Standard Test Method for Respirator Fit Capability for Negative-Pressure Half-Facepiece Particulate Respirators

SIGNIFICANCE AND USE
5.1 In the U.S., when 42 Code of Federal Regulations Part 84 (42 CFR 84) was promulgated in 1995, the isoamyl acetate tightness test as described in 30 Code of Federal Regulations Part 11 for certain particulate-removing respirators was removed. These particulate-removing respirators were designed as protection against: (1) fumes of various metals having an air contamination level not less than 0.05 mg/m3, and (2) dusts, fumes, and mists having an air contamination level less than 0.05 mg/m3 or radionuclides. The isoamyl acetate test was removed because particulate respirators had to be modified before they could be tested and there were no other available fit tests suitable to the National Institute for Occupational Safety and Health (NIOSH) for approval testing at the time (1).4 There was a concern that the modified respirators may have had different fitting characteristics from the versions marketed. According to NIOSH, removing this requirement also allowed for further research on the effectiveness of certification fit testing methods (1).  
5.2 NIOSH conducted benchmark testing of 101 respirator models on the market during 2008 and 2009, using a similar test to that described herein (2). The results were analyzed to develop key test parameters and pass/fail criteria options for a respirator fit capability test for half-facepiece air-purifying particulate respirators (3). According to NIOSH, approximately 30 % of the models tested did not have good fitting characteristics (2). This was also supported by published research (4, 5). This standard establishes a performance requirement called respirator fit capability to assess respirator face-sealing characteristics.  
5.3 This standard can be used to evaluate all particulate-removing respirators on a population of wearers. A respirator model meeting the fit capability requirement will be capable of fitting the facial sizes and shapes for which it was designed. To achieve this goal, it is necessary for the method ...
SCOPE
1.1 This standard provides detailed instructions for performing a respirator fit capability test to determine the fit of air-purifying, half-facepiece respirators, which will include both filtering facepiece respirators and elastomeric respirators equipped with any type of particulate filter. The purpose is to increase the probability that available respirators fit a general worker population. The standard provides increased assurance to respirator purchasers and users that respirators that meet the requirement of this standard can be expected to effectively fit persons with various lengths and widths of faces, such as long and narrow or short and wide, when fit tested in the workplace as part of a complete respiratory protection program in accordance with 29 CFR 1910.134.  
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.3 It is the responsibility of the investigator to determine whether good laboratory practices (GLP standards—40 CFR, Part 160 of FIFRA) are required and to follow them when appropriate.  
1.4 This standard does not address specific product performance standards established by regulatory authorities; see 2.2 for details.  
1.5 This standard does not eliminate the need for every wearer to undergo a personal respirator fit test.  
1.6 This standard does not guarantee that every respirator wearer will be able to achieve the required fit factor on a particular manufacturer’s single-size or multi-size respirator model. Respirator wearers must always be given the opportunity to try other models or other manufacturers’ respirators.  
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior t...

General Information

Status
Published
Publication Date
31-Oct-2021
Drafting Committee
F23.65 - Respiratory

Relations

Effective Date
01-Nov-2023
Effective Date
01-Aug-2019

Overview

ASTM F3407-21: Standard Test Method for Respirator Fit Capability for Negative-Pressure Half-Facepiece Particulate Respirators is an internationally recognized protocol developed by ASTM International. This standard provides detailed guidance for evaluating the fit capability of air-purifying, half-facepiece particulate respirators, including both filtering facepiece and elastomeric models. The primary objective is to ensure that respirators fit a broad cross-section of the intended user population, thereby improving workplace safety and compliance with regulatory requirements.

Key Topics

  • Fit Capability Test: The standard outlines methods to determine a respirator’s ability to effectively fit users with varying face shapes and sizes, following protocols such as the use of the NIOSH Bivariate Panel.
  • Population Representation: The standard uses a diverse group of test subjects representing various facial dimensions to assess whether a respirator can fit its intended audience.
  • Performance Requirements: It defines pass/fail criteria based on quantitative fit factors, improving confidence that respirators meeting this standard will reliably fit most workers when used as part of a comprehensive respiratory protection program.
  • Test Protocols: The document details precise procedures for conducting fit tests, including equipment calibration, facial measurement, and standardized fit test exercises.
  • Limitations: While the standard improves the likelihood of proper fit across a workforce, it does not replace individual fit testing for each wearer, nor does it guarantee fit for all possible facial structures.

Applications

ASTM F3407-21 is highly relevant for:

  • Respirator Manufacturers: Provides a benchmark for designing and verifying new models, ensuring they are suitable for worker populations and align with market and regulatory expectations.
  • Workplace Safety Managers: Offers greater confidence in respirator selection, supporting compliance with OSHA’s 29 CFR 1910.134 respiratory protection standard and improving workplace health and safety outcomes.
  • Regulatory Bodies: Supports certification and approval processes for negative-pressure, half-facepiece particulate respirators, filling the gap left after the removal of the isoamyl acetate test in prior regulations.
  • Occupational Health Professionals: Enhances the respiratory protection program by ensuring a higher probability of achieving proper fit when selecting respirators that comply with this ASTM standard.

Related Standards

  • ASTM F3387: Practice for Respiratory Protection - foundational practices for respiratory protective equipment.
  • 29 CFR 1910.134: Occupational Safety and Health Administration (OSHA) Respiratory Protection Standard - establishes mandatory workplace requirements for respirator programs.
  • 42 CFR Part 84: NIOSH Respiratory Protective Devices - certification criteria for respirators used in the United States.
  • 30 CFR Part 11: Historical context for respirator testing and approval.

Practical Value

Adopting ASTM F3407-21 provides significant practical benefits:

  • Increases worker protection by helping ensure that available respirators will fit a wide variety of facial sizes and shapes.
  • Facilitates compliance with regulatory requirements through standardized, reproducible fit test results.
  • Reduces risk for employers by standardizing fit capability assessments before respirators are placed into use.
  • Supports product development by offering objective, validated criteria for respirator design evaluation.

