Standard Test Method for Base Number of Petroleum Products by Potentiometric Perchloric Acid Titration

SIGNIFICANCE AND USE
5.1 New and used petroleum products can contain basic constituents that are present as additives. The relative amounts of these materials can be determined by titration with acids. The base number is a measure of the amount of basic substance in the oil, always under the conditions of the test. It is sometimes used as a measure of lubricant degradation in service; however, any condemning limits must be empirically established.
SCOPE
1.1 This test method covers the determination of basic constituents in petroleum products by titration with perchloric acid in glacial acetic acid.  
1.2 Procedures A and B use different titration solvent volumes and sample weights.  
Note 1: A round robin on a series of new and used oils and additive concentrates has shown that the two procedures give statistically equivalent results.  
1.3 Appendix X2 provides the use of an alternative solvent system which eliminates the use of chlorobenzene in this test method. The use of the alternative solvent gives statistically equivalent results; however, the precision is worse. Paragraph X2.5.5 provides guidance when comparing results using the two different solvents.  
1.4 The constituents that may be considered to have basic characteristics include organic and inorganic bases, amino compounds, salts of weak acids (soaps), basic salts of polyacidic bases, and salts of heavy metals.  
Note 2: This test method is applicable to both fresh oils and used oils as described in Sections 16, 17, and 19 and Appendix X1.  
1.5 This test method can be used to determine base number >300 mg KOH/g. However, the precision statement in Section 19 has been obtained only on base number ≤300 mg KOH/g.  
1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. For specific warning statements, see Section 7, Section 10, and X2.2.  
1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
31-Oct-2021

Relations

Effective Date
01-Jun-2011
Effective Date
01-Mar-2006
Effective Date
10-Apr-2000
Effective Date
10-Feb-1999
Effective Date
10-Feb-1999

Overview

ASTM D2896-21: Standard Test Method for Base Number of Petroleum Products by Potentiometric Perchloric Acid Titration is an internationally recognized standard developed by ASTM International. It specifies procedures for determining the base number of petroleum products, including both fresh and used oils, via potentiometric titration using perchloric acid in glacial acetic acid. The base number quantifies the amount of basic substances, such as additives, present in an oil sample. Regular testing using this method helps monitor lubricant condition, assess additive depletion, and support oil maintenance strategies for engines and industrial equipment.

Key Topics

  • Base Number Definition: The base number is the amount of acid, expressed as milligrams of potassium hydroxide (mg KOH) per gram of sample, required to titrate the oil to a specified endpoint using potentiometric detection.
  • Applicability: Suitable for new and used petroleum products, as well as additive concentrates. Results offer insight into the composition and degradation status of lubricants.
  • Titration Procedures: The standard presents two procedures (A and B) utilizing different solvent volumes and sample sizes. Both methods yield statistically equivalent results for a wide range of oil types and additives.
  • Alternative Solvent Option: Appendix X2 allows for the use of a mixed xylenes solvent in place of chlorobenzene, eliminating the need for chlorobenzene in the test. The results remain equivalent, although precision is slightly reduced.
  • Analyte Types: The test detects a wide array of basic constituents, including organic and inorganic bases, amino compounds, soaps (salts of weak acids), basic salts of polyacidic bases, and heavy metal salts.
  • Calibration and Quality Control: The method outlines thorough calibration, electrode maintenance, and quality control procedures to ensure accuracy and repeatability.

Applications

ASTM D2896-21 is widely adopted in the oil and petroleum industry for several practical applications:

  • Condition Monitoring of Lubricants: By measuring the base number, users can assess the depletion of alkaline additives and monitor oil degradation in engines, turbines, and compressors.
  • Quality Control of Additives: The method is essential for verifying the additive content in new lubricants and formulation concentrates, ensuring product consistency.
  • Used Oil Analysis: Used oils from engines and machinery are analyzed to determine remaining additive reserves and to decide on oil replacement schedules.
  • Research and Development: Laboratories apply ASTM D2896-21 in developing and benchmarking new petroleum products, especially where additive performance is crucial.
  • Regulatory Compliance and Certification: Compliance with ASTM D2896 is often required in contracts and product specifications within international oil trade and government procurement.

Related Standards

Several ASTM and industry standards complement or are referenced in ASTM D2896-21:

  • ASTM D1193 – Specification for Reagent Water
  • ASTM D4057 – Practice for Manual Sampling of Petroleum and Petroleum Products
  • ASTM D4177 – Practice for Automatic Sampling of Petroleum and Petroleum Products
  • Other base number and acid number determination methods, such as ASTM D974 (acid number), may be used in parallel for comprehensive lubricant analysis.

Keywords: ASTM D2896-21, base number, petroleum products, potentiometric titration, perchloric acid, lubricant additives, oil analysis, used oil, quality control, lubricant monitoring, oil condition, additive depletion, international standards.

Buy Documents

Standard

ASTM D2896-21 - Standard Test Method for Base Number of Petroleum Products by Potentiometric Perchloric Acid Titration

English language (10 pages)
sale 15% off
sale 15% off
Standard

REDLINE ASTM D2896-21 - Standard Test Method for Base Number of Petroleum Products by Potentiometric Perchloric Acid Titration

English language (10 pages)
sale 15% off
sale 15% off

Get Certified

Connect with accredited certification bodies for this standard

ABS Quality Evaluations Inc.

American Bureau of Shipping quality certification.

ANAB United States Verified

Element Materials Technology

Materials testing and product certification.

UKAS United Kingdom Verified

ABS Group Brazil

ABS Group certification services in Brazil.

