ASTM F2423-11(2016)
(Guide)Standard Guide for Functional, Kinematic, and Wear Assessment of Total Disc Prostheses
Standard Guide for Functional, Kinematic, and Wear Assessment of Total Disc Prostheses
SIGNIFICANCE AND USE
4.1 This guide can be used to determine the fatigue and wear behavior of IVD prostheses subjected to functional and kinematic cyclic loading/motion for relatively large numbers of cycles (for example, various designs of IVD prostheses, as well as the effects of materials, manufacturing techniques and other design variables on one particular design can be determined using this guide).
4.2 This guide is intended to be applicable to IVD prostheses, that support load and transmit motion by means of an articulating joint or by use of compliant materials. Ceramics, metals, or polymers, or combination thereof, are used in IVD prostheses, and it is the goal of this guide to enable a kinematic wear and/or fatigue comparison of these devices, regardless of material and type of device.
SCOPE
1.1 This guide provides guidance for wear and/or fatigue testing of total disc prostheses under functional and kinematic conditions and, to this end, describes test methods for assessment of the wear or functional characteristics, or both, of total disc prostheses.
1.2 Both lumbar and cervical prostheses are addressed.
1.3 Load and kinematic profiles for lumbar and cervical devices are not identical and, therefore, are addressed separately in the guide.
1.4 Partial disc replacements, such as nucleus replacements or facet joint replacements, are not intended to be addressed.
1.5 Wear is assessed using a weight loss method in a testing medium as defined in this guide.
1.6 This guide does not address any potential failure mode as it relates to the fixation of the implant to its bony interfaces.
1.7 It is the intent of this guide to enable comparison of intervertebral disc (IVD) prostheses with regard to wear and fatigue characteristics when tested under the specified conditions. It must be recognized, however, that there are many possible variations in in vivo conditions. A single laboratory simulation with a fixed set of parameters might not be universally representative.
1.8 Most IVD prostheses primarily fall into two classifications: articulating ball-in-socket type prostheses, and elastomeric or compliant type prostheses. For the former, this guide primarily addresses Mode 1 wear (defined herein); whereas for the latter, this guide addresses potential failure of the prosthesis when the implant is subjected to a range of motion and/or loads that fall within the full range of possible physiologic motions and loads.
1.9 For articulating components, this guide predominantly describes a Mode 1 test. The user is cautioned that other modes of wear may occur and may have significant influence on the functionality and performance of an articulating IVD prosthesis, and therefore the user should consider the effects of other wear modes on the performance of the prosthesis.
1.10 In order that the data be reproducible and comparable within and between laboratories, it is essential that uniform procedures are established. This guide is intended to facilitate uniform methods for testing and reporting of data for total disc replacement prostheses.
1.11 Without a substantial clinical retrieval history of IVD prostheses, actual loading profiles and patterns cannot be delineated at the time of the writing of this guide. It therefore follows that the load and motion conditions specified by this guide do not necessarily accurately reproduce those occurring in vivo. Rather, this guide provides useful boundary/endpoint conditions for evaluating prosthesis designs in a functional manner.
1.12 The values stated in SI units are to be regarded as the standard with the exception of angular measurements, which may be reported in either degrees or radians.
1.13 This guide is not intended to be a performance standard. It is the responsibility of the user of this guide to characterize the safety and effectiveness of the prosthesis under evaluation.
1.14 This standard does not purport to address all of th...
General Information
Buy Standard
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: F2423 − 11 (Reapproved 2016)
Standard Guide for
Functional, Kinematic, and Wear Assessment of Total Disc
Prostheses
This standard is issued under the fixed designation F2423; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope of wear may occur and may have significant influence on the
functionality and performance of an articulating IVD
1.1 This guide provides guidance for wear and/or fatigue
prosthesis,andthereforetheusershouldconsidertheeffectsof
testing of total disc prostheses under functional and kinematic
other wear modes on the performance of the prosthesis.
conditions and, to this end, describes test methods for assess-
ment of the wear or functional characteristics, or both, of total 1.10 In order that the data be reproducible and comparable
disc prostheses. within and between laboratories, it is essential that uniform
procedures are established. This guide is intended to facilitate
1.2 Both lumbar and cervical prostheses are addressed.
uniformmethodsfortestingandreportingofdatafortotaldisc
1.3 Load and kinematic profiles for lumbar and cervical
replacement prostheses.