By following ASTM F3407-21, organizations strengthen their respiratory protection programs, manufacturers enhance the credibility of their products, and safety professionals gain confidence in the reliable fit of half-facepiece particulate respirators. This contributes to healthier, safer work environments across industries where exposure to airborne particulates is a concern.

Buy Documents

Standard

ASTM F3407-21 - Standard Test Method for Respirator Fit Capability for Negative-Pressure Half-Facepiece Particulate Respirators

English language (12 pages)
sale 15% off
sale 15% off
Standard

REDLINE ASTM F3407-21 - Standard Test Method for Respirator Fit Capability for Negative-Pressure Half-Facepiece Particulate Respirators

English language (12 pages)
sale 15% off
sale 15% off

Get Certified

Connect with accredited certification bodies for this standard

NSF International

Global independent organization facilitating standards development and certification.

ANAB United States Verified

CIS Institut d.o.o.

Personal Protective Equipment (PPE) certification body. Notified Body NB-2890 for EU Regulation 2016/425 PPE.

SA Slovenia Verified

Kiwa BDA Testing

Building and construction product certification.

RVA Netherlands Verified

Sponsored listings

Frequently Asked Questions

ASTM F3407-21 is a standard published by ASTM International. Its full title is "Standard Test Method for Respirator Fit Capability for Negative-Pressure Half-Facepiece Particulate Respirators". This standard covers: SIGNIFICANCE AND USE 5.1 In the U.S., when 42 Code of Federal Regulations Part 84 (42 CFR 84) was promulgated in 1995, the isoamyl acetate tightness test as described in 30 Code of Federal Regulations Part 11 for certain particulate-removing respirators was removed. These particulate-removing respirators were designed as protection against: (1) fumes of various metals having an air contamination level not less than 0.05 mg/m3, and (2) dusts, fumes, and mists having an air contamination level less than 0.05 mg/m3 or radionuclides. The isoamyl acetate test was removed because particulate respirators had to be modified before they could be tested and there were no other available fit tests suitable to the National Institute for Occupational Safety and Health (NIOSH) for approval testing at the time (1).4 There was a concern that the modified respirators may have had different fitting characteristics from the versions marketed. According to NIOSH, removing this requirement also allowed for further research on the effectiveness of certification fit testing methods (1). 5.2 NIOSH conducted benchmark testing of 101 respirator models on the market during 2008 and 2009, using a similar test to that described herein (2). The results were analyzed to develop key test parameters and pass/fail criteria options for a respirator fit capability test for half-facepiece air-purifying particulate respirators (3). According to NIOSH, approximately 30 % of the models tested did not have good fitting characteristics (2). This was also supported by published research (4, 5). This standard establishes a performance requirement called respirator fit capability to assess respirator face-sealing characteristics. 5.3 This standard can be used to evaluate all particulate-removing respirators on a population of wearers. A respirator model meeting the fit capability requirement will be capable of fitting the facial sizes and shapes for which it was designed. To achieve this goal, it is necessary for the method ... SCOPE 1.1 This standard provides detailed instructions for performing a respirator fit capability test to determine the fit of air-purifying, half-facepiece respirators, which will include both filtering facepiece respirators and elastomeric respirators equipped with any type of particulate filter. The purpose is to increase the probability that available respirators fit a general worker population. The standard provides increased assurance to respirator purchasers and users that respirators that meet the requirement of this standard can be expected to effectively fit persons with various lengths and widths of faces, such as long and narrow or short and wide, when fit tested in the workplace as part of a complete respiratory protection program in accordance with 29 CFR 1910.134. 1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.3 It is the responsibility of the investigator to determine whether good laboratory practices (GLP standards—40 CFR, Part 160 of FIFRA) are required and to follow them when appropriate. 1.4 This standard does not address specific product performance standards established by regulatory authorities; see 2.2 for details. 1.5 This standard does not eliminate the need for every wearer to undergo a personal respirator fit test. 1.6 This standard does not guarantee that every respirator wearer will be able to achieve the required fit factor on a particular manufacturer’s single-size or multi-size respirator model. Respirator wearers must always be given the opportunity to try other models or other manufacturers’ respirators. 1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior t...

SIGNIFICANCE AND USE 5.1 In the U.S., when 42 Code of Federal Regulations Part 84 (42 CFR 84) was promulgated in 1995, the isoamyl acetate tightness test as described in 30 Code of Federal Regulations Part 11 for certain particulate-removing respirators was removed. These particulate-removing respirators were designed as protection against: (1) fumes of various metals having an air contamination level not less than 0.05 mg/m3, and (2) dusts, fumes, and mists having an air contamination level less than 0.05 mg/m3 or radionuclides. The isoamyl acetate test was removed because particulate respirators had to be modified before they could be tested and there were no other available fit tests suitable to the National Institute for Occupational Safety and Health (NIOSH) for approval testing at the time (1).4 There was a concern that the modified respirators may have had different fitting characteristics from the versions marketed. According to NIOSH, removing this requirement also allowed for further research on the effectiveness of certification fit testing methods (1). 5.2 NIOSH conducted benchmark testing of 101 respirator models on the market during 2008 and 2009, using a similar test to that described herein (2). The results were analyzed to develop key test parameters and pass/fail criteria options for a respirator fit capability test for half-facepiece air-purifying particulate respirators (3). According to NIOSH, approximately 30 % of the models tested did not have good fitting characteristics (2). This was also supported by published research (4, 5). This standard establishes a performance requirement called respirator fit capability to assess respirator face-sealing characteristics. 5.3 This standard can be used to evaluate all particulate-removing respirators on a population of wearers. A respirator model meeting the fit capability requirement will be capable of fitting the facial sizes and shapes for which it was designed. To achieve this goal, it is necessary for the method ... SCOPE 1.1 This standard provides detailed instructions for performing a respirator fit capability test to determine the fit of air-purifying, half-facepiece respirators, which will include both filtering facepiece respirators and elastomeric respirators equipped with any type of particulate filter. The purpose is to increase the probability that available respirators fit a general worker population. The standard provides increased assurance to respirator purchasers and users that respirators that meet the requirement of this standard can be expected to effectively fit persons with various lengths and widths of faces, such as long and narrow or short and wide, when fit tested in the workplace as part of a complete respiratory protection program in accordance with 29 CFR 1910.134. 1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.3 It is the responsibility of the investigator to determine whether good laboratory practices (GLP standards—40 CFR, Part 160 of FIFRA) are required and to follow them when appropriate. 1.4 This standard does not address specific product performance standards established by regulatory authorities; see 2.2 for details. 1.5 This standard does not eliminate the need for every wearer to undergo a personal respirator fit test. 1.6 This standard does not guarantee that every respirator wearer will be able to achieve the required fit factor on a particular manufacturer’s single-size or multi-size respirator model. Respirator wearers must always be given the opportunity to try other models or other manufacturers’ respirators. 1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior t...