CGCRE Brazil Verified

Sponsored listings

Frequently Asked Questions

ASTM D2896-21 is a standard published by ASTM International. Its full title is "Standard Test Method for Base Number of Petroleum Products by Potentiometric Perchloric Acid Titration". This standard covers: SIGNIFICANCE AND USE 5.1 New and used petroleum products can contain basic constituents that are present as additives. The relative amounts of these materials can be determined by titration with acids. The base number is a measure of the amount of basic substance in the oil, always under the conditions of the test. It is sometimes used as a measure of lubricant degradation in service; however, any condemning limits must be empirically established. SCOPE 1.1 This test method covers the determination of basic constituents in petroleum products by titration with perchloric acid in glacial acetic acid. 1.2 Procedures A and B use different titration solvent volumes and sample weights. Note 1: A round robin on a series of new and used oils and additive concentrates has shown that the two procedures give statistically equivalent results. 1.3 Appendix X2 provides the use of an alternative solvent system which eliminates the use of chlorobenzene in this test method. The use of the alternative solvent gives statistically equivalent results; however, the precision is worse. Paragraph X2.5.5 provides guidance when comparing results using the two different solvents. 1.4 The constituents that may be considered to have basic characteristics include organic and inorganic bases, amino compounds, salts of weak acids (soaps), basic salts of polyacidic bases, and salts of heavy metals. Note 2: This test method is applicable to both fresh oils and used oils as described in Sections 16, 17, and 19 and Appendix X1. 1.5 This test method can be used to determine base number >300 mg KOH/g. However, the precision statement in Section 19 has been obtained only on base number ≤300 mg KOH/g. 1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. For specific warning statements, see Section 7, Section 10, and X2.2. 1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

SIGNIFICANCE AND USE 5.1 New and used petroleum products can contain basic constituents that are present as additives. The relative amounts of these materials can be determined by titration with acids. The base number is a measure of the amount of basic substance in the oil, always under the conditions of the test. It is sometimes used as a measure of lubricant degradation in service; however, any condemning limits must be empirically established. SCOPE 1.1 This test method covers the determination of basic constituents in petroleum products by titration with perchloric acid in glacial acetic acid. 1.2 Procedures A and B use different titration solvent volumes and sample weights. Note 1: A round robin on a series of new and used oils and additive concentrates has shown that the two procedures give statistically equivalent results. 1.3 Appendix X2 provides the use of an alternative solvent system which eliminates the use of chlorobenzene in this test method. The use of the alternative solvent gives statistically equivalent results; however, the precision is worse. Paragraph X2.5.5 provides guidance when comparing results using the two different solvents. 1.4 The constituents that may be considered to have basic characteristics include organic and inorganic bases, amino compounds, salts of weak acids (soaps), basic salts of polyacidic bases, and salts of heavy metals. Note 2: This test method is applicable to both fresh oils and used oils as described in Sections 16, 17, and 19 and Appendix X1. 1.5 This test method can be used to determine base number >300 mg KOH/g. However, the precision statement in Section 19 has been obtained only on base number ≤300 mg KOH/g. 1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. For specific warning statements, see Section 7, Section 10, and X2.2. 1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

ASTM D2896-21 is classified under the following ICS (International Classification for Standards) categories: 75.080 - Petroleum products in general. The ICS classification helps identify the subject area and facilitates finding related standards.

ASTM D2896-21 has the following relationships with other standards: It is inter standard links to ASTM D4057-06(2011), ASTM D1193-06, ASTM D4057-95(2000), ASTM D1193-99e1, ASTM D1193-99. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

