ASTM E341-08(2015)
(Practice)Standard Practice for Measuring Plasma Arc Gas Enthalpy by Energy Balance
Standard Practice for Measuring Plasma Arc Gas Enthalpy by Energy Balance
SIGNIFICANCE AND USE
3.1 The purpose of this practice is to measure the total or stagnation gas enthalpy of a plasma-arc gas stream in which nonreactive gases are heated by passage through an electrical discharge device during calibration tests of the system.
3.2 The plasma arc represents one heat source for determining the performance of high temperature materials under simulated hyperthermal conditions. As such the total or stagnation enthalpy is one of the important parameters for correlating the behavior of ablation materials.
3.3 The most direct method for obtaining a measure of total enthalpy, and one which can be performed simultaneously with each material test, if desired, is to perform an energy balance on the arc chamber. In addition, in making the energy balance, accurate measurements are needed since the efficiencies of some plasma generators are low (as low as 15 to 20 % or less in which case the enthalpy depends upon the difference of two quantities of nearly equal magnitude). Therefore, the accuracy of the measurements of the primary variables must be high, all energy losses must be correctly taken into account, and steady-state conditions must exist both in plasma performance and fluid flow.
3.4 In particular it is noted that total enthalpy as determined by the energy balance technique is most useful if the plasma generator design minimizes coring effects. If nonuniformity exists the enthalpy determined by energy balance gives only the average for the entire plasma stream, whereas the local enthalpy experienced by a model in the core of the stream may be much higher. More precise methods are needed to measure local variations in total enthalpy.
SCOPE
1.1 This practice covers the measurement of total gas enthalpy of an electric-arc-heated gas stream by means of an overall system energy balance. This is sometimes referred to as a bulk enthalpy and represents an average energy content of the test stream which may differ from local values in the test stream.
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
Buy Standard
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E341 − 08 (Reapproved 2015)
Standard Practice for
Measuring Plasma Arc Gas Enthalpy by Energy Balance
This standard is issued under the fixed designation E341; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope This is a direct application of the First Law of Thermodynam-
ics and, for the particular control volume cited here, can be
1.1 This practice covers the measurement of total gas
written as follows:
enthalpy of an electric-arc-heated gas stream by means of an
overallsystemenergybalance.Thisissometimesreferredtoas EnergyIn 2 EnergyOut 5 EnergytoGas (1)
abulkenthalpyandrepresentsanaverageenergycontentofthe
n p
test stream which may differ from local values in the test
¯
EI 2 Q 2 W C ∆T 2∆T 2 M H
~ !
CR ( p 0 1 ( j j
H O H O
2 i 2 i
i51 j51
stream.
1.2 This standard does not purport to address all of the
5W H 2 H
~ !
g g in
safety concerns, if any, associated with its use. It is the
where:
responsibility of the user of this standard to establish appro-
C = water, specific heat,
priate safety and health practices and determine the applica-
p
E = plasma arc voltage,
bility of regulatory limitations prior to use.
H = exhaust gas enthalpy,
g
H = inlet gas enthalpy,
in
2. Summary of Test Method
H = heat of vaporization corresponding to the material
j
2.1 A measure of the total or stagnation gas enthalpy of
M ,
j
plasma-arc heated gases (nonreacting) is based upon the
I = plasma arc current,
following measurements:
M = mass loss rate of electrode insulator, interior metal
j
2.1.1 Energy input to the plasma arc,
surface, etc.
2.1.2 Energy losses to the plasma arc hardware and cooling Q = energy convected and radiated from external sur-
CR
face of plasma generator,
water, and
∆T = T − T =water temperature rise during plasma
2.1.3 Gas mass flow. 0 0 0
H2O 2 1
arc operation,
2.2 The gas enthalpy is determined numerically by dividing
∆T = T −T =water temperature rise before plasma arc
1 2 1
H2O
the gas mass flow into the net power input to the plasma arc
operation,
(power to plasma arc minus the energy losses).
