Standard Test Method for Measurement of Hindered Phenolic and Aromatic Amine Antioxidant Content in Non-zinc Turbine Oils by Linear Sweep Voltammetry

SIGNIFICANCE AND USE
4.1 The quantitative determination of hindered phenol and aromatic amine antioxidants in a new turbine oil measures the amount of these compounds that has been added to the oil as protection against oxidation. Beside phenols, turbine oils can be formulated with other antioxidants such as amines which can extend the oil life. In in-service oil, the determination measures the amount of original (hindered phenol and aromatic amine) antioxidants remaining after oxidation has reduced its initial concentration. This test method is not designed or intended to detect all of the antioxidant intermediates formed during the thermal and oxidative stressing of the oils, which are recognized as having some contribution to the remaining useful life of the in-service oil. Nor does it measure the overall stability of an oil, which is determined by the total contribution of all species present. Before making final judgment on the remaining useful life of the in-service oil, which might result in the replacement of the oil reservoir, it is advised to perform additional analytical techniques (as in accordance with Test Methods D6224 and D4378; see also Test Method D2272), having the capability of measuring remaining oxidative life of the in-service oil.  
4.1.1 This test method is applicable to non-zinc type of turbine oils as defined by ISO 6743 Part 4, Table 1. These are refined mineral oils containing rust and oxidation inhibitors, but not antiwear additives.  
4.2 The test is also suitable for manufacturing control and specification acceptance.  
4.3 When a voltammetric analysis is obtained for a turbine oil inhibited with a typical synergistic mixture of hindered phenol and aromatic amine antioxidants, there is an increase in the current of the produced voltammogram between 8 to 12 s (or 0.8 to 1.2 V applied voltage) (see Note 1) for the aromatic amines, and an increase in the current of the produced voltammogram between 13 and 16 s (or 1.3 to 1.6 V applied voltage) (see Note 1...
SCOPE
1.1 This test method covers the voltammetric determination of hindered phenol and aromatic amine antioxidants in new or in-service type non-zinc turbine oils in concentrations from 0.0075 mass % up to concentrations found in new oils by measuring the amount of current flow at a specified voltage in the produced voltammogram.  
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
30-Apr-2014
Current Stage
Ref Project

Buy Standard

Standard
ASTM D6971-09(2014) - Standard Test Method for Measurement of Hindered Phenolic and Aromatic Amine Antioxidant Content in Non-zinc Turbine Oils by Linear Sweep Voltammetry
English language
6 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM D6971-09(2014) - Standard Test Method for Measurement of Hindered Phenolic and Aromatic Amine Antioxidant Content in Non-zinc Turbine Oils by Linear Sweep Voltammetry
English language
6 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: D6971 − 09 (Reapproved 2014)
Standard Test Method for
Measurement of Hindered Phenolic and Aromatic Amine
Antioxidant Content in Non-zinc Turbine Oils by Linear
Sweep Voltammetry
This standard is issued under the fixed designation D6971; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 3. Summary of Test Method
1.1 This test method covers the voltammetric determination
3.1 A measured quantity of sample is dispensed into a vial
of hindered phenol and aromatic amine antioxidants in new or
containing a measured quantity of acetone based electrolyte
in-service type non-zinc turbine oils in concentrations from
test solution and a layer of sand. When the vial is shaken, the
0.0075 mass % up to concentrations found in new oils by
hinderedphenolandaromaticamineantioxidantsandothertest
measuring the amount of current flow at a specified voltage in
solution soluble oil components present in the sample are
the produced voltammogram.
extracted into the test solution and the remaining droplets
1.2 This standard does not purport to address all of the
suspended in the test solution are agglomerated by the sand.
safety concerns, if any, associated with its use. It is the The sand/droplet suspension is allowed to settle out and the
responsibility of the user of this standard to establish appro-
hindered phenol and aromatic amine antioxidants dissolved in
priate safety and health practices and determine the applica-
the test solution are quantified by voltammetric analysis. The
bility of regulatory limitations prior to use.
results are calculated and reported as mass % of antioxidant or
as millimoles (mmol) of antioxidant per litre of sample for
2. Referenced Documents
prepared and fresh oils and as a percent remaining antioxidant
2.1 ASTM Standards:
for in-service oils.