ASTM F3407-21 is classified under the following ICS (International Classification for Standards) categories: 13.340.30 - Respiratory protective devices. The ICS classification helps identify the subject area and facilitates finding related standards.

ASTM F3407-21 has the following relationships with other standards: It is inter standard links to ASTM F3387-23, ASTM F3387-19. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

ASTM F3407-21 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: F3407 − 21
Standard Test Method for
Respirator Fit Capability for Negative-Pressure Half-
Facepiece Particulate Respirators
This standard is issued under the fixed designation F3407; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope priate safety, health, and environmental practices and deter-
mine the applicability of regulatory limitations prior to use.
1.1 Thisstandardprovidesdetailedinstructionsforperform-
1.8 This international standard was developed in accor-
ing a respirator fit capability test to determine the fit of
dance with internationally recognized principles on standard-
air-purifying, half-facepiece respirators, which will include
ization established in the Decision on Principles for the
both filtering facepiece respirators and elastomeric respirators
Development of International Standards, Guides and Recom-
equipped with any type of particulate filter. The purpose is to
mendations issued by the World Trade Organization Technical
increase the probability that available respirators fit a general
Barriers to Trade (TBT) Committee.
worker population. The standard provides increased assurance
to respirator purchasers and users that respirators that meet the
2. Referenced Documents
requirement of this standard can be expected to effectively fit
2.1 ASTM Standards:
persons with various lengths and widths of faces, such as long
F3387 Practice for Respiratory Protection
and narrow or short and wide, when fit tested in the workplace
as part of a complete respiratory protection program in
2.2 Federal Standards:
accordance with 29 CFR 1910.134.
29 CFR Part 1910.134 Respiratory Protection
30 CFR Part 11 Respiratory Protective Apparatus, Tests for
1.2 The values stated in SI units are to be regarded as
Permissibility, Fees
standard. No other units of measurement are included in this
42 CFR Part 84 Respiratory Protective Devices
standard.
1.3 It is the responsibility of the investigator to determine
3. Terminology
whether good laboratory practices (GLP standards—40 CFR,
3.1 Definitions:
Part 160 of FIFRA) are required and to follow them when
3.1.1 fit test, n—the use of a protocol to qualitatively or
appropriate.
quantitatively evaluate the fit of a particular respirator on an
1.4 This standard does not address specific product perfor-
individual.
mance standards established by regulatory authorities; see 2.2
3.1.2 high-effıciency particulate air (HEPA) filter, n—a filter
for details.
with a minimum particle removal efficiency of no less than
1.5 This standard does not eliminate the need for every
99.97 % for monodisperse particles having an aerodynamic
wearer to undergo a personal respirator fit test.
diameter of 0.3 µm.
1.6 This standard does not guarantee that every respirator
3.1.3 individual exercise RFC result, n—a numeric assess-
wearer will be able to achieve the required fit factor on a
ment of how well a tight-fitting respirator facepiece fits a test
particular manufacturer’s single-size or multi-size respirator
subject during each exercise performed during a subject
model. Respirator wearers must always be given the opportu-
respirator fit capability (RFC) test. It is the ratio of the
nity to try other models or other manufacturers’ respirators.
concentration outside the facepiece (C ) to the concentration
out
inside the facepiece (C ) not adjusted for respiratory tract
1.7 This standard does not purport to address all of the
in
deposition. (C /C ).
safety concerns, if any, associated with its use. It is the
out in
responsibility of the user of this standard to establish appro-
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
ThistestmethodisunderthejurisdictionofASTMCommitteeF23onPersonal contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Protective Clothing and Equipment and is the direct responsibility of Subcommittee Standards volume information, refer to the standard’s Document Summary page on
F23.65 on Respiratory. the ASTM website.
Current edition approved Nov. 1, 2021. Published November 2021. Originally Available from U.S. Government Printing Office, Superintendent of
approved in 2020. Last previous edition approved in 2020 as F3407 – 20. DOI: Documents, 732 N. Capitol St., NW, Washington, DC 20401-0001, http://
10.1520/F3407-21. www.access.gpo.gov.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
F3407 − 21
3.1.4 respirator fit capability (RFC) test, n—an assessment fumes, and mists having an air contamination level less than
of a respirator model’s ability to achieve passing face seal 0.05 mg⁄m or radionuclides. The isoamyl acetate test was
performance on either the complete NIOSH Bivariate Panel or removed because particulate respirators had to be modified
a specified subset of the panel representing the population of beforetheycouldbetestedandtherewerenootheravailablefit
respiratorwearerswhenthewearersareproperlytrainedandfit tests suitable to the National Institute for Occupational Safety
tested in compliance with the manufacturer’s user instructions andHealth(NIOSH)forapprovaltestingatthetime (1). There
and Practice F3387 and the Occupational Safety and Health was a concern that the modified respirators may have had
Administration (OSHA) standard 29 CFR 1910.134. different fitting characteristics from the versions marketed.
According to NIOSH, removing this requirement also allowed
3.1.5 respirator model, n—a group or series of identical
for further research on the effectiveness of certification fit
half-facepiece respirators (that is, utilizing the same compo-
testing methods (1).
nents such as facepiece blank, head straps, exhalation valves,
and composed of the same construction materials) differing
5.2 NIOSH conducted benchmark testing of 101 respirator
only in the size of the facepiece in order to fit the NIOSH
models on the market during 2008 and 2009, using a similar
Bivariate Model. A respirator model may consist of only one
test to that described herein (2). The results were analyzed to
unique half-facepiece respirator.
develop key test parameters and pass/fail criteria options for a
respirator fit capability test for half-facepiece air-purifying
3.1.6 subject RFC result, n—the harmonic mean of the
particulaterespirators (3).AccordingtoNIOSH,approximately
seven individual exercise RFC results for a particular subject
30 % of the models tested did not have good fitting character-
and respirator model.
istics (2).This was also supported by published research (4, 5).
3.1.7 subject RFC test, n—an RFC test performed by one
This standard establishes a performance requirement called