devices are not identical and, therefore, are addressed sepa-
1.11 Without a substantial clinical retrieval history of IVD
rately in the guide.
prostheses, actual loading profiles and patterns cannot be
1.4 Partial disc replacements, such as nucleus replacements
delineated at the time of the writing of this guide. It therefore
or facet joint replacements, are not intended to be addressed.
follows that the load and motion conditions specified by this
guide do not necessarily accurately reproduce those occurring
1.5 Wearisassessedusingaweightlossmethodinatesting
medium as defined in this guide. in vivo. Rather, this guide provides useful boundary/endpoint
conditions for evaluating prosthesis designs in a functional
1.6 This guide does not address any potential failure mode
manner.
as it relates to the fixation of the implant to its bony interfaces.
1.12 The values stated in SI units are to be regarded as the
1.7 It is the intent of this guide to enable comparison of
standard with the exception of angular measurements, which
intervertebral disc (IVD) prostheses with regard to wear and
may be reported in either degrees or radians.
fatigue characteristics when tested under the specified condi-
1.13 This guide is not intended to be a performance stan-
tions. It must be recognized, however, that there are many
possible variations in in vivo conditions. A single laboratory dard. It is the responsibility of the user of this guide to
characterizethesafetyandeffectivenessoftheprosthesisunder
simulation with a fixed set of parameters might not be
universally representative. evaluation.
1.14 This standard does not purport to address all of the
1.8 Most IVD prostheses primarily fall into two classifica-
safety concerns, if any, associated with its use. It is the
tions: articulating ball-in-socket type prostheses, and elasto-
responsibility of the user of this standard to establish appro-
meric or compliant type prostheses. For the former, this guide
priate safety and health practices and determine the applica-
primarilyaddressesMode1wear(definedherein);whereasfor
bility of regulatory limitations prior to use.
thelatter,thisguideaddressespotentialfailureoftheprosthesis
whentheimplantissubjectedtoarangeofmotionand/orloads
2. Referenced Documents
that fall within the full range of possible physiologic motions
and loads.
2.1 ASTM Standards:
F561 Practice for Retrieval and Analysis of Medical
1.9 For articulating components, this guide predominantly
Devices, and Associated Tissues and Fluids
describesaMode1test.Theuseriscautionedthatothermodes
F1582Terminology Relating to Spinal Implants
F1714GuideforGravimetricWearAssessmentofProsthetic
This guide is under the jurisdiction ofASTM Committee F04 on Medical and
Surgical Materials and Devices and is the direct responsibility of Subcommittee
F04.25 on Spinal Devices. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved Dec. 1, 2016. Published December 2016. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 2005. Last previous edition approved in 2011 as F2423–11. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/F2423-11R16. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
F2423 − 11 (2016)
Hip Designs in Simulator Devices 3.2.2.3 Y-axis, n—positive Y-axis is a global fixed axis
F1877Practice for Characterization of Particles relative to the test machine’s stationary base, and is directed
F2077TestMethodsForIntervertebralBodyFusionDevices laterally relative to the specimen’s initial unloaded position.
2.2 ISO Standard:
3.2.2.4 Z-axis, n—positive Z-axis is a global fixed axis
ISO 18192–1Implants for Surgery—Wear ofTotal Interver-
relative to the test machine’s stationary base, and is to be
tebral Spinal Disc Prostheses—Part 1: Loading and Dis-
directed superiorly relative to the specimen’s initial unloaded
placement Parameters for Wear Testing and Correspond-
position.
ing Environmental Conditions for Test
3.2.2.5 x-axis, n—positive x-axis is a fixed axis relative to
3. Terminology
the IVD prosthesis and a moving axis relative to the global
coordinate system, and is directed anteriorly relative to the
3.1 All functional and kinematic testing terminology is
prosthesis.
consistent with the referenced standards (for example, Test
Methods F2077, Terminology F1582, and so forth), unless
3.2.2.6 y-axis, n—positive y-axis is a fixed axis relative to
otherwise stated.
the IVD prosthesis and a moving axis relative to the global
coordinate system, and is directed laterally relative to the
3.2 Definitions:
prosthesis.