D1193 Specification for Reagent Water
3.2 Voltammetric analysis is a technique that applies
D2272 Test Method for Oxidation Stability of Steam Tur-
electro-analytic methods wherein a sample to be analyzed is
bine Oils by Rotating Pressure Vessel
mixedwithanelectrolyteandatestsolution,andplacedwithin
D4057 Practice for Manual Sampling of Petroleum and
an electrolytic cell. Data is obtained by measuring the current
Petroleum Products
passing through the cell as a function of the potential applied,
D4378 Practice for In-Service Monitoring of Mineral Tur-
and test results are based upon current, voltage, and time
bine Oils for Steam, Gas, and Combined Cycle Turbines
relationships at the cell electrodes. The cell consists of a fluid
D6224 PracticeforIn-ServiceMonitoringofLubricatingOil
container into which is mounted a small, easily polarized,
for Auxiliary Power Plant Equipment
working electrode, and a large, non-polarizable, reference
D6810 Test Method for Measurement of Hindered Phenolic
electrode. The reference electrode should be massive relative
Antioxidant Content in Non-Zinc Turbine Oils by Linear
totheworkingelectrodesothatitsbehaviorremainsessentially
Sweep Voltammetry
constant with the passage of small current; that is, it remains
2.2 ISO Standards:
unpolarized during the analysis period. Additional electrodes,
ISO 6743 Part 4, Lubricants, Industrial Oils, and Related
such as auxiliary electrodes, can be added to the electrode
Products
system to eliminate the effects of resistive drop for high
resistance test solutions. In performing a voltammetric
This test method is under the jurisdiction of ASTM Committee D02 on
analysis, the potential across the electrodes is varied linearly
Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of
Subcommittee D02.09.0C on Oxidation of Turbine Oils. with time, and the resulting current is recorded as a function of
CurrenteditionapprovedMay1,2014.PublishedJuly2014.Originallyapproved
the potential. As the increasing voltage is applied to the
in 2004. Last previous edition approved in 2009 as D6971 – 09. DOI: 10.1520/
prepared sample within the cell, the various additive species
D6971-09R14.
under investigation within the oil are caused to electrochemi-
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
cally oxidize. The data recorded during this oxidation reaction
Standards volume information, refer to the standard’s Document Summary page on
can then be used to determine the remaining useful life of the
the ASTM website.
3 oil type. A typical current-potential curve produced during the
Available fromAmerican National Standards Institute (ANSI), 25 W. 43rd St.,
4th Floor, New York, NY 10036, http://www.ansi.org. practice of the voltammetric test can be seen by reference to
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D6971 − 09 (2014)
NOTE 1—x-axis = time (seconds) and y-axis is current (arbitrary units). Top line in Fig. 1 is voltammogram of a fresh R&O turbine oil showing valley
indicators before and after antioxidant valleys.
FIG. 1 Aromatic Amine and Hindered Phenol Voltammetric Response in the Neutral Test Solution with Blank Response Zeroed
Fig. 1. Initially the applied potential produces an electrochemi- 4.1.1 This test method is applicable to non-zinc type of
calreactionhavingaratesoslowthatvirtuallynocurrentflows turbine oils as defined by ISO 6743 Part 4, Table 1. These are
through the cell. As the voltage is increased, as shown in Fig. refined mineral oils containing rust and oxidation inhibitors,
1, the electro-active species (for example, substituted phenols) but not antiwear additives.
begin to oxidize at the working electrode surface, producing an
4.2 The test is also suitable for manufacturing control and
anodic rise in the current.As the potential is further increased,
specification acceptance.
the decrease in the electro-active species concentration at the
4.3 When a voltammetric analysis is obtained for a turbine
electrode surface and the exponential increase of the oxidation
oil inhibited with a typical synergistic mixture of hindered
rate lead to a maximum in the current-potential curve shown in
phenol and aromatic amine antioxidants, there is an increase in
Fig. 1.