ASTM D2896-21 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: D2896 − 21
Standard Test Method for
Base Number of Petroleum Products by Potentiometric
Perchloric Acid Titration
This standard is issued under the fixed designation D2896; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the U.S. Department of Defense.
1. Scope* mine the applicability of regulatory limitations prior to use.
Forspecificwarningstatements,seeSection7,Section10,and
1.1 This test method covers the determination of basic
X2.2.
constituents in petroleum products by titration with perchloric
1.8 This international standard was developed in accor-
acid in glacial acetic acid.
dance with internationally recognized principles on standard-
1.2 Procedures A and B use different titration solvent
ization established in the Decision on Principles for the
volumes and sample weights.
Development of International Standards, Guides and Recom-
mendations issued by the World Trade Organization Technical
NOTE 1—A round robin on a series of new and used oils and additive
concentrates has shown that the two procedures give statistically equiva-
Barriers to Trade (TBT) Committee.
lent results.
2. Referenced Documents
1.3 Appendix X2 provides the use of an alternative solvent
system which eliminates the use of chlorobenzene in this test
2.1 ASTM Standards:
method. The use of the alternative solvent gives statistically
D1193Specification for Reagent Water
equivalent results; however, the precision is worse. Paragraph
D4057Practice for Manual Sampling of Petroleum and
X2.5.5 provides guidance when comparing results using the
Petroleum Products
two different solvents.
D4177Practice for Automatic Sampling of Petroleum and
Petroleum Products
1.4 The constituents that may be considered to have basic
characteristics include organic and inorganic bases, amino
3. Terminology
compounds, salts of weak acids (soaps), basic salts of poly-
acidic bases, and salts of heavy metals. 3.1 Definitions:
3.1.1 base number, n—the quantity of a specified acid,
NOTE 2—This test method is applicable to both fresh oils and used oils
expressed in terms of the equivalent number of milligrams of
as described in Sections 16, 17, and 19 and Appendix X1.
potassium hydroxide per gram of sample, required to titrate a
1.5 This test method can be used to determine base number
sample in a specified solvent to a specified endpoint using a
>300mg KOH/g. However, the precision statement in Section
specified detection system.
19 has been obtained only on base number ≤300mg KOH/g.
4. Summary of Test Method
1.6 The values stated in SI units are to be regarded as
standard. No other units of measurement are included in this
4.1 The sample is dissolved in an essentially anhydrous
standard.
mixture of chlorobenzene and glacial acetic acid and titrated
1.7 This standard does not purport to address all of the with a solution of perchloric acid in glacial acetic acid using
safety concerns, if any, associated with its use. It is the potentiometric titrimeter. A glass indicating electrode and a
responsibility of the user of this standard to establish appro- reference electrode are used, the latter being connected with
priate safety, health, and environmental practices and deter- the sample solution by means of a salt bridge. The meter
readings are plotted against the respective volumes of titrating
solution, and the end point is taken at the inflection in the
This test method is under the jurisdiction of ASTM Committee D02 on
resulting curve.
Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of
Subcommittee D02.06 on Analysis of Liquid Fuels and Lubricants.
Current edition approved Nov. 1, 2021. Published November 2021. Originally
approved in 1970. Last previous edition approved in 2015 as D2896–15. DOI: For referenced ASTM standards, visit the ASTM website, www.astm.org, or
10.1520/D2896-21. contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
This test method has been approved by the sponsoring committees and accepted Standards volume information, refer to the standard’s Document Summary page on
by the cooperating societies in accordance with established procedures. the ASTM website.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D2896 − 21
4.2 ProcedureAuses120mLoftitrationsolvent.Procedure the removal of the beaker without disturbing the electrodes,
B uses 60mL of titration solvent. In addition, the two buret, and stirrer is desirable.
proceduresusedifferentequationsforthecalculationofappro-
NOTE 5—Some apparatus may be sensitive to interference by static
priate sample weights. Since many portions of the test method
electricity,shownbyerraticmovementsofrecorderpenormeterindicator,
are identical for ProceduresAand B, only the unique sections
when the titration assembly (beaker and electrodes) is approached by the
operator.Inthiscasesurroundthebeakercloselywithacylinderofcopper
will be described separately for the two versions of the test
gauze that is electrically grounded.
method.
4.3 Occasionally certain used oils give no inflection in the 7. Reagents and Materials
forward titration mode, in which case a back titration modifi-
7.1 Purity of Reagents—Reagent grade chemicals shall be
cation with sodium acetate titrant is employed.
used in all tests. Unless otherwise indicated, it is intended that
all reagents shall conform to the specifications of the Commit-
5. Significance and Use
tee onAnalytical Reagents of theAmerican Chemical Society,
where such specifications are available. Other grades may be
5.1 New and used petroleum products can contain basic
used, provided it is first ascertained that the reagent is of
constituents that are present as additives.The relative amounts
sufficiently high purity to permit its use without lessening the
of these materials can be determined by titration with acids.
accuracy of the determination.
Thebasenumberisameasureoftheamountofbasicsubstance
in the oil, always under the conditions of the test. It is
7.2 Purity of Water—Unless otherwise indicated, references
sometimes used as a measure of lubricant degradation in
to water shall be understood to mean reagent water that meets
service; however, any condemning limits must be empirically
the requirement of either Type I, II, or III of Specification
established.
D1193.
7.3 Acetic Acid, glacial (Warning—Toxic and irritant).
6. Apparatus
7.4 Acetic Anhydride (Warning—Toxic and irritant).
6.1 Potentiometric Titrimeters, either automatic recording
7.5 Chlorobenzene (Warning—Toxic and irritant).
or manual.
7.6 Perchloric Acid, Standard Solution in Acetic Acid
6.2 Glass Electrode, pH 0 to 11, general-purpose type.
(0.1 N) (Warning—Powerful oxidant when dry or heated.
6.3 Reference Electrode, silver/silver chloride (Ag/AgCl)
Greatcareshouldbetakentoavoidcontactwithorganicmatter
reference electrode with a nonaqueous bridge as described in
under conditions that may result in subsequent drying or
Section 10. (See also 19.1.)
heating, and spills should be washed immediately and thor-
NOTE 3—Some reference electrodes with fritted or fiber diaphragms
oughly with water)—Mix 8.5mL of 70% to 72% perchloric
and some combined glass plus reference electrodes systems are commer-
acid (HClO , 70% to 72%) (or 10.2mL of 60% to 62%
cially available, such as the single-rod glass plus silver/silver chloride
HClO solution) with 500mL of glacial acetic and 30mL (or
