ASTM F1983-23
(Practice)Standard Practice for Assessment of Selected Tissue Effects of Absorbable Biomaterials for Implant Applications
Standard Practice for Assessment of Selected Tissue Effects of Absorbable Biomaterials for Implant Applications
SIGNIFICANCE AND USE
5.1 This practice is a guideline for a screening test of candidate materials or assessment of local tissue response to absorbable medical devices which are expected to undergo complete absorption within three years.
5.2 This practice is similar to those for studies on candidate materials or medical devices that are not absorbable, such as those specified in Practices F763, F981, and F1408; however, analysis of the host response must take into account the effect of degradation and degradation products on the inflammatory response at the local tissue site and on subsequent healing of the implantation site, as well as the potential for adverse distal tissue effects.
5.3 For testing of absorbable medical devices, the test article for implantation should be in the final finished form as for intended use, including packaging and sterilization (if applicable). Configurations specific to the animal study may be needed. The test article’s surface-area-to-body mass or mass-to-body mass ratios within the animal model should be established by calculating based on surface-area-to-body mass or mass-to-body mass ratios in humans during the device’s intended clinical use. Worst-case clinical dose should be considered in the study design. For implantation studies incorporating evaluation of both local tissue responses and systemic toxicity, exaggerated material surface area or mass-to-body mass ratios (for example, a 2X to 10X safety factor to assess implant safety for regulatory submissions) compared to clinical use (for example, largest device size, maximum number of devices) should be considered, unless otherwise justified. For example, implantation of exaggerated doses may not be feasible in the selected animal model. For some devices, additional animal group(s) for exaggerated conditions should be considered if dose response information is needed. Additionally, for some devices, exaggerated dose at a specific implantation site can also be used to evaluate local tissue res...
SCOPE
1.1 This practice provides experimental protocols for biological assays of tissue reactions to absorbable biomaterials for implant applications. This practice applies only to absorbable materials with projected clinical applications in which the materials will reside in bone or soft tissue longer than 30 days and less than three years. Other standards with designated implantation times are available to address shorter time periods. Careful consideration should be given to the appropriateness of this practice for slowly degrading materials that will remain for longer than three years. It is anticipated that the tissue response to degrading biomaterials will be different from the response to nonabsorbable materials. In many cases, a chronic inflammatory response may be observed during the degradation phase, but the local histology should return to normal after absorption; therefore, the minimal tissue response usually equated with biocompatibility may require long implantations.
1.2 The time period for implant absorption can depend on variables of chemical composition, implant size, implant location, and animal models. Therefore, the selected time points for assessing tissue effects may be selected based on the rate of absorption.
1.3 These protocols assess the effects of the material on the animal tissue in which it is implanted. They do not fully assess systemic toxicity, carcinogenicity, reproductive and development toxicity, or mutagenicity of the material. Other standards are available to address these issues.
1.4 To maximize use of the animals in the study protocol, some aspects of systemic toxicity, including effects of degradation products on different organs and tissues downstream of or surrounding the target site, can be addressed with this practice.
1.5 Because animal models are not identical to human biology, this practice cannot account for all potential biological hazards, for example the effect o...
General Information
- Status
- Published
- Publication Date
- 31-Mar-2023
- Technical Committee
- F04 - Medical and Surgical Materials and Devices
- Drafting Committee
- F04.16 - Biocompatibility Test Methods
Relations
- Effective Date
- 01-Aug-2020
- Effective Date
- 01-Jun-2020
- Effective Date
- 01-Jan-2019
- Effective Date
- 01-Oct-2018
- Effective Date
- 01-Apr-2018
- Effective Date
- 01-Dec-2016
- Effective Date
- 01-Dec-2016
- Effective Date
- 01-Dec-2016
- Effective Date
- 01-Oct-2013
- Effective Date
- 01-Sep-2013
- Effective Date
- 01-Dec-2012
- Effective Date
- 01-Mar-2011
- Effective Date
- 01-Sep-2010
- Refers
ASTM F1903-10 - Standard Practice for Testing For Biological Responses to Particles <i>In Vitro</i> - Effective Date
- 01-Jun-2010
- Effective Date
- 01-Jun-2010
Overview
ASTM F1983-23: Standard Practice for Assessment of Selected Tissue Effects of Absorbable Biomaterials for Implant Applications provides protocols for evaluating the biological tissue reactions to absorbable materials intended for long-term implantation. This ASTM standard is crucial for stakeholders in medical device development, regulatory affairs, and research as it outlines experimental approaches to assess local tissue response, degradation, and healing following implantation. The primary focus is on absorbable biomaterials with intended clinical residency in bone or soft tissue for periods ranging from over 30 days to less than three years.
Key Topics
- Absorbable Biomaterials: Materials expected to fully degrade in the body within a specified timeframe.
- Implantation Protocols: Detailed guidance on animal model selection, implant site choice, and handling of final finished form devices, including packaging and sterilization.
- Assessment Points: Recommendations on time points for evaluating tissue response, based on absorption rates influenced by the biomaterial's composition, size, and implantation site.
- Local and Systemic Effects: Emphasis on analyzing both localized tissue responses and, when appropriate, systemic effects from degradation products.
- Histological Evaluation: Standardized criteria for assessing acute and chronic inflammation, granulation tissue, necrosis, and foreign body reactions across multiple stages of material degradation.
- Controls and Study Design: Guidance on the selection of control materials, including requirements for sham surgeries and matched material-to-host ratios to ensure study validity.
Applications
ASTM F1983-23 is instrumental for:
- Medical Device Manufacturers: Assisting in the preclinical evaluation and safety assessment of absorbable implants such as sutures, meshes, and orthopedic devices.
- Regulatory Submissions: Supporting data packages for FDA and international regulatory bodies with validated protocols for animal testing of absorbable materials.
- Research and Development: Guiding researchers in designing animal studies that accurately represent clinical conditions, material absorption profiles, and tissue compatibility.
- Quality Assurance and Compliance: Ensuring consistent, reproducible methodologies for testing biocompatibility, reducing the risk of adverse tissue responses and improving patient safety.