the current of the produced voltammogram between 8 to 12 s
(or 0.8 to 1.2 V applied voltage) (see Note 1) for the aromatic
4. Significance and Use
amines, and an increase in the current of the produced
4.1 The quantitative determination of hindered phenol and
voltammogram between 13 and 16 s (or 1.3 to 1.6 V applied
aromatic amine antioxidants in a new turbine oil measures the
voltage) (see Note 1) for the hindered phenols in the neutral
amount of these compounds that has been added to the oil as
acetone test solution (Fig. 1: x-axis1s=0.1V). Hindered
protection against oxidation. Beside phenols, turbine oils can
phenol antioxidants detected by voltammetric analysis include,
be formulated with other antioxidants such as amines which
but are not limited to, 2,6-di-tert-butyl-4-methylphenol; 2,6-di-
can extend the oil life. In in-service oil, the determination
tert-butylphenol; and 4,4’-Methylenebis (2,6-di-tert-
measurestheamountoforiginal(hinderedphenolandaromatic
butylphenol). Aromatic amine antioxidants detected by volta-
amine) antioxidants remaining after oxidation has reduced its
mmetric analysis include, but are not limited to, phenyl alpha
initial concentration. This test method is not designed or
naphthylamines, and alkylated diphenylamines.
intended to detect all of the antioxidant intermediates formed
NOTE 1—Voltages listed with respect to reference electrode. The
duringthethermalandoxidativestressingoftheoils,whichare
voltammograms shown in Figs. 1 and 2 were obtained with a platinum
recognizedashavingsomecontributiontotheremaininguseful
reference electrode and a voltage scan rate of 0.1 V/s.
life of the in-service oil. Nor does it measure the overall
4.4 For turbine oil containing only aromatic amines as
stabilityofanoil,whichisdeterminedbythetotalcontribution
antioxidants, there will only be an increase in the current of the
of all species present. Before making final judgment on the
produced voltammogram between 8 to 12 seconds (or 0.8 to
remainingusefullifeofthein-serviceoil,whichmightresultin
1.2Vapplied voltage) (see Note 1) for the aromatic amines, by
the replacement of the oil reservoir, it is advised to perform
using the neutral acetone test solution (first peak in Fig. 1).
additional analytical techniques (as in accordance with Test
Methods D6224 and D4378; see also Test Method D2272), 4.5 For turbine oils containing only hindered phenolic
having the capability of measuring remaining oxidative life of antioxidants, it is preferable to use a basic alcohol test solution
the in-service oil. rather than the neutral acetone test solutions, as there is an
D6971 − 09 (2014)
NOTE 1—x-axis = time (seconds) and y-axis is current (arbitrary units) with top line in Fig. 2 showing the fresh oil.
FIG. 2 Hindered Phenol Voltammetric Response in Basic Test Solution with Blank Response Zeroed
increaseinthecurrentoftheproducedvoltammogrambetween 6. Reagents
3 to 6 seconds (or 0.3 to 0.6 V applied voltage) (see Note 1)in
6.1 Purity of Reagents—Reagent-grade chemicals shall be
basic alcohol test solution (Fig. 2: x-axis 1 second = 0.1 V) in
used in all tests. Unless otherwise indicated, it is intended that
accordance with Test Method D6810.
all reagents shall conform to the specifications of the Commit-
tee onAnalytical Reagents of theAmerican Chemical Society,
5. Apparatus
where such specifications are available. Other grades may be
used, provided it is first ascertained that the reagent’s purity
5.1 Voltammetric Analyzer—The instrument used to quan-
suffices to permit its use without lessening the accuracy of the
tify the hindered phenol and aromatic amine antioxidants is a
determination.
voltammograph equipped with a three-electrode system and a
digital or analog output. The combination electrode system
6.2 Purity of Water—Unless otherwise specified, references
consists of a glassy carbon disc (3 mm diameter) working
to water that conforms to Specification D1193, Type II.
electrode, a platinum wire (0.5 mm diameter) auxiliary
6.3 Analysis Materials:
electrode, and a platinum wire (0.5 mm diameter) reference
6.3.1 Acetone Test Solution (Neutral)—Proprietary Green
electrode, as described inTest Method D6810.The voltammet-
Test Solution, Acetone test solution (1:10 distilled water/
ric analyzer applies a linear voltage ramp (0 to -1.8 V range
acetone test solution) containing a dissolved neutral electro-
with respect to the reference electrode) at a rate of 0.01 to
lyte. (Warning—Corrosive, poisonous, flammable, and a skin
0.5 V⁄s (0.1 optimum) to the auxiliary electrode. The current
irritant. Harmful if inhaled.)