electrodeassembly.Duringthedevelopmentofthistestmethod,theuseof
35mL if the 60% to 62% HClO solution is used) of acetic
electrodes of these types gave problems in some laboratories, but not in 4
anhydride. Dilute to 1L with glacial acetic acid. Allow the
others. Accordingly, these electrodes are permitted in this test method,
provided that the sodium perchlorate bridge is used; however, when
solution to stand for 24h.
stability or other problems arise with their use, the sleeve-type electrode
NOTE 6—Excess acetic anhydride should be avoided to prevent acety-
should be used.
lation of any primary or secondary amines that may be present.
6.4 Stirrer, either mechanical or electrical, with variable
7.7 Potassium Hydrogen Phthalate—(KHC H O ).
8 4 4
speeds and with propeller or paddle of chemically inert
material.Whenanelectricalstirrerisused,itmustbegrounded 7.8 Sodium Perchlorate Electrolyte—(Warning—Sodium
sothatdisconnectingorconnectingthepowertothemotorwill
perchlorate is toxic and an irritant. It is also a powerful
not produce a permanent change in meter reading during the oxidizing agent when heated. Great care should be taken to
course of a titration.Amagnetic stirrer with stirring bar can be
avoid contact with organic matter under conditions that may
used provided it meets these conditions. result in subsequent drying or heating, and spills should be
washed immediately and thoroughly with water.) Prepare a
6.5 Buret, 10mLor 20mL, graduated in 0.05mLdivisions
saturated solution of sodium perchlorate (NaClO ) in glacial
and calibrated with an accuracy of 60.02mL, or an automatic
acetic acid.An excess of undissolved NaClO shall always be
buret of similar accuracy.
present at the bottom of the solution.
6.6 Titration Beaker, made of borosilicate glass or other
7.9 Titration Solvent—Add one volume of glacial acetic
suitable titration beaker, tall form recommended.
acid to two volumes of chlorobenzene.
6.6.1 For Procedure A, use a beaker of 250mL or 300mL
capacity. For Procedure B, use a beaker of about 150mL
capacity such that 60mL of titration solvent will cover the
ACS Reagent Chemicals, Specifications and Procedures for Reagents and
Standard-Grade Reference Materials, American Chemical Society, Washington,
electrodes.
DC. For suggestions on the testing of reagents not listed by theAmerican Chemical
Society, see Analar Standards for Laboratory Chemicals, BDH Ltd., Poole, Dorset,
NOTE 4—Other beakers of suitable size capacity may be used.
U.K., and the United States Pharmacopeia and National Formulary, U.S. Pharma-
6.7 Titration Stand, suitable to support the beaker,
copeial Convention, Inc. (USPC), Rockville, MD.
electrodes, stirrer, and buret. An arrangement that allows for Available commercially for purchase already prepared.
D2896 − 21
7.10 Sodium Carbonate, anhydrous (Na CO ).
N = normality of the HClO solution, and
2 3
A 4
G = volume of standard sodium acetate used in the
7.11 Sodium Acetate Solution, 0.1 N in acetic acid (for back
standardization, mL.
titration,seeSections16and17)—Dissolve5.3gofanhydrous
Na CO in 300mL of glacial acetic acid. Dilute to 1L with
8.2.2 Procedure B (60mL)—Use 60mLof titration solvent
2 3
acetic acid after solution is complete.
and 4.00mL of 0.1 N HClO solution. Titrate as described in
8.2.1. Calculate the normality, N , of the sodium acetate
B
8. Standardization of Reagents
solution as follows:
8.1 Perchloric Acid Solution—The standardization of the
N 5 @~4.00 2 b!N #/G (3)
B A
perchloricacidsolution(HClO )differsforthetwoprocedures
where:
as follows:
b = volume corresponding to V for the blank titration,
8.1.1 Procedure A (120mL)—Heat a quantity of potassium
N = normality of the HClO solution, and
hydrogenphthalateinanovenat120°Cfor2handallowitto A 4
G = volumeofstandardacetoussodiumacetateusedinthe
cool. Take 0.1g to 0.2g of the potassium hydrogen phthalate
standardization, mL.
weighedtothenearest0.1mganddissolveitin40mLofwarm
glacial acetic acid. Add 80mL of chlorobenzene, cool, and
9. Preparation of Sample
titrate with 0.1 N HClO solution, using the electrode system
and procedures given in 10.1 to 10.4 and 11.4 to 11.7. Detect
9.1 When applicable, refer to Practice D4057 (Manual
the end point by the same procedure used for base number
Sampling)orPracticeD4177(AutomaticSampling)forproper
determination (see 14.2). Carry out a blank titration on 40mL
sampling techniques.
of glacial acetic acid plus 80mL of chlorobenzene (see 11.8).
9.1.1 When sampling used lubricants, the specimen shall be
8.1.2 Procedure B (60mL)—Heat a quantity of potassium
representative of the system sampled and shall be free of
hydrogenphthalateinanovenat120°Cfor2handallowitto
contamination from external sources.
cool.Take 0.05g to 0.1g of the potassium hydrogen phthalate
9.1.2 Agitateusedoilsamplesthoroughlytoensurethatany
weighedtothenearest0.1mganddissolveitin20mLofwarm
sedimentpresentishomogeneouslysuspendedbeforeanalysis,
glacial acetic acid. Add 40 mL of chlorobenzene, cool, and
as the sediment can be acidic or basic or have adsorbed acidic
titrate with 0.1 N HClO solution as described in 8.1.1. Carry
or basic material from the sample. When necessary, samples
out a blank titration on 20mLof glacial acetic acid and 40 mL are warmed to aid mixing.
of chlorobenzene (see 11.8). NOTE 8—As used oils can change appreciably in storage, samples
should be tested as soon as possible after removal from the lubricating
8.1.3 Calculate the normality, N , of the HClO solution as
A 4
system and the dates of sampling and testing, if known, should be noted.
follows:
N 51000W/ 204.23· V 2 b (1)
@ ~ !#
A
10. Preparation of Electrode System
where:
10.1 Preparation of Electrodes—When the reference elec-
trode is to be changed from aqueous bridge to nonaqueous,
W = potassium hydrogen phthalate, g,
V = HClO solution used, mL, and drain out the aqueous solution, wash out all crystals of KCl
b = volume corresponding to V for the blank titration, mL.
with water, then rinse the outer jacket (salt bridge) several
times with NaClO electrolyte solution. Finally fill the outer
NOTE 7—Because of the relatively large coefficient of volumetric
jacket with NaClO electrolyte solution up to the filling hole.
expansion of organic liquids, the acetous HClO solution should be used
When using the sleeve-type electrode, carefully remove the
within 65°C of the temperature at which it was standardized. If used at
ground-glasssleeveandthoroughlywipebothgroundsurfaces.
a temperature more than 5°C higher, multiply the volume used by the
Replacethesleevelooselyandallowafewdropsofelectrolyte
factor 1−(t·0.001). If used at a temperature more than 5°C lower,
to drain through to flush the ground-glass joint and to wet the
multiply by the factor 1+(t·0.001), where t is the difference in degrees
Celsius between temperatures of standardization and use and is always
ground surfaces thoroughly with electrolyte. Set the sleeve
positive.
firmly in place, refill the outer jacket with the NaClO
8.2 Sodium Acetate Solution—The standardization of the
electrolyte solution, and rinse the electrode with chloroben-
sodium acetate solution (Na CO ) differs for the two proce- zene. When in use, the electrolyte level in the reference
2 3
dures as follows:
electrodeshouldbekeptabovethatoftheliquidinthetitration
8.2.1 Procedure A (120mL)—Use 120mL of titration sol- beaker to prevent entry of contaminants into the salt bridge.