- Comparative Studies: Enabling side-by-side assessments of different absorbable biomaterials or comparison with nonabsorbable standards to optimize material selection.
Related Standards
ASTM F1983-23 references and aligns with a range of international and ASTM standards for comprehensive biological evaluation:
- ASTM F763 - Short-Term Intramuscular Screening of Implantable Medical Device Materials
- ASTM F981 - Assessment of Compatibility of Biomaterials for Surgical Implants
- ASTM F1408 - Subcutaneous Screening Test for Implant Materials
- ISO 10993 series (notably Parts 6, 9, 11, 18, and 19) - Biological evaluation of medical devices including tests for local effects after implantation, and chemical characterization procedures
- ASTM F1635, F1903, F1904 - Degradation and biological response testing methods
- ISO/TS 17137 and ISO/TS 37137-1 - Requirements and guidance for absorbable device evaluation
Practical Value
Adhering to ASTM F1983-23 ensures manufacturers and researchers apply validated, internationally recognized procedures for assessing the safety and biocompatibility of absorbable implants. This standard promotes:
- Accurate Prediction of Tissue Response: Accounts for unique biological dynamics of degrading biomaterials.
- Risk Mitigation: Early detection of adverse local or systemic effects supports device safety and performance optimization.
- Regulatory Confidence: Using recognized procedures streamlines regulatory approval processes for new implantable medical devices.
Keywords: absorbable biomaterials, tissue response, implant applications, ASTM F1983-23, biocompatibility testing, medical device standards, degradation products, animal model studies
Buy Documents
ASTM F1983-23 - Standard Practice for Assessment of Selected Tissue Effects of Absorbable Biomaterials for Implant Applications
REDLINE ASTM F1983-23 - Standard Practice for Assessment of Selected Tissue Effects of Absorbable Biomaterials for Implant Applications
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
ASTM F1983-23 is a standard published by ASTM International. Its full title is "Standard Practice for Assessment of Selected Tissue Effects of Absorbable Biomaterials for Implant Applications". This standard covers: SIGNIFICANCE AND USE 5.1 This practice is a guideline for a screening test of candidate materials or assessment of local tissue response to absorbable medical devices which are expected to undergo complete absorption within three years. 5.2 This practice is similar to those for studies on candidate materials or medical devices that are not absorbable, such as those specified in Practices F763, F981, and F1408; however, analysis of the host response must take into account the effect of degradation and degradation products on the inflammatory response at the local tissue site and on subsequent healing of the implantation site, as well as the potential for adverse distal tissue effects. 5.3 For testing of absorbable medical devices, the test article for implantation should be in the final finished form as for intended use, including packaging and sterilization (if applicable). Configurations specific to the animal study may be needed. The test article’s surface-area-to-body mass or mass-to-body mass ratios within the animal model should be established by calculating based on surface-area-to-body mass or mass-to-body mass ratios in humans during the device’s intended clinical use. Worst-case clinical dose should be considered in the study design. For implantation studies incorporating evaluation of both local tissue responses and systemic toxicity, exaggerated material surface area or mass-to-body mass ratios (for example, a 2X to 10X safety factor to assess implant safety for regulatory submissions) compared to clinical use (for example, largest device size, maximum number of devices) should be considered, unless otherwise justified. For example, implantation of exaggerated doses may not be feasible in the selected animal model. For some devices, additional animal group(s) for exaggerated conditions should be considered if dose response information is needed. Additionally, for some devices, exaggerated dose at a specific implantation site can also be used to evaluate local tissue res... SCOPE 1.1 This practice provides experimental protocols for biological assays of tissue reactions to absorbable biomaterials for implant applications. This practice applies only to absorbable materials with projected clinical applications in which the materials will reside in bone or soft tissue longer than 30 days and less than three years. Other standards with designated implantation times are available to address shorter time periods. Careful consideration should be given to the appropriateness of this practice for slowly degrading materials that will remain for longer than three years. It is anticipated that the tissue response to degrading biomaterials will be different from the response to nonabsorbable materials. In many cases, a chronic inflammatory response may be observed during the degradation phase, but the local histology should return to normal after absorption; therefore, the minimal tissue response usually equated with biocompatibility may require long implantations. 1.2 The time period for implant absorption can depend on variables of chemical composition, implant size, implant location, and animal models. Therefore, the selected time points for assessing tissue effects may be selected based on the rate of absorption. 1.3 These protocols assess the effects of the material on the animal tissue in which it is implanted. They do not fully assess systemic toxicity, carcinogenicity, reproductive and development toxicity, or mutagenicity of the material. Other standards are available to address these issues. 1.4 To maximize use of the animals in the study protocol, some aspects of systemic toxicity, including effects of degradation products on different organs and tissues downstream of or surrounding the target site, can be addressed with this practice. 1.5 Because animal models are not identical to human biology, this practice cannot account for all potential biological hazards, for example the effect o...