output of the working electrode is converted to voltage by the
6.3.2 Alcohol Test Solution (Basic)—Proprietary Yellow
voltammetric analyzer, using the gain ratio of 1 V/20 µA, and
Test Solution, Ethanol test solution (1:10 distilled water/
is outputted to an analog or digital recording device (0 to 1 V
ethanol test solution) containing a dissolved base electrolyte.
full scale) as shown in Figs. 1 and 2.
(Warning—Corrosive, poisonous, flammable, and a skin irri-
5.2 Vortex Mixer, with a 2800 to 3000 rpm motor and a pad
tant. Harmful if inhaled.)
suitable for mixing test tubes and vials.
6.3.3 Alcohol Cleansing Pads—70 % isopropyl alcohol
saturatedcleansingpads(alcoholpreparedskincleansingpads,
5.3 Pipette, or equivalent, capable of delivering sample
for the preparation of the skin prior to injection (antiseptic)).
volumes required in the test method, from 0.10 to 0.50 mL.
5.4 Test Solution Dispenser, or equivalent, capable of deliv-
ering volumes of analysis test solution (see 6.3) required in the
Reagent Chemicals, American Chemical Society Specifications, American
test method, such as 3.0 and 5.0 mL.
Chemical Society, Washington, DC. For Suggestions on the testing of reagents not
listed by the American Chemical Society, see Annual Standards for Laboratory
5.5 Glass Vials, with caps, 4 or 7 mL capacity and contain-
Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia
ing1gof sand. White quartz suitable for chromatography,
and National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville,
within the size range of 200 to 300 µm 6 100 µm. MD.
D6971 − 09 (2014)
NOTE 1—Standard (top line) and sample in-service oil (lower line).
FIG. 3 Voltammetric Reading for a In-service Oil Sample Comparing Hindered Phenols and Aromatic Amines Peaks (in the Neutral Test
Solution)
7. Sampling measurement gives you a voltammetric reading (standard
reading) that indicates the voltammetric response for the
7.1 Obtain the sample in accordance with Practice D4057.
concentration hindered phenol and aromatic amines antioxi-
dants being analyzed for the oil being tested.
8. Procedure
8.2.3 Sample (In-service Oil) Reading.
8.1 The voltammetric analyzer used in this test method
8.2.3.1 The sample reading is a measurement of a fresh or
gives linear results between 2 to 50 mmol for hindered phenols
in-serviceoilmixedwiththesametypeofanalysistestsolution
and aromatic amines using an oil sample size of 0.40 mL and
as the standard. This measurement will provide voltammetric
5.0 mL of the analysis test solution. The corresponding range
readings that normally range between the blank and standard
of mass % depends on the molecular weight of the hindered
measurements,andreflecttheconcentrationofhinderedphenol
phenol and aromatic amine, and the density of the base oil. For
andaromaticamineantioxidantpresent(freshoil)orremaining
instance, the mass % range of 0.044 to 1.1 is equal to 2 to
(in-service oil) in the oil sample. Voltammetric readings for
50 mmol⁄L for a hindered phenol containing one hydroxyl
in-service oils will decrease as hindered phenol and aromatic
group and with a molecular weight of 220 g/mol (2,6-di-tert-
amine antioxidants are depleted.
butyl-4-methylphenol) and an oil density of 1 g/mL. Below 2
8.3 Voltammetric Reading—After the operator has selected
mmol, the noise to signal ratio becomes large, decreasing the
the valleys before and after the antioxidant peaks (as shown in
accuracy of the measurements. For measurements below 2
Fig. 1), the software (R-DMS ) will automatically identify and
mmol or for fresh oils with hig
...


