of mineral or synthetic oils compounded with oxidation and rust inhibitors. Depending
upon their application and the performance level desired, specific required amounts of other additives such as metal deactivators,
pour depressants, extreme pressure additives, and foam suppressants can also be present.
3.2 With modern formulations of industrial lubricants, the antioxidants play a major role in protecting the base-oil against
excessive degradation. To prevent this base-oil degradation, resulting in the eventual build-up of deposits, varnish and sludge, the
monitoring of the antioxidants represents a proactive information on the remaining oxidative health of the in-service lubricant.
Oxidation is a chemical reaction between oxygen atoms with the base oil hydrocarbon molecules, which are converting the
hydrocarbon molecules into oxidation products and subsequently weak organic acids. The rate of oxidation depends on the
presence of antioxidant additives, which controls the speed of oxidation, but eventually the antioxidants are consumed.
Consequently as part of modern proactive maintenance strategies, it is vital to know at any time during the operating cycle of the
lubricants, its condition by assessing the remaining activity of antioxidants, to prevent the oxidative degradation of the base oil.
3.3 Antioxidant monitoring guidelines have been part of International Standards such as Practice D4378, Practice D6224, and
VGB Guideline VGB-M 416 M, as well International OEM Maintenance Specifications. This guide presents guidelines for the
lubricant professionals using voltammetric techniques as part of their regular maintenance strategies, such as data interpretation,
oil analysis frequency, combination with other condition monitoring tests, etc.
4. Summary of Linear Sweep Voltammetric (LSV) Test Method
4.1 Linear Sweep Voltammetric (LSV) test can be performed on any type of industrial lubricant containing at least one type of
antioxidant. The voltammetric test is a comparative test method. By establishing a comparison between its reference oil (fresh oil
or standard) and its used oil, this guide can be used without the specific knowledge on the category to which the antioxidants
belong.
4.2 ASTM International has two standards, Test Method D6810 and D6971, that shall enable the measurement of the remaining
phenolic and aminic type of antioxidants. No standard test method has been developed for the detection of other type of
antioxidants by linear voltammetry, although LSV also has detection capabilities for these types of secondary antioxidants (such
as zinc dialkyl dithiophosphates).
4.3 A measured quantity of sample is dispensed into a vial containing a measured quantity of a selected test solution and
containing a layer of sand. When the vial is shaken, the antioxidants and other solution soluble oil components present in the
sample are extracted into the electrolytic test solution and the remaining droplets suspended in the test solution are agglomerated
by the sand. The sand/droplet suspension is allowed to settle out and the antioxidants dissolved in the test solution are quantified
by voltammetric analysis. The results are calculated and reported as mass percent of antioxidant or as millimoles (mmol) of
antioxidant per litre of sample for prepared and fresh oils and as a percent remaining antioxidant for in-service oils.
4.4 Voltammetric analysis is a technique that applies electroanalytic methods wherein a sample to be analyzed is mixed with
an electrolyte and a solvent (acetone or ethanol based), and placed within an electrolytic cell. Data is obtained by measuring the
current passing through the cell as a function of the potential applied, and test results are based upon current, voltage and time
relationships at the cell electrodes. The cell consists of a fluid container into which is mounted a small, easily polarized working
electrode, and a large non-polarizable reference electrode. The reference electrode should be massive relative to the working
electrode so that its behavior remains essentially constant with the passage of small current; that is, it remains unpolarized during
the analysis period. Additional electrodes, auxiliary electrodes, can be added to the electrode system to eliminate the effects of
Available from International Organization for Standardization (ISO), 1, ch. de la Voie-Creuse, Case postale 56, CH-1211, Geneva 20, Switzerland, http://www.iso.org.
Available from VGB PowerTech e.V., P. O. Box 10 39 32, D-45039 Essen, Klinkestraße 27 - 31, D-45136 Essen, http://www.vgb.org.
“Remaining Useful Life Measurements of Diesel Engine Oils, Automotive Engine Oils, Hydraulic Fluids, and Greases Using Cyclic Voltammetric Methods,” STLE,
Lubrication Engineering, Vol 51, 3, pp. 223 –229.