subject wearing a particular respirator model.
respirator fit capability to assess respirator face-sealing char-
3.1.8 test panel, n—an organized group of people with
acteristics.
varying facial dimensions representing the respirator wearer
5.3 This standard can be used to evaluate all particulate-
population. The RFC test uses the NIOSH Bivariate Panel,
removing respirators on a population of wearers. A respirator
which is based on face length and face width (Fig. A1.1).
model meeting the fit capability requirement will be capable of
3.1.9 user, n—person or organization who makes use of the
fitting the facial sizes and shapes for which it was designed.To
respirator; for example, one involved in selecting, maintaining,
achieve this goal, it is necessary for the method to reject
or wearing the respirator.
poor-fitting respirators, while still passing well-fitting respira-
3.1.10 wearer, n—the person who actually wears the respi-
tors meeting the pass/fail criteria established in this standard. It
rator in the workplace.
is thought that this standard will increase the likelihood that
respirators meeting this requirement will fit a wide variety of
3.1.11 wearer seal check (namely, user seal check), n—a
their prospective wearers when properly fit tested, donned, and
procedure conducted by the wearer to determine if a tight-
used.
fitting respirator is properly donned. This is consistent with
Practice F3387 and the Occupational Safety and Health Ad-
6. Interferences
ministration 29 CFR 1910.134 user seal check.
6.1 ParticlesintheTestSubject’sExhaledBreath—Eachtest
4. Summary of Test Method
wearershallnotbepermittedtoeatorsmokeforatleast30min
before the start of the test.
4.1 This standard defines performance requirements for
ensuring that a respirator model, available in either a unique
6.2 Facial Hair—Each test subject shall be cleanly shaven
single size or multiple sizes, is capable of achieving the
before being able to participate in the test. Mustaches are
pass/fail criteria fit on a specified percentage of the NIOSH
permitted if they do not interfere with the facepiece seal as
Bivariate Test Panel (NIOSH Panel) representing a range of
assessed by the test administrator.
face sizes. These performance requirements will increase the
6.3 Other Facial Characteristics—Any condition that could
likelihood that most respirator wearers will be able to achieve
potentially interfere with the face-to-facepiece seal or valve
the required pass/fail criteria when fit tested on either a
functionsuchasjewelry,scars,etc.,willbepermittedifthetest
manufacturer’s unique single-size respirator model or on at
administrator determines that it will not interfere.
least one unique size of a manufacturer’s multi-size respirator
model. The exercises and pass/fail criteria are based on those
7. Apparatus
found in Title 29, Code of Federal Regulations, Part 1910.134.
7.1 Condensation nuclei counter with particle classifier
5. Significance and Use
technology (for example, a differential mobility analyzer). The
particle classifier technology shall only allow nominal 55 nm
5.1 In the U.S., when 42 Code of Federal Regulations Part
negatively charged particles to pass through to the condensa-
84 (42 CFR 84) was promulgated in 1995, the isoamyl acetate
tion nuclei counter for counting while eliminating the zero-
tightness test as described in 30 Code of Federal Regulations
charge and positive-charge particles from the sample.
Part 11 for certain particulate-removing respirators was re-
moved. These particulate-removing respirators were designed
as protection against: (1) fumes of various metals having an air
The boldface numbers in parentheses refer to a list of references at the end of
contamination level not less than 0.05 mg⁄m , and (2) dusts, this standard.
F3407 − 21
7.2 Software to control the condensation nuclei counter. 7.7.2 Elastomeric facepiece respirators: three complete res-
pirator assemblies with ten sets of filters for a unique one-size
7.3 High-efficiency particulate (HEPA) filter for diagnostic
model,andthreecompleterespiratorassembliesineachunique
checks recommended by the instrument manufacturer.
size and ten sets of filters when a respirator is designed and
7.4 Test Chamber:
manufactured in two or more unique sizes.
7.4.1 Size—Large enough to permit each of the test subjects
7.8 Facial Size Measurement Calipers—To measure the test
conducting an RFC test to freely perform all required exercises
subject.
without disturbing the positioning of the facepiece or the
7.8.1 Calibrated sliding measurement calipers, 0 mm to
measurement apparatus, or interfering with the movements of
200 mm length and 0 mm to 50 mm depth.
any other test subjects in the chamber.
7.8.2 Calibrated spreading measurement calipers capable of
7.4.2 Isolation—The test chamber shall be equipped and
measuring 0 mm to 300 mm width.
constructed so that the chamber air containing the sodium
chloride test agent is effectively isolated from the ambient air 7.9 Other accessories and supplies required by the conden-
outside the chamber.The access door to the chamber should be sationnucleicountermanufacturerinordertoperformtheRFC
test.
tighttoavoidleakage.Thetestsubject(s)mustbeabletosafely
enter and exit from the chamber.
7.4.3 Subject Visibility—The test chamber shall have a 8. Reagents and Materials
window or other means for the test subject(s) to be visible to
8.1 Sodium chloride solution, 2 % NaCl solution in distilled
the test administrator at all times.
water.
7.4.4 Aerosol Concentration—The aerosol concentration
8.2 Isopropyl alcohol, or other working fluid as specified by
shall be well mixed (that is, uniformly distributed) throughout
the instrument manufacturer. Reagent grade (>99.5 %).
the chamber (610 %) where the test subject(s) will be per-
forming the test. For N99 testing, the concentration shall be
9. Hazards
stable (that is, 610 % of the initial concentration of between
2000 and 8000 particles/cm ) for the duration of the test. For 9.1 Working Fluid—Avoid eye and skin contact.
N95 mode testing, the concentration shall be stable (that is,
610 % of the initial concentration of between 200 and 800 10. Sampling, Test Specimens, and Test Units
particles/cm ) for the duration of the test.
10.1 Test Subjects—All human subject testing conducted in
7.4.5 Particle Size—The particles in the chamber should be
accordance with this RFC standard will be guided by a
between0.02 µmand1 µmwithageometricstandarddeviation
statement of principles governing the institution in the dis-
≤2.2.
charge of its responsibilities for protecting the rights and
7.4.6 Temperature and Humidity—The airflow through the
welfare of human subjects conducting this testing, including
test chamber shall be sufficient to maintain the temperature
obtaining informed consent and screening the subjects to
between 21 °C and 24 °C, the relative humidity below 40 % to
ensure their safety while performing the RFC test. If
prevent agglomeration of the sodium chloride test agent, and