This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: D6971 − 09 D6971 − 09 (Reapproved 2014)
Standard Test Method for
Measurement of Hindered Phenolic and Aromatic Amine
Antioxidant Content in Non-zinc Turbine Oils by Linear
Sweep Voltammetry
This standard is issued under the fixed designation D6971; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This test method covers the voltammetric determination of hindered phenol and aromatic amine antioxidants in new or
in-service type non-zinc turbine oils in concentrations from 0.0075 mass % up to concentrations found in new oils by measuring
the amount of current flow at a specified voltage in the produced voltammogram.
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Referenced Documents
2.1 ASTM Standards:
D1193 Specification for Reagent Water
D2272 Test Method for Oxidation Stability of Steam Turbine Oils by Rotating Pressure Vessel
D4057 Practice for Manual Sampling of Petroleum and Petroleum Products
D4378 Practice for In-Service Monitoring of Mineral Turbine Oils for Steam, Gas, and Combined Cycle Turbines
D6224 Practice for In-Service Monitoring of Lubricating Oil for Auxiliary Power Plant Equipment
D6810 Test Method for Measurement of Hindered Phenolic Antioxidant Content in Non-Zinc Turbine Oils by Linear Sweep
Voltammetry
2.2 ISO Standards:
ISO 6743 Part 4, Lubricants, Industrial Oils, and Related Products
3. Summary of Test Method
3.1 A measured quantity of sample is dispensed into a vial containing a measured quantity of acetone based electrolyte test
solution and a layer of sand. When the vial is shaken, the hindered phenol and aromatic amine antioxidants and other test solution
soluble oil components present in the sample are extracted into the test solution and the remaining droplets suspended in the test
solution are agglomerated by the sand. The sand/droplet suspension is allowed to settle out and the hindered phenol and aromatic
amine antioxidants dissolved in the test solution are quantified by voltammetric analysis. The results are calculated and reported
as mass % of antioxidant or as millimoles (mmol) of antioxidant per litre of sample for prepared and fresh oils and as a percent
remaining antioxidant for in-service oils.
3.2 Voltammetric analysis is a technique that applies electro-analytic methods wherein a sample to be analyzed is mixed with
an electrolyte and a test solution, and placed within an electrolytic cell. Data is obtained by measuring the current passing through
the cell as a function of the potential applied, and test results are based upon current, voltage, and time relationships at the cell
electrodes. The cell consists of a fluid container into which is mounted a small, easily polarized, working electrode, and a large,
non-polarizable, reference electrode. The reference electrode should be massive relative to the working electrode so that its
behavior remains essentially constant with the passage of small current; that is, it remains unpolarized during the analysis period.
This test method is under the jurisdiction of ASTM Committee D02 on Petroleum Products Products, Liquid Fuels, and Lubricants and is the direct responsibility of
Subcommittee D02.09.0C on Oxidation of Turbine Oils.
Current edition approved June 1, 2009May 1, 2014. Published June 2009July 2014. Originally approved in 2004. Last previous edition approved in 20042009 as
D6971D6971 – 09.–04. DOI: 10.1520/D6971-09.10.1520/D6971-09R14.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Available from American National Standards Institute (ANSI), 25 W. 43rd St., 4th Floor, New York, NY 10036, http://www.ansi.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D6971 − 09 (2014)
NOTE 1—x-axis = time (seconds) and y-axis is current (arbitrary units). Top line in Fig. 1 is voltammogram of a fresh R&O turbine oil showing valley
indicators before and after antioxidant valleys.
FIG. 1 Aromatic Amine and Hindered Phenol Voltammetric Response in the Neutral Test Solution with Blank Response Zeroed
Additional electrodes, such as auxiliary electrodes, can be added to the electrode system to eliminate the effects of resistive drop
for high resistance test solutions. In performing a voltammetric analysis, the potential across the electrodes is varied linearly with
time, and the resulting current is recorded as a function of the potential. As the increasing voltage is applied to the prepared sample
within the cell, the various additive species under investigation within the oil are caused to electrochemically oxidize. The data
recorded during this oxidation reaction can then be used to determine the remaining useful life of the oil type. A typical
current-potential curve produced during the practice of the voltammetric test can be seen by reference to Fig. 1. Initially the applied
potential produces an electrochemical reaction having a rate so slow that virtually no current flows through the cell. As the voltage
is increased, as shown in Fig. 1, the electro-active species (for example, substituted phenols) begin to oxidize at the working
electrode surface, producing an anodic rise in the current. As the potential is further increased, the decrease in the electro-active
species concentration at the electrode surface and the exponential increase of the oxidation rate lead to a maximum in the
current-potential curve shown in Fig. 1.