D7590 − 09 (2014)
FIG. 1 Zinc Dialkyl Dithiophosphate (ZDDP) Voltammetric Response in the Neutral Test Solution with Blank Response Zeroed
resistive drop for high resistance solutions. In performing a voltammetric analysis, the potential across the electrodes is varied
linearly with time, and the resulting current is recorded as a function of the potential. As the increasing voltage is applied to the
prepared sample within the cell, the various additive species under investigation within the oil are caused to electrochemically
oxidize. The data recorded during this oxidation reaction can then be used to determine the remaining useful life of the oil type.
A typical current-potential curve produced during the practice of the voltammetric test can be seen by reference to Fig. 1. Initially
the applied potential produces an electrochemical reaction having a rate so slow that virtually no current flows through the cell.
As the voltage is increased, as shown in Fig. 1, the electroactive species (for example, substituted phenols) begin to oxidize at the
working electrode surface, producing an anodic rise in the current. As the potential is further increased, the decrease in the
electroactive species concentration at the electrode surface and the exponential increase of the oxidation rate lead to a maximum
in the current-potential curve shown in Fig. 1.
5. Significance and Use
5.1 The quantitative determination of remaining antioxidants for in-service industrial oils by measuring the amount of these
additives that have been added to the oil as protection against oxidation. Industrial lubricants, such as turbine oils, compressor oils,
gear oils, hydraulic oils, bearing lubricants and greases can be formulated with a wide variety of antioxidants types such as phenols
and amines (as primary antioxidants), which are working synergistically and therefore all important to be monitored individually.
For in-service oils, the LSV determines and compares the amount of original primary antioxidants remaining after oxidation have
reduced its initial concentration.
5.2 This guide covers procedures for primary antioxidants such as amines and phenols, as described by Test Method D6971 and
D6810.
5.3 LSV is not designed or intended to detect all of the antioxidant intermediates formed during the thermal and oxidative
stressing of the oils, which are recognized as having some contribution to the remaining useful life of the used or in-service oil.
In order to measure the overall stability of an oil (including contribution of intermediates present), and before making final
judgment on the remaining useful life of the used oil (which might result in the replacement of the oil reservoir), it is advised to
perform additional analytical techniques (in accordance with Practice D4378 and Practice D6224).
5.4 This guide is applicable to a wide range of industrial oils, both mineral or synthetic based, which can contain rust and
oxidation inhibitors, antiwear additives such as zinc dialkyl dithiophosphates on gear oils, circulating oils, transmission oils and
other industrial lubricating oils.
5.5 The test is also suitable for manufacturing control and specification acceptance.
5.6 When a voltammetric analysis is obtained for a industrial lubricant inhibited with at least one type of antioxidant, there is
an increase in the current of the produced voltammogram between 5 to 8 s (or 0.5 to 0.8 V applied voltage) (see Note 1) for the
zinc dialkyl dithiophosphate type of antioxidant (Fig. 1), an increase in the current of the produced voltammogram between 8 to
12 s (or 0.8 to 1.2 V applied voltage) (Fig. 2) (see Note 1) for the aromatic amines, and increase in the current of the produced
voltammogram between 13 and 16 s (or 1.3 to 1.6 V applied voltage) (see Note 1) for the hindered phenols or carbamates in the
D7590 − 09 (2014)
FIG. 2 Aromatic Amine and Hindered Phenol Voltammetric Response in the Neutral Test Solution with Blank Response Zeroed
neutral acetone solution (Fig. 2: x-axis 1 s = 0.1 V), or both. Hindered phenol antioxidants detected by voltammetric analysis
include, but are not limited to, 2,6-di-tert -butyl-4-methylphenol; 2,6-di-tert-butylphenol and 4,4’-Methylenebis(2,6-di- tert-
butylphenol). Aromatic amine antioxidants detected by voltammetric analysis include, but are not limited to, phenyl alpha
naphthylamines, and alkylated diphenylamines.
NOTE 1—Voltages listed with respect to reference electrode. The voltammograms shown in Figs. 1-6 were obtained with a platinum reference electrode
and a voltage scan rate of 0.1 V/s.
5.7 For industrial lubricants containing zinc dialkyl dithiophosphate type of antioxidants, there is an increase in the current of
the produced voltammogram between
...










Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.