applicable, it must be approved by an Institutional Review
the oxygen level above 19.5 %.
Board or other appropriate body. Twenty-five test subjects
7.5 Particle Generator—An aerosol generator capable of meeting the participation and qualification criteria for the
producingthesodiumchlorideconcentrationspecifiedin7.4.4. testing of respirators having facial dimensions falling within
the requirements of the NIOSH Bivariate Panel for the size of
7.6 In-Facepiece Sampling Apparatus—For filtering face-
the respirator to be recruited. See AnnexA1. If any test subject
piece respirators, a flush-mounted probe equipped with a push
is measured and the facial dimensions are not within the
nut specifically designed to be attached to this type of
boundaries of the NIOSH Bivariate Panel, the test subject shall
respirator. For elastomeric respirators, a fit test adapter placed
not be tested.
between the facepiece and the filter can be used. The flush-
mounted filtering facepiece probe (N95 probe) shall not be
11. Preparation of Apparatus
used for elastomeric respirators.
7.6.1 The probe should be placed on the midline between 11.1 Several diagnostic checks shall be performed at least
the nose and the mouth, whenever possible. If a different daily. These shall include:
position is necessary, every effort should be made to avoid 11.1.1 Chamber Particle Concentration Check—To ensure
contact with the face, placement on a seam, or interference the chamber concentration is within the range specified in
with other features of the respirator. The sampling apparatus 7.4.4.
should be supported in a way that it does not affect or interfere 11.1.2 Particle classifier check.
with the fit of the respirator (that is, the respirator with the 11.1.3 Zero Check—To provide assurance that there are no
sampling apparatus connected must fit the test subject in the leaks in the system.
same manner as it would without the sampling apparatus). 11.1.4 Maximum Fit Factor Check—To ensure the conden-
Annex A2 contains more information on probe location.
sation nuclei counter is capable of measuring high fit factors.
7.7 Respirators—The number of respirators required for 11.2 Follow the condensation nuclei counter and particle
testing is:
classifier manufacturer’s instructions for performing these
7.7.1 Filtering facepiece respirators: 35. checks.
F3407 − 21
12. Calibration and Standardization 14.2.4.1 To reduce the inter- and intra-wearer variability,
each test wearer shall don the respirator under the supervision
12.1 At a minimum, all measuring equipment utilized for
of the test administrator. The test wearer and test administrator
this testing must have been calibrated by the manufacturer
will be permitted to make adjustments to the facepiece until
within the timeframe specified in the equipment’s operation
he/she and the test administrator are satisfied that the respirator
manual. The calibration shall use methods traceable to Na-
is being worn in compliance with the manufacturer’s UI,
tional Institute of Standards and Technology (NIST) standards.
including passing the appropriate seal checks.
12.2 Equipment calibration records shall be available for
14.2.4.2 Have the test subject wearing the respirator do the
examination at each testing facility.
appropriate seal check. Every test subject must pass a seal
12.3 Prior to beginning any testing, a statement that all test
check before performing the subject RFC test. If the seal check
equipment is within calibration shall be attested by the lab
fails (that is, leakage is detected), the test administrator will
technician, laboratory manager, or other designated person on
allow the test subject to reposition the respirator and redo the
each test report.
seal check after explaining the seal check procedure again. For
a unique single-size respirator model, if the second seal check
13. Conditioning
is a failure, the test subject will not be allowed to perform the
13.1 The respirators received for testing shall be free of
RFC test. Another test subject in the same test panel cell will
visual damage such as distorted face seals. There are no other
be recruited and the procedure will be started over. For
specific conditioning requirements.
multiple size respirator models, the test subject will be allowed
to don a different size of the respirator and perform the seal
14. Procedure
check. The test subject will continue to don a different size of
14.1 Facial Measurement:
the respirator until either the seal check is passed or all sizes
14.1.1 Face width is the maximum horizontal breadth of the
are tried. If the test subject cannot pass the seal check with any
face as measured with a spreading caliper between the zygo-
of the unique sizes, he/she will not be allowed to perform the
matic arches (AnnexA1). The test subject sits looking straight
subject RFC test. Another test subject in the same test panel
ahead with teeth together (lightly occluded). Only enough
cell will be recruited and the procedure will be started over.
pressure is exerted to ensure that the caliper tips are on the
zygomatic arches. The test administrator will take two mea- 14.2.4.3 The test subject shall wear the respirator for 1 min
surements of face width, removing the caliper between mea- priortothesubjectRFCtesttoensurethehe/shecanwearitfor
surements. If the two measurements for face width are within
the duration of the test and purge the particles trapped inside
2 mm, then stop. Otherwise, take a third measurement. the facepiece during donning.
14.1.2 Face length is the distance in the mid-sagittal plane
14.2.4.4 AfterverifyingthefitoftherespiratorusingtheUI,
between the menton landmark at the bottom of the chin and the
no further adjustments to the facepiece are permitted before
sellion landmark at the deepest point of the nasal root depres-
commencing and during the subject RFC test.
sion and is measured with a sliding caliper (Annex A1). The
14.2.5 Subject RFC Testing Procedure:
test subject sits looking straight ahead with teeth together
14.2.5.1 The test subject shall enter the chamber. The
(lightly occluded). The fixed blade of the caliper is placed on
concentration of particles in the chamber shall be measured
the sellion. Only enough pressure is exerted to obtain contact
prior to the subject entering the chamber to ensure that it is
between the caliper and the skin is exerted. The test adminis-
between 2000 and 8000 particles/cm for N99 mode testing
trator will take two measurements of face length, removing the
and between 200 and 800 particles/cm for N95 mode testing.
caliper between measurements. If the two measurements for
face width are within 3 mm, then stop. Otherwise, take a third The concentration of particles in the chamber shall remain
measurement.
within 610 % of the concentration measured at the beginning
14.1.3 The average of the face width and face length
of the test for the duration of the test. This can be verified by
measurements will be used to determine the placement of the
havi
...