T = water exhaust temperature during plasma arc
operation,
2.3 Thetechniqueforperformingtheoverallenergybalance
T = inlet water temperature during plasma arc
is illustrated schematically in Fig. 1. The control volume for
operation,
theenergybalancecanberepresentedbytheentireenvelopeof
T = water exhaust temperature before plasma arc
this drawing. Gas enters at an initial temperature, or enthalpy, 2
operation,
andemergesatahigherenthalpy.Waterorothercoolantenters
T = inlet water temperature before plasma arc
the control volume at an initial temperature and emerges at a
operation,
higher temperature. Across the arc, electrical energy is dissi-
W = gas flow rate,
g
pated by virtue of the resistance and current in the arc itself.A
W O = mass flow rate of coolant water, and
H
heat balance of the system requires that the energy gained by 2
¯
= averageoftheproductofvoltage, E,andcurrent, I.
EI
thegasmustbedefinedbythedifferencebetweentheincoming
energy (electrical input) and total coolant and external losses.
2.4 AnexaminationofEq1showsthat,inordertoobtainan
evaluation of the energy content of the plasma for a specified
setofoperatingconditions,measurementsmustbemadeofthe
voltage and current, the mass-flow rate and temperature rise of
This practice is under the jurisdiction of ASTM Committee E21 on Space
Simulation andApplications of SpaceTechnology and is the direct responsibility of
the coolant, the mass-flow rate and inlet ambient temperature
Subcommittee E21.08 on Thermal Protection.
of the test gas, and the external surface temperature and
Current edition approved May 1, 2015. Published June 2015. Originally
housing of the arc chamber. For all practical purposes, the
approved in 1968. Last previous edition approved in 2008 as E341–08. DOI:
10.1520/E0341-08R15. external surface temperature of the water-cooled plasma arc is
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E341 − 08 (2015)
FIG. 1 Schematic Energy Balance Method for Determining Gas
Enthalpy
minimum. Consequently, it will be assumed throughout this each material test, if desired, is to perform an energy balance
discussion that negligible energy (compared to the input on the arc chamber. In addition, in making the energy balance,
energy) is lost from the external plasma generator surface by accurate measurements are needed since the efficiencies of
convectiveorradiativemechanismsandthattheinternallossof
some plasma generators are low (as low as 15 to 20% or less
electrode or plasma generator material is small compared with in which case the enthalpy depends upon the difference of two
theenergyinput.Inaddition,assomeplasmageneratorsutilize
quantities of nearly equal magnitude). Therefore, the accuracy
magneticfieldsintheirdesign,themagneticfieldcoilelectrical
of the measurements of the primary variables must be high, all
power and ohmic-heating dissipation should be included in the
energy losses must be correctly taken into account, and
over-all heat balance. Precautions should be taken to assure
steady-state conditions must exist both in plasma performance
that only a negligible portion of magnetic energy is being
and fluid flow.
dissipated in hardware not within the heat balance circuit. For
3.4 In particular it is noted that total enthalpy as determined
the purposes of this discussion, the magnetic field power input
by the energy balance technique is most useful if the plasma
and loss aspects have been omitted because of their unique
generator design minimizes coring effects. If nonuniformity
applicability to specific plasma generator designs.
exists the enthalpy determined by energy balance gives only
2.5 The energy balance is given by Eq 2 when these factors
the average for the entire plasma stream, whereas the local
are taken into account:
enthalpyexperiencedbyamodelinthecoreofthestreammay
n
be much higher. More precise methods are needed to measure
¯
EI 2 W C ∆T 2∆T 5 W H 2 H (2)
~ ! ~ !