3.2.1 axial load, n—the resultant force F applied to the
axial
superior or inferior fixture-end plate that simulates the in vivo
3.2.2.7 z-axis, n—positive z-axis is a fixed axis relative to
load that an IVD prosthesis (original healthy disc) must resist.
the IVD prosthesis and a moving axis relative to the global
3.2.1.1 Discussion—Based on a healthy disc, the primary
coordinate system, and is directed superiorly relative to the
component would be an axial compressive force F in the
prosthesis.
Z
direction of the negative global Z axis, and it would pass
3.2.3 degradation, n—loss of material or function or mate-
through the origin of the IVD prosthesis. Shear components in
rial properties as a result of causes other than that associated
the XYplanewouldbe F and F .Lateralbendingmoment M
X Y X
with wear.
and flexion/extension moment M components would be cre-
Y
3.2.4 fluid absorption, n—fluid absorbed by the device
atedabouttheoriginwhentheaxialloaddoesnotpassthrough
material during testing.
it.
3.2.5 functional failure, n—permanent deformation or wear
3.2.2 coordinate system/axes, n—global XYZ orthogonal
that renders the IVD prosthesis assembly ineffective or unable
axesaredefinedfollowingaright-handedCartesiancoordinate
to resist load/motion or any secondary effects that result in a
system in which the XY plane is to bisect the sagittal plane
reduction of clinically relevant motions or the motions in-
angle between the superior and inferior surfaces that are
tended by the design of the device.
intended to simulate the adjacent vertebral end plates. The
global axes are stationary relative to the IVD prosthesis’s
3.2.6 interval net volumetric wear rate VR during cycle
i
inferior end plate fixture, which, in this guide, is also consid-
interval i (mm /million cycles), n—VR = WR/ρ, where ρ =
i i
ered to be stationary with respect to the test machine’s frame.
mass density (for example, units of g/mm ) of the wear
Lower case letters, xyz, denote a local, moving orthogonal
material.
coordinate system attached to the superior end plate fixture
3.2.7 interval net wear rate WR during cycle interval i
i
withdirectionsinitiallycoincidentwiththoseoftheglobalXYZ
(g/million cycles), n—WR =((NW – NW )/(number of cycles
i i i-1
axes, respectively. The 3-D motion of the superior relative to
in interval i))×10 .
the inferior end plate fixture is specified and is to be measured
3.2.7.1 Discussion—For i=1, NW =0.
i-1
in terms of sequential Eulerian angular rotations about the xyz
axes, respectively (z, axial rotation; x, lateral bending; and y, 3.2.8 intervertebral disc (IVD) prosthesis, n—nonbiologic
structure intended to restore the support and motion or a
flexion-extension).
portion thereof between adjacent vertebral bodies.
3.2.2.1 origin, n—center of the global coordinate system
which is located at the initial position of the total disc 3.2.9 kinematic profile, n—relativemotionbetweenadjacent
replacement’s instantaneous center of rotation (COR). vertebral bodies that the IVD prosthesis is subjected to while
3.2.2.1 Discussion—Some articulating devices do not have
being tested.
a single center of rotation, but instead have either a mobile
3.2.10 limit, n—a significant change in stiffness during a
center of rotation or multiple distinct centers of rotation,
given motion, indicating the implant has reached its designed
depending on the direction of movement. In this case, the
endpoint in range of motion.
origin should be explicitly defined by the user with a rationale
3.2.11 load profile, n—loading that the device experiences
for that definition.
while being tested under an applied kinematic profile or the
3.2.2.2 X-axis, n—positive X-axis is a global fixed axis
loading that the IVD prosthesis is subjected to if tested in load
relative to the test machine’s stationary base, and is to be
control.
directed anteriorly relative to the specimen’s initial unloaded
3.2.12 mechanical failure, n—failure associated with a de-
position.
fect in the material (for example, fatigue crack) or of the
bonding between materials that may or may not produce
Available fromAmerican National Standards Institute (ANSI), 25 W. 43rd St.,
4th Floor, New York, NY 10036, http://www.ansi.org. functional failure.