4. Significance and Use
4.1 The quantitative determination of hindered phenol and aromatic amine antioxidants in a new turbine oil measures the
amount of these compounds that has been added to the oil as protection against oxidation. Beside phenols, turbine oils can be
formulated with other antioxidants such as amines which can extend the oil life. In in-service oil, the determination measures the
amount of original (hindered phenol and aromatic amine) antioxidants remaining after oxidation has reduced its initial
concentration. This test method is not designed or intended to detect all of the antioxidant intermediates formed during the thermal
and oxidative stressing of the oils, which are recognized as having some contribution to the remaining useful life of the in-service
oil. Nor does it measure the overall stability of an oil, which is determined by the total contribution of all species present. Before
making final judgment on the remaining useful life of the in-service oil, which might result in the replacement of the oil reservoir,
it is advised to perform additional analytical techniques (as in accordance with Test Methods D6224 and D4378; see also Test
Method D2272), having the capability of measuring remaining oxidative life of the in-service oil.
4.1.1 This test method is applicable to non-zinc type of turbine oils as defined by ISO 6743 Part 4, Table 1. These are refined
mineral oils containing rust and oxidation inhibitors, but not antiwear additives.
4.2 The test is also suitable for manufacturing control and specification acceptance.
4.3 When a voltammetric analysis is obtained for a turbine oil inhibited with a typical synergistic mixture of hindered phenol
and aromatic amine antioxidants, there is an increase in the current of the produced voltammogram between 8 to 12 s (or 0.8 to
1.2 V applied voltage) (see Note 1) for the aromatic amines, and an increase in the current of the produced voltammogram between
13 and 16 s (or 1.3 to 1.6 V applied voltage) (see Note 1) for the hindered phenols in the neutral acetone test solution (Fig. 1: x-axis
1 s = 0.1 V). Hindered phenol antioxidants detected by voltammetric analysis include, but are not limited to, 2,6-di-tert-butyl-
4-methylphenol; 2,6-di-tert-butylphenol; and 4,4’-Methylenebis (2,6-di-tert-butylphenol). Aromatic amine antioxidants detected
by voltammetric analysis include, but are not limited to, phenyl alpha naphthylamines, and alkylated diphenylamines.
D6971 − 09 (2014)
NOTE 1—x-axis = time (seconds) and y-axis is current (arbitrary units) with top line in Fig. 2 showing the fresh oil.
FIG. 2 Hindered Phenol Voltammetric Response in Basic Test Solution with Blank Response Zeroed
NOTE 1—Voltages listed with respect to reference electrode. The voltammograms shown in Figs. 1 and 2 were obtained with a platinum reference
electrode and a voltage scan rate of 0.1 V/s.
4.4 For turbine oil containing only aromatic amines as antioxidants, there will only be an increase in the current of the produced
voltammogram between 8 to 12 seconds (or 0.8 to 1.2 V applied voltage) (see Note 1) for the aromatic amines, by using the neutral
acetone test solution (first peak in Fig. 1).
4.5 For turbine oils containing only hindered phenolic antioxidants, it is preferable to use a basic alcohol test solution rather
than the neutral acetone test solutions, as there is an increase in the current of the produced voltammogram between 3 to 6 seconds
(or 0.3 to 0.6 V applied voltage) (see Note 1) in basic alcohol test solution (Fig. 2: x-axis 1 second = 0.1 V) in accordance with
Test Method D6810.
5. Apparatus
5.1 Voltammetric Analyzer—The instrument used to quantify the hindered phenol and aromatic amine antioxidants is a
voltammograph equipped with a three-electrode system and a digital or analog output. The combination electrode system consists
of a glassy carbon disc (3 mm diameter) working electrode, a platinum wire (0.5 mm diameter) auxiliary electrode, and a platinum
wire (0.5 mm diameter) reference electrode, as described in Test Method D6810. The voltammetric analyzer applies a linear
voltage ramp (0 to -1.8 V range with respect to the reference electrode) at a rate of 0.01 to 0.50.5 V V/s ⁄s (0.1 optimum) to the
auxiliary electrode. The current output of the working electrode is converted to voltage by the voltammetric analyzer, using the
gain ratio of 1 V/20 μA, and is outputted to an analog or digital recording device (0 to 1 V full scale) as shown in Figs. 1 and 2.