SIGNIFICANCE AND USE 5.1 This practice is a guideline for a screening test of candidate materials or assessment of local tissue response to absorbable medical devices which are expected to undergo complete absorption within three years. 5.2 This practice is similar to those for studies on candidate materials or medical devices that are not absorbable, such as those specified in Practices F763, F981, and F1408; however, analysis of the host response must take into account the effect of degradation and degradation products on the inflammatory response at the local tissue site and on subsequent healing of the implantation site, as well as the potential for adverse distal tissue effects. 5.3 For testing of absorbable medical devices, the test article for implantation should be in the final finished form as for intended use, including packaging and sterilization (if applicable). Configurations specific to the animal study may be needed. The test article’s surface-area-to-body mass or mass-to-body mass ratios within the animal model should be established by calculating based on surface-area-to-body mass or mass-to-body mass ratios in humans during the device’s intended clinical use. Worst-case clinical dose should be considered in the study design. For implantation studies incorporating evaluation of both local tissue responses and systemic toxicity, exaggerated material surface area or mass-to-body mass ratios (for example, a 2X to 10X safety factor to assess implant safety for regulatory submissions) compared to clinical use (for example, largest device size, maximum number of devices) should be considered, unless otherwise justified. For example, implantation of exaggerated doses may not be feasible in the selected animal model. For some devices, additional animal group(s) for exaggerated conditions should be considered if dose response information is needed. Additionally, for some devices, exaggerated dose at a specific implantation site can also be used to evaluate local tissue res... SCOPE 1.1 This practice provides experimental protocols for biological assays of tissue reactions to absorbable biomaterials for implant applications. This practice applies only to absorbable materials with projected clinical applications in which the materials will reside in bone or soft tissue longer than 30 days and less than three years. Other standards with designated implantation times are available to address shorter time periods. Careful consideration should be given to the appropriateness of this practice for slowly degrading materials that will remain for longer than three years. It is anticipated that the tissue response to degrading biomaterials will be different from the response to nonabsorbable materials. In many cases, a chronic inflammatory response may be observed during the degradation phase, but the local histology should return to normal after absorption; therefore, the minimal tissue response usually equated with biocompatibility may require long implantations. 1.2 The time period for implant absorption can depend on variables of chemical composition, implant size, implant location, and animal models. Therefore, the selected time points for assessing tissue effects may be selected based on the rate of absorption. 1.3 These protocols assess the effects of the material on the animal tissue in which it is implanted. They do not fully assess systemic toxicity, carcinogenicity, reproductive and development toxicity, or mutagenicity of the material. Other standards are available to address these issues. 1.4 To maximize use of the animals in the study protocol, some aspects of systemic toxicity, including effects of degradation products on different organs and tissues downstream of or surrounding the target site, can be addressed with this practice. 1.5 Because animal models are not identical to human biology, this practice cannot account for all potential biological hazards, for example the effect o...
ASTM F1983-23 is classified under the following ICS (International Classification for Standards) categories: 11.040.40 - Implants for surgery, prosthetics and orthotics. The ICS classification helps identify the subject area and facilitates finding related standards.
ASTM F1983-23 has the following relationships with other standards: It is inter standard links to ASTM F1408-20a, ASTM F1408-20, ASTM F561-19, ASTM F1903-18, ASTM F3268-18, ASTM F2902-16, ASTM F2902-16e1, ASTM F1635-16, ASTM F1408-97(2013), ASTM F561-13, ASTM F2902-12, ASTM F1635-11, ASTM F561-05a(2010), ASTM F1903-10, ASTM F981-04(2010). Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ASTM F1983-23 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: F1983 − 23
Standard Practice for
Assessment of Selected Tissue Effects of Absorbable
Biomaterials for Implant Applications
This standard is issued under the fixed designation F1983; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope hazards, for example the effect of the oligosaccharide a-Gal
(Gala 1,3-Galb1-4GlcNAc-R), known as the “a-Gal” epitope
1.1 This practice provides experimental protocols for bio-
present in xenogeneic materials on humans. See ISO 22442.
logical assays of tissue reactions to absorbable biomaterials for
1.6 This standard does not purport to address all of the
implant applications. This practice applies only to absorbable
safety concerns, if any, associated with its use. It is the
materials with projected clinical applications in which the
responsibility of the user of this standard to establish appro-
materials will reside in bone or soft tissue longer than 30 days
priate safety, health, and environmental practices and deter-
and less than three years. Other standards with designated
mine the applicability of regulatory limitations prior to use.
implantation times are available to address shorter time peri-
1.7 This international standard was developed in accor-
ods. Careful consideration should be given to the appropriate-
dance with internationally recognized principles on standard-
ness of this practice for slowly degrading materials that will
ization established in the Decision on Principles for the
remain for longer than three years. It is anticipated that the
Development of International Standards, Guides and Recom-
tissue response to degrading biomaterials will be different from
mendations issued by the World Trade Organization Technical
the response to nonabsorbable materials. In many cases, a
Barriers to Trade (TBT) Committee.
chronic inflammatory response may be observed during the
degradation phase, but the local histology should return to
2. Referenced Documents
normal after absorption; therefore, the minimal tissue response
usually equated with biocompatibility may require long im-
2.1 ASTM Standards:
plantations.
F561 Practice for Retrieval and Analysis of Medical
Devices, and Associated Tissues and Fluids
1.2 The time period for implant absorption can depend on
F763 Practice for Short-Term Intramuscular Screening of
variables of chemical composition, implant size, implant
Implantable Medical Device Materials
location, and animal models. Therefore, the selected time
F981 Practice for Assessment of Compatibility of Biomate-
points for assessing tissue effects may be selected based on the
rials for Surgical Implants with Respect to Effect of
rate of absorption.
Materials on Muscle and Insertion into Bone
1.3 These protocols assess the effects of the material on the
F1408 Practice for Subcutaneous Screening Test for Implant
animal tissue in which it is implanted. They do not fully assess
Materials
systemic toxicity, carcinogenicity, reproductive and develop-
F1635 Test Method for in vitro Degradation Testing of
ment toxicity, or mutagenicity of the material. Other standards
Hydrolytically Degradable Polymer Resins and Fabricated
are available to address these issues.
Forms for Surgical Implants
1.4 To maximize use of the animals in the study protocol,
F1903 Practice for Testing for Cellular Responses to Par-
some aspects of systemic toxicity, including effects of degra-
ticles in vitro
dation products on different organs and tissues downstream of
F1904 Practice for Testing the Biological Responses to
or surrounding the target site, can be addressed with this
Particles in vivo
practice.
F2902 Guide for Assessment of Absorbable Polymeric Im-
plants
1.5 Because animal models are not identical to human
F3268 Guide for in vitro Degradation Testing of Absorbable
biology, this practice cannot account for all potential biological
Metals
This practice is under the jurisdiction of ASTM Committee F04 on Medical and
Surgical Materials and Devices and is the direct responsibility of Subcommittee
F04.16 on Biocompatibility Test Methods. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved April 1, 2023. Published April 2023. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 1999. Last previous edition approved in 2014 as F1983 – 14. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/F1983-23. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
F1983 − 23
2.2 ISO Standards: 5.2 This practice is similar to those for studies on candidate
ISO 10993-6 Biological evaluation of medical devices—Part materials or medical devices that are not absorbable, such as
6: Tests for local effects after implantation those specified in Practices F763, F981, and F1408; however,
ISO 10993-9 Biological evaluation of medical devices—Part analysis of the host response must take into account the effect
9: Framework for identification and quantification of of degradation and degradation products on the inflammatory
potential degradation products response at the local tissue site and on subsequent healing of
ISO 10993-11 Biological evaluation of medical devices— the implantation site, as well as the potential for adverse distal
Part 11: Tests for systemic toxicity tissue effects.