H O p 0 1 g g in
( H O
2 i 2 i
local variations in total enthalpy.
i51
The exhaust enthalpy, H , of the effluent as defined by Eq 1
g
4. Apparatus
and 2 is a measure of the average total (stagnation) enthalpy at
the nozzle exit plane of the plasma-arc heater. This enthalpy
4.1 General—The apparatus shall consist of the plasma-arc
does not necessarily apply to the plasma downstream of the
facility and the necessary instrumentation to measure the
nozzle exit plane.
power input to the arc, gas stream and coolant flow rates, inlet
gas temperature and net coolant temperature rise of the plasma
3. Significance and Use
generator hardware. Although the recommended instrumenta-
3.1 The purpose of this practice is to measure the total or
tion accuracies are state-of-the-art values, higher accuracy
stagnation gas enthalpy of a plasma-arc gas stream in which
instruments (than those recommended) may be required for
nonreactive gases are heated by passage through an electrical
low efficiency plasma generators.
discharge device during calibration tests of the system.
4.2 Input Energy Measurements—The energy input term,
3.2 Theplasmaarcrepresentsoneheatsourcefordetermin-
EI, to a large degree may be time dependent. Fluctuations in
ing the performance of high temperature materials under
the power input can produce errors as large as 50% under
simulated hyperthermal conditions. As such the total or stag-
certain conditions. The magnitude of the error will depend on
nation enthalpy is one of the important parameters for corre-
theamplitudeoftheunsteadycomparedwiththesteadyportion
lating the behavior of ablation materials.
of the current and voltage and also on the instantaneous phase
3.3 The most direct method for obtaining a measure of total relationship between current and voltage. The power input
enthalpy,andonewhichcanbeperformedsimultaneouslywith portion term should be written:
E341 − 08 (2015)
t
4.3.2 Coolant Temperature Measurement—The method of
¯
EI 51/t * EIdt (3)
temperature measurement must be sufficiently sensitive and
reliable to ensure accurate measurement of the coolant water
Asaconsequenceeachplasmageneratorshouldmakeuseof
oscilloscopic voltage-current traces during operation in order temperature rise. Procedures similar to those given in the
Annual Book of ASTM Standards, Part 44, and Ref (3) should
to ascertain the time variation of the voltage-current input. If
these traces show significant unsteadiness it is recommended be adhered to in the calibration and preparation of temperature
sensors. The bulk or average temperature of the coolant shall
that additional methods of input power measurements be
pursued, such as an integrating device if available. In order to be measured at the inlet and output lines of each cooled unit.
The error in measurement of temperature difference between
measurepowerdirectly,awattmeterascitedbyDawes(1) can
be employed. As a precaution in the use of the wattmeter, inlet and outlet shall be not more than 61 %. The water
temperature-indicating devices shall be placed as close as
reversed readings of current and voltage should be taken and
the average of the two readings used. For those plasma practical to the plasma arc in the inlet and outlet lines. No
additional apparatus shall be between the temperature sensor
generator facilities which operate under known and steady
input power the use of a voltmeter and ammeter is recom- and the plasma arc. The temperature measurements shall be
recorded continuously. Ref (2) lists a variety of commercially
mended owing to their high degree of accuracy.
available temperature sensors. During the course of operation
4.2.1 Voltage Measurement—The determination of power
oftheplasmaarc,careshouldbetakentominimizedepositson
input to the plasma generator requires the measurement of the
the sensors and to eliminate any possibility of sensor heating
voltageac
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E341 − 08 E341 − 08 (Reapproved 2015)
Standard Practice for
Measuring Plasma Arc Gas Enthalpy by Energy Balance
This standard is issued under the fixed designation E341; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This practice covers the measurement of total gas enthalpy of an electric-arc-heated gas stream by means of an overall
system energy balance. This is sometimes referred to as a bulk enthalpy and represents an average energy content of the test stream
which may differ from local values in the test stream.
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Summary of Test Method
2.1 A measure of the total or stagnation gas enthalpy of plasma-arc heated gases (nonreacting) is based upon the following
measurements:
2.1.1 Energy input to the plasma arc,
2.1.2 Energy losses to the plasma arc hardware and cooling water, and
2.1.3 Gas mass flow.
2.2 The gas enthalpy is determined numerically by dividing the gas mass flow into the net power input to the plasma arc (power
to plasma arc minus the energy losses).