This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: F3407 − 20 F3407 − 21
Standard Test Method for
Respirator Fit Capability for Negative-Pressure Half-
Facepiece Particulate Respirators
This standard is issued under the fixed designation F3407; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This standard provides detailed instructions for performing a respirator fit capability test to determine the fit of air-purifying,
half-facepiece respirators, which will include both filtering facepiece respirators and elastomeric respirators equipped with any type
of particulate filter. The purpose is to increase the probability that available respirators fit a general worker population. The standard
provides increased assurance to respirator purchasers and users that respirators that meet the requirement of this standard can be
expected to effectively fit persons with various lengths and widths of faces, such as long and narrow or short and wide, when fit
tested in the workplace as part of a complete respiratory protection program in accordance with 29 CFR 1910.134.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 It is the responsibility of the investigator to determine whether good laboratory practices (GLP standards—40 CFR, Part 160
of FIFRA) are required and to follow them when appropriate.
1.4 This standard does not address specific product performance standards established by regulatory authorities; see 2.2 for details.
1.5 This standard does not eliminate the need for every wearer to undergo a personal respirator fit test.
1.6 This standard does not guarantee that every respirator wearer will be able to achieve the required fit factor on a particular
manufacturer’s single-size or multi-size respirator model. Respirator wearers must always be given the opportunity to try other
models or other manufacturers’ respirators.
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of
regulatory limitations prior to use.
1.8 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
This test method is under the jurisdiction of ASTM Committee F23 on Personal Protective Clothing and Equipment and is the direct responsibility of Subcommittee
F23.65 on Respiratory.
Current edition approved Oct. 1, 2020Nov. 1, 2021. Published October 2020November 2021. Originally approved in 2020. Last previous edition approved in 2020 as
F3407 – 20. DOI: 10.1520/F3407-20.10.1520/F3407-21.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
F3407 − 21
2. Referenced Documents
2.1 ASTM Standards:
F3387 Practice for Respiratory Protection
2.2 Federal Standards:
29 CFR Part 1910.134 Respiratory Protection
30 CFR Part 11 Respiratory Protective Apparatus, Tests for Permissibility, Fees
42 CFR Part 84 Respiratory Protective Devices
3. Terminology
3.1 Definitions:
3.1.1 fit test, n—the use of a protocol to qualitatively or quantitatively evaluate the fit of a particular respirator on an individual.
3.1.2 high-effıciency particulate air (HEPA) filter, n—a filter with a minimum particle removal efficiency of no less than 99.97 %
for monodisperse particles having an aerodynamic diameter of 0.3 μm.
3.1.3 individual exercise RFC result, n—a numeric assessment of how well a tight-fitting respirator facepiece fits a test subject
during each exercise performed during a subject respirator fit capability (RFC) test. It is the ratio of the concentration outside the
facepiece (C ) to the concentration inside the facepiece (C ) not adjusted for respiratory tract deposition. (C /C ).
out in out in
3.1.4 respirator fit capability (FRC)(RFC) test, n—an assessment of a respirator model’s ability to achieve passing face seal
performance on either the complete NIOSH Bivariate Panel or a specified subset of the panel representing the population of
respirator wearers when the wearers are properly trained and fit tested in compliance with the manufacturer’s user instructions and
Practice F3387 and the Occupational Safety and Health Administration (OSHA) standard 29 CFR 1910.134.
3.1.5 respirator model, n—a group or series of identical half-facepiece respirators (that is, utilizing the same components such as
facepiece blank, head straps, exhalation valves, and composed of the same construction materials) differing only in the size of the
facepiece in order to fit the NIOSH Bivariate Model. A respirator model may consist of only one unique half-facepiece respirator.
3.1.6 subject RFC result, n—the harmonic mean of the seven individual exercise RFC results for a particular subject and respirator
model.
3.1.7 subject RFC test, n—an RFC test performed by one subject wearing a particular respirator model.
3.1.8 test panel, n—an organized group of people with varying facial dimensions representing the respirator wearer population.
The RFC test uses the NIOSH Bivariate Panel, which is based on face length and face width (Fig. A1.1).
3.1.9 user, n—person or organization who makes use of the respirator; for example, one involved in selecting, maintaining, or
wearing the respirator.
3.1.10 wearer, n—the person who actually wears the respirator in the workplace.
3.1.11 wearer seal check (namely, user seal check), n—a procedure conducted by the wearer to determine if a tight-fitting
respirator is properly donned. This is consistent with Practice F3387 and the Occupational Safety and Health Administration 29
CFR 1910.134 user seal check.
4. Summary of Test Method
4.1 This standard defines performance requirements for ensuring that a respirator model, available in either a unique single size
or multiple sizes, is capable of achieving the pass/fail criteria fit on a specified percentage of the NIOSH Bivariate Test Panel
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Available from U.S. Government Printing Office, Superintendent of Documents, 732 N. Capitol St., NW, Washington, DC 20401-0001, http://www.access.gpo.gov.
F3407 − 21
(NIOSH Panel) representing a range of face sizes. These performance requirements will increase the likelihood that most respirator
wearers will be able to achieve the required pass/fail criteria when fit tested on either a manufacturer’s unique single-size respirator
model or on at least one unique size of a manufacturer’s multi-size respirator model. The exercises and pass/fail criteria are based
on those found in Title 29, Code of Federal Regulations, Part 1910.134.
5. Significance and Use
5.1 In the U.S., when 42 Code of Federal Regulations Part 84 (42 CFR 84) was promulgated in 1995, the isoamyl acetate tightness
test as described in 30 Code of Federal Regulations Part 11 for certain particulate-removing respirators was removed. These
particulate-removing respirators were designed as protection against: (1) fumes of various metals having an air contamination level