F2423 − 11 (2016)
TABLE 1 Test Profiles and Associated Parameters for Cervical
3.2.13 Wear modes (1) for articulating type designs:
IVD Prostheses
3.2.13.1 Mode 1 refers to the articulation between two
Preferred Alternate
primary bearing surfaces only.
Axial
Displacement Control: Load Control:
Test Profile Load, N
3.2.13.2 Mode2occurswheneveraprimarysurfacearticu-
Range of Motion Applied Moment
(2-4)
A
lates directly against a secondary, nonbearing surface.
(ROM), degree (3) Ranges, Nm (3)
3.2.13.3 Mode 3 occurs when the two primary bearing Flexion/extension 100 ±7.5 ±2.0
Lateral bend/ 100 ±6 ±2.0
surfaces are still articulating together, but third-body particles
rotation ±6 ±4.0
have become entrapped between them.
A
TheuseroftheguidemustdeterminewhethertheROM(rangeofmotion)willbe
3.2.13.4 Mode4referstoanycontactandmotionoccurring
equally divided between flexion and extension or weighted more toward one of the
between two secondary, nonbearing surfaces.
motion directions.
3.2.14 net wear NW of wear specimen (g), n—NW 5 W
~
i i 0
2W !1~S 2S ! ; loss in weight of the wear specimen corrected
i i 0
for fluid absorption at end of cycle interval i.
joints, biconcave joints having a free-floating or semi-
3.2.15 net volumetric wear NV of wear specimen (mm ),
i constrained third body, metallic endplates bonded to elastomer
n—NV 5NW/ρ atendofcycleinterval iwhere ρ=massdensity
i i
cores, and single-axis hinge joints.
(for example, units of g/mm ) of the wear material.
5.2 Spinal Testing Apparatus:
3.2.16 run out (cycles), n—maximum number of cycles that
5.2.1 Test Chambers—Incaseofamulti-specimenmachine,
a test needs to be carried to if functional failure has not yet
each chamber shall be isolated to prevent cross-contamination
occurred.
of the test specimens. The chamber shall be made entirely of
3.2.17 wear, n—progressive loss of material from the de-
noncorrosive components, such as acrylic plastic or stainless
vice(s)asaresultofrelativemotionatthesurfacesasmeasured
steel, and shall be easily removable from the machine for
by the change in mass of the IVD prosthesis or components of
thorough cleaning between tests.
the IVD prosthesis. 5.2.2 Component Clamping/Fixturing—Since the purpose
3.2.17.1 Discussion—In the case of a nonarticulating, com-
of the test is to characterize the wear and/or fatigue properties
pliant IVD prosthesis, wear is defined simply as the loss of of the IVD prosthesis under functional and kinematic
material from the prosthesis. Note that inferior and superior
conditions, the method for mounting components in the test
bone interface components are excluded from this definition; chamber shall not compromise the accuracy of assessment of
see 5.2.2.
the weight loss or stiffness variation during the test. For
example, prostheses having complicated superior and inferior
3.2.18 weight S of soak control specimen (g), n—S initial
i 0
surfaces for contacting bone (for example, sintered beads,
and S at end of cycle interval i.
i
hydroxylapatite (HA) coating, plasma spray) may be specially
3.2.19 weight W of wear specimen (g), n—W initial and
i 0
manufactured to modify that surface in a manner that does not
W at end of cycle interval i.
i
affect the wear simulation.
5.2.3 The device should be securely (rigidly) attached at its
4. Significance and Use
bone-implant interface to the mating test fixtures.
4.1 This guide can be used to determine the fatigue and
5.2.4 The motion of the superior test fixture relative to the
wear behavior of IVD prostheses subjected to functional and
inferior testing fixture shall be unconstrained in three-
kinematiccyclicloading/motionforrelativelylargenumbersof
dimensional space except for the components in the direction
cycles(forexample,variousdesignsofIVDprostheses,aswell
of specified test motions/loads.
as the effects of materials, manufacturing techniques and other
5.2.5 Load and Motion (components in Table 1 and Table
design variables on one particular design can be determined
2):
using this guide).