2.3 The technique for performing the overall energy balance is illustrated schematically in Fig. 1. The control volume for the
energy balance can be represented by the entire envelope of this drawing. Gas enters at an initial temperature, or enthalpy, and
emerges at a higher enthalpy. Water or other coolant enters the control volume at an initial temperature and emerges at a higher
temperature. Across the arc, electrical energy is dissipated by virtue of the resistance and current in the arc itself. A heat balance
of the system requires that the energy gained by the gas must be defined by the difference between the incoming energy (electrical
input) and total coolant and external losses. This is a direct application of the First Law of Thermodynamics and, for the particular
control volume cited here, can be written as follows:
Energy In 2 Energy Out 5 Energy to Gas (1)
n p
¯
EI 2 Q 2 W C ΔT 2 ΔT 2 M H
~ !
CR ( p 0 1 ( j j
H O H O
2 i 2 i
i51 j51
5W H 2 H
~ !
g g in
where:
C = water, specific heat,
p
E = plasma arc voltage,
H = exhaust gas enthalpy,
g
H = inlet gas enthalpy,
in
H = heat of vaporization corresponding to the material M ,
j j
I = plasma arc current,
M = mass loss rate of electrode insulator, interior metal surface, etc.
j
Q = energy convected and radiated from external surface of plasma generator,
CR
This practice is under the jurisdiction of ASTM Committee E21 on Space Simulation and Applications of Space Technology and is the direct responsibility of
Subcommittee E21.08 on Thermal Protection.
Current edition approved May 1, 2008May 1, 2015. Published May 2008June 2015. Originally approved in 1968. Last previous edition approved in 20022008 as E341 – 96
(2002).E341 – 08. DOI: 10.1520/E0341-08.10.1520/E0341-08R15.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E341 − 08 (2015)
FIG. 1 Schematic Energy Balance Method for Determining Gas
Enthalpy
ΔT = T − T = water temperature rise during plasma arc operation,
0 0 0
H2O 2 1
ΔT = T − T = water temperature rise before plasma arc operation,
1 2 1
H2O
T = water exhaust temperature during plasma arc operation,
T = inlet water temperature during plasma arc operation,
T = water exhaust temperature before plasma arc operation,
T = inlet water temperature before plasma arc operation,
W = gas flow rate,
g
W = mass flow rate of coolant water, and
H O
¯
= average of the product of voltage, E, and current, I.
EI
2.4 An examination of Eq 1 shows that, in order to obtain an evaluation of the energy content of the plasma for a specified set
of operating conditions, measurements must be made of the voltage and current, the mass-flow rate and temperature rise of the
coolant, the mass-flow rate and inlet ambient temperature of the test gas, and the external surface temperature and housing of the
arc chamber. For all practical purposes, the external surface temperature of the water-cooled plasma arc is minimum. Consequently,
it will be assumed throughout this discussion that negligible energy (compared to the input energy) is lost from the external plasma
generator surface by convective or radiative mechanisms and that the internal loss of electrode or plasma generator material is
small compared with the energy input. In addition, as some plasma generators utilize magnetic fields in their design, the magnetic
field coil electrical power and ohmic-heating dissipation should be included in the over-all heat balance. Precautions should be
taken to assure that only a negligible portion of magnetic energy is being dissipated in hardware not within the heat balance circuit.
For the purposes of this discussion, the magnetic field power input and loss aspects have been omitted because of their unique
applicability to specific plasma generator designs.
2.5 The energy balance is given by Eq 23 when these factors are taken into account:
n
¯
EI 2 W C ΔT 2 ΔT 5 W H 2 H (2)
~ ! ~ !