3 3
not less than 0.05 mg ⁄m , and (2) dusts, fumes, and mists having an air contamination level less than 0.05 mg ⁄m or radionuclides.
The isoamyl acetate test was removed because particulate respirators had to be modified before they could be tested and there were
no other available fit tests suitable to the National Institute for Occupational Safety and Health (NIOSH) for approval testing at
the time (1). There was a concern that the modified respirators may have had different fitting characteristics from the versions
marketed. According to NIOSH, removing this requirement also allowed for further research on the effectiveness of certification
fit testing methods (1).
5.2 NIOSH conducted benchmark testing of 101 respirator models on the market during 2008 and 2009, using a similar test to
that described herein (2). The results were analyzed to develop key test parameters and pass/fail criteria options for a respirator
fit capability test for half-facepiece air-purifying particulate respirators (3). According to NIOSH, approximately 30 % of the
models tested did not have good fitting characteristics (2). This was also supported by published research (4, 5). This standard
establishes a performance requirement called respirator fit capability to assess respirator face-sealing characteristics.
5.3 This standard can be used to evaluate all particulate-removing respirators on a population of wearers. A respirator model
meeting the fit capability requirement will be capable of fitting the facial sizes and shapes for which it was designed. To achieve
this goal, it is necessary for the method to reject poor-fitting respirators, while still passing well-fitting respirators meeting the
pass/fail criteria established in this standard. It is thought that this standard will increase the likelihood that respirators meeting this
requirement will fit a wide variety of their prospective wearers when properly fit tested, donned, and used.
6. Interferences
6.1 Particles in the Test Subject’s Exhaled Breath—Each test wearer shall not be permitted to eat or smoke for at least 30 min
before the start of the test.
6.2 Facial Hair—Each test subject shall be cleanly shaven before being able to participate in the test. Mustaches are permitted
if they do not interfere with the facepiece seal as assessed by the test administrator.
6.3 Other Facial Characteristics—Any condition that could potentially interfere with the face-to-facepiece seal or valve function
such as jewelry, scars, etc., will be permitted if the test administrator determines that it will not interfere.
7. Apparatus
7.1 Condensation nuclei counter with particle classifier technology (for example, a differential mobility analyzer). The particle
classifier technology shall only allow nominal 55 nm negatively charged particles to pass through to the condensation nuclei
counter for counting while eliminating the zero-charge and positive-charge particles from the sample.
7.2 Software to control the condensation nuclei counter.
7.3 High-efficiency particulate (HEPA) filter for diagnostic checks recommended by the instrument manufacturer.
7.4 Test Chamber:
The boldface numbers in parentheses refer to a list of references at the end of this standard.
F3407 − 21
7.4.1 Size—Large enough to permit each of the test subjects conducting an RFC test to freely perform all required exercises
without disturbing the positioning of the facepiece or the measurement apparatus, or interfering with the movements of any other
test subjects in the chamber.
7.4.2 Isolation—The test chamber shall be equipped and constructed so that the chamber air containing the sodium chloride test
agent is effectively isolated from the ambient air outside the chamber. The access door to the chamber should be tight to avoid
leakage. The test subject(s) must be able to safely enter and exit from the chamber.
7.4.3 Subject Visibility—The test chamber shall have a window or other means for the test subject(s) to be visible to the test
administrator at all times.
7.4.4 Aerosol Concentration—The aerosol concentration shall be well mixed (that is, uniformly distributed) throughout the
chamber (610 %) where the test subject(s) will be performing the test. The For N99 testing, the concentration shall be stable (that
is, 610 % of the initial concentration of between 2000 and 8000 particles/cm ) for the duration of the test. For N95 mode testing,
the concentration shall be stable (that is, 610 % of the initial concentration of between 200 and 800 particles/cm ) for the duration
of the test.
7.4.5 Particle Size—The particles in the chamber should be between 0.02 μm and 1 μm with a geometric standard deviation ≤2.2.
7.4.6 Temperature and Humidity—The airflow through the test chamber shall be sufficient to maintain the temperature between
21 °C and 24 °C, the relative humidity below 40 % to prevent agglomeration of the sodium chloride test agent, and the oxygen
level above 19.5 %.
7.5 Particle Generator—An aerosol generator capable of producing the sodium chloride concentration specified in 7.4.4.
7.6 In-Facepiece Sampling Apparatus—For filtering facepiece respirators, a flush-mounted probe equipped with a push nut
specifically designed to be attached to this type of respirator. For elastomeric respirators, a fit test adapter placed between the
facepiece and the filter can be used. The flush-mounted filtering facepiece probe (N95 probe) shall not be used for elastomeric
respirators.
7.6.1 The probe should be placed on the midline between the nose and the mouth, whenever possible. If a different position is
necessary, every effort should be made to avoid contact with the face, placement on a seam, or interference with other features of
the respirator. The sampling apparatus should be supported in a way that it does not affect or interfere with the fit of the respirator
(that is, the respirator with the sampling apparatus connected must fit the test subject in the same manner as it would without the
sampling apparatus). Annex A2 contains more information on probe location.
7.7 Respirators—The number of respirators required for testing is:
7.7.1 Filtering facepiece respirators: 35.
7.7.2 Elastomeric facepiece respirators: three complete respirator assemblies with ten sets of filters for a unique one-size model,
and three complete respirator assemblies in each unique size and ten sets of filters when a respirator is designed and manufactured
in two or more unique sizes.
7.8 Facial Size Measurement Calipers—To measure the test subject.
7.8.1 Calibrated sliding measurement calipers, 0 mm to 200 mm length and 0 mm to 50 mm depth.