ISO 10993-18 Biological evaluation of medical devices—
5.3 For testing of absorbable medical devices, the test article
Part 18: Chemical characterization of medical device
for implantation should be in the final finished form as for
materials within a risk management process
intended use, including packaging and sterilization (if appli-
ISO/TS 10993-19 Biological evaluation of medical
cable). Configurations specific to the animal study may be
devices—Part 19: Physico-chemical, morphological and
needed. The test article’s surface-area-to-body mass or mass-
topographical characterization of materials
to-body mass ratios within the animal model should be
ISO 13781 Implants for surgery—Homopolymers, copoly-
established by calculating based on surface-area-to-body mass
mers and blends on poly(lactide)—In vitro degradation
or mass-to-body mass ratios in humans during the device’s
testing
intended clinical use. Worst-case clinical dose should be
ISO/TS 17137 Cardiovascular implants and extracorporeal
considered in the study design. For implantation studies
systems—Cardiovascular absorbable implants
incorporating evaluation of both local tissue responses and
ISO 22442 Medical devices utilizing animal tissues and their
systemic toxicity, exaggerated material surface area or mass-
derivatives
to-body mass ratios (for example, a 2X to 10X safety factor to
ISO/TS 37137-1 Biological evaluation of absorbable medi-
assess implant safety for regulatory submissions) compared to
cal devices—Part 1: General requirements
clinical use (for example, largest device size, maximum num-
ber of devices) should be considered, unless otherwise justi-
3. Terminology
fied. For example, implantation of exaggerated doses may not
3.1 Definitions:
be feasible in the selected animal model. For some devices,
3.1.1 final finished form—a device or device component that
additional animal group(s) for exaggerated conditions should
includes all manufacturing processes for the “to be marketed”
be considered if dose response information is needed.
device including packaging and sterilization, if applicable.
Additionally, for some devices, exaggerated dose at a specific
implantation site can also be used to evaluate local tissue
4. Summary of Practice
responses.
4.1 Under strict aseptic conditions, sterile test articles (for
5.4 Materials that are designed for use in devices with in
example, final device) are implanted into a relevant animal
situ polymerization shall be introduced in a manner such that in
model and at a clinically relevant anatomical tissue site.
situ polymerization occurs. Additional testing of individual
However, for screening candidate materials, testing in a clini-
precursor components or partially polymerized materials may
cally relevant animal model and anatomical tissue site may not
be needed in some cases (for example, if testing of the final
be necessary. Small laboratory animals such as mice, rats,
implant indicates an adverse response or incomplete polymer-
hamsters, or rabbits are preferred. In addition, the use of larger
ization).
animals, such as the dog, goat, pig, or sheep may be justified
based upon special considerations of the particular study.
6. Animal Model
Choice of the animal model should also consider the availabil-
ity of historical data on biological responses of these animals to
6.1 The choice of animal model shall take into consideration
similar devices to aid in analysis and comparison of the data the normal life span of the animal, the clinical use conditions,
obtained. device absorption kinetics, and the length of the implantation
study, and shall be justified. The strain, sex, age, weight, origin,
4.2 All animal studies shall be done in a facility in accor-
and general health of the animals used should be recorded.
dance with all appropriate regulations.
Institutional and government animal use and care policies and
5,6,7,8,9
regulations shall be followed.
5. Significance and Use
5.1 This practice is a guideline for a screening test of
candidate materials or assessment of local tissue response to
Animal Welfare Act, 7 U.S.C. § 2131 et seq., as amended. 2013.
absorbable medical devices which are expected to undergo
Animal Welfare Regulations, 9 CFR Chapter 1, Subchapter A, Parts 1, 2, and
complete absorption within three years.
3. 2004.
Health Research Extension Act of 1985, Public Law 99-158 November 20,
1985.
3 8
Available from American National Standards Institute (ANSI), 25 W. 43rd St., Office of Laboratory Animal Welfare, National Institutes of Health. Public
4th Floor, New York, NY 10036, http://www.ansi.org. Health Service Policy on Humane Care and Use of Laboratory Animals, Bethesda,
FDA Biocompatibility Guidance, “Use of International Standard ISO 10993-1, MD, 2015.
‘Biological evaluation of medical devices—Part 1: Evaluation and testing within a National Research Council, Guide for the Care and Use of Laboratory Animals:
risk management process’” (https://www.fda.gov/media/85865/download). Eighth Edition. Washington, DC, The National Academies Press; 2011.
F1983 − 23
6.2 The number of implant sites shall depend on the size of consideration for clinical best practice, availability, and dimen-
the implant and the animal. The distance between implants sional suitability for the intended implantation site:
shall be sufficient so that separate tissue blocks are prepared (1) Absorbable device/material with similar expected ab-
easily for each implant and that the local biological reactions sorption profile.
do not overlap or interfere with each other. Implants may be (2) Absorbable device/material with different absorption
placed bilaterally in soft tissue, including muscle. Bilateral profile or non-absorbable device/material.
implantation into bone should be considered carefully and
NOTE 3—Absorbable device/material controls that possess a different
justification given. In general, mice, rats, hamsters, and other
degradation rate than the test implant may require retrieval at additional
small laboratory animal species should receive no more than
intervals to allow assessment of tissue response at an equivalent stage in
the control material’s degradation/absorption process. Use of an absorb-
one implant on each side. Larger animals, including rabbits,
able device/material with an absorption profile different from the test
may receive up to five implants on each side. When the implant
implant may not allow a bilateral implantation and additional animals may
is composed of a collection of particles, pellets, and so forth,
be used for the implantation of the control device/material.
each collection is considered one implant site.