vent and 8.00mL of 0.1 N HClO solution. Titrate with 0.1 N
When not in use, fill the reference electrode with the NaClO
sodium acetate solution, using the electrode system and pro- electrolyte solution, leave the bung in the filling orifice, and
cedure given in 10.1 to 10.4 and 11.4 to 11.7. Detect the end
immersebothelectrodesindistilledwater,keepingthelevelof
point by the same procedure as will be used for base number
the electrolyte above that of the distilled water.
determination (see 14.2). Calculate the normality, N ,ofthe
B
10.2 Testing of Electrodes—Test the meter-electrode com-
sodium acetate solution as follows:
bination when first put into use or when new electrodes are
N 5 @~8.00 2 b!N #/G (2)
installed and retest at intervals thereafter as follows:
B A
10.2.1 Procedure A—Dip the electrodes into a well-stirred
where:
mixture of 100mL of glacial acetic acid plus 0.2g of
b = volume corresponding to V for the blank titration,
KHC H O and record the reading given by the meter. Rinse
8 4 4
D2896 − 21
the electrodes with chlorobenzene and immerse in 100mL of 11.3 Place the beaker on the titration stand and stir the
glacial acetic acid plus 1.5mL of 0.1 N HClO solution. The solution until the sample is dissolved.
difference between readings is to be at least 0.3 V.
NOTE11—Ifsolutionofthesampleprovesdifficult,dissolveitin80mL
10.2.2 Procedure B—Dip the electrodes into a well-stirred
of chlorobenzene in the titration beaker, then add 40mL of glacial acetic
mixture of 60 mL of glacial acetic acid plus 0.1 g of acid. Many used oils contain some solid materials that will not dissolve.
This is a frequently observed condition.
KHC H O andrecordthereadingveinbythemeter.Rinsethe
8 4 4
electrodeswithchlorobenzeneandimmersein50mLofglacial
11.4 Prepare the electrodes as directed in 10.1, 10.2, and
acetic acid plus 0.75mL of 0.1 N HClO solution. The
10.3. Position the electrodes in the solution so that they are
difference between readings is to be at least 0.3 V.
immersed as far as possible. Continue stirring throughout the
determination at a rate sufficient to produce vigorous agitation
NOTE 9—See Appendix X4 for a possible procedure to check the
without spattering and without stirring air into the solution.
electrode performance.
Adjust the meter so that it reads in the upper part of the
10.3 Cleaning of Electrodes—Following a titration, it is
millivolt scale; for example, 700mV. For simple meters
necessary that the electrodes are properly cleaned before
without this adjustment, it is necessary to incorporate a source
proceeding with a subsequent titration. One such way that has
of potential in series with the electrode. A 1.5V dry cell and
beenfoundsuitableistofirstwashtheelectrodeswithtitration
potential divider is suitable.
solventtoremoveanyadheringoilymaterialfromtheprevious
11.5 Fill the buret with 0.1 N HClO solution and place the
titration. Then wash the electrodes with water to dissolve any 4
buret in position in the titration assembly, taking care that the
NaClO that may have formed around the sleeve of the
tip is immersed below the level of the surface of the liquid in
reference electrode and to restore the aqueous gel layer of the
the beaker. Record the initial buret and meter (cell potential)
glass electrode. Rinse again with the titration solvent. Before
readings.
starting a series of sample titrations, follow this rinsing
procedure, then run one or two blank titrations on the solvent
11.6 Titration:
to condition the electrodes. Repeat the blank titrations if
11.6.1 Manual Titration—Add suitable small portions of
necessary.
titrant and, after waiting until a constant potential has been
established (Note 12), record the buret and meter readings.At
10.4 Maintenance of Electrodes—When there is reason to
the start of the titration and in any subsequent regions
believe that the glass electrode has become contaminated, it
(inflections) where 0.1mL of titrant consistently produces a
canbecleanedbyimmersionincoldchromicacid(Warning—
total change of more than 0.03V (corresponding to 0.5pH
Corrosive and carcinogenic) or an alternative non-chromium-
scale unit) in the cell potential, add 0.05mL portions. In the
containing strongly-oxidizing acid cleaning solution for 5 min,
intermediate regions (plateaus) where 0.1mL increments
followed by thorough water washing. After this cleaning
change the potential by less than 0.03V, add large portions
treatment,testtheelectrodeasdescribedin10.2.Thereference
sufficient to produce a total potential change approximately
electrode can be cleaned by draining and refilling with fresh
equal to, but not greater than, 0.03V. Titrate in this manner
NaClO solution.Maintaintheelectrolytelevelinthereference
until the potential changes less than 0.005V(corresponding to
electrode above that of the liquid in the titration beaker at all
0.1 pH scale unit) per 0.1 mL.
times. Do not allow the electrodes to remain immersed in
titration solvent for any appreciable period of time between
NOTE12—Considerthecellpotentialconstantwhenitchangeslessthan
titrations. While the electrodes are not extremely fragile, 0.005V⁄min.
handle them carefully at all times and particularly avoid
11.6.2 Automatic Recording Titration—Adjust the instru-
scratching the glass electrode.
ment in accordance with the manufacturer’s instructions and
set the titration speed at 1.0mL⁄min maximum.
11. Procedure A (120 mL)
11.7 On completion of the titration, remove the beaker and
11.1 Calculate the quantity of sample required from its
rinse the electrodes and buret tip with titration solvent, then
expected base number, BN, as follows:
withwater,thenagainwithtitrationsolvent(see10.3).Storein
water when not in use (see 10.1).
Approximateweightofsample, g 528/expected BN (4)
NOTE 10—For the back titration procedure (see 16.2), or when
11.8 For each set of samples make a blank titration using
analyzing used oils, it may be necessary to use a smaller sample weight.
120mL of titration solvent. For a manual titration add 0.1 N
11.1.1 Weigh the sample into the titration beaker, applying
HClO solution in 0.05mL increments, waiting between each
the limits shown as follows. A maximum of 20g should be addition until a constant cell potential is established. Record
taken for analysis.
meter and buret readings after each increment. Follow the
procedure in 11.6.2 for an automatic titration.
Precision of
Sample Weight, g Weighing, g
12. Procedure B (60 mL)
10 to 20 0.05
5to10 0.02
12.1 Calculate the quantity of sample required from its
1 to 5 0.005
expected base number as follows:
0.25 to 1.0 0.001
0.1 to 0.25 0.0005
Approximateweightofsample, g 510/expected BN (5)
11.2 Add 120mL of titration solvent to the sample. NOTE 13—For the back titration procedure (see 17.2) it may be
D2896 − 21
necessary to use a smaller sample weight.
solutionin0.05mLincrements,waitingbetweeneachaddition
until a constant cell potential is established. Record meter and
12.1.1 Weigh the sample into the titration beaker, applying
buret readings after each increment. Follow the procedure in
the limits shown as follows. A maximum of 10 g should be
12.6.2 for an automatic titration.
taken for analysis.
Precision of
13. Quality Control Checks
Sample Weight, g Weighing, g
13.1 Confirm the performance of the equipment or the
5to10 0.02
1 to 5 0.005
procedure each day it
...