7.8.2 Calibrated spreading measurement calipers capable of measuring 0 mm to 300 mm width.
7.9 Other accessories and supplies required by the condensation nuclei counter manufacturer in order to perform the RFC test.
8. Reagents and Materials
8.1 Sodium chloride solution, 2 % NaCl solution in distilled water.
F3407 − 21
8.2 Isopropyl alcohol, or other working fluid as specified by the instrument manufacturer. Reagent grade (>99.5 %).
9. Hazards
9.1 Working Fluid—Avoid eye and skin contact.
10. Sampling, Test Specimens, and Test Units
10.1 Test Subjects—All human subject testing conducted in accordance with this RFC standard will be guided by a statement of
principles governing the institution in the discharge of its responsibilities for protecting the rights and welfare of human subjects
conducting this testing, including obtaining informed consent and screening the subjects to ensure their safety while performing
the RFC test. If applicable, it must be approved by an Institutional Review Board or other appropriate body. Twenty-five test
subjects meeting the participation and qualification criteria for the testing of respirators having facial dimensions falling within the
requirements of the NIOSH Bivariate Panel for the size of the respirator to be recruited. See Annex A1. If any test subject is
measured and the facial dimensions are not within the boundaries of the NIOSH Bivariate Panel, the test subject shall not be tested.
11. Preparation of Apparatus
11.1 Several diagnostic checks shall be performed at least daily. These shall include:
11.1.1 Chamber Particle Concentration Check—To ensure the chamber concentration is within the range specified in 7.4.4.
11.1.2 Particle classifier check.
11.1.3 Zero Check—To provide assurance that there are no leaks in the system.
11.1.4 Maximum Fit Factor Check—To ensure the condensation nuclei counter is capable of measuring high fit factors.
11.2 Follow the condensation nuclei counter and particle classifier manufacturer’s instructions for performing these checks.
12. Calibration and Standardization
12.1 At a minimum, all measuring equipment utilized for this testing must have been calibrated by the manufacturer within the
timeframe specified in the equipment’s operation manual. The calibration shall use methods traceable to National Institute of
Standards and Technology (NIST) standards.
12.2 Equipment calibration records shall be available for examination at each testing facility.
12.3 Prior to beginning any testing, a statement that all test equipment is within calibration shall be attested by the lab technician,
laboratory manager, or other designated person on each test report.
13. Conditioning
13.1 The respirators received for testing shall be free of visual damage such as distorted face seals. There are no other specific
conditioning requirements.
14. Procedure
14.1 Facial Measurement:
14.1.1 Face width is the maximum horizontal breadth of the face as measured with a spreading caliper between the zygomatic
arches (Annex A1). The test subject sits looking straight ahead with teeth together (lightly occluded). Only enough pressure is
exerted to ensure that the caliper tips are on the zygomatic arches. The test administrator will take two measurements of face width,
removing the caliper between measurements. If the two measurements for face width are within 2 mm, then stop. Otherwise, take
a third measurement.
F3407 − 21
14.1.2 Face length is the distance in the mid-sagittal plane between the menton landmark at the bottom of the chin and the sellion
landmark at the deepest point of the nasal root depression and is measured with a sliding caliper (Annex A1). The test subject sits
looking straight ahead with teeth together (lightly occluded). The fixed blade of the caliper is placed on the sellion. Only enough
pressure is exerted to obtain contact between the caliper and the skin is exerted. The test administrator will take two measurements
of face length, removing the caliper between measurements. If the two measurements for face width are within 3 mm, then stop.
Otherwise, take a third measurement.
14.1.3 The average of the face width and face length measurements will be used to determine the placement of the test subject
in the NIOSH Bivariate Panel (Annex A1) (6).
14.2 Conducting the Subject RFC Test:
14.2.1 Check the respirator to make sure the sampling probe and line are securely attached to the facepiece and that the particle
classifier technology is being used with the condensation nuclei counter.
14.2.2 The test administrator will familiarize the test subjects on the donning procedures, completing the appropriate seal checks,
and doffing procedures as specified by the manufacturer’s user instructions (UI).
14.2.3 The test administrator shall provide a description of the exercises and demonstrate them to the test subjects before the test
begins.
14.2.4 Donning:
14.2.4.1 To reduce the inter- and intra-wearer variability, each test wearer shall don the respirator under the supervision of the test
administrator. The test wearer and test administrator will be permitted to make adjustments to the facepiece until he/she and the
test administrator are satisfied that the respirator is being worn in compliance with the manufacturer’s UI, including passing the
appropriate seal checks.
14.2.4.2 Have the test subject wearing the respirator do the appropriate seal check. Every test subject must pass a seal check before
performing the subject RFC test. If the seal check fails (that is, leakage is detected), the test administrator will allow the test subject
to reposition the respirator and redo the seal check after explaining the seal check procedure again. For a unique single-size
respirator model, if the second seal check is a failure, the test subject will not be allowed to perform the RFC test. Another test
subject in the same test panel cell will be recruited and the procedure will be started over. For multiple size respirator models, the
test subject will be allowed to don a different size of the respirator and perform the seal check. The test subject will continue to
don a different size of the respirator until either the seal check is passed or all sizes are tried. If the test subject cannot pass the
seal check with any of the unique sizes, he/she will not be allowed to perform the subject RFC test. Another test subject in the
same test panel cell will be recruited and the procedure will be started over.
14.2.4.3 The test subject shall wear the respirator for 1 min prior to the subject RFC test to ensure the he/she can wear it for the
duration of the test and purge the p
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...