5.2.5.1 An axial load is to be a compressive load applied in
4.2 This guide is intended to be applicable to IVD thedirectionofthenegative Z-axis.Deviationsfromthisasthe
IVD moves from its initial position are to be reported as shear
prostheses, that support load and transmit motion by means of
an articulating joint or by use of compliant materials. components F , F , and moments
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: F2423 − 11 F2423 − 11 (Reapproved 2016)
Standard Guide for
Functional, Kinematic, and Wear Assessment of Total Disc
Prostheses
This standard is issued under the fixed designation F2423; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This guide provides guidance for wear and/or fatigue testing of total disc prostheses under functional and kinematic
conditions and, to this end, describes test methods for assessment of the wear or functional characteristics, or both, of total disc
prostheses.
1.2 Both lumbar and cervical prostheses are addressed.
1.3 Load and kinematic profiles for lumbar and cervical devices are not identical and, therefore, are addressed separately in the
guide.
1.4 Partial disc replacements, such as nucleus replacements or facet joint replacements, are not intended to be addressed.
1.5 Wear is assessed using a weight loss method in a testing medium as defined in this guide.
1.6 This guide does not address any potential failure mode as it relates to the fixation of the implant to its bony interfaces.
1.7 It is the intent of this guide to enable comparison of intervertebral disc (IVD) prostheses with regard to wear and fatigue
characteristics when tested under the specified conditions. It must be recognized, however, that there are many possible variations
in the in vivo conditions. A single laboratory simulation with a fixed set of parameters might not be universally representative.
1.8 Most IVD prostheses primarily fall into two classifications: articulating ball-in-socket type prostheses, and elastomeric or
compliant type prostheses. For the former, this guide primarily addresses Mode 1 wear (defined herein); whereas for the latter, this
guide addresses potential failure of the prosthesis when the implant is subjected to a range of motion and/or loads that fall within
the full range of possible physiologic motions and loads.
1.9 For articulating components, this guide predominantly describes a Mode 1 test. The user is cautioned that other modes of
wear may occur and may have significant influence on the functionality and performance of an articulating IVD prosthesis, and
therefore the user should consider the effects of other wear modes on the performance of the prosthesis.
1.10 In order that the data be reproducible and comparable within and between laboratories, it is essential that uniform
procedures are established. This guide is intended to facilitate uniform methods for testing and reporting of data for total disc
replacement prostheses.
1.11 Without a substantial clinical retrieval history of IVD prostheses, actual loading profiles and patterns cannot be delineated
at the time of the writing of this guide. It therefore follows that the load and motion conditions specified by this guide do not
necessarily accurately reproduce those occurring in vivo. Rather, this guide provides useful boundary/endpoint conditions for
evaluating prosthesis designs in a functional manner.
1.12 The values stated in SI units are to be regarded as the standard with the exception of angular measurements, which may
be reported in either degrees or radians.
1.13 This guide is not intended to be a performance standard. It is the responsibility of the user of this guide to characterize
the safety and effectiveness of the prosthesis under evaluation.
1.14 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
This guide is under the jurisdiction of ASTM Committee F04 on Medical and Surgical Materials and Devices and is the direct responsibility of Subcommittee F04.25
on Spinal Devices.
Current edition approved July 1, 2011Dec. 1, 2016. Published August 2011December 2016. Originally approved in 2005. Last previous edition approved in 20052011 as
F2423 – 05.F2423 – 11. DOI: 10.1520/F2423-11.10.1520/F2423-11R16.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
F2423 − 11 (2016)
2. Referenced Documents
2.1 ASTM Standards:
F561 Practice for Retrieval and Analysis of Medical Devices, and Associated Tissues and Fluids
F1582 Terminology Relating to Spinal Implants
F1714 Guide for Gravimetric Wear Assessment of Prosthetic Hip Designs in Simulator Devices
F1877 Practice for Characterization of Particles
F2077 Test Methods For Intervertebral Body Fusion Devices
2.2 ISO Standard:
ISO 18192–1 Implants for Surgery—Wear of Total Intervertebral Spinal Disc Prostheses—Part 1: Loading and Displacement
Parameters for Wear Testing and Corresponding Environmental Conditions for Test
3. Terminology
3.1 All functional and kinematic testing terminology is consistent with the referenced standards (for example, Test Methods
F2077, Terminology F1582, and so forth), unless otherwise stated.