This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: D2896 − 15 D2896 − 21
Standard Test Method for
Base Number of Petroleum Products by Potentiometric
Perchloric Acid Titration
This standard is issued under the fixed designation D2896; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the U.S. Department of Defense.
1. Scope*
1.1 This test method covers the determination of basic constituents in petroleum products by titration with perchloric acid in
glacial acetic acid.
1.2 Procedures A and B use different titration solvent volumes and sample weights.
NOTE 1—A round robin on a series of new and used oils and additive concentrates has shown that the two procedures give statistically equivalent results.
1.3 Appendix X2 provides the use of an alternative solvent system which eliminates the use of chlorobenzene in this test method.
The use of the alternative solvent gives statistically equivalent results; however, the precision is worse. Paragraph X2.5.5 provides
guidance when comparing results using the two different solvents.
1.4 The constituents that may be considered to have basic characteristics include organic and inorganic bases, amino compounds,
salts of weak acids (soaps), basic salts of polyacidic bases, and salts of heavy metals.
NOTE 2—This test method is applicable to both fresh oils and used oils as described in Sections 16, 17, and 19 and Appendix X1.
1.5 This test method can be used to determine base number >300 mg KOH/g. However, the precision statement in Section 19 has
been obtained only on base number ≤300 mg KOH/g.
1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety safety, health, and healthenvironmental practices and determine the
applicability of regulatory limitations prior to use. For specific warning statements, see Section 7, Section 10, and X2.2.
1.8 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
This test method is under the jurisdiction of ASTM Committee D02 on Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of Subcommittee
D02.06 on Analysis of Liquid Fuels and Lubricants.
Current edition approved Dec. 1, 2015Nov. 1, 2021. Published February 2016November 2021. Originally approved in 1970. Last previous edition approved in 20112015
as D2896 – 11.D2896 – 15. DOI: 10.1520/D2896-15.10.1520/D2896-21.
This test method has been approved by the sponsoring committees and accepted by the cooperating societies in accordance with established procedures.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D2896 − 21
2. Referenced Documents
2.1 ASTM Standards:
D1193 Specification for Reagent Water
D4057 Practice for Manual Sampling of Petroleum and Petroleum Products
D4177 Practice for Automatic Sampling of Petroleum and Petroleum Products
3. Terminology
3.1 Definitions:
3.1.1 base number—number, n—the quantity of a specified acid, expressed in terms of the equivalent number of milligrams of
potassium hydroxide per gram of sample, required to titrate a sample in a specified solvent to a specified endpoint using a specified
detection system.
4. Summary of Test Method
4.1 The sample is dissolved in an essentially anhydrous mixture of chlorobenzene and glacial acetic acid and titrated with a
solution of perchloric acid in glacial acetic acid using potentiometric titrimeter. A glass indicating electrode and a reference
electrode are used, the latter being connected with the sample solution by means of a salt bridge. The meter readings are plotted
against the respective volumes of titrating solution, and the end point is taken at the inflection in the resulting curve.
4.2 Procedure A uses 120 mL of titration solvent. Procedure B uses 60 mL of titration solvent. In addition, the two procedures use
different equations for the calculation of appropriate sample weights. Since many portions of the test method are identical for
Procedures A and B, only the unique sections will be described separately for the two versions of the test method.
4.3 Occasionally certain used oils give no inflection in the forward titration mode, in which case a back titration modification with
sodium acetate titrant is employed.
5. Significance and Use
5.1 New and used petroleum products can contain basic constituents that are present as additives. The relative amounts of these
materials can be determined by titration with acids. The base number is a measure of the amount of basic substance in the oil,
always under the conditions of the test. It is sometimes used as a measure of lubricant degradation in service; however, any
condemning limits must be empirically established.
6. Apparatus
6.1 Potentiometric Titrimeters, either automatic recording or manual.
6.2 Glass Electrode, pH 0 to 11, general-purpose type.
6.3 Reference Electrode, silver/silver chloride (Ag/AgCl) reference electrode with a nonaqueous bridge as described in Section 10.
(See also 19.1.)
NOTE 3—Some reference electrodes with fritted or fiber diaphragms and some combined glass plus reference electrodes systems are commercially
available, such as the single-rod glass plus silver/silver chloride electrode assembly. During the development of this test method, the use of electrodes
of these types gave problems in some laboratories, but not in others. Accordingly, these electrodes are permitted in this test method, provided that the
sodium perchlorate bridge is used; however, when stability or other problems arise with their use, the sleeve-type electrode should be used.
6.4 Stirrer, either mechanical or electrical, with variable speeds and with propeller or paddle of chemically inert material. When
an electrical stirrer is used, it must be grounded so that disconnecting or connecting the power to the motor will not produce a
permanent change in meter reading during the course of a titration. A magnetic stirrer with stirring bar can be used provided it
meets these conditions.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
D2896 − 21
6.5 Buret, 10 mL or 20 mL, graduated in 0.05 mL divisions and calibrated with an accuracy of 60.02 mL, or an automatic buret
of similar accuracy.
6.6 Titration Beaker, made of borosilicate glass or other suitable titration beaker, tall form recommended.
6.6.1 For Procedure A, use a beaker of 250 mL or 300 mL capacity. For Procedure B, use a beaker of about 150 mL capacity such
that 60 mL of titration solvent will cover the electrodes.
NOTE 4—Other beakers of suitable size capacity may be used.
6.7 Titration Stand, suitable to support the beaker, electrodes, stirrer, and buret. An arrangement that allows for the removal of the
beaker without disturbing the electrodes, buret, and stirrer is desirable.
NOTE 5—Some apparatus may be sensitive to interference by static electricity, shown by erratic movements of recorder pen or meter indicator, when the
titration assembly (beaker and electrodes) is approached by the operator. In this case surround the beaker closely with a cylinder of copper gauze that
is electrically grounded.
7. Reagents and Materials
7.1 Purity of Reagents—Reagent grade chemicals shall be used in all tests. Unless otherwise indicated, it is intended that all
reagents shall conform to the specifications of the Committee on Analytical Reagents of the American Chemical Society, where
such specifications are available. Other grades may be used, provided it is first ascertained that the reagent is of sufficiently high