(3) Sham sites / sham animals—A sham surgical site (to
assess local effects), or a sham surgical animal (to assess local
6.3 Scientifically established analytical methods should be
and systemic effects) may be helpful. If a sham site or sham
used in the identification and quantification of degradation
animal is used, the same implantation procedure without the
products (ISO 10993-9, ISO 10993-18, ISO 10993-19, ISO
test or control should be used.
13781, Test Method F1635, Guide F2902, Guide F3268).
NOTE 4—Such sites may be used to assess the impact of surgical
Literature information (if available) on the fate of the absorb-
procedures but may not enable a direct comparison of tissue responses to
able material’s degradation products can be used to address
the ongoing presence of an implant (absorbable or nonabsorbable).
their absorption, distribution, metabolism, and excretion
7.3.1 The material/host (material surface area or mass-to-
(ADME) and identify the potential organs involved. Literature
body mass) ratio of any control material should be comparable
evaluations should focus on all degradation products, including
to the material/host ratio used for the test implant as described
those from major compositional components as well as any
in 5.3. The selection of the control shall be justified. Guidance
other constituents with known or suspected toxicities, at the
regarding considerations prior to commencing an in vivo study
amounts present that could impact tissue response.
of absorbable materials can be found in the following stan-
NOTE 1—A pilot study in vitro or in small animals may be undertaken
dards:
to assess the rate of degradation which can potentially be used to select
(1) Guide F2902—For absorbable polymeric devices, this
estimated time points for evaluating degradation in large animal studies.
standard describes the manufacturing, characterization,
NOTE 2—In some cases where degradation products or metabolites of
packaging, sterilization, and biocompatibility aspects and the
the candidate material are not known or well established, it may be
related testing that should be considered prior to undertaking in
possible to assess the quantitation and distribution of degradation products
or metabolites using in vivo radio-labeling methods following administra- vivo evaluations.
tion of radio-labeled parent material for ADME assessments. However, if
(2) ISO 10993-6—Provides absorbable-specific consider-
radio-labeling is used, a justification should be included to explain why
ations when evaluating a material’s biological safety through
radio-labeling is not expected to impact ADME results.
implantation, which includes guidance for selecting appropri-
ate animal retrieval intervals (see Clause 5 of ISO 10993-
7. Test Article and Implant Placement Considerations
6:2016).
7.1 Test Article—May be devices in their final finished form
(3) ISO/TS 17137—Provides recommendations on in vitro
or made from candidate materials in configurations specific for
and in vivo assessments of absorbable (test or control) implants
the animal study. Photograph(s) of the implant articles should
(see the stages of degradation depicted in Figure 2 and the
be taken prior to implantation. As described in 5.3, the
supporting discussions contained within subclauses 5.1, 5.3,
material/host ratio should be selected based on clinical use,
5.4, and 5.6 in ISO/TS 17137:2021).
with material/body mass ratios of 2X and 10X, if applicable.
(4) ISO/TS 37137-1—Provides supplemental absorbable-
Relevant configurations of implant articles, such as cylinders,
specific considerations when biologically evaluating a device
flat cloths, amorphous gels, and polymerizable liquids may be
in accordance with the ISO 10993 series.
used for material screening studies.
7.3.2 If assessing systemic endpoints as part of the implan-
tation study, it is essential that separate groups of animals be
7.2 The implantation site of the absorbable device or can-
used for test and control groups.
didate material shall be described and recorded with anatomic
landmarks and include adequate means to identify the specific 7.4 The material used shall be in its final finished form and
implant sites, including during and after advanced stages of sterilized as indicated for its ultimate use. It shall be handled
degradation. Such means of site identification may include use for implantation in a manner analogous to that for intended
of an implanted non-absorbable marker or other permanent final use (for example, special forceps, special cannulas or
method, such as a template.
needles, special syringes, and so forth).
7.3 Control materials shall be implanted using the same NOTE 5—If this method is used for material research, testing for
endotoxin prior to implantation should be considered.
placement techniques as the test material to allow the compari-
son of the tissue response. Choice of control devices/materials 7.5 The candidate material shall be described thoroughly to
with established biocompatibility and clinical relevance shall facilitate development of a suitable implant application proto-
adhere to the following selection priority with appropriate col. The ADME of the material and its degradation products
F1983 − 23
should be described. The information shall include, but is not 8.2.3 If infection or accidental injury of the test implant site
limited to, the following: occurs, record the information and process the implant site and
7.5.1 Expected mechanism of degradation (for example, tissues and organs as described in 8.3 and 9.1. Exclusion of this
hydrolysis, enzymatic, phagocytosis, and so forth). data in the final analysis shall be justified, as infection/injury
7.5.2 Expected nonabsorbable degradation products (for could be implant-related. A replacement animal may be added,
example, fibrils, particles from composites). if desired.
7.5.3 Expected stages and rate of degradation.
8.2.4 If an animal dies or is euthanized before the scheduled
7.5.4 Expected target organ effects (for example, eliminated termination, record the information and process the implant
in the kidney, stored in the liver, stored in the spleen or lymph
site and tissues and organs as described in 8.3 and 9.1.
nodes). Exclusion of this data in the final analysis of results shall be
justified, as death could be implant-related. The cause of death
7.6 For each time period, at least six small laboratory
shall be investigated and reported. If the death is related to
animals shall be used with either unilateral or bilateral implants
anesthesia, a replacement animal may be selected.
to assess local responses per test and control groups. For larger
animals, including rabbits, at least four animals shall be used
8.3 Euthanasia and Post-Mortem Assessments:
per time period per test and control group. It is recommended
8.3.1 The euthanasia method shall be the one recommended
that additional animals be included in the protocol to address
for the particular animal species according to local and
assessment of systemic responses (for example, per ISO
government regulations. The termination time points shall be
10993-11) and to accommodate any differences in in vitro and
based on the expected degradation rate of the implant, and
in vivo degradation rates of the material.