( H O p 0 1 H O g g in
2 i 2 i
i51
The exhaust enthalpy, H , of the effluent as defined by Eq 1 and 23 is a measure of the average total (stagnation) enthalpy at
g
the nozzle exit plane of the plasma-arc heater. This enthalpy does not necessarily apply to the plasma downstream of the nozzle
exit plane.
3. Significance and Use
3.1 The purpose of this practice is to measure the total or stagnation gas enthalpy of a plasma-arc gas stream in which
nonreactive gases are heated by passage through an electrical discharge device during calibration tests of the system.
3.2 The plasma arc represents one heat source for determining the performance of high temperature materials under simulated
hyperthermal conditions. As such the total or stagnation enthalpy is one of the important parameters for correlating the behavior
of ablation materials.
3.3 The most direct method for obtaining a measure of total enthalpy, and one which can be performed simultaneously with each
material test, if desired, is to perform an energy balance on the arc chamber. In addition, in making the energy balance, accurate
measurements are needed since the efficiencies of some plasma generators are low (as low as 15 to 20 % or less in which case the
E341 − 08 (2015)
enthalpy depends upon the difference of two quantities of nearly equal magnitude). Therefore, the accuracy of the measurements
of the primary variables must be high, all energy losses must be correctly taken into account, and steady-state conditions must exist
both in plasma performance and fluid flow.
3.4 In particular it is noted that total enthalpy as determined by the energy balance technique is most useful if the plasma
generator design minimizes coring effects. If nonuniformity exists the enthalpy determined by energy balance gives only the
average for the entire plasma stream, whereas the local enthalpy experienced by a model in the core of the stream may be much
higher. More precise methods are needed to measure local variations in total enthalpy.
4. Apparatus
4.1 General—The apparatus shall consist of the plasma-arc facility and the necessary instrumentation to measure the power
input to the arc, gas stream and coolant flow rates, inlet gas temperature and net coolant temperature rise of the plasma generator
hardware. Although the recommended instrumentation accuracies are state-of-the-art values, higher accuracy instruments (than
those recommended) may be required for low efficiency plasma generators.
4.2 Input Energy Measurements—The energy input term, EI, to a large degree may be time dependent. Fluctuations in the power
input can produce errors as large as 50 % under certain conditions. The magnitude of the error will depend on the amplitude of
the unsteady compared with the steady portion of the current and voltage and also on the instantaneous phase relationship between
current and voltage. The power input portion term should be written:
t
¯
EI 5 1/t EI dt (3)
*
As a consequence each plasma generator should make use of oscilloscopic voltage-current traces during operation in order to
ascertain the time variation of the voltage-current input. If these traces show significant unsteadiness it is recommended that
additional methods of input power measurements be pursued, such as an integrating device if available. In order to measure power
directly, a wattmeter as cited by Dawes (1) can be employed. As a precaution in the use of the wattmeter, reversed readings of
current and voltage should be taken and the average of the two readings used. For those plasma generator facilities which operate
under known and steady input power the use of a voltmeter and ammeter is recommended owing to their high degree of accuracy.
4.2.1 Voltage Measurement—The determination of power input to the plasma generator requires the measurement of the voltage
across the circuit. Suitable instruments for such voltage measurements are presented by the Instrument Society of America (ISA)
(2). The measurement techniques to be used can be either a voltage divider network or a direct reading instrument. It is highly
desirable to be able to record the voltage such that time variations are a part of the test data. Accuracy of the voltage measurements
shall be within 61 %. The voltage measurement shall be taken at the electrode terminals of the plasma generator circuit.
4.2.2 Current Measurement—The measurement of plasma arc current shall be accomplished with an ammeter equipped with a
precision shunt and the reading shall be within 61 %. Ref (2) lists other instruments suitable for measuring arc current. If a
precision shunt is utilized, the temperature across the shunt shall be constant and within the stated limits as given by the
manufacturer. Arc current shall be measur
...










Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.