purity to permit its use without lessening the accuracy of the determination.
7.2 Purity of Water—Unless otherwise indicated, references to water shall be understood to mean reagent water that meets the
requirement of either Type I, II, or III of Specification D1193.
7.3 Acetic Acid, glacial (Warning—Toxic and irritant).
7.4 Acetic Anhydride (Warning—Toxic and irritant).
7.5 Chlorobenzene (Warning—Toxic and irritant).
7.6 Perchloric Acid, Standard Solution in Acetic Acid (0.1 N) (Warning—Powerful oxidant when dry or heated. Great care
should be taken to avoid contact with organic matter under conditions that may result in subsequent drying or heating, and spills
should be washed immediately and thoroughly with water)—Mix 8.5 mL of 70 % to 72 % perchloric acid (HClO , 70 % to 72 %)
(or 10.2 mL of 60 % to 62 % HClO solution) with 500 mL of glacial acetic and 30 mL (or 35 mL if the 60 % to 62 % HClO
4 4
solution is used) of acetic anhydride. Dilute to 1 L with glacial acetic acid. Allow the solution to stand for 24 h.
NOTE 6—Excess acetic anhydride should be avoided to prevent acetylation of any primary or secondary amines that may be present.
7.7 Potassium Hydrogen Phthalate—(KHC H O ).
8 4 4
7.8 Sodium Perchlorate Electrolyte—(Warning—Sodium perchlorate is toxic and an irritant. It is also a powerful oxidizing agent
when heated. Great care should be taken to avoid contact with organic matter under conditions that may result in subsequent drying
or heating, and spills should be washed immediately and thoroughly with water.) Prepare a saturated solution of sodium perchlorate
(NaClO ) in glacial acetic acid. An excess of undissolved NaClO shall always be present at the bottom of the solution.
4 4
Reagent Chemicals, American Chemical Society SpecificationsACS Reagent Chemicals, Specifications and Procedures for Reagents and Standard-Grade Reference
Materials, , American Chemical Society, Washington, DC. For suggestions on the testing of reagents not listed by the American Chemical Society, see Analar Standards for
Laboratory Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia and National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC),
Rockville, MD.
Available commercially for purchase already prepared.
D2896 − 21
7.9 Titration Solvent—Add one volume of glacial acetic acid to two volumes of chlorobenzene.
7.10 Sodium Carbonate, anhydrous (Na CO ).
2 3
7.11 Sodium Acetate Solution, 0.1 N in acetic acid (for back titration, see Sections 16 and 17)—Dissolve 5.3 g of anhydrous
Na CO in 300 mL of glacial acetic acid. Dilute to 1 L with acetic acid after solution is complete.
2 3
8. Standardization of Reagents
8.1 Perchloric Acid Solution—The standardization of the perchloric acid solution (HClO ) differs for the two procedures as
follows:
8.1.1 Procedure A (120 mL)—Heat a quantity of potassium hydrogen phthalate in an oven at 120 °C for 2h2 h and allow it to cool.
Take 0.1 g to 0.2 g of the potassium hydrogen phthalate weighed to the nearest 0.1 mg and dissolve it in 40 mL of warm glacial
acetic acid. Add 80 mL of chlorobenzene, cool, and titrate with 0.1 N HClO solution, using the electrode system and procedures
given in 10.1 to 10.4 and 11.4 to 11.7. Detect the end point by the same procedure used for base number determination (see 14.2).
Carry out a blank titration on 40 mL of glacial acetic acid plus 80 mL of chlorobenzene (see 11.8).
8.1.2 Procedure B (60 mL)—Heat a quantity of potassium hydrogen phthalate in an oven at 120 °C for 2 h and allow it to cool.
Take 0.05 g to 0.1 g of the potassium hydrogen phthalate weighed to the nearest 0.1 mg and dissolve it in 20 mL of warm glacial
acetic acid. Add 40 mL of chlorobenzene, cool, and titrate with 0.1 N HClO solution as described in 8.1.1. Carry out a blank
titration on 20 mL of glacial acetic acid and 40 mL of chlorobenzene (see 11.8).
8.1.3 Calculate the normality, N , of the HClO solution as follows:
A 4
N 5 1000W/@204.23·~V 2 b!# (1)
A
where:
W = potassium hydrogen phthalate, g,
V = HClO solution used, mL, and
b = volume corresponding to V for the blank titration, mL.
NOTE 7—Because of the relatively large coefficient of volumetric expansion of organic liquids, the acetous HClO solution should be used within 65 °C
of the temperature at which it was standardized. If used at a temperature more than 5 °C higher, multiply the volume used by the factor 1 − (t·0.001).
If used at a temperature more than 5 °C lower, multiply by the factor 1 + (t·0.001), where t is the difference in degrees Celsius between temperatures of
standardization and use and is always positive.
8.2 Sodium Acetate Solution—The standardization of the sodium acetate solution (Na CO ) differs for the two procedures as
2 3
follows:
8.2.1 Procedure A (120 mL)—Use 120 mL of titration solvent and 8.00 mL of 0.1 N HClO solution. Titrate with 0.1 N sodium
acetate solution, using the electrode system and procedure given in 10.1 to 10.4 and 11.4 to 11.7. Detect the end point by the same
procedure as will be used for base number determination (see 14.2). Calculate the normality, N , of the sodium acetate solution
B
as follows:
N 5 8.00 2 b N /G (2)
@~ ! #
B A
where:
b = volume corresponding to V for the blank titration,
N = normality of the HClO solution, and
A 4
G = volume of standard sodium acetate used in the standardization, mL.
8.2.2 Procedure B (60 mL)—Use 60 mL of titration solvent and 4.00 mL of 0.1 N HClO solution. Titrate as described in 8.2.1.
Calculate the normality, N , of the sodium acetate solution as follows:
B
N 5 4.00 2 b N /G (3)
@~ ! #
B A
D2896 − 21
where:
b = volume corresponding to V for the blank titration,
N = normality of the HClO solution, and
A 4
G = volume of standard acetous sodium acetate used in the standardization, mL.
9. Preparation of Sample
9.1 It is essential to ensure thatWhen applicable, refer to Practice D4057 the sample is representative since any sediment can be
acidic or basic or(Manual Sampling) or Practice D4177 have adsorbed acidic or basic material from the sample. When necessary,
samples are warmed to aid mixing. Used oils should be vigorously shaken to ensure homogeneity before sampling. (Automatic
Sampling) for proper sampling techniques.
NOTE 8—As used oils can change appreciably in storage, samples should be tested as soon as possible after removal from the lubricating system and the
dates of sampling and testing, if known, should be noted.
9.1.1 When sampling used lubricants, the specimen shall be representative of the system sampled and shall be free of
contamination from external sources.
9.1.2 Agitate used oil samples thoroughly to ensure that any sediment present is homogeneously suspended before analysis, as the
sediment can be acidic or basic or have adsorbed acidic or basic material from the sample. When necessary, samples are warmed
to aid mixing.
NOTE 8—As used oils can change appreciably in storage, samples should be tested as soon as possible after removal from the lubricating system and the
dates of sampling and testing, if known, should be noted.
10. Preparation of Electrode System
10.1 Preparation of Electrodes—When the reference electrode is to be changed from aqueous bridge to nonaqueous, drain out the
aqueous solution, wash out all crystals of KCl with water, then rinse the outer jacket (salt bridge) several times with NaClO
electrolyte solution. Finally fill the outer jacket with NaClO electrolyte solution up to the filling hole. When using the sleeve-type