5.2 Vortex Mixer, with a 2800 to 3000 rpm motor and a pad suitable for mixing test tubes and vials.
5.3 Pipette, or equivalent, capable of delivering sample volumes required in the test method, from 0.10 to 0.50 mL.
5.4 Test Solution Dispenser, or equivalent, capable of delivering volumes of analysis test solution (see 6.3) required in the test
method, such as 3.0 and 5.0 mL.
5.5 Glass Vials, with caps, 4 or 7 mL capacity and containing 1 g of sand. White quartz suitable for chromatography, within
the size range of 200 to 300 μm 6 100 μm.
6. Reagents
6.1 Purity of Reagents—Reagent-grade chemicals shall be used in all tests. Unless otherwise indicated, it is intended that all
reagents shall conform to the specifications of the Committee on Analytical Reagents of the American Chemical Society, where
D6971 − 09 (2014)
such specifications are available. Other grades may be used, provided it is first ascertained that the reagent’s purity suffices to
permit its use without lessening the accuracy of the determination.
6.2 Purity of Water—Unless otherwise specified, references to water that conforms to Specification D1193, Type II.
6.3 Analysis Materials:
6.3.1 Acetone Test Solution (Neutral)—Proprietary Green Test Solution, Acetone test solution (1:10 distilled water/acetone test
solution) containing a dissolved neutral electrolyte. (Warning—Corrosive, poisonous, flammable, and a skin irritant. Harmful if
inhaled.)
6.3.2 Alcohol Test Solution (Basic)—Proprietary Yellow Test Solution, Ethanol test solution (1:10 distilled water/ethanol test
solution) containing a dissolved base electrolyte. (Warning—Corrosive, poisonous, flammable, and a skin irritant. Harmful if
inhaled.)
6.3.3 Alcohol Cleansing Pads—70 % isopropyl alcohol saturated cleansing pads (alcohol prepared skin cleansing pads, for the
preparation of the skin prior to injection (antiseptic)).
7. Sampling
7.1 Obtain the sample in accordance with Practice D4057.
8. Procedure
8.1 The voltammetric analyzer used in this test method gives linear results between 2 to 50 mmol 50 mmol for hindered phenols
and aromatic amines using an oil sample size of 0.40 mL and 5.0 mL of the analysis test solution. The corresponding range of mass
% depends on the molecular weight of the hindered phenol and aromatic amine, and the density of the base oil. For instance, the
mass % range of 0.044 to 1.1 is equal to 2 to 5050 mmol mmol/L ⁄L for a hindered phenol containing one hydroxyl group and
with a molecular weight of 220 g/mol (2,6-di-tert-butyl-4-methylphenol) and an oil density of 1 g/mL. Below 2 mmol, the noise
to signal ratio becomes large, decreasing the accuracy of the measurements. For measurements below 2 mmol or for fresh oils with
high noise to signal ratios, the sample size should be increased to 0.60 mL and the volume of analysis test solution decreased to
3.0 mL.
8.2 General Voltammetric Test Procedure—The test procedure for voltammetric analysis will consist of the blank reading
(calibration), followed by a standard reading, and finally the sample (in-service oil) reading.
8.2.1 Blank Reading—(0 mmol/L = 0 mass %).
8.2.1.1 The blank reading (voltammetric number) is a measurement of the analysis test solution by itself. The blank
measurement gives a reference number with no antioxidant present (the zero baseline).
8.2.2 Standard Reading—(30 to 150 mmol/L—mass % dependent on density of fresh oil and molecular weight of antioxidant).
8.2.2.1 The standard reading is a measurement of a fresh, unused oil (containing hindered phenol and amines antioxidants)
mixed with an appropriate analysis test solution. This measurement gives you a voltammetric reading (standard reading) that
indicates the voltammetric response for the concentration hindered phenol and aromatic amines antioxidants
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.