3.2 Definitions:
3.2.1 axial load, n—the resultant force F applied to the superior or inferior fixture-end plate that simulates the in vivo load
axial
that an IVD prosthesis (original healthy disc) must resist.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’sstandard’s Document Summary page on the ASTM website.
Available from American National Standards Institute (ANSI), 25 W. 43rd St., 4th Floor, New York, NY 10036, http://www.ansi.org.
3.2.1.1 Discussion—
Based on a healthy disc, the primary component would be an axial compressive force F in the direction of the negative global
Z
Z axis, and it would pass through the origin of the IVD prosthesis. Shear components in the XY plane would be F and F . Lateral
X Y
bending moment M and flexion/extension moment M components would be created about the origin when the axial load does
X Y
not pass through it.
3.2.2 coordinate system/axes, n—global XYZ orthogonal axes are defined following a right-handed Cartesian coordinate system
in which the XY plane is to bisect the sagittal plane angle between the superior and inferior surfaces that are intended to simulate
the adjacent vertebral end plates. The global axes are stationary relative to the IVD prostheses’prosthesis’s inferior end plate
fixture, which, in this guide, is also considered to be stationary with respect to the test machine’s frame. Lower case letters, xyz,
denote a local, moving orthogonal coordinate system attached to the superior end plate fixture with directions initially coincident
with those of the global XYZ axes, respectively. The 3-D motion of the superior relative to the inferior end plate fixture is specified
and is to be measured in terms of sequential Eulerian angular rotations about the xyz axes, respectively (z, axial rotation; x, lateral
bending; and y, flexion-extension).
3.2.2.1 origin, n—center of the global coordinate system which is located at the initial position of the total disc replacement’s
instantaneous center of rotation (COR).
3.2.2.1 Discussion—
Some articulating devices do not have a single center of rotation, but instead have either a mobile center of rotation or multiple
distinct centers of rotation, depending on the direction of movement. In this case, the origin should be explicitly defined by the
user with a rationale for that definition.
3.2.2.2 X-axis, n—positive X-axis is a global fixed axis relative to the test machine’s stationary base, and is to be directed
anteriorly relative to the specimen’s initial unloaded position.
3.2.2.3 Y-axis, n—positive Y-axis is a global fixed axis relative to the test machine’s stationary base, and is directed laterally
relative to the specimen’s initial unloaded position.
3.2.2.4 Z-axis, n—positive Z-axis is a global fixed axis relative to the test machine’s stationary base, and is to be directed
superiorly relative to the specimen’s initial unloaded position.
3.2.2.5 x-axis, n—positive x-axis is a fixed axis relative to the IVD prosthesis and a moving axis relative to the global coordinate
system, and is directed anteriorly relative to the prosthesis.
3.2.2.6 y-axis, n—positive y-axis is a fixed axis relative to the IVD prosthesis and a moving axis relative to the global coordinate
system, and is directed laterally relative to the prosthesis.
F2423 − 11 (2016)
3.2.2.7 z-axis, n—positive z-axis is a fixed axis relative to the IVD prosthesis and a moving axis relative to the global coordinate
system, and is directed superiorly relative to the prosthesis.
3.2.3 degradation, n—loss of material or function or material properties as a result of causes other than that associated with
wear.
3.2.4 fluid absorption, n—fluid absorbed by the device material during testing.
3.2.5 functional failure, n—permanent deformation or wear that renders the IVD prosthesis assembly ineffective or unable to
resist load/motion or any secondary effects that result in a reduction of clinically relevant motions or the motions intended by the
design of the device.
3.2.6 interval net volumetric wear rate VR during cycle interval i (mm /million cycles), n—VR = WR /ρ, where ρ = mass
i i i
density (for example, units of g/mm ) of the wear material.
3.2.7 interval net wear rate WR during during cycle interval i (g/million cycles), n—WR = ((NW – NW )/(number of cycles
i i i i-1
in interval i))×10 .
3.2.7.1 Discussion—
For i = 1, NW = 0.
i-1
3.2.8 intervertebral disc (IVD) prosthesis, n—nonbiologic structure intended to restore the support and motion or a portion
thereof between adjacent vertebral bodies.
3.2.9 kinematic profile, n—relative motion between adjacent vertebral bodies that the IVD prosthesis is subjected to while being
tested.