electrode, carefully remove the ground-glass sleeve and thoroughly wipe both ground surfaces. Replace the sleeve loosely and
allow a few drops of electrolyte to drain through to flush the ground-glass joint and to wet the ground surfaces thoroughly with
electrolyte. Set the sleeve firmly in place, refill the outer jacket with the NaClO electrolyte solution, and rinse the electrode with
chlorobenzene. When in use, the electrolyte level in the reference electrode should be kept above that of the liquid in the titration
beaker to prevent entry of contaminants into the salt bridge. When not in use, fill the reference electrode with the NaClO
electrolyte solution, leave the bung in the filling orifice, and immerse both electrodes in distilled water, keeping the level of the
electrolyte above that of the distilled water.
10.2 Testing of Electrodes—Test the meter-electrode combination when first put into use or when new electrodes are installed and
retest at intervals thereafter as follows:
10.2.1 Procedure A—Dip the electrodes into a well-stirred mixture of 100 mL of glacial acetic acid plus 0.2 g of KHC H O and
8 4 4
record the reading given by the meter. Rinse the electrodes with chlorobenzene and immerse in 100 mL of glacial acetic acid plus
1.5 mL of 0.1 N HClO solution. The difference between readings is to be at least 0.3 V.
10.2.2 Procedure B—Dip the electrodes into a well-stirred mixture of 60 mL of glacial acetic acid plus 0.1 g of KHC H O and
8 4 4
record the reading vein by the meter. Rinse the electrodes with chlorobenzene and immerse in 50 mL of glacial acetic acid plus
solution. The difference between readings is to be at least 0.3 V.
0.75 mL of 0.1 N HClO
NOTE 9—See Appendix X4 for a possible procedure to check the electrode performance.
10.3 Cleaning of Electrodes—Following a titration, it is necessary that the electrodes are properly cleaned before proceeding with
a subsequent titration. One such way that has been found suitable is to first wash the electrodes with titration solvent to remove
any adhering oily material from the previous titration. Then wash the electrodes with water to dissolve any NaClO that may have
formed around the sleeve of the reference electrode and to restore the aqueous gel layer of the glass electrode. Rinse again with
the titration solvent. Before starting a series of sample titrations, follow this rinsing procedure, then run one or two blank titrations
on the solvent to condition the electrodes. Repeat the blank titrations if necessary.
D2896 − 21
10.4 Maintenance of Electrodes—When there is reason to believe that the glass electrode has become contaminated, it can be
cleaned by immersion in cold chromic acid (Warning—Corrosive and carcinogenic) or an alternative non-chromium-containing
strongly-oxidizing acid cleaning solution for 5 min, followed by thorough water washing. After this cleaning treatment, test the
electrode as described in 10.2. The reference electrode can be cleaned by draining and refilling with fresh NaClO solution.
Maintain the electrolyte level in the reference electrode above that of the liquid in the titration beaker at all times. Do not allow
the electrodes to remain immersed in titration solvent for any appreciable period of time between titrations. While the electrodes
are not extremely fragile, handle them carefully at all times and particularly avoid scratching the glass electrode.
11. Procedure A (120 mL)
11.1 Calculate the quantity of sample required from its expected base number, BN, as follows:
Approximate weight of sample, g5 28/expected BN (4)
NOTE 10—For the back titration procedure (see 16.2), or when analyzing used oils, it may be necessary to use a smaller sample weight.
11.1.1 Weigh the sample into the titration beaker, applying the limits shown as follows. A maximum of 20 g should be taken for
analysis.
Precision of
Sample Weight, g Weighing, g
10 to 20 0.05
5 to 10 0.02
1 to 5 0.005
0.25 to 1.0 0.001
0.1 to 0.25 0.0005
11.2 Add 120 mL of titration solvent to the sample.
11.3 Place the beaker on the titration stand and stir the solution until the sample is dissolved.
NOTE 11—If solution of the sample proves difficult, dissolve it in 80 mL of chlorobenzene in the titration beaker, then add 40 mL of glacial acetic acid.
Many used oils contain some solid materials that will not dissolve. This is a frequently observed condition.
11.4 Prepare the electrodes as directed in 10.1, 10.2, and 10.3. Position the electrodes in the solution so that they are immersed
as far as possible. Continue stirring throughout the determination at a rate sufficient to produce vigorous agitation without
spattering and without stirring air into the solution. Adjust the meter so that it reads in the upper part of the millivolt scale; for
example, 700 mV. For simple meters without this adjustment, it is necessary to incorporate a source of potential in series with the
electrode. A 1.5 V dry cell and potential divider is suitable.
11.5 Fill the buret with 0.1 N HClO solution and place the buret in position in the titration assembly, taking care that the tip is
immersed below the level of the surface of the liquid in the beaker. Record the initial buret and meter (cell potential) readings.
11.6 Titration:
11.6.1 Manual Titration—Add suitable small portions of titrant and, after waiting until a constant potential has been established
(Note 12), record the buret and meter readings. At the start of the titration and in any subsequent regions (inflections) where 0.1 mL
of titrant consistently produces a total change of more than 0.03 V (corresponding to 0.5 pH scale unit) in the cell potential, add
0.05 mL portions. In the intermediate regions (plateaus) where 0.1 mL increments change the potential by less than 0.03 V, add
large portions sufficient to produce a total potential change approximately equal to, but not greater than, 0.03 V. Titrate in this
manner until the potential changes less than 0.005 V (corresponding to 0.1 pH scale unit) per 0.1 mL.
NOTE 12—Consider the cell potential constant when it changes less than 0.005 V ⁄min.
11.6.2 Automatic Recording Titration—Adjust the instrument in accordance with the manufacturer’s instructions and set the
titration speed at 1.0 mL ⁄min maximum.
D2896 − 21
11.7 On completion of the titration, remove the beaker and rinse the electrodes and buret tip with titration solvent, then with water,
then again with titration solvent (see 10.3). Store in water when not in use (see 10.1).
11.8 For each set of samples make a blank titration using 120 mL of titration solvent. For a manual titration add 0.1 N HClO
solution in 0.05 mL increments, waiting between each addition until a constant cell potential is established. Record meter and buret
readings after each increment. Follow the procedure in 11.6.2 for an automatic titration.
12. Procedure B (60 mL)
12.1 Calculate the quantity of sample required from its expected base number as follows:
Approximate weight of sample, g5 10/expected BN (5)
NOTE 13—For the back titration procedure (see 17.2) it may be necessary to use a smaller sample weight.
12.1.1 Weigh the sample into the titration beaker, applying the limits shown as follows. A maximum of 10 g should be taken for
analysis.
Precision of
Sample Weight, g Weighing, g
5 to 10 0.02
1 to 5
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...