include early,
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: F1983 − 14 F1983 − 23
Standard Practice for
Assessment of Selected Tissue Effects of Absorbable
Biomaterials for Implant Applications
This standard is issued under the fixed designation F1983; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This practice provides experimental protocols for biological assays of tissue reactions to absorbable biomaterials for implant
applications. This practice applies only to absorbable materials with projected clinical applications in which the materials will
reside in bone or soft tissue longer than 30 days and less than three years. Other standards with designated implantation times are
available to address shorter time periods. Careful consideration should be given to the appropriateness of this practice for slowly
degrading materials that will remain for longer than three years. It is anticipated that the tissue response to degrading biomaterials
will be different from the response to nonabsorbable materials. In many cases, a chronic inflammatory response may be observed
during the degradation phase, but the local histology should return to normal after absorption; therefore, the minimal tissue
response usually equated with “biocompatibility”biocompatibility may require long implantations.
1.2 The time period for implant absorption will vary depending on chemical compositioncan depend on variables of chemical
composition, implant size, implant location, and test subject species; therefore, the implantation times for examination of tissue
response will be linked to animal models. Therefore, the selected time points for assessing tissue effects may be selected based
on the rate of absorption. No single implantation time is indicated in this practice.
1.3 These protocols assess the effects of the material on the animal tissue in which it is implanted. The experimental protocols
They do not fully assess systemic toxicity, carcinogenicity, teratogenicity, reproductive and development toxicity, or mutagenicity
of the material. Other standards are available to address these issues.
1.4 To maximize use of the animals in the study protocol, all toxicological findings should be recorded. There are some aspects
of systemic toxicity, including effects of degradation products on the target organs, that different organs and tissues downstream
of or surrounding the target site, can be addressed with this practice, and these effects should be documented fully. practice.
1.5 Because animal models are not identical to human biology, this practice cannot account for all potential biological hazards,
for example the effect of the oligosaccharide a-Gal (Gala 1,3-Galb1-4GlcNAc-R), known as the “a-Gal” epitope present in
xenogeneic materials on humans. See ISO 22442.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and healthsafety, health, and environmental practices and determine
the applicability of regulatory limitations prior to use.
1.7 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
This practice is under the jurisdiction of ASTM Committee F04 on Medical and Surgical Materials and Devices and is the direct responsibility of Subcommittee F04.16
on Biocompatibility Test Methods.
Current edition approved Nov. 1, 2014April 1, 2023. Published January 2015April 2023. Originally approved in 1999. Last previous edition approved in 20082014 as
F1983 – 99 (2008).F1983 – 14. DOI: 10.1520/F1983-14.10.1520/F1983-23.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
F1983 − 23
2. Referenced Documents
2.1 ASTM Standards:
F561 Practice for Retrieval and Analysis of Medical Devices, and Associated Tissues and Fluids
F750 Practice for Evaluating Acute Systemic Toxicity of Material Extracts by Systemic Injection in the Mouse
F763 Practice for Short-Term Intramuscular Screening of Implantable Medical Device Materials
F981 Practice for Assessment of Compatibility of Biomaterials for Surgical Implants with Respect to Effect of Materials on
Muscle and Insertion into Bone
F1408 Practice for Subcutaneous Screening Test for Implant Materials
F1635 Test Method for in vitro Degradation Testing of Hydrolytically Degradable Polymer Resins and Fabricated Forms for
Surgical Implants
F1903 Practice for Testing for Cellular Responses to Particles in vitro
F1904 Practice for Testing the Biological Responses to Particles in vivo
F1905F2902 Practice For Selecting Tests for Determining the Propensity of Materials to Cause ImmunotoxicityGuide for
Assessment of Absorbable Polymeric Implants (Withdrawn 2011)
F1906F3268 PracticeGuide for Evaluation ofin vitro Immune Responses In Biocompatibility Testing Using ELISA Tests,
Lymphocyte Proliferation, and Cell MigrationDegradation Testing of Absorbable Metals (Withdrawn 2011)
2.2 ISO Standards:
ISO 10993-6 Biological evaluation of medical devices—Part 6: Tests for local effects after implantation
ISO 10993-9 Biological evaluation of medical devices—Part 9: Framework for identification and quantification of potential
degradation products
ISO 10993-11 Biological evaluation of medical devices—Part 11: Tests for systemic toxicity
ISO 10993-18 Biological evaluation of medical devices—Part 18: Chemical characterization of medical device materials within
a risk management process
ISO/TS 10993-19 Biological evaluation of medical devices—Part 19: Physico-chemical, morphological and topographical
characterization of materials
ISO 13781 Implants for surgery—Homopolymers, copolymers and blends on poly(lactide)—In vitro degradation testing
ISO/TS 17137 Cardiovascular implants and extracorporeal systems—Cardiovascular absorbable implants
ISO 22442 Medical devices utilizing animal tissues and their derivatives
ISO/TS 37137-1 Biological evaluation of absorbable medical devices—Part 1: General requirements
3. Terminology
3.1 Definitions:
3.1.1 final finished form—a device or device component that includes all manufacturing processes for the “to be marketed” device
including packaging and sterilization, if applicable.
4. Summary of Practice
3.1 Under strict aseptic conditions, specimens of the sterile final implant form candidate material are implanted into the most
relevant anatomical tissue site in small laboratory animals, preferably mice, rats, hamsters, or rabbits.
4.1 The Under strict aseptic conditions, sterile test articles (for example, final device) are implanted into a relevant animal model
and at a clinically relevant anatomical tissue site. However, for screening candidate materials, testing in a clinically relevant animal
model and anatomical tissue site may not be necessary. Small laboratory animals such as mice, rats, hamsters, or rabbits are
preferred. In addition, the use of larger animals, such as the dog, goat, pig, or sheep may be justified based upon special
considerations of the particular study. Choice of species also should the animal model should also consider the availability of
historical data on biological responses of these animals to similar devices to aid in analysis and comparison of the data obtained.
4.2 All animal studies shall be done in a facility approved by a nationally recognized organization and in accordance with all
appropriate regulations.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Available from American National Standards Institute (ANSI), 25 W. 43rd St., 4th Floor, New York, NY 10036, http://www.ansi.org.
FDA Biocompatibility Guidance, “Use of International Standard ISO 10993-1, ‘Biological evaluation of medical devices—Part 1: Evaluation and testing within a risk
management process’” (https://www.fda.gov/media/85865/download).
F1983 − 23
5. Significance and Use
5.1 This practice is a guideline for a screening test for the evaluation of the of candidate materials or assessment of local tissue
response to materials that may be selected for implantation into the human body and absorbable medical devices which are
expected to undergo complete absorption within three years.
5.2 This practice is similar to thatthose for studies on candidate materials or medical devices that are not absorbable, such as those
specified in Practices F763, F981, and F1408; however, analysis of the host response must take into account the effect of
degradation and degradation products on the inflammatory response at the local tissue site and on subsequent healing of the
implantation site. site, as well as the potential for adverse distal tissue effects.
5.3 The material to be tested For testing of absorbable medical devices, the test article for implantation should be in the final
finished form as for intended use, including sterilization. Material/body ratios should be relevant to that of intended device use.
Material packaging and sterilization (if applicable). Configurations specific to the animal study may be needed. The test article’s
surface-area-to-body mass or mass-to-body mass ratios within the animal model should be established by calculating based on
surface-area-to-body mass or mass-to-body mass ratios in humans during the device’s intended clinical use. Worst-case clinical
dose should be considered in the study design. For implantation studies incorporating evaluation of both local tissue responses and
systemic toxicity, exaggerated material surface area or mass to body mass ratios of 1X, 10X, and 50X if applicable, are
recommended. mass-to-body mass ratios (for example, a 2X to 10X safety factor to assess implant safety for regulatory
submissions) compared to clinical use (for example, largest device size, maximum number of devices) should be considered, unless
otherwise justified. For example, implantation of exaggerated doses may not be feasible in the selected animal model. For some
devices, additional animal group(s) for exaggerated conditions should be considered if dose response information is needed.
Additionally, for some devices, exaggerated dose at a specific implantation site can also be used to evaluate local tissue responses.
5.4 Materials that are designed for use in devices with in situ polymerization shall be introduced in a manner such that in situ
polymerization occurs. Testing Additional testing of individual precursor components is not recommended.or partially polymerized
materials may be needed in some cases (for example, if testing of the final implant indicates an adverse response or incomplete
polymerization).
6. Test Animals and SitesAnimal Model
6.1 Choice of test animal The choice of animal model shall take into consideration the normal life span of the animal animal, the
clinical use conditions, device absorption kinetics, and the length of the implantation study. Small laboratory animals are preferred.
study, and shall be justified. The strain, sex, age, and origin weight, origin, and general health of the animals used should be noted.
If larger animals are used, justification for their use should be provided. The source of the animals, species/strain, weight, age
(where known or approximate if not known), general health, and boarding conditions should be recorded. Animal recorded.
5,6,7,8,9
Institutional and government animal use and care policies and regulations shall be followed.
6.2 The number of implant sites shall depend on the size of the implant and the animal. The distance between implants shall be
sufficient so that separate tissue blocks are prepared easily for each implant and that the local biological reactions do not overlap
or interfere with each other. Implants may be placed bilaterally in soft tissue, including muscle. Bilateral implantation into bone
should be considered carefully and justification given. In general, mice, rats, hamsters, and other similarly sized rodents small
laboratory animal species should receive no more than one implant on each side. Larger animals, including rabbits, may receive
up to five implants on each side. When the implant is composed of a collection of particles, pellets, and so forth, each collection
is considered one implant site.
6.3 Before embarking on studies in large animals, it is recommended that a pilot study Scientifically established analytical
methods should in vitro or in rodents be undertaken to determine the expected rate of degradation and assist in the selection of
study periods in long-term animal studies. During analysis of study results, the distribution and metabolism of the degradation
products should be determined by available analytical methods, such as massbe used in the identification and quantification of
degradation products (ISO 10993-9, ISO 10993-18, ISO 10993-19, ISO 13781, Test Method F1635spectrometry. Alternatively,,
Animal Welfare Act, 7 U.S.C. § 2131 et seq., as amended. 2013.
Animal Welfare Regulations, 9 CFR Chapter 1, Subchapter A, Parts 1, 2, and 3. 2004.
Health Research Extension Act of 1985, Public Law 99-158 November 20, 1985.
Office of Laboratory Animal Welfare, National Institutes of Health. Public Health Service Policy on Humane Care and Use of Laboratory Animals, Bethesda, MD, 2015.
National Research Council, Guide for the Care and Use of Laboratory Animals: Eighth Edition. Washington, DC, The National Academies Press; 2011.
F1983 − 23
Guide F2902prediction may, Guide F3268be done by radio-labeling the material and following the loss of radioactivity; however,
radioactive specimens shall not be used for biocompatibility testing. Other methods of characterizing the absorption are acceptable.
The target organs of the metabolism and excretion of the products should be identified. It is recommended that acute systemic
studies with material extracts according to Practice). Literature information (if available) on the fate of the absorbable material’s
degradation products can be used to address their absorption, distribution, metabolism, and excretion (ADME) and identify the
potential organs involved. Literature evaluations should focus on all degradation products, including those from major
compositional components as well as any other constituents with known or suspected toxicities, at F750 be completed prior to the
initiation of the implantation study.the amounts present that could impact tissue response.
NOTE 1—A pilot study in vitro or in small animals may be undertaken to assess the rate of degradation which can potentially be used to select estimated
time points for evaluating degradation in large animal studies.
NOTE 2—In some cases where degradation products or metabolites of the candidate material are not known or well established, it may be possible to assess
the quantitation and distribution of degradation products or metabolites using in vivo radio-labeling methods following administration of radio-labeled
parent material for ADME assessments. However, if radio-labeling is used, a justification should be included to explain why radio-labeling is not expected
to impact ADME results.
7. Implant Specimens Test Article and Implant Placement Considerations
7.1 Design of the Implant—Test Article—Specimens may be made from theMay be devices in their final finished form candidate
materialor made from candidate materials in configurations specific for the animal study. Photograph(s) of the implant articles
should be taken prior to implantation. As described in 4.35.3, the material/host ratio should be available and referrable to ultimate
use in the human selected based on clinical use, with material/body mass ratios of 1X, 10X, and 50X, if applicable, recommended.
2X and 10X, if applicable. Relevant configurations of implant specimens,articles, such as cylinders, flat cloth,cloths, amorphous
gels, and polymerizable liquids may be used. used for material screening studies.
7.2 The implantation site of the absorbable device or candidate material shall be described and recorded with anatomic landmarks
and include adequate means to identify the specific implant sites, including during and after advanced stages of degradation. Such
means of site identification may include use of an implanted non-absorbable marker or other permanent method, such as a template.
7.3 The implantation site of the candidate material shall be accompanied by the use of an implanted marker or other permanent
method, such as a template, to mark the implant site to allow identification of the implant site at the various time periods. In
additional animals, control materials shall be implanted by the same techniques, to allow the Control materials shall be implanted
using the same placement techniques as the test material to allow the comparison of the tissue response. When assessing systemic
endpoints, it is essential that separate groups of animals be used for test and comparator groups. A sham surgical site, or a sham
surgical animal, is necessary.Choice of control devices/materials with established biocompatibility and clinical relevance shall
adhere to the following selection priority with appropriate consideration for clinical best practice, availability, and dimensional
suitability for the intended implantation site:
(1) Absorbable device/material with similar expected absorption profile.
(2) Absorbable device/material with different absorption profile or non-absorbable device/material.
NOTE 3—Absorbable device/material controls that possess a different degradation rate than the test implant may require retrieval at additional intervals
to allow assessment of tissue response at an equivalent stage in the control material’s degradation/absorption process. Use of an absorbable device/material
with an absorption profile different from the test implant may not allow a bilateral implantation and additional animals may be used for the implantation
of the control device/material.
(3) Sham sites / sham animals—A sham surgical site (to assess local effects), or a sham surgical animal (to assess local and
systemic effects) may be helpful. If a sham site or sham animal is used, the same implantation procedure without the test or control
should be used.
NOTE 4—Such sites may be used to assess the impact of surgical procedures but may not enable a direct comparison of tissue responses to the ongoing
presence of an implant (absorbable or nonabsorbable).
7.3.1 The material/host (material surface area or mass-to-body mass) ratio of any control material should be comparable to the
material/host ratio used for the test implant as described in 5.3. The selection of the control shall be justified. Guidance regarding
considerations prior to commencing an in vivo study of absorbable materials can be found in the following standards:
(1) Guide F2902—For absorbable polymeric devices, this standard describes the manufacturing, characterization, packaging,
sterilization, and biocompatibility aspects and the related testing that should be considered prior to undertaking in vivo evaluations.
(2) ISO 10993-6—Provides absorbable-specific considerations when evaluating a material’s biological safety through
implantation, which includes guidance for selecting appropriate animal retrieval intervals (see Clause 5 of ISO 10993-6:2016).
F1983 − 23
(3) ISO/TS 17137—Provides recommendations on in vitro and in vivo assessments of absorbable (test or control) implants (see
the stages of degradation depicted in Figure 2 and the supporting discussions contained within subclauses 5.1, 5.3, 5.4, and 5.6
in ISO/TS 17137:2021).
(4) ISO/TS 37137-1—Provides supplemental absorbable-specific considerations when biologically evaluating a device in
accordance with the ISO 10993 series.
7.3.2 If assessing systemic endpoints as part of the implantation study, it is essential that separate groups of animals be used for
test and control groups.
7.4 The material used shall be in its final finished form and sterilized as indicated for its ultimate use. It shall be handled for
implantation in a manner analogous to that for intended final use, foruse (for example, special forceps, special cannulas or needles,
special syringes, and so forth.forth).
NOTE 5—If this method is used for material research, testing for endotoxin prior to implantation should be considered.
7.5 The candidate material shall be described thoroughly to facilitate development of a suitable implant application protocol. The
absorption, distribution, metabolism, and excretion ADME of the material and its degradation products should be described. The
information shall include, but is not limited to, the following:
7.5.1 Expected method of degradation, formechanism of degradation (for example, hydrolysis, enzymatic, phagocytosis, and so
forth. forth).
7.5.2 Expected nonabsorbable degradation products, forproducts (for example, fibrils, particles from composites.composites).
7.5.3 Expected stages and rate of degradation.
7.5.4 Expected target organ effects where known or expected, for (for example, eliminated in the kidney, stored in the liver, stored
in the spleen or lymph nodes.nodes).
7.6 For each time period, at least six rodents small laboratory animals shall be used with either single or bilateral implants. For
the larger animals, unilateral or bilateral implants to assess local responses per test and control groups. For larger animals,
including rabbits, at least four animals shall be used per time period. period per test and control group. It is recommended that
additional animals be included in the initial protocol to accommodate any unexpected changesprotocol to address assessment of
systemic responses (for example, per ISO 10993-11) and to accommodate any differences in in vitro and in vivo degradation rates
of the material.
8. Procedure
8.1 Implantation:
8.1.1 Implant the specimen under sterile conditions in anesthetized animals. Where possible, implant the specimentest and control
under aseptic conditions in animals that are under surgical plane of anesthesia. For screening studies with subcutaneous
implantations, place the articles using a trochar method to avoid the need for an incision. If an incision is needed, insert the implant
as far from the incision site as possible. Close the insertion site with a suitable suture material.
7.1.1.1 A sham site or sham animal with the identical implantation procedure, but not the test material, should be included in the
protocol. If animals are to be used as part of a systemic toxicity study, the sham shall be a separate animal.
8.1.2 The implantation site shall be described and recorded with anatomic landmarks and shall be marked in a manner suitable
for identification of the site at the designated time periods. The use of a permanent skin marker and a template marking the
placement of the specimentest and the shamcontrol/sham site i
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...