3.2.10 limit, n—a significant change in stiffness during a given motion, indicating the implant has reached its designed endpoint
in range of motion.
3.2.11 load profile, n—loading that the device experiences while being tested under an applied kinematic profile or the loading
that the IVD prosthesis is subjected to if tested in load control.
3.2.12 mechanical failure, n—failure associated with a defect in the material (for example, fatigue crack) or of the bonding
between materials that may or may not produce functional failure.
3.2.13 Wear modes (1) for articulating type designs:
3.2.13.1 Mode 1 refers to the articulation between two primary bearing surfaces only.
3.2.13.2 Mode 2 occurs whenever a primary surface articulates directly against a secondary, nonbearing surface.
3.2.13.3 Mode 3 occurs when the two primary bearing surfaces are still articulating together, but third-body particles have
become entrapped between them.
3.2.13.4 Mode 4 refers to any contact and motion occurring between two secondary, nonbearing surfaces.
3.2.14 net wear NW of wear specimen (g), n—NW 5 W 2W 1 S 2S ; loss in weight of the wear specimen corrected for fluid
~ ! ~ !
i i 0 i i 0
absorption at end of cycle interval i.
3.2.15 net volumetric wear NV of wear specimen (mm ),n—NV 5NW /ρ at end of cycle interval i where ρ = mass density (for
i i i
example, units of g/mm ) of the wear material.
3.2.16 run out (cycles), n—maximum number of cycles that a test needs to be carried to if functional failure has not yet occurred.
3.2.17 wear, n—progressive loss of material from the device(s) as a result of relative motion at the surfaces as measured by the
change in mass of the IVD prosthesis or components of the IVD prosthesis.
The boldface numbers in parentheses refer to the list of references at the end of this standard.
3.2.17.1 Discussion—
Or in In the case of a nonarticulating, compliant IVD prosthesis, wear is defined simply as the loss of material from the prosthesis.
Note that inferior and superior bone interface components are excluded from this definition; see 5.2.2.
3.2.18 weight S of soak control specimen (g), n—S initial and S at end of cycle interval i.
i 0 i
3.2.19 weight W of wear specimen (g), n—W initial and W at end of cycle interval i.
i 0 i
4. Significance and Use
4.1 This guide can be used to determine the fatigue and wear behavior of IVD prostheses subjected to functional and kinematic
cyclic loading/motion for relatively large numbers of cycles (for example, various designs of IVD prostheses, as well as the effects
of materials, manufacturing techniques and other design variables on one particular design can be determined using this guide).
F2423 − 11 (2016)
TABLE 1 Test Profiles and Associated Parameters for Cervical
IVD Prostheses
Preferred Alternate
Axial
Displacement Control: Load Control:
Test Profile Load, N
Range of Motion Applied Moment
(2-4)
A
(ROM), degree (3) Ranges, Nm (3)
Flexion/extension 100 ±7.5 ±2.0
Lateral bend/ 100 ±6 ±2.0
rotation ±6 ±4.0
A
The user of the guide must determine whether the ROM (range of motion) will be
equally divided between flexion and extension or weighted more toward one of the
motion directions.
4.2 This guide is intended to be applicable to IVD prostheses, that support load and transmit motion by means of an articulating
joint or by use of compliant materials. Ceramics, metals, or polymers, or combination thereof, are used in IVD prosthesis,
prostheses, and it is the goal of this guide to enable a kinematic wear and/or fatigue comparison of these devices, regardless of
material and type of device.
5. Apparatus
5.1 Total Disc Prosthesis Components—The total disc replacement may comprise a variety of shapes and configurations. Some
known forms include ball-and-socket articulating joints, biconcave joints having a free-floating or semi-constrained third body,
metallic endplates bonded to elastomer cores, and single-axis hinge joints.
5.2 Spinal Testing Apparatus:
5.2.1 Test Chambers—In case of a multispecimenmulti-specimen machine, each chamber shall be isolated to prevent
cross-contamination of the test specimens. The chamber shall be made entirely of noncorrosive components, such as acrylic plastic
or stainless steel, and shall be easily removable from the machine for thorough cleaning between tests.
5.2.2 Component Clamping/Fixturing—Sinc